Jennie Siva Keita Mayank
Following our previous attempt here . We opened a new ISS PD array (S.N. 1202965).
This unit is in great condition. i.e.
1) No sign of contamination.
2) All the optics are intact (No chipping)
We tried interfacing the QPD-cable S1203257 with the QPD but it turned out that they are not compatible.
We will look for the updated version of the QPD cable.
More photos I took of the unboxed unit,
Keita holding part of QPD connector that connects to cable,
zoom in of part of prisms close to PD array to show they don't look damaged like the previous one we unboxed,
dcc and serial number of baseplate (this is different part for each observatory due to differing beam heights).
Keita explaining the QPD cable clamp to Shiva (right) and Mayank (left).
View of optics with periscope upper mirror on the left.
View of part of prisms close to periscope.
View of back of array and strain relief.
plus a picture of an packaged coptic that was sitting on top of this capsule while it was in the storage cupboard.
For future reference all the ISS arrays and there serial numbers are listed in the dcc entry for the assembly drawing LIGO-D1101059-v5.
[Matthew Mayank Siva Keita]
On Friday (2025-02-28) we moved the optics onto taller posts so that we did not have to pitch the beam up to much (in hind-sight, we probably would've been okay doing this) when we align the beam into the input port of the ISS array. We have not aligned the beam yet and most likely should re-profile it(may not need to) to ensure that the planned lens position is correct.
We also spent some time checking the electronics box for proper connections and polarity; then we tested the upper row of PDs (4 top ones) by plugging in each cathode/anode to the respective port. The output DSUB we used a breakout board and threw each channel onto an oscilloscope -- it seems that all four of the top row of PDs are functioning as anticipated.
Important Note:
Keita and I looked at the "blue glass" plates that serve as beam dumps, but just looking at the ISS array we do not know how to mount them properly. We think there may be some component missing that clamps them to the array. So we repackaged the blue-glass in its excessive lens paper.
02:44 UTC lockloss
I stopped at CHECK_ASC and reduced the gain of SRC2P from 60 to 40 and we were able to stay locked as the oscillation came and went, I ran the olg measurement. After the measurement I tried to reduce the gain down to 30 but I fat fingered it and made it bigger and we lost lock.
Ibrahim, Oli
Attached are the most recent BBSS Model Comparisons with most recent BBSS parameter settings (+/- from FDR):
d0=+3.0mm
l1 = -3.0mm
m2 = +0.3kg
FDR, d1 = +3.5mm (which is equivalent to BP = -4mm)
TITLE: 02/27 Day Shift: 1530-0030 UTC (0730-1630 PST), all times posted in UTC
STATE of H1: Observing at 145Mpc
INCOMING OPERATOR: Ryan C
SHIFT SUMMARY:
LOG:
15:30UTC Relocking
- ISC_LOCK in PREP_FOR_LOCKING; H1_MANAGER had gone into error less than a minute before (15:29 UTC)
- I ran an initial alignment
- We got those saturations again during OFFLOAD_DRMI_ASC (ndscope)
- Lockloss during LOWNOISE_ESD_ETMX due to earthquake coming in
- Holding in DOWN until earth settles
- Started another initial alignment
- Another set of saturations during OFFLOAD_DRMI_ASC - this time we lost lock from them (ndscope2)
- Saturations again but got through them this time
18:15 NOMINAL_LOW_NOISE
18:20 Observing
21:16 Commissioning to adjust A2L and PRCL FF
21:51 Back to Observing
22:48 Superevent S250226dl
00:13 Superevent S250227e
| Start Time | System | Name | Location | Lazer_Haz | Task | Time End |
|---|---|---|---|---|---|---|
| 16:13 | FAC | Nellie | OpticsLab | n | Tech clean | 16:28 |
| 16:19 | FAC | Tyler | MX, MY | n | 3IFO | 18:19 |
| 17:20 | FAC | Kim | MY, MX | n | Tech clean | 20:47 |
| 17:29 | VAC | Janos | EY | N | Parts Inventory | 17:44 |
| 20:40 | FIT | Matt | XARM | n | Runnin | 21:31 |
| 22:49 | OPT | Jennie, Keita, Siva, Mayank | OpticsLab | n | Unwrapping class A parts (Jennie out 00:10) | 00:36 |
TITLE: 02/27 Eve Shift: 0030-0600 UTC (1630-2200 PST), all times posted in UTC
STATE of H1: Observing at 143Mpc
OUTGOING OPERATOR: Oli
CURRENT ENVIRONMENT:
SEI_ENV state: CALM
Wind: 11mph Gusts, 6mph 3min avg
Primary useism: 0.03 μm/s
Secondary useism: 0.43 μm/s
QUICK SUMMARY:
This weird noise was first noticed in 82997 during OFFLOAD_DRMI_ASC. The first instance of this happening while relocking was three days ago 2025-02-23 03:59:19 UTC. It started showing up LSC-PR_GAIN and LSC-POPAIR_B_RF90_I_NORM and the lower stages of SR2 and SRM. Almost every relock since then has seen these glitches. They almost always start showing up in OFFLOAD_DRMI_ASC, but I've found a few instances(one, two, three) where the noise clearly starts during DRMI_LOCKED_CHECK_ASC, the state before OFFLOAD_DRMI_ASC. Next time we lose lock we will be sitting in DRMI_LOCKED_CHECK_ASC, so we will probably see this ringup then.
Based on Oli's range checks (alog83063) we decided to run the a2l script and a PRCLFF measurement after getting run coordinator approval. Below are the results of the a2l script, a few moving quite a bit. These have been added into lscparams.py and they are not in SDF. Jenne will post the PRCLFF results.
RESULTS
| Initial | Final | Diff | ||
| ETMX | P | 3.34 | 3.23 | -0.11 |
| Y | 4.9 | 4.91 | 0.01 | |
| ETMY | P | 5.56 | 5.49 | -0.07 |
| Y | 1.28 | 1.35 | 0.07 | |
| ITMX | P | -0.66 | -0.53 | 0.13 |
| Y | 2.97 | 3.21 | 0.24 | |
| ITMY | P | -0.06 | 0.06 | 0.12 |
| Y | -2.51 | -2.74 | -0.23 |
We reran the PRCLFF injection template to try and improve the range in the 20-50 Hz range. The template was at /opt/rtcds/userapps/release/lsc/h1/scripts/feedforward/PRCL_excitation_gain_adjust.xml.
The right two panels are the magnitude and phase of the PRCL to DARM transfer function.
We tried gains of 0.9 (starting value)
1.0
1.3
1.5
We found the TF was minimised in the 20-50Hz range using a gain of 1.3 in PRCLFF (pink trace).
The live trace was our last step (red) which was a gain of 1.5.
The coherences are shown in the bottom left corner (labels correct but colours don't correlate with the TF colours).
(NB: Ref 1 darm spectra in top left is old reference - ignore).
TJ added the value to lsc params and I accepted it in the OBSERVE snap file.
At the EY station the compressor is being replaced - after the one at EX is done. In this aLog, in the comments, the progress of this operation is tracked continuously, until the 1st startup by the supplier, Rogers Machinery. Another important consideration here, is that the purge line at EY needs to be replaced (based on an FTIR test - see DCC LIGO-E2300222-v2. As it can be seen, the level of contamination reaches even the 10 ug/cm2 value). This operation will be done after the April-May vent, so the EY station will be ready to be vented after O4. 02-25 (maintenance Tuesday): the old compressor was pulled out (it is temporarily stored in the EY receiving area). The beginning of the purge and TMDS lines with the associated brackets and unistruts were taken off. The new compressor unit and dryer skid were anchored in the mechanical room. Here it is important to mention that the orientation of the inlet was brought closer to the purge line inlet into the VEA, so the overall length of the associated circulation lines will be much shorter. Next is the electrical and pneumatic installation, which will be completed in the next 1-2 weeks.
The filter tree was installed and supports anchored to the slab. Ken also reports that electrical installation is complete. Connection to the purge air header is awaiting CF fittings from the supplier. However, startup testing can continue prior to header connection.
The 1st startup of the compressor was carried out by Rogers Machinery on March 18th, during maintenance hours.
Ran a low range check coherence check for a time during the early morning lock after we were thermalized (2025-02-26 10:05 UTC). Many of the channels came back clearly showing coupling between it and darm.
In almost all of them you can see that the coherence is higher in the last lock than the references, but the ones that stood out to me were:
Sheila, Matt, Camilla
We repeated the HD data set taken in 82202 as in 83022 we found we had pump depletion during these measurements so that the OPO trans powers were actually lower than reported. We had some issues at first which ended up being that with HAM7 ISI tripped the optics were moving around too much, plot of optics here. After we fixed that, we retook the data, so that our real data set is in 83039. Below documents the steps taken and initial data that we ended up not trusting.
Then started data taking:
Sheila and Matt measured the LO OLG as 2.5kHz 80deg phase margin is LO loop with gain -12 LO on LOCKED_HD with opo_grTrans_setpoint_uW = 80uW before we started.
One we got to the 110uW measurements, we felt the squeezing was unstable, moving independently of the SQZ angle, steps taken to help this:
Data saved to camilla.compton/Documents/sqz/templates/dtt/20250225_HD.xml
FC2 is misaligned during this dataset.
| Type | NLG | SQZ dB @ 1kHz | Angle | DTT Ref | Notes |
| Dark Noise | N/A | N/A | ref 0 | This and shot noise was noisy <300Hz at times when the OPO was locking or scanning, unsure why. | |
| Shot Noise | N/A | N/A | ref 10 |
Blocked SEED, LO only.
|
|
| SQZ | 12 | -7.1 | 157 | ref1 |
opo_grTrans_setpoint_uW = 80uW
Amplified Max = 0.0107mW
Amplified Min = 0.00037mW
UnAmp = 0.00095
Dark = 6.5e-5
NLG = 11.9(max/min) and12.0 (usual method)
|
| ASQZ | 12 | 14.7 | 243 | ref2 | |
| Mean SQZ | 12 | 11.8 | N/A | ref3 | |
| SQZ | 66 | -6.6 | 167 | ref4 |
opo_grTrans_setpoint_uW = 120uW
Amplified Max = 0.05347mW
Amplified Min = 0.0002933mW
UnAmp = 0.00095
Dark = 6.5e-5
NLG = 66 (max/min) and 60.3 (usual method)
|
| ASQZ | 66 | 22.6 | 229 | ref5 | |
| Mean SQZ | 66 | 19.7 | N/A | ref6 | |
| SQZ | 23 | -7.0 | 156 | ref7 |
opo_grTrans_setpoint_uW = 100uW
Amplified Max = 0.0210425mW
Amplified Min = 0.00034922mW
NLG = 23.0 (max/min) and 23.7 (usual method)
|
| ASQZ | 23 | 18.1 | 236 | ref8 | |
| Mean SQZ | 23 | 15 | N/A | ref9 | |
| SQZ | 36 | -7.2 | 165 | ref11 |
opo_grTrans_setpoint_uW = 110uW.
Amplified Max = 0.033029mW
Amplified Min = 0.00033752mW
NLG = 36.0 (max/min) and 37.2 (usual method)
OPO gain reduced -8 to -14. HAM7 WD untripped. Visibility again measured. |
| ASQZ | 36 | 19.9 | 230 | ref12 | Rechecked after visibility adjusted, no change |
| Mean SQZ | 36 | 16.9 | N/A | ref13 | |
| SQZ | 36 | -7.2 | 165 | ref14 | Remeasured after visibility again measured. |
Wed Feb 26 10:16:27 2025 INFO: Fill completed in 16min 24secs
This is the second run of new rate-of-change code. I have created a new set of MEDMs and a new Plot format to show the rate-of-change channels and the new trip level.
MEDM:
The main medm shows the six rate-of-change channels as bar charts (TC-A = blue, TC-B = purple). The trip level (60degc) is shown as a grey bar, the first ROC channel to exceed this stops the fill.
The related trend MEDM has been extended to show the six ROC channels, keeping with the same color scheme.
PLOT:
The plot has an additional panel showing the six ROC channels. The trip level is shown as a horizontal line. Both panels show start/stop times as vertical lines.
TITLE: 02/26 Day Shift: 1530-0030 UTC (0730-1630 PST), all times posted in UTC
STATE of H1: Aligning
OUTGOING OPERATOR: TJ
CURRENT ENVIRONMENT:
SEI_ENV state: CALM
Wind: 7mph Gusts, 3mph 3min avg
Primary useism: 0.02 μm/s
Secondary useism: 0.31 μm/s
QUICK SUMMARY:
Currently in PREP_FOR_LOCKING, looks like H1 MANAGER went into error at 15:29 UTC. I reloaded it and we've just started an initial alignment.
This is another instance of guardian issue #83. I think this is a race condition that we run into about two times a year, I need to find a way around it.
Summary
Initial call was because the IFO was waiting to turn on the OMC whitening since the violins are elevated, though I thought this was something we fixed. The whitening was safe to turn on minutes after I started looking at it, but the SQZer couldn't lock the FC. Once I fixed the FC alignment, I had to futz with the SQZ ASC, then all was well. There were some SDF diffs. Violin modes 5&6 of ITMY aren't damping well, not a new issue. Back to Observing at 0905UTC.
SQZ
H1:SQZ-FC_TRANS_C_OUTPUT wasn't getting flashes above 17 (the wiki says it should be above 100). I followed the instructions in the wiki and adjusted FC2 to get good flashes in the cavity and eventually got it to lock. It was left with a "FC ASC not on??" notification. Sure enough, H1:SQZ-FC_ASC_TRIGGER_INMON was at 1.5, still lower than the recently lower 2.0 trigger threshold(alog82853). I tried to touch up FC2 a bit more and got the trans output up to 112, but the trigger input was still at 1.5. I cycled SQZ_FC guardian node and was back in the same place. I then lowered the threshold to 1.5 and it turned on the ASC, temporarily raising the trans output and bringing the trigger input to 4.5. Range was at 146Mpc but we were still thermalizing so I didn't want to tune much else.
SDFs
There were two SDFs - H1:LSC-LOCKIN_1_OSC_TRAMP and H1:SYS-MOTION_C_PICO_I_CURRENT_ENABLE, both reverted.
Violins
While I was here I checked to make sure the violins were damping well. All seemed good except maybe ITMY 5&6. We've recently gone through some gain and phase changes to this mode (alog82977 alog82963). At this point mode 5 looked to be slowly decreasing, but mode 6 wasn't really moving. This was better than increasing though, so I left the settings as they were.
Violin mode ITMY mode 05/06 is damping fine using the nominal settings - FM6 + FM8 + FM10 Gain +0.01 (+30deg phase) during this (Wed morning) lock stretch.
TITLE: 02/26 Eve Shift: 0030-0600 UTC (1630-2200 PST), all times posted in UTC
STATE of H1: Lock Acquisition
INCOMING OPERATOR: TJ
SHIFT SUMMARY: The winds calmed down to the 20s and lower around 6 but locking issues persisted, the 2ndary microseism is still at or above the 90th percentile, so that wasn't helping us. There was a dolphin glitch but Dave helped me to get everything up and running. I just finished an IA as of the end of the shift.
LOG:
Sheila, Matt, Camilla
We aligned and took some SQZ HD data this morning but realized that the ZM optics were noisy with HAM7 tripped and the increased (60mph!! wind). After un-tripping HAM7, Matt and Sheila rechecked balancing and visibility, started at 98.3% but noisy, improved to 99.0%. Since then HAM7 tripped and was un-tripped again but assumed nothing would have changed.
NLG Measurements:
| opo_grTrans_ setpoint_uW | Amplified Max | Amplified Min | UnAmp | Dark | NLG (usual) |
| 120 | 0.108354 | 0.000573291 | 0.0018491 | 6.6e-5 | 60 |
| 110 | 0.0656167 | 0.000599855 | 36.8 | ||
| 100 | 0.0413918 | 0.00062871 | 23.2 | ||
| 80 | 0.0209032 | 0.000681607 | 11.7 | ||
| 60 | 0.0121804 | 0.000745147 | 0.0018428 | 6.2e-5 | 6.8 |
| 40 | 0.00710217 | 0.000852173 | 3.98 |
Data saved to camilla.compton/Documents/sqz/templates/dtt/20250225_GOOD_HD.xml
FC2 is misaligned during this dataset.
| Type | NLG | SQZ dB @ 1kHz | Angle | DTT Ref | Notes |
| Dark Noise | N/A | N/A | ref 0 | This and shot noise was noisy <300Hz at times when the OPO was locking or scanning, unsure why. | |
| Shot Noise | N/A | N/A | ref 10 |
Blocked SEED, LO only.
|
|
| SQZ | 60 | -7.4 | 167 | ref1 | opo_grTrans_setpoint_uW = 120uW, OPO gain -14 |
| ASQZ | 60 | 22.6 | 225 | ref2 | |
| MSQZ | 60 | 19.6 | N/A | ref3 | |
| SQZ | 37 | -7.3 | 164 | ref4 | opo_grTrans_setpoint_uW = 110uW |
| ASQZ | 37 | 20.2 | 229 | ref5 | |
| MSQZ | 37 | 17.2 | N/A | ref6 | |
| SQZ | 23 | -7.4 | 161 | ref7 | opo_grTrans_setpoint_uW = 100uW |
| ASQZ | 23 | 18.0 | 233 | ref8 | |
| MSQZ | 23 | 15.0 | N/A | ref9 | |
| SQZ | 12 | -7.3 | 151 | ref11 | opo_grTrans_setpoint_uW = 80uW, OPO gain -8 |
| ASQZ | 12 | 14.5 | 246 | ref12 | |
| MSQZ | 12 | 11.7 | N/A | ref13 | |
| SQZ | 7 | -7.0 | 143 | ref14 | opo_grTrans_setpoint_uW = 60uW |
| ASQZ | 7 | 11.7 | 253 | ref15 | |
| MSQZ | 7 | 8.7 | N/A | ref16 | |
| SQZ | 4 | -5.9 | 132 | ref17 | opo_grTrans_setpoint_uW = 40uW |
| ASQZ | 4 | 8.9 | (-)85 | ref18 | |
| MSQZ | 4 | 6.0 | N/A | ref19 |
I'm also attaching pictures of SR785 measurements of various loop gains.
| Measurement | Time | Notes |
| LO OLG | 11:28 | 2.5kHz UGF |
| OPO OLG | 12:06 PST | -8dB gain OPO loop gain |
| OPO OLG | 12:09 PST | -14 dB on OPO loop gain |
| CLF OLG | 12:10 PST | 13.kHz UGF |
This data set is nicely fit by standard squeezing equations, and suggests we have 6% unexplained losses and very little phase noise.
In the attached PDF I took the median of the ASD from 500-700 Hz, subtracted dark noise in quadrature from each ASD, then calculated the dB relative to shot noise for squeezing, anti-squeezing and mean squeeze. I used the squeezing and anti-squeezing and opo transmission numbers to fit for the OPO threshold (in units of transmitted power), total efficiency of squeezing, and phase noise. The attached plot shows the resulting model plotted against the data, including mean squeezing and NLG measurements that were not used for fitting. The NLG plot does suggest that we are slightly underestimating our NLG with our measurements.
This suggests that the OPO threshold is at 156uW transmitted power, and that the totall efficiency is 83%. This can be compared to 73365 and to the expected losses from the loss tracking sheet and sqz wiki. Expected losses:
| opo escape efficiency | 0.985 | |
| 3 SFI passes | (0.99)^3 = 0.97 | |
| B:BS1 | 0.9897 | HAM7 total = 0.946 |
| SQZT7 | 0.98 | |
| visibilty | 0.99^2 | |
| homodyne QE | 0.977 | |
| total expected in homodyne | 0.887 |
With the measured efficiency of 0.833, this means we have 6% unexplained losses in HAM7 or SQZT7.