Attached are plots of dust counts > .3 microns and > .5 microns in particles per cubic foot from approximately 7 PM Nov. 13 to 7 PM Nov. 14. Also attached are plots of the modes to show when they were running/acquiring data.
[Paul, Giacomo]
After moving the HAUX on the HAM table, we observed an overall reduction in OSEMs open light values (OLV), with relative changes spanning from a few % increase to up to almost 30% decrease (see entry 4603). This triggered the need for some debigging, that is summarized here.
First of all, we discovered that there was an error in the out of vacuum connections: as a results, different electronics channels were reading different OSEMs, making the measurement not directly comparable with the ones taken chamber-side (channels and OSEMs have all different "gains", so when you change the pairing...).
Once proper connections were restored we took a series of OLV measurements in which each OSEM of a single HAUX was read by each of the same HAUX's electronic channels. Nothing magic in this particular combination, just a way of having an "OSEM vs channels" matrix to use for some statistics. We obtained a matrix like the following for each suspension:
| UL OSEM | LL OSEM | UR OSEM | LR OSEM | |
| UL CH | ||||
| LL CH | ||||
| UR CH | ||||
| LR CH |
As we wanted to calculate a sort of "gain" for each OSEM, independent from the channel it was connected to, we proceeded as follows:
- took the OLV reading of UL_OSEM with UL_CH, and divided it by the average OLV read by UL_CH on all 4 OSEMs (same row)
- repeated the above calculation using UL_OSEM and LL_CH, UR_CH and LR_CH
- averaged these 4 values (same columns) to obtain the "gain" for the UL_OSEM
We repeated the exact procedure for all 4 OSEMs of a single suspension.
We then calculated a "gain" for each channel as well, using the exact same procedere but swapping rows with columns. The following table reports the OSEMs and channels "gains" calculated this way (the full set of data is available in the attached pdf):
| UL OSEM | LL OSEM | UR OSEM | LR OSEM | UL CH | LL CH | UR CH | LR CH | ||
| IM1 | 1.078 | 1.003 | 0.826 | 1.093 | 1.013 | 0.989 | 0.989 | 1.014 | |
| IM2 | 0.957 | 1.083 | 0.959 | 1.001 | 0.993 | 0.969 | 0.996 | 1.042 | |
| IM3 | 1.010 | 1.031 | 1.055 | 0.904 | 1.006 | 1.006 | 1.009 | 0.979 | |
| IM4 | 0.981 | 1.056 | 1.006 | 0.958 | 0.999 | 1.003 | 1.000 | 0.999 |
From this table, the channels seems to have very consistent "gains" (to within a few percent). The OSEMs show somehow more variability, as probably expected. particularly "bad" is the spread in the IM1 OSEMs...
As of now, the OLVs are as follows (with the "Aug-12" column reporting the values read in august during chamber-side testing):
| Aug-12 | Nov-12 | Change | ||
| IM1 | UL | 29075 | 27637 | -4.94% |
| LL | 27325 | 25614 | -6.26% | |
| UR | 26650 | 22636 | -15.06% | |
| LR | 29380 | 28125 | -4.27% | |
| IM2 | UL | 25155 | 23480 | -6.66% |
| LL | 29040 | 27816 | -4.22% | |
| UR | 26220 | 25102 | -4.27% | |
| LR | 28165 | 25690 | -8.79% | |
| IM3 | UL | 25440 | 24174 | -4.98% |
| LL | 26175 | 24416 | -6.72% | |
| UR | 26940 | 25144 | -6.67% | |
| LR | 26870 | 22569 | -16.01% | |
| IM4 | UL | 27790 | 25763 | -7.29% |
| LL | 29485 | 27886 | -5.42% | |
| UR | 27765 | 26393 | -4.94% | |
| LR | 26990 | 24959 | -7.52% |
Overall, there is a reduction in the 5-10% range, maybe due to increase resisitance in the new cables (as observed by Jeff in entry 4541); I don't have the data for the previous and current lenght and/or resistance of the cables, so I cannot confirm this.
In addition, IM1_UR and IM3_LR both show a reduction significantly bigger than the others (and the lowest absolute OLV values); also, for some reason they were reading even lower values (by about 10-15%) during the OLV matrix measurements than reported in the above table. A closer look at the recorded data (see attached png) shows that the change happened "during" the OLV matrix measurements: they were both reading about 20k before the OLV matrix measurement (also taken as the first value for the diagonal elements in that measurement); after being disconnected and reconnected several times, they were put back in their original positions and the readings increased to about 22k. The reason for this is unclear and may require further investigation.
We put ITMY in stops for the BSC2 cartridge install. We had a small issue with the bump stops between the CP and Test mass touching. We could not engage the face stops enough to hold the Teflon strips in place without the inside bump stops coming together. This occurred even when the barrel stops were in so tight the fluorel cap distorted. Therefore, we put Teflon transport caps on the CP face stops to cover the fluorel & fused silica tips. We put the Teflon strips, freed up from the CP face stops, between the bottom and both side bump stops between the CP and Test Masses. This will prevent glass to silica contact between the two masses during transport.
M. Barton, (J. Kissel) After the IAS/SUS team performed the final tweaks on H1SUSITMY, Mark (late last night!) remeasured each chain's TOP to TOP transfer functions. I attach the processed results. Baring noise arising from the in-air measurement, the mechanics look as good as they did at-vacuum in chamber, back in late May. In fact, it appears as though the alignment tweaks had little to know affect on the dynamics. Nice work all! I approve this QUAD for re-re-re-installation. (Don't worry, we've still got one more!) Interesting note (which is of no consequence but academic): this QUAD's Thin CP has been flipped, as per E1200943. Because we've assembled the TCP and ERM prisms to have a physical d4 of 0 mm (i.e. the prism break off is "exactly" ([+/- 1 [mm]) parallel to the optic's horizontal centerline, see T1200487), the dynamics should not have changed after flipping the TCP. However, as can be seen on pg 11 (R0 P to P) of allquads_121115_H1SUSITMY_Phase2a_ALL_ZOOMED_TFs.pdf, the (poorly modeled) P/L mode that was (measured) formerly at 0.74 Hz has moved up to 0.77 Hz. This mode is known to vary almost entirely due to changes in effective d4 ( = physical d4 + flexure correction). Assuming that the flexure correction (dependent on wire parameters and the tension on them from the optic mass) has not changed substantially (if at all), this implies that the physical d has changed with the flip, further implying that indeed the prisms are not bonded exactly along the optic centerline, but slightly below the center line (where "below" is defined by its original orientation, yet presumably still within the +/- 1 [mm] tolerance). That pitch is too sensitive, I tell you!
Note, this entry was posted simultaneously to Stuart's LHO aLOG 4694, and conclusion is the same between us.
[Joe Gleason, Chris Mueller, Alberto]
Yesterday we set the lenses on the ALS path on the PSL for mode matching to the ISCT1 table.
The ideal solution had a R=75mm lens and a R=257mm lens placed after the ALS Farady and separated by about 60cm (see LIGO-T1200471-v5 for details). Ideally this should have given us a waist of 68um radius between thw two lenses and a collimated beam of 3mm radius after the second lens. In actuality we measured a waist radius of 75um between the two lenses and becasue of this the second lens was placed at a different distance than designed. The beam coming out of the second lens had the desired size, however, due to the limited room on the table to do a long range beam scan, it was not easy to assess whether the beam was actually colimated, rather than diverging/diverging.
To be sure that the beam would not further diverge on its path to ISCT1, we set the second lens such that the beam would slightly converge over the 50cm or so that it covered over the table.
Mode matching on ISCT1 is designed to accept basically any relatively large beam (~>1mm) which arrives to the table almost collimated.
Attached are plots of dust counts > .3 microns and > .5 microns in particles per cubic foot from approximately 8 PM Nov. 12 to 8 PM Nov. 13. Also attached are plots of the modes to show when they were running/acquiring data.
Apollo, Jax, Thomas We completed the mechanical installation of the H1 ITMY Optical Lever, all of the components fit together as designed and the QPD currently reads back a signal into EPICS via these channels: H1:SUS-QUADTST_L3_OPLEV_PIT_OUTMON H1:SUS-QUADTST_L3_OPLEV_YAW_OUTMON (with H2 ITMY calibration parameters but these are no longer valid, will re-run test for new parameters once the cartridge is installed, the optic damped, and the chamber pumped down) These channels run into the test stand electronics in the LVEA at the moment but will soon be plugged into the permanent H1 OptLev channel in the H1 electronics bay, cables to be ran. Next we have to align, measure the waist and calibrate the signal.
[Joe G., Luke W., Cheryl V., Chris M.] Today we realigned the paths on the PSL table according to D0902114 with a few minor exceptions. The first exception is that the ALS beam path was moved two inches closer to the reference cavity to give some more room between the EOM and ALS Faraday. The second exception is that the optics which were moved to accommodate a possible periscope shift (alog 4662) were left in place even though the periscope was not shifted significantly. In addition the periscope was removed in order to fix the damping goo problem and put back in place. As of now the PSL is ready to inject a beam into the HAM 2 chamber for IMC flashing and alignment, which we expect to do tomorrow evening.
Purge air depressurized for ~15 minutes around 1615 hrs. local
BSC1 (Cartridge)
PSL/IO:
Septum viewport install HAM1/2
Viewports installed on top of HAM2 (center & north)
BSC2 Cartridge
Optical Lever (need for Robert's test)
DAQ Work
EX--clean & dome ready to be removed (maybe this afternoon)
Bubba said, "Clean stainless is being loaded into bin"
The LVEA is now in LASER HAZARD from about 10am - 1pm.
EY Dust Alarm (gerardo, worden, jodi)---note: Access System is not communicating to EY & EX
EY Chiller Yard alarm (due to Robert measurements)
Maintenance:
Pallets picked up near Staging (two loads)
Fire Dept on site.
A staging cleanroom was moved from the HAM1/2 area to BSC1/HAM3 area and then the central curtain partition was installed to create a garbing/staging cleanroom. Cris cleaned once the cleanroom was in place. Folks working in BSC1 (Lower Level) and in HAM3 should be able to garb and stage from this cleanroom.
Mark B. Commencing DTT TFs on ITMy. Will get as far as possible this afternoon and then continue this evening. Priority is M0 and R0 undamped, then M0 and R0 damped.
Mark B. Took M0 and R0 undamped TFs. All data is in the SVN as 2012-11-13_1515* (M0) and 2012-11-14_0015* (R0). plotquad_dtttfs.m has been run successfully. plotallquad_dttfs.m has been run and generated some comparison plots, but crashed while trying to merge them. All plots that were generated have been committed. Damping was restored at 02:49 am. No more tonight.
Thanks for Mark's late night efforts! I attach a merged plot of TFs showing:- Orange trace = L1 ITMX (thincp 2012−11−03), Phase 2a Black trace = L1 ITMY (thincp 2012−09−06),Phase 2a Magenta trace = H2 ITMY (thincp 2012−05−29), Phase 3b Dark cyan = H1 QUADTST (thincp 2012−11−14), Phase 2a
Added memory to the h1broadcast0 computer to avoid memory swapping. Fixed error in configuration file for monit, then started the daqd process to start broadcasting to the DMT computers.
As part of WP#3559 I changed the h1iopsusquadtst model to add a user switch to disable the watchdog. I also renamed B478 and ITMY to QUADTST in this model.
I then stopped the QUADTST user model and the phone promptly rang in the control room, it was Doug, who was using h1susquadtst to check out the ITMY suspension attached to the test stand, oops.
After a get together with Mike, Mark, Doug and Dave we figured out where the confusion was. We have stopped the IOP upgrade work for now so Doug can complete his work this afternoon.
We restarted H1DAQ to investigate "invisible" disconnected slow channels which the EDCU could not list by name. Restarting the DAQ caused h1sush2a (MC1,MC3) to go into a bad loop, constantly printing an error messge to the console. It was not ping'able or ssh'able. The error message was
nf_conntrack: table full, dropping packet
...ad infinitum
This is the network filter connection tracking system. We presume that the DAQ restart did something bad at the mx_stream FEDAQ network port on this machine. We had to power cycle to recover h1sush2a. Unfortunately h1sush2a also took down all MSR Dolphin machines, and all mx_streamers needed a restart.
At the end of this bootfest, the H1DAQ EDCU was now connected to all channels, so we could not reproduce the "invisible" disconnected channel issue. Patrick did some dataviewer tests for slow channels and did see some inconsistencies in minute trend data.
The IOChassis for H1 LSC and ASC were connected to the front ends and their timing was connected. We powered up h1asc0 first, and as expected it killed all user models on Dolphin connected computers and timing glitched their IOPs (a standard Dolphin glitch). Before we powered up h1lsc0 for the first time, we tried to disconnect its node from the dolphin fabric, but when it started we again dolphin glitched the MSR.
At the end of this bootfest, I copied the IOP models from L1 to run as h1ioplsc0 and h1iopasc0. There is an IRIG-B timing error on h1lsc0 which will look at later.
[Giacomo, Paul]
Yesterday we made some preliminary tests of the HAM AUX suspensions in the HAM2 chamber. We released the earthquake stops and checked that we could read sensible data from the OSEMs, and actuate the mirrors too - all of them looked fine, so the questionable cable connection mentioned in entry 4593 doesn't appear to be a problem. We then put the earthquake stops back in place and moved the HAM AUXs into the middle of the table to keep them out of the way for the moment (since we don't have dog-clamps to hold them down). We pulled the OSEMs out to make open light measurements (see comment from Giacomo in the near future). At first glance these look fine.
These are the measured open light values for the HAUX AOSEMs (10 seconds average):
| IM1 | IM2 | IM3 | IM4 | |
| UL | 23592 | 27614 | 25743 | 24264 |
| LL | 27862 | 25716 | 27778 | 24565 |
| UR | 25194 | 20038 | 26476 | 25148 |
| LR | 25681 | 28232 | 25051 | 19580 |
These are to be compared with the (approximate) ones recorded at the end of August (see post 4004). It appears that most of them have gon down in magnitude, with a few going below the 25000 threshold and one even below 20000. Not sure what the reason for and impact of this is...
Also, we measured the open light noise to look for excess noise of other strange thing, but we found none (good!). See attached image (input filters have for this measurements have offsets and gains based on the above open light values, as well as de-withening filters and nominal counts->um conversion factor).
It turns out that all the out of vacuum cables were swapped during this measurements, so these values cannot be directly compared to the ones measured in August, as different channels were reading different physical OSEMs (see entry 4697)