[Jeff B, Stuart A] After a few power glitches etc, the triple test-stand was in need of some attention prior to hooking-up OMC suspensions for Phase 1b testing. Logging in remotely, I noticed that the incorrect model was running, x1sushxts27 instead of x1sushxts05, so I re-started the *05 model. I then BURT restored the safe snapshot (/opt/rtcds3/tst/x1/userapps/release/sus/x1/burtfiles/20130911_x1sushxts05_OMC.snap) I had already previously taken of the OMC configuration (see LHO aLOG entry 7728 for further details). The triple test-stand should now be ready to accept OMCS for testing.
Before taking the beam jitter measurements last Friday (aLOG post coming soon), I adjusted the input filter offsets for each individual MC2trans and IM4trans DCQPD quadrant such that even with the shutter closed they always read a (small) positive number of counts. This should ensure that the SUM channels never swing through zero. The purpose of this was to avoid the "divide by zero" errors in the pitch and yaw channels, which are normalized by the SUM values.
After adjusting these offsets, the DCQPD SUM dark levels were:
MC2trans DC SUM dark level: ~1.5 counts
IM4trans DC SUM dark level: ~1.3 counts
This is a write up of various measurements of the MC2 violin modes done from 7/29 to 9/12.
Much of the procedure was based on Keiko's LLO alog 5097. All excitations were done at H1:SUS-MC2_M2_DRIVEALIGN_L2L_EXC, and readback at H1:IMC-F_OUT (later H1:IMC-F_OUT_DQ).
For initial mode ID (7/29) I used awggui for excitation, with amplitude of 1000000, and a cheby1("BandPass",1,1,620,680) filter, and searched with DTT in the same range (620-680 Hz). I tried measurement bandwidths of 0.01 and 0.001 Hz but 0.01 Hz turned out to be plenty. For the higher harmonics I scaled the range by 2, 3 and 4. With this approach I was easily able to find all n=1 and n=2 modes, and 3 of n=3, but no n=4.
Later (9/5) I went back and did narrow band searches where I had previously found peaks, and where I suspected n=3 and n=4 peaks. I used DTT for excitation, with Gaussian noise of amplitude 5e6 over a range of 1 Hz for n=1, 2 Hz for n=2, etc, with filters like cheby1(BandPass",1,1,644,645). I labelled the four wires "a", "b", "c" and "d" in increasing order of mode frequency. Because the main focus was to find the n=4 peaks, I did 1a, 1b, 1c, 1d, 2a, and then skipped straight to the ones I hadn't found before:3a, 4a, 4b, 4c, and 4d, and this time I found them easily.
On 9/6 I started trying ringdown measurements. I adapted the DTT files from 9/5 as monitors, switching off the excitations (e.g., . To excite the modes I reverted to awggui, with a sine excitation with amplitude 30000 at the previously identified frequencies. I then did a parallel set of Triggered Time Response measurements, with 1 average of 500 s.
On 9/8 I went back and dide mode ID for the ones I skipped on 9/5, 2b, 2c, 2d, 3b, 3c, 3d.
On 9/12 I did ringdown measurements for all 16 modes, except that to work with Keiko's analysis script I used 10 averages of 1/10 the length. Since the Q's for the higher harmonics were expected to be similar but the frequencies were higher, I used 10*24 s for n=2, 10*16 s for n=3 and 10*12 s for n=4. This was nearly a disaster because I didn't export as I went and DTT has a bug that it doesn't save the separate segments in the .xml file for a triggered time response with averaging. Fortunately the start times were store in the files so I was able to rerun them retrospectively and export them as .txt files.
I adapted Keiko's scripts to process all the files as a batch and save the plots as a PDF (see first attachment) in the manner of the many other SUS scripts such as plotquad_dtttfs.m. The adapted scripts are in ^/trunk/Common/MatlabTools/ViolinModes and the primary script is plot_vm.m. It writes a summary PDF to ^/trunk/HSTS/H1/MC2/SAGM2/Results (or the like, depending on the settings at the start of the script). It also generates a summary table like the following (except that the notes in parentheses have been added by hand):
1a: f = 644.641 Hz, Q = 227793.765
1b: f = 648.734 Hz, Q = 223441.7436
1c: f = 654.758 Hz, Q = 225393.2642
1d: f = 661.492 Hz, Q = 220474.6201
2a: f = 1289.38 Hz, Q = 158588.9576
2b: f = 1297.58 Hz, Q = 160877.9141
2c: f = 1309.7 Hz, Q = 145061.1334
2d: f = 1323.02 Hz, Q = 155196.8117
3a: f = 1934.4 Hz, Q = 169442.0117 (poor S/N)
3b: f = 1946.62 Hz, Q = 142432.5364
3c: f = 1964.75 Hz, Q = 131593.7049
3d: f = 1984.85 Hz, Q = 131407.1051
4a: f = 2579.77 Hz, Q = 1454265.3967 (very poor S/N)
4b: f = 2596.1 Hz, Q = 2302393.4224 (very poor S/N)
4c: f = 2620.29 Hz, Q = 255698.7597 (poor S/N)
4d: f = 2647.04 Hz, Q = 161984.0682 (poor S/N)
All the n=1 and n=2 ringdowns had excellent S/N, as did 3 of the n=3. The n=4 ringdowns ranged from poor to very poor. I tried running the data analysis from T1200418 on the data through n=3 (see second attachment), but the fit is poor and results in unphysical negative values of structural phi.
When the modecleaner is next available, I will redo the n=3 and n=4 modes, experimenting with different actuation frequencies to get higher levels of excitation of the modes.
h1fw0 is regularly crashing, log files suggest the SATABOY file system is sluggish. Dan is coming up to the MSR to investigate.
Providing h1nds1 is the default NDS, most users of the DAQ wont notice this problem.
Vern, Cyrus, Jim and Dave.
The problem: we are running PEM front ends in the mid stations, but dont want to install a complete timing sytem (fanout and IRIG-B units) to service just one front end computer per building.
One possible solution is to use the atomic clock/MCA distribution system to provide a local IRIG-B signal at the mid stations. To test this, we took a Time Code Translator (TCT) which used to be installed at the mid station and fired it up in the MSR. It was connected to the MCA via a short LC-LC single mode fiber optics cable. Its rear IRIG-B (the square wave version) was viewed on a scope by Vern, and then plugged into the front end computer h1susauxb123. When restarted, this machine's GPS time was off by 1year and 1day. Cyrus tracked this to a setting on the TCT card which was not outputting the year info in the IRIG-B code, verified by the scope. This was fixed and when h1susauxb123 was again rebooted it was 16 seconds off. This is the accumulated leap seconds between UTC and GPS time.
Thinking about it, we realized we should have expected this. The LIGO Timing System is GPS based. It is actually GPS time which is provided by the IRIG-B unit. Whereas all commercial timing systems, including the Symmetricom NTP server and the Timing Solutions MCA, use UTC time encoded into IRIG-B. Therefore the independent Symmetricom/Datum/TimingSolutions will differ from GPS by the leap seconds. We certainly dont wish to maintain a leap seconds lookup table, so we are abandoning the TCT as an IRIG-B source for the mid stations. This closes out WP#4175.
To proceed onwards, Rolf has suggested an RCG code change whereby a front end which is missing an IRIG-B card will use its system time instead.
When a solution has been found, the timing units in MX can be removed and MY PEM can be constructed.
Position loops were installed on HAM3-HEPI last week. UUG is 2Hz.
I installed the same controllers on HAM2 HEPI this morning. The phase margin is really high (above 60 degrees on every degree of freedom), so we can certainly raise the UUG and thus, push the performance.
Performance spectra were collected. They are attached to this post.
Controllers and simulations are also attached.
Spectra saved under the svn at:
SeiSVN/seismic/HEPI/H1/HAM2/Misc/DTT_Plots/
- 2013_10_08_HAM2_HEPI_Pos_Loops_Perf.xml
- 2013_10_08_HAM2_HEPI_Pos_Loops_Perf.pdf
revision# 7617
The ITMx Quad has been fully locked down, protective lens caps installed on the optics, and bagged in C3. It is now ready for cartridge flight/install into BSC3.
Today, Travis and I loaded the EMT07 PUM into the custom optics air bake oven for it's 34 degC 12 hour cure cycle. It had it's prisms and discs glued on ~1-2 weeks ago, so this cure cycle is a little late due to manpower constraints.
Alexa and Sheila have turned on the laser in the HAM 6 bay.
The ref cav power has been dropping over the last two weeks. I lowered the resonant threshold so that we could lock the MC for work in HAM1, but we probaby need to look at the alignment.
Attached are plots of dust counts requested from 4 PM October 6 to 4 PM October 7.
Will finish tomorrow
Currently RM1 and RM2 are still in acquire mode. Attached is a coil output signal plot in cts, with a calibration that mimics the the anti-dewhitening filter is applied. The RMS will still be <200cts, and that's with the purge air running. This will require putting jumpers in the RM1 and RM2 coil drivers (4 each). At the same time the OM1,OM2 and OM3 coild drivers can also be jumper-ed for RUN mode and to permanently enable them (2x4 jumpers per coil driver).
930 SnoValley to EX 930 Sheila transistioning LVEA to laser safe 1000 Thomas working on ACB and MCB in LVEA 1038 Apollo begins removing extra piping along X-manifold (WP 4171) 1100 Jeff K restarting H1ASC 1250 Filiberto and Aaron to EX to work on cabling 1300 Cheryl to EX to make everything peachy in the TMS lab 1440 Rich and Alexa to EX to inspect cabling 1515 T-Vo and P-Tho restarting IO/PSL environmental chassis
on Friday Kiwamu and stefan moved the tip tilts according to the calculation. Yesterday Stefan and I both remeasured these positions, what we got is that the distance from RM1 to RM2 is sqrt((17*2)^2+1^2)=34 inches. The distance from RM2 to M5 is sqrt((18*2)^2+6.5^2)=36.6 inches. We moved the BS for the refl LSC detector to 31 inches from M5.
When we first started making beam measurements we got some not repeatable results, because the beam was moving alot. With no one in the chamber or on the platform and the tip tilts damped we got repeatable results. With the mode master placed 22 inches from the beam splitter in this position we measured the beam using mode master ( we got a maximum gamma error). The result is in the attached photo P1000690.png. We also made measurements with the mode master 31 and 3/4 inches from the BS, this measurement is also attached as P1000689.png
This afternoon we also used the nanoscan to measure the beam waist (with the BS in the same position). These are saved on the Nanoscan computer and I will add them to this post in the morning.
I have also attached some plots to help compare our measurements to each other and to Paul's calculated beam (and the script). Summary_Hor and Summary_Vert show Paul's calculation, our measurement before the TTs, and our two measurements after the TTs after we moved them to their current locations, with all of these beams propagated through the refl WFS path in HAM1. The magenta and green curves are the two mode master measurements made about a foot apart after the tip tilts, the green line is almost not visible because these two measurements agreed with each other. The plot shows each of these four beams propagated back to before any of the tip tilts, and compared to the nanoscan measurement we made before the tip tilts (the first tip tilt has been deleted to allow the comparison). The last plot shows all four of these beams compared to the nano scan measurements we made after the tip tilts (with L101 deleted). The nanoscan results are intermediate between Paul's calculation and our mode master measurement before the tip tilts.
The gouy phase separations based on each of these beams is:
| TT seperation hor | TT seperation vert | WFS seperation hor | WFS seperation vert | |
| Pauls calc | 85 | 84 | 89 | 90 |
| MM before TTs | 87 | 79 | 75 | 77 |
| MM after TTs1 | 57 | 43 | 88 | 81 |
| MM after TT2 | 57 | 42 | 88 | 81 |
So, although the agreement between our measurements (and calculation) aren't great, at least it seems clear that the gouy phase separations were improved by moving the tip tilts.
Open light values have been measured on RM1 and RM2
| OPTIC | OSEM | Open Ligh Value (cts) | Gain | Offset | Serial Number |
| RM1 | M1UL | -32768 | -0.916 | 16384 | 1105240 a |
| M1LL | -32768 | -0.916 | 16384 | 1105240 b | |
| M1UR | -22268 | -1.347 | 11134 | 1105240 c | |
| M1LR | -29171 | -1.028 | 14586 | 1105240 d | |
| RM2 | M1UL | -26656 | -1.125 | 13328 | 1105238 a |
| M1LL | -32768 | -0.916 | 16384 | 1105238 b | |
| M1UR | -32708 | -0.917 | 16354 | 1105238 c | |
| M1LR | -32768 | -0.916 | 16384 | 1105238 d |
damping gains have been set to
-30 in long
-0.02 in Pitch
-0.02 in Yaw
the safe snapshot has been commited to the svn
I used the script "[OSEM2EUL, EUL2OSEM, SENSALIGN, DRIVEALIGN] = make_sushtts_projections" in /HTTS/Common/MatlabTools/ to generate the matrices for RM1 and RM2
They have been automatically filled with fill_matrix_values.m
The values are
OSEM2EUL:
0.2500 0.2500 0.2500 0.2500
10.3681 -10.3681 10.3681 -10.3681
-10.3681 -10.3681 10.3681 10.3681
EUL2OSEM matrix :
0.2500 10.3681 -10.3681
0.2500 -10.3681 -10.3681
0.2500 10.3681 10.3681
0.2500 -10.3681 10.3681
Safe snapshot has been saved for both RM1 and RM2
Joe, Pablo, Cheryl, Sheila
We measured the beams coming out of the HAM2 viewport, with the nanoscan head 2 feet 6 inches from the viewport. (in the first version of this alog I wrote the wrong distance)
For the brighter beam we measured beam diameters: 2549 um horizontal, 2632 um vertical. after rorating the scan head 90 degrees we measured 2650um vertical, 2561 um horizontal.
For the dim beam we measured 7160um vertical, 5972 um horizontal after rotating by 90 degrees we got 6077um horizontal 5865um vertical.
We also measured the beam in HAM1, in the same location as wensday (we move the BS for the RF PD to 16 inches after M2, we put the nanoscan 40 inches after the BS, there we got 4025 um vertical, 4126 horizontal. After rotating the head 90 degrees we got 4045 um vertical 4020um horizontal.
Hi Sheila, Can you be a little more specific which beam you measure where? The 4000um diameters sound good but I don't know what you mean with the brighter beam and what with the dimer beam.
Hi Guido-
I believe the brighter beam is the reflection off of PRM, although I haven't looked at the layout to double check. As you face the viewport, near the veiwport the dim beam is on the right while the bright beam is on the left. If you move to about 3 feet away from the viewport they cross, and the dim one is on the left if you are facing the veiwport. Pablo watched the spots on the wall as I moved the PRM alignment, and the bright one moved.
I should have been more clear, the first two measurements are of beam sizes for SM2 trans, coming out of the HAM2 veiwport where IOT2R would normally be.
Looking at the table layout for HAM2 (D0901083), I would expect the SM1 forward trans beam (from PMMT2 a.k.a. IM3) to be the dimmer beam, since this beam is split twice before making its way to the viewport. The SM1 return trans beam is not split at all before getting to the viewport. (I think you already came to this conclusion).
From the model, using rather rough estimates for the distances after IM4 transmission (+/- 2 inches), we expect that the SM1 forward trans beam (dimmer beam) where you measured it should have diameters of 6.20mm horizontal and vertical. The SM1 return trans beam (brighter beam) where you measured it should have diameters of 2.68mm horizonal and 2.67mm vertical. Not so far away from what you measured...
I'll try to get more precise numbers for the after-IM4-trans distance from Luke and update accordingly.
Apologies, everywhere I wrote "SM1" I meant "SM2". Or IM4, they are the same mirror.
SM2TransReturn1.png is a screen shot of the nanoscan program with the bright beam, when apperature 1 veritcal.
SM2TransForward1.png the dimmer beam, first with Apperature 1 horizontal then SM2TransForward2.png is with the apperature 1 vertical.
I haven't figured out how to export data from the software, and I can't uplaod .nsd files to the alog, but if anyone wants the data it is available at http://www.ligo-wa.caltech.edu/~sheila.dwyer/NanoScanData/IO/
You can use the free Nanosan software from Ophir/ Newport to open these files.
We saw that the Forward beam looks like it could possibly be clipping, Joe tried to move the location of the nanoscan head but the beam profile didn't get any better.
Just an update on the expected beam sizes with more accurate distances used in the model (good to 1/4" or so, inlcuding IM4 substrate and viewport substrate effect):
SM2trans forward 2wx=6.349mm, 2wy=6.3455mm
SM2trans return 2wx=2.6813mm, 2wy=2.6697mm
We placed Mode Master downstream of three-mirror Gouy phase matching telescope comprising two tip-tilts and one fixed mirror that is used for REFL WFS. (See the last picture for layout and distances.)
Note that the measured TT1-TT2 distance is about 1cm shorter than nominal described in Sam Waldman's document (http://dcc.ligo.org/T1000247), TT2-M5 distance is about 14mm longer than nominal, both of which should have been quite acceptable.
Anyway, we made this measurement and the beam was much smaller than what was expected. The first plot as well as the table below show the measured VS the expected mode profile coming out of HAM1 propagated through the telescope with the measured mirror distances.
| measured, x | measured, y | expected | |
| M^2 | 1.04 | 0.98 | 1-ish |
| Waist radius | 1.38 mm | 1.15 mm | 1.92mm |
| Waist position (away from MM head into HAM1) | 4.31 m | 4.35 m | 1.78 m |
| Mode overlap between measured and expected | 0.872 | 0.753 | 1 |
The total mode overlap between the actual beam and what is expected is somewhere between 0.75 and 0.87 (sort of tedious to do the real calculation so I leave it).
The 2nd plot shows that IF the incoming beam from HAM1 is as expected, in order to explain the measured mode the TT1-TT2 distance labeled as delta1 should be shorter by 4.5cm than was measured for X, or by 5.7cm for Y. This is a huge number, there's no way my distance measurement was that much off.
The 3rd plot shows the Gouy shift between TTs (i.e. actuation orthogonality) and WFSs for the WFS sled (i.e. sensing), and it seems like both are quite poor for the measured mode, 26deg for actuation and 35 for sensing are sad though not a complete disaster.
Anyway, since it's hard to imagine that the ROC of TT1 (+1.7m), TT2 (-0.6m) and M5 (+1.7m) are grossly wrong, and since it's hard to imagine that the distance measurement has a 5 to 6cm error, this should mean either (or some) of the followings:
The third one doesn't sound likely, but neither Sam nor I have thought about this.
One quick thing to do is to measure the beam before it gets to the telescope by inserting M6 upstream of the TT1 to direct the beam to the Mode Master.
Yes, please, measure the beam before the TTs.
The original calculations were done by assuming that beam reaching HAM1 was perfectly matched to PRM.
I don't think we have reasons to believe that's true..
The "nominal" q of the beam right before the first tip-tilt RM1 is:
% REFL in-vacuum path beam propagation, HAM1 drawing v10
% https://dcc.ligo.org/LIGO-D1000313-v10
% LisaBar, August 14, 2013
q_in = 1.03+13.1i; % Beam on HAM1 calculated from CalculatePRM.m
% Lisa: we don't have a measurement yet which confirms
% this number!
I don't think the table in T1000247 is correct. The beam from PMMT2 goes through the Faraday, hits PMMT1 and is then send to HAM1. This is a) longer than 2.5m and b) adds PMMT1's curvature to it. Did you include this?
Yes, PMMT1 is included, it is just a typo in the note (there are two PMMT2!). Anyway, let's redo the calculations with the as-built parameters, and cross check with the measurements before the REFL telescope.
Kiwamu and Pablo and I measured the mode before the TTs by moving the BS for the RF detector to 16 inches after M2, with the front of the mode master 40 inches from the BS we measured:
| x | y | r | |
| M^2 | 0.97 | 1.03 | 1.00 |
| 2Wo (mm) | 3.588 | 3.532 | 3.567 |
| Z0 (m) | -2.387 | -3.578 | -3.026 |
The overlap between the mode measured before the Tip tilts and the mode measured after is 93% for X, 87% for Y. I used Lisa's alm mode model attached to D1000313, added Keita's measurements of the distances from RM1 to RM2 and M5, but didn't include the tilt of the optics. From this measurement before the tip tilts (projected through Keita's measurements of distances), the gouy phase separation is a little better than from Keita's measurement, WFS X=65 degrees, WFSY 60 degrees, TT x=56 TT y 50 degrees.
I checked to see how far wrong things in HAM2 would have to be in order to explain the beam waist sizes measured by Sheila before the tip-tilts.
I used the design parameters for IM2 and IM3 Rcs (except where varied), design parameters for HAM2 optics placement as found in E1200616 (except where varied), the measured value of PRM HR Rc of -10.9478m, and the design IMC parameters to get the starting beam parameter. The attached plots show the forward beam waist size (identical to the IMC waist size) and the return from PRM beam waist size, over variations in IM2 and IM3 Rc, and IM2->IM3 and IM4->PRM distance. At the design values (at the x-axis midpoint), the return x-waist size matches the forward waist size.
It looks like things in HAM2 would have to be further off from the design than is probably likely, in order to explain the measured beam waist size before the tip-tilts.
Propagating the IMC transmission beam through the "as-built" IMs, back from the PRM, off the FI rejected beam pick off mirror and onto HAM1 to the location where Sheila measured gives:
| axis | parameter | value |
| x | w0 | 2.123mm |
| y | w0 | 2.101mm |
| x | z | -2.469m |
| y | z | -2.150m |
| x | q | -2.469+13.310i |
| y | q | -2.150+13.035i |
| x | w | 2.159mm |
| y | w | 2.129mm |
| x | Rc | -74.21m |
| y | Rc | -81.16m |
I did not yet consider the calcite wedge polarizer effect on the beam parameter, and I didn't account for the thickness of the septum viewport.
The overlap of these beam parameters with Sheila's measured parameters are:
x overlap = 0.945
y overlap = 0.934
I'm including the Finesse kat file I used for the calculation, which has a list of all the parameters I used at the top. I also include that list here for convenience:
# H1_IMCtoPRC_matching.kat
# A file for checking the expected beam parameter in direct reflection from the PRM
# as a function of HAM2 optic RCs and placement positions
#
# Mirror curvature parameters taken from the nebula page, except IM2 and IM3 for
# which the design values were taken
#
# Distances taken from E1200616_v7 except where otherwise noted
#
# IMCC Curvature = 27275mm
# MC1->MC2 = 16240.6mm
# MC2->MC3 = 16240.6mm
# MC3->MC1 = 465mmm
# MC3 substrate path length = 84.5mm
# MC3-AR surface to IM1 = 428.2mm
# IM1->IM2 = 1293.8mm
# IM2 Rc = 12800mm
# IM2 AOI = 7deg
# IM2->IM3 = 1170.4mm
# IM3 Rc = -6240mm
# IM3 AOI = 7.1deg
# IM3->IM4 = 1174.5mm
# IM4->PRM-AR surface = 413.5mm
# PRM substrate path length = 73.7mm
# PRM Rc = 10947.8mm (from Rodica's measurement value)
# IM2->FIrejected pick off mirror = 1.012m (From Luke Williams)
# FI rejected pick off mirror->HAM1 mode master location =3.0175m (Estimated from
# Sheila's alog entry, HAM2 drawing, and 27.6" for HAM1 table edge to HAM2 table edge)
#####################################################################################
If anything, my measurement is a bit more suspicious than Sheila's, as mine is downstream of TTs in air and they are moving (mostly in PIT).
| axis | parameter | value |
| x | w0 | 2.121mm |
| y | w0 | 2.101mm |
| x | z | -2.583m |
| y | z | -2.154m |
| x | q | -2.583 + 13.29i |
| y | q | -2.154 + 13.04i |
| x | w | 2.161mm |
| y | w | 2.130mm |
| x | Rc | -70.95m |
| y | Rc | -81.05 |
Apparently I posted the last comment as Giacomo, sorry about that!
After discussing with Lisa about sign conventions for the beam waist position parameter, I realised that there are errors in some of the parameters I posted above. The mode master gives results as "z0" for waist position relative to the measurement position (z0-z), whereas Finesse gives results as "z" for the measurement position relative to the waist position (z-z0).
I had thought the convention was different, so I flipped my results to match the mode master convention. This was a mistake, because the conventions are the same, they just give different outputs. To get the q-parameter from the mode master results one should use the formula q = -z0 + i*zR. From the Finesse results one should use the formula q = z + i*zR.
This means that the z-values, Rc values and the real part of the q-parameters I posted should all have their signs flipped. Apologies for any confusion I caused here.