Reports until 18:36, Friday 04 April 2014
H1 ISC
keita.kawabe@LIGO.ORG - posted 18:36, Friday 04 April 2014 (11185)
Reduced diode current for ISCTEY Prometheus

On the test sheet of this unit (SN 2240A), the current was specified at 1.61A. Since the potentiometer for the current was maxed out and the current clamp LED was on, though the current was 1.63 I turned it down to 1.61. No current clamp LED any more.

H1 ISC
stefan.ballmer@LIGO.ORG - posted 17:40, Friday 04 April 2014 - last comment - 12:04, Saturday 05 April 2014(11182)
Today's WFS battle

 

Sheila, Chris, Stefan

In order to tackle the changing sensing matrix problem today we tried to use the soft-hard degree of freedoms. The theory
is that the beam overlap is much more sensitive to the hard DoF, thus this should be the one that is easily sensed independent
of the sensing-matrix-of-the-day. Then a gain hierarchy could be used to lock down everything.

Thus we started with calculation the theoretical hard-soft output matrix. Details can be found at
https://gwic.ligo.org/thesisprize/2011/dooley_thesis.pdf
For reference, the desin RoCs are ETM: 2245m, ITM: 1934m, giving (with L=3994.5m) g_ETM =-0.779, g_ITM =-1.065. 

We messed up the initial calculation and worked with
ITM =   .571  .495  * SOFT 
ETM    -.495  .571    HARD 
for yaw, and opposite off-diagonal sign for pitch.

while we should have used
ITM =   1.000 -0.867  * SOFT 
ETM     0.867  1.000   HARD 

We verified that the corresponding modes were mostly hard/soft by observing the power drop for a fixed offset in soft/hard.
The hard mode has the bigger power drop. 


Next we engaged DoF's one at a time, starting with the hard modes.

For yaw, DoF2 (hard mode) we inverted the last measured sensing matrix and converted it into the soft/hard basis. We got
SOFT =   5.14  11.7 *  WFS A 
HARD    -.272  1.04    WFS B
The hard loop easily closed that way. 

For the pitch DoF's we first tried the same scheme, but noticed that the hard mode ended up much stronger in what should have
been the soft signal. So instead we used the diagonal matrix WFS A -> hard, WFS B -> soft
SOFT =   0 1 *  WFS A 
HARD     1 0    WFS B
Again closing that loop was straight forward.

Next we closed the soft loops with low gain. That seemed straight forward. But then we started noticing that we saturated the PUM
with slow signals because the top-mass bleed-off was too slow. (We moved the bleed-off filters to the unused UIM filters to keep
the reliev script working.)

Before we could fix the saturation issue we noticed that the cavity transmitted spot on the camera did no longer line up
with the straight shot beam once all loops were on. So we manually realigned the cavity and tried to reengage the WFS, loop-by-loop.
Yaw hard worked fine, but to our surprise pitch hard didn't work at all - the error signal behaviour was completely different.

So our tentative conclusion is that our "sensing matrix of the day" doesn't seem provide a clean signal for the supposedly easy
hard pitch mode.


Comments related to this report
rainer.weiss@LIGO.ORG - 08:13, Saturday 05 April 2014 (11186)ISC
I have been following the struggle to get a stable sensing matrix for the WFS and urge you to make sure
the WFS head isn't driven into oscillation by the rf on the optical beam hitting the detector. This was a central
problem in the WFS heads in initial LIGO and led to the rewiring of input leads in the WFS head. The
symptom was that at some critical intensity into a vulnerable quadrant (not all quadrants in a given WFS head
did this), the amplifier attached to the offending quadrant would begin to oscillate at the drive frequency and
produce an output signal with a different RF phase, the transition from the correct phase to the phase when
oscillating was quite sudden as a function of the intensity. The rewiring and the removal of the gain adjust
chip in each channel stopped this. Do you suppose this is happening now? A good test is to illuminate the 
quadrant with RF modulated laser light and display the output of all four quadrants on an oscilloscope while
one varies the laser intensity and the amplitude modulation index. One of the traces will shift in phase relative 
to the others on the screen.  
christopher.wipf@LIGO.ORG - 12:04, Saturday 05 April 2014 (11188)

(Sheila, Stefan, Chris)

We again closed 3 WFS loops (the hard pitch loop, and both yaw loops), after decentering the beam on the WFS heads.  In this state, I compared Rich's new blend filters (Tbetter) and the standard (Tcrappy), finding a modest but repeatable improvement.

The reason we investigated the centering was that, with the beam spot well centered, the response of individual WFS quadrants to pitch drives on ETMX and ITMX was bogus. All quadrants responded with the same phase, as if it had been a length drive. We then suspected that an ugly beam profile on the WFS could make the DC light levels on the quadrants unreliable for centering. Moving the beam spot down on WFS_A, sure enough, there was a "happy place" where a usable error signal for the hard pitch DOF was recovered. We will check the beam quality on the WFS again to see if it can explain this.

The attached spectra show the ALS COMM noise with the ISI RY blends on Tcrappy (refs 10/11) and Tbetter (refs 8/9). Also, Rich advised us of a problem with the RZ blends, so those were left on Start for all measurements.

Images attached to this comment
H1 SUS (SUS)
giles.hammond@LIGO.ORG - posted 17:23, Friday 04 April 2014 (11184)
ITMY modal frequencies
Today we finished off measuring the modal/violin frequencies of ITMY. Everything looks fine. These are briefly summarised below:

Modal
			frequency (Hz)			bin width (Hz)
pendulum		0.65				1.56E-02
transverse		0.656				1.56E-02
pitch			1.109				1.56E-02
yaw			1.094				1.56E-02
bounce (*)		6.87				1.25E-01
roll   (*)		9.7				1.25E-01
	
* Note: bounce and roll will be sqrt(2) higher when both PUM and ITM are free (9.7Hz/13.7Hz)

Violin modes						
			frequency (Hz)			bin width (Hz)
Back Left		501.5	          		2.50E-01
Back Right		501.3		        	2.50E-01
Front Left		504.2			        2.50E-01
Front Right		502.8			        2.50E-01

The average violin mode is 502.4Hz with a standard deviatioin of 1.3Hz (+1.7Hz, -1.1Hz)

Fibre guards have now been installed.

Giles, Travis, Jason, Gary, Betsy 
H1 CDS (CDS)
david.barker@LIGO.ORG - posted 17:15, Friday 04 April 2014 (11183)
medm launcher icon and guardian

I have been trying different things to fix the problem of starting of guardmedm from a sitemap which in turn was started from a panel icon. Following Keith's system I configured the icon to run a script to start MEDM instead of running it directly and it seem to have fixed the problem. If you are logged in as controls and need the new configuration you will have to log out and log back in again. 

I was not able to print the error of why a directly launched sitemap was unable to run guardmedm. I tried using fie paths in full or redirecting stderr to a file to no avail. If I ran within a terminal to get stderr printed it started working.

LHO VE
kyle.ryan@LIGO.ORG - posted 16:50, Friday 04 April 2014 (11181)
Sprayed new joints @ Y-end with helium
Sprayed "around" each new joint with an audible flow of helium for ~25 seconds duration -> Access to several joints was impeded by various obstructions and, in many cases, I could not apply the helium to the test slots etc. -> No flanges were bagged and this exercise should not be considered a definitive test but should have revealed any significant offenders.  

LD indicated baseline signal was "offscale" < 10-11 torr*L/sec for the 1st half of testing -> I guessed the time constant to be ~100 seconds for this V/S -> to validate the less than one time constant application duration I verified that the baseline helium indication was legitimate by spraying the O-ring at the inlet to the LD and observing the permeation.  The signal rose to a high of 8 x 10-9 torr*L/sec over the course of 60 minutes -> I let the signal decay to < 8 x 10-10 torr*L/sec before resuming and testing the remaining joints
H1 ISC
alexan.staley@LIGO.ORG - posted 16:22, Friday 04 April 2014 (11180)
EY PLL OLTF

End Y PLL Open Loop TF results:

Servo Board Settings:

 

The UGF is at 23kHz with a phase margin of 52 deg (consistent with HIFOX). Using the AG spectrum analyzer, I tried finding the PZT resonance. The data is a bit ratty, but there appears to be a resonance a little below 200kHz (see the picture -- the two pictures were collected with different BWs; the first at 300Hz and the second at 2Hz; this notch-looking feature seemed to be repeatable...).  Meanwhile, the board has a notch at around 280kHz, which also seems to appear in the pictures.

 

I have also uploaded and plotted the data I collected.

Images attached to this report
Non-image files attached to this report
H1 SEI
patrick.thomas@LIGO.ORG - posted 16:03, Friday 04 April 2014 - last comment - 10:39, Saturday 05 April 2014(11174)
Plots of SEI watchdog trips
Attached are plots of SEI watchdog trips.
Images attached to this report
Comments related to this report
jameson.rollins@LIGO.ORG - 10:39, Saturday 05 April 2014 (11187)

It would be useful if there was some context for each of these.  Do we know what was going on at the time of each trip?  It looks like there are four distinct events:

  • 09:15 ITMY ISI stage 1/2 drives
  • 14:52 BS ISI GS13
  • 15:10 BS HPI L4C, BS ISI T240
  • 15:21 ITMY HPI L4C

It would be useful to know what was going on around each of these times.

LHO General
patrick.thomas@LIGO.ORG - posted 16:00, Friday 04 April 2014 (11173)
Ops summary
8:15 morning meeting:

HAM4 HEPI replacement of Parker valve
Jim W. working on HAM4/5 SEI model update
Jim W. troubleshooting isolation filters at end Y
Grout cleaning at end Y
Kyle helium leak testing at end Y


08:36 Jeff B. to end Y for grout removal
09:00 David H. working on TCSX
09:00 Keita transitioned end Y to laser safe
09:12 Filiberto getting TCS electronics from mid Y
09:32 Karen to end Y to clean grout removal area
09:46 Filiberto and Jim W. to end Y, troubleshooting seismic rack components for misbehaving seismometer
10:19 Grouting removal finished, end Y transitioned to laser hazard
10:55 Karen back from end Y
12:04 David H. and Alastair stopping work on TCSX for lunch
12:47 Thomas restarting h1ecatc1 Beckhoff system manager
13:08 David H. and Alastair working on TCSX
14:09 Filiberto and Aaron to end Y to replace chassis
15:06 h1isiham5 model restarted

I removed the WD SUM and DC components from the end Y SUS IOP watchdog alarm handler. These were no longer valid.
H1 SEI (DAQ)
david.barker@LIGO.ORG - posted 15:28, Friday 04 April 2014 (11179)
Restart h1isiham5 model

WP4548

Jim, Rich M and Dave.

We installed a new h1isiham5 model on h1seih45. Jim corrected a cut-and-paste error with the BIO parts.

No DAQ restart was required, restart went smoothly. This closes WP4548.

H1 ISC
keita.kawabe@LIGO.ORG - posted 11:41, Friday 04 April 2014 - last comment - 13:49, Friday 04 April 2014(11176)
More EY PLL work (it fully works now). (Keita, Daniel)

Opened the PLL CM board chassis (S1002641) and Installed a notch filter daughter board S1300952 in the fast option slot. The notch frequency is about 290kHz now.

Put a timing comparator in the remote rack.

Recalibrated the DC diodes (fiber power, fiber wrong pol monitor, BBPD).

Auto locker worked with these things in place.

One odd thing was, when we started working the beat note was -40dBm or something, a big decrease from yesterday.

Then we started calibrating things, touched some steering mirrors to center the beam on DCPDs and such.

Then something happened, suddenly we're back to -10dBm beat note. We couldn't find out any obvious cause for this.

Comments related to this report
keita.kawabe@LIGO.ORG - 11:56, Friday 04 April 2014 (11177)

Power levels:

Fiber output is about 60uW.

About 3uW is sent to the fiber power monitor.

About 55uW goes to the PBS.

This morning 31uW was on the wrong polarization monitor PD (it was less yesterday).

About 24uW goes to 50:50 splitter and is combined with Prometheus IR.

From the Prometheus, about 39mW is coming to the PBS.

HWP was adjusted to send 1mW to the IR monitor diode path and 38mW to the 50:50 splitter that combines Prometheus and fiber.

In the Prometheus IR monitor path that receives 1mW, there is a pick off mirror that throws away most of the power and send 10% or something to the actual PD, but I didn't measure the splitting ratio there.

The BBPD receives 19mW.

daniel.sigg@LIGO.ORG - 13:49, Friday 04 April 2014 (11178)
Images attached to this comment
H1 General
jeffrey.bartlett@LIGO.ORG - posted 11:28, Friday 04 April 2014 (11175)
Grout Removal from EY Op-Lev Pier
   The excess grout found on the Op-Lev receiver pier at End-Y has been removed. The grout was kept damp during the removal process to hold down dust contamination. Dust counts were monitored during the process and although elevated no readings over 9000 0.3 micron particles per sample time were recorded. The surrounding floor area has been recleaned.   
Images attached to this report
LHO VE
kyle.ryan@LIGO.ORG - posted 09:21, Friday 04 April 2014 (11172)
~0915 hrs. local -> Switched IP6 from 5000V to 7000V


			
			
H1 SUS
betsy.weaver@LIGO.ORG - posted 16:18, Thursday 03 April 2014 - last comment - 09:10, Friday 04 April 2014(11160)
ITMy fiber welding completed

Travis, Giles, Gary, Jason, Norna, Betsy

This afternoon, Travis, Gary, and Giles finished welding in fibers to the ITMy (ITM011) and it's glass PUM.  At some point during the welding process on Wed a roll of the Test Mass was introduced.  It was noticed this morning before the final destress/annealing took place.  The team attempted a correction to alleviate ~1/2 of the roll error which was somewhat successful.  After reevaluating the tolerance for roll with Dennis and Calum it was decided to leave the roll error as is.  The Test Mass is now hanging from the fibers.

 

Note, the roll error observed this morning was +/- 1.2mm (~7mRad).  We determined that this roll occured between welding the 2 sides of the suspension with the masses locked down the entire time.  The PUM did not show this large roll error.  After correcting and suspending the ITMy, the roll was found to be +/- 0.75mm (~4mRad).  While the roll "tolerance" has never really been set, we attempt to get the roll error within +/- 0.2 mm because that is what we have found we are able to do in all of the previous other monolithic suspensions.

Comments related to this report
betsy.weaver@LIGO.ORG - 16:26, Thursday 03 April 2014 (11162)

Email from Dennis regarding acceptable roll tolerances of an ITM:

Calum,
Using the Zemax model to ray trace, I find that a 7 mrad roll rotation
of the ITM causes only ~0.03 mm radial decentering of the beam at the BS
and a ~0.8 mm radial decentering of the beam at the PRM and at the SRM.
I think that this is acceptable.

Other effects of a DC roll error in the ITM:
- increased coupling of vertical ISI motion to roll motion of the ITM
... but so what?
- slight rotation in the of the phase maps of the ITM ... but so what?
- slight shift in the violin mode frequencies due to the length change
(I think the force in each is not changed due to the roll bias error),
but this is ~1 part in 1000 ... so likely within the violin mode
repeatability anyway?

-- 
Dennis Coyne
giles.hammond@LIGO.ORG - 18:07, Thursday 03 April 2014 (11164)SUS
Attached are some photos of the welds and the final suspension. As mentioned by Betsy, a +/-1.2mm (~7mRad) roll appeared between welding and with the ITM and PUM locked. As a result, during the destress, the PUM right welds developed a slight necking. The stock still has a large radius at this point and we will run some thermal noise models to verify that no significant change in performance is expected.
Images attached to this comment
mark.barton@LIGO.ORG - 08:58, Friday 04 April 2014 (11169)

To see if the residual roll was problematic, I prepared a case mark.barton/20140304TMproductionTMrollpert of the QuadLite2Lateral model with the latest monolithic parameters but with a perturbation to d4 (the height of the fibre attachment point above the optic COM) at each of the fibres, positive on one side and negative on the other. Even a perturbation of ±1.2 mm (the value reported to me before the extra de-stressing reduced it to 0.75 mm), there was no visible difference in any of the top mass force/torque to optic TFs and especially no extra peaks from cross coupling. This is to be expected because the optic is almost perfectly symmetrical in roll and the compliance of the fibres was not changed (the de-stressing did not touch the central section of the fibres so this was not changed in the model). So the performance of the suspension should not be degraded.

jason.oberling@LIGO.ORG - 09:10, Friday 04 April 2014 (11170)

Final alignment numbers for the monolithic (all directions/rotations reported from the view of the IAS equipment, i.e. looking at the HR face of the ITMy):

  • Pitch measurements
    • PUM: 740 µrad up
    • ITMy: 400 µrad down
    • Differential: 1.14 mrad down
    • Differential Spec: 2.0 mrad
  • Roll
    • PUM: ±0.15 mm CW
    • ITMy: ±0.9 mm CW
    • Differential: ±0.75 mm CW
  • Center of Mass Separation
    • Left: 600.5 mm
    • Right: 602.0 mm
    • Target (for both sides): 602.0 mm
H1 SUS
travis.sadecki@LIGO.ORG - posted 16:32, Thursday 27 March 2014 - last comment - 09:19, Friday 04 April 2014(11044)
ITMx Monolithic Welding

The new ITMx monolithic welding has been successfully completed and was left suspended overnight last night.  The majority of the modal measurements have also been completed today, with the exception of the violin modes, which are proving to be trickier to measure than in previous suspensions.  We will continue with these measurements tomorrow with some different equipment.  Stay tuned for subsequent posts with IAS and modal measurement numbers.

Comments related to this report
angus.bell@LIGO.ORG - 12:28, Friday 28 March 2014 (11062)
Modal frequencies of the test mass (ITMX) with the PUM nominally locked are:

ITMX				
          value	resolution		
Pendulum   0.649	0.002						
Pitch	   1.1153	0.002		
Yaw	   1.0928	0.002		
Bounce	   6.74 	0.016		
Roll	   9.195	0.016		

Violin modes				fibre serial number
BR	   499.9	0.06		S1400155
BL	   501.2	0.06		S1400156
FR	   500.8	0.06		S1400138
FL	   502.2	0.06		S1400153

[just as a note to self as much as anything - remember to use the Uniform window on the spectrum analyser for the violin modes when using the chirped sweep output]
Non-image files attached to this comment
jason.oberling@LIGO.ORG - 09:19, Friday 04 April 2014 (11171)

Forgot to post this last week.  Final alignment numbers from the ITMx welding (all directions/rotations reported from the view of the IAS equipment, i.e. looking at the HR face of the ITMx):

  • Pitch
    • PUM: 10 µrad down
    • ITMx: 2.12 mrad down
    • Differential: 2.11 mrad down
      • The pitch of the ITMx in chamber is 619 µrad down, so we align the ITMx for the welding at this pitch.  Since we are targeting a differential pitch of 619 µrad down, we only missed our target by 1.49 mrad, which puts us within the 2.0 mrad differential pitch spec
    • Differential Spec: 2.0 mrad
  • Roll
    • PUM: ±0.05 mm CW
    • ITMx: ±0.25 mm CW
    • Differential: ±0.2 mm CW
  • Center of Mass Separation
    • Left: 601.3 mm
    • Right: 601.7 mm
    • Target (for both sides): 602.0 mm