FAMIS25985
ITMX ST1 was looking a bit higher, but perhaps not an issue yet. Everything else looks good.
Sheila, Jennie W
We have some a series of peaks inthe DARM spectrum from 25-34 Hz, some of which have broad shoulders, which have been with us for a long time but are now more prominent due to improvements in low frequency sensitivity.
I took a 2mHz resolution spectrum from a time with good range last night (165 Mpc). I tried searching all of these frequencies on the alog but found none of them.
frequency (Hz) | |
25.223 | |
26.475 | maybe BS roll mode (0.3 Hz off) Jennie found HAM6 channels in Gabriele's BRUCO from last night around 26.5 Hz. This frequency shifted since March 31st when it was at 26.512 Hz. Jim found these peaks in HAM6 GS13s and HPI L4Cs, and sees them when the IFO is locked and unlocked. He also sees that the frequency of this line shifted slightly since March 31st in the seismic sensors, as in DARM |
27.418 | HSTS bounce modes could be candidates (these look closest): SRM, SR2, PR2, MC1, MC2 Jim also sees this in HAM6 GS13s and L4Cs. |
27.504 | SR2/SRM bounce modes are closest |
27.707 | SR3 bounce? |
28.217 | PR3 bounce BRUCO shows coherence with PRCL, which |
29.000 | suspicously close to exactly 29 Hz, Jim also sees this in HAM6 GS13s and L4Cs. |
29.971 | |
32.301 | broad shoulder, dramatic change with PRCL offset |
33.334 | |
33.428 | |
33.639 | PR2 coherence |
33.639 | |
33.513 | |
35.712 | |
40.938 | HSTS and HLTS roll modes, PR2 and MC2 are closest. 40.938*0.6666 is 27.291 Hz, not lining up exactly with any of the above. (see similar discussion here: 21696) |
In the resonances wiki there is a ERM mode listed at 25.6281 (calculated not measured). There is a BS roll mode (R3) calculated as 25.9175 Hz with a link to 49643, which suggests it is 26.06 Hz
Jennie searched Gabriele's Bruco from last night, and found that there are PRCL coherences for some of these peaks. I retook the 2mHz spectra for a comparison of the times before and after the PRCL offset was first tuned, you can toggle between this and this screenshot to see the difference. Since this PRCL offset was first put in, we have changed the BS camera spot position which brought back the board PRCL coherence this offset was getting rid of (76814). The PRCL offset did seem to reduce the broad noise in DARM around these peaks, as you can see in the third attachment with with more binning.
Here's a similar effort from a few weeks ago: 76505
Interestingly, some of the peaks appear to have moved:
25.194 to 25.225 Hz
26.536 to 26.477 Hz
27.041 Hz is gone
29.500 Hz is gone
33.591 to 33.642 Hz
35.221 to 35.400 Hz
39.800 Hz is gone
Following up from my alog77064 saying that this needed to change, today I moved SRY triggering from POP_A_DC to AS_C_NSUM with new trigger threshold of upper=0.005 and lower=0.004. This worked well and gave us a quick, well aligned DRMI during main locking. The Acquire_SRY state used to lower the trigger thresholds after a few minutes for times that we had less power on POP_A, I've removed the lowering but added a notification to check the alignment or triggering.
I've updated ALIGN_IFO to reflect this and committed to the SVN. I anticipate we might need to fine tune these thresholds over the coming weeks.
The ETMY HEPI pump tripped this morning around 14:35 UTC (7:35am PST) potentially from the CS power glitch? I ended up going out to the Yend mech room and reset the VFH panel to reset the HEPI pump controller. The light was still on when I first checked the panel, I pressed the green ON button first which did not fix the issue, I then cycled it off and back on, pressing the red then green button, confirming that the light went off and on which Jim confirmed to have fixed it. Jim then brought back the pressure to 70PSI from the Beckoff_Pump_Controller medm.
Verified the HEPI Beckhoff electronics are powered by 24V from the VAC-R1 rack. Vacuum rack is on UPS.
Thu Apr 11 10:12:01 2024 INFO: Fill completed in 11min 57secs
Jordan confirmed a good fill curbside
BruCo scan from a good range time from last night: https://ldas-jobs.ligo-wa.caltech.edu/~gabriele.vajente/bruco_1396823046_GDS_CALIB_CLEAN/
Some highlights:
The last two plots show what improvement in sensitivity we should be able to get by reducing the above noise couplings, and which of the above noises contribute most at each frequency. Those plots are produced by offline subtraction with NonSENS
The MSR UPS went onto battery backup for 2 seconds at 07:35:02 due to an AC power glitch.
TITLE: 04/11 Day Shift: 15:00-23:00 UTC (08:00-16:00 PST), all times posted in UTC
STATE of H1: Aligning
OUTGOING OPERATOR: Ryan C
CURRENT ENVIRONMENT:
SEI_ENV state: CALM
Wind: 3mph Gusts, 2mph 5min avg
Primary useism: 0.02 μm/s
Secondary useism: 0.15 μm/s
QUICK SUMMARY: 17 hour lock ended about an hour ago. H1 is currently running an alignment, but stuck at green locking with no light on Y arm. HEPI and both ISI statges at EY have tripped during locking, not sure of the cause yet.
HEPI EY pump controller looks to have tripped off for some reason. I'll get ahold of Jim and continue investigating.
TITLE: 04/11 Eve Shift: 23:00-07:00 UTC (16:00-00:00 PST), all times posted in UTC
STATE of H1: Observing at 167Mpc
INCOMING OPERATOR: Ryan C
SHIFT SUMMARY: Very quiet shift (expect for the air handler) with H1 observing for all but 2 minutes.
H1 has now been locked for 10 hours.
LOG:
No log for this shift.
State of H1: Observing at 161Mpc
H1 has been locked for 6 hours, things are running smoothly.
Starting around 00:20 UTC (5:20pm PDT) this evening, the air has been much louder in the control room. Called Bubba to ask about it, and he says it's related to Eric's work today on the AHU-3 VAV (alog77092) and that the temporary ducting may have come undone. Other than it being louder in the control room, the duct should be fine and will be addressed in the morning.
I have not noticed a noticeable change in the IFO range since this noise increase, but I'll tag PEM and DetChar just in case.
This particular issue seems to have been caused by tubing coming loose from the pressure transducer for the supply fan on the hot deck side of AHU-3. The control system no longer had a reference for the fan static pressure and opened the vanes to maximum which sent the duct static pressure to over 3" W.C. The air handler ran this way all night until I was able to reconnect the tubing and reset the air handler.
TITLE: 04/10 Eve Shift: 23:00-07:00 UTC (16:00-00:00 PST), all times posted in UTC
STATE of H1: Observing at 157Mpc
OUTGOING OPERATOR: TJ
CURRENT ENVIRONMENT:
SEI_ENV state: CALM
Wind: 6mph Gusts, 3mph 5min avg
Primary useism: 0.02 μm/s
Secondary useism: 0.23 μm/s
QUICK SUMMARY: H1 has been locked and observing for 3 hours.
TITLE: 04/10 Day Shift: 15:00-23:00 UTC (08:00-16:00 PST), all times posted in UTC
STATE of H1: Observing at 160Mpc
INCOMING OPERATOR: Ryan S
SHIFT SUMMARY: One lockloss today with an easy relock. Some calibration work on the HVAC system that caused more fan noise in the OSB from 15:00-16:00UTCish. Rest of the day was quiet. We hit our current highest O4b range, 167.6Mpc!
LOG:
15:00ish-15:30ish Calibration of the CS HVAC system(77092) - fans much louder than normal - tagging Detchar
15:00UTC Detector relocking and at LOWNOISE_ASC
15:06 NOMINAL_LOW_NOISE
15:09 Observing
15:45ish HVAC calibration done but fans still slightly louder than normal while cooling the CS to correct temps - detchar
18:15-18:19UTC Richard and Eric going into LVEA to check for HVAC noise - detchar
19:46 Lockloss
20:41 NOMINAL_LOW_NOISE
20:56 Observing
Start Time | System | Name | Location | Lazer_Haz | Task | Time End |
---|---|---|---|---|---|---|
14:59 | FAC | Karen | MY | n | Tech cleaning | 16:00 |
16:02 | VAC | Janos+others | MY | n | Vacuum tour | 18:17 |
18:15 | FAC | Richard, Eric | LVEA | n | Quick hvac noise check | 18:19 |
20:17 | VAC | Gerardo, Jordan | FCES | n | Checking cable racks | 21:00 |
21:39 | VAC | Gerardo | OSB receiving door | n | Moving stuff | 22:09 |
late alog from work done yesterday:
I bumped up the ETMX L2 line to keep the uncertainty in the kappa PUM SUS line below a threshold of 0.01. Uncertainties above 0.01 clash with the hourly uncertainty services that run on the ldas clusters.
command and output:
gpstime;val=1.7 && caput H1:SUS-ETMX_L2_CAL_LINE_CLKGAIN $val && caput H1:SUS-ETMX_L2_CAL_LINE_SINGAIN $val && caput H1:SUS-ETMX_L2_CAL_LINE_COSGAIN $val && caput H1:CAL-CS_TDEP_SUS_LINE2_COMPARISON_OSC_CLKGAIN $val && caput H1:CAL-CS_TDEP_SUS_LINE2_COMPARISON_OSC_SINGAIN $val && caput H1:CAL-CS_TDEP_SUS_LINE2_COMPARISON_OSC_COSGAIN $val
PDT: 2024-04-09 15:42:06.558045 PDT
UTC: 2024-04-09 22:42:06.558045 UTC
GPS: 1396737744.558045
Old : H1:SUS-ETMX_L2_CAL_LINE_CLKGAIN 1.5
New : H1:SUS-ETMX_L2_CAL_LINE_CLKGAIN 1.7
Old : H1:SUS-ETMX_L2_CAL_LINE_SINGAIN 1.5
New : H1:SUS-ETMX_L2_CAL_LINE_SINGAIN 1.7
Old : H1:SUS-ETMX_L2_CAL_LINE_COSGAIN 1.5
New : H1:SUS-ETMX_L2_CAL_LINE_COSGAIN 1.7
Old : H1:CAL-CS_TDEP_SUS_LINE2_COMPARISON_OSC_CLKGAIN 1.5
New : H1:CAL-CS_TDEP_SUS_LINE2_COMPARISON_OSC_CLKGAIN 1.7
Old : H1:CAL-CS_TDEP_SUS_LINE2_COMPARISON_OSC_SINGAIN 1.5
New : H1:CAL-CS_TDEP_SUS_LINE2_COMPARISON_OSC_SINGAIN 1.7
Old : H1:CAL-CS_TDEP_SUS_LINE2_COMPARISON_OSC_COSGAIN 1.5
New : H1:CAL-CS_TDEP_SUS_LINE2_COMPARISON_OSC_COSGAIN 1.7
The ETMX L2 cal line gains are set by ISC_LOCK during the TURN_ON_CALIBRATION_LINES state, so Louis's gain changes had been reverted when we relocked. I dropped H1 out of observing at 00:07 UTC to re-run the command from Louis's alog above, update the gain value in lscparams.py, load ISC_LOCK, and accept the associated SDF diffs (screenshot attached).
PDT: 2024-04-10 17:08:06.083606 PDT
UTC: 2024-04-11 00:08:06.083606 UTC
GPS: 1396829304.083606
Old : H1:SUS-ETMX_L2_CAL_LINE_CLKGAIN 1.5
New : H1:SUS-ETMX_L2_CAL_LINE_CLKGAIN 1.7
Old : H1:SUS-ETMX_L2_CAL_LINE_SINGAIN 1.5
New : H1:SUS-ETMX_L2_CAL_LINE_SINGAIN 1.7
Old : H1:SUS-ETMX_L2_CAL_LINE_COSGAIN 1.5
New : H1:SUS-ETMX_L2_CAL_LINE_COSGAIN 1.7
Old : H1:CAL-CS_TDEP_SUS_LINE2_COMPARISON_OSC_CLKGAIN 1.7
New : H1:CAL-CS_TDEP_SUS_LINE2_COMPARISON_OSC_CLKGAIN 1.7
Old : H1:CAL-CS_TDEP_SUS_LINE2_COMPARISON_OSC_SINGAIN 1.7
New : H1:CAL-CS_TDEP_SUS_LINE2_COMPARISON_OSC_SINGAIN 1.7
Old : H1:CAL-CS_TDEP_SUS_LINE2_COMPARISON_OSC_COSGAIN 1.7
New : H1:CAL-CS_TDEP_SUS_LINE2_COMPARISON_OSC_COSGAIN 1.7
WP 11805
ECR E2400083
Lengths for possible SPI Pick-off fiber. Part of ECR E2400083.
PSL Enclosure to PSL-R2 - 50ft
PSL-R2 to SUS-R2 - 100ft
SUS-R2 to Top of HAM3 (flange D7/D8) - 25ft
SUS-R2 to HAM3 (flange D5) - 20ft
J. Kissel [for J. Oberling] Jason also took the opportunity during his dust monitoring PSL incursion today to measure the distance between where the new fiber collimator would go on the PSL table to the place where it would exit at the point Fil calls the PSL enclosure. He says SPI Fiber Collimator to PSL Enclosure = 9ft.
J. Kissel [for F. Clara, J. Oberling] After talking with Fil I got some clarifications on how he defines/measures his numbers: - They *do* include any vertical traversing that the cable might need to go through, - Especially for rack-to-rack distances, always assumes that the cable will go to the bottom of the rack (typically 10 ft height from cable tray to rack bottom), - He adds two feet (on either end) such that we can neatly strain relieve and dress the cable. So -- the message -- Fil has already built in some contingency into the numbers above. (More to the point: we should NOT consider them "uncertain" and in doing so add an addition "couple of feet here" "couple of feet there" "just in case.") Thanks Fil! P.S. We also note that, at H1, the optical fibers exit the PSL at ground level on the +X wall of the enclosure between the enclosure and HAM1, underneath the light pipes. Then the immediately shoot up to the cable trays, then wrap around the enclosure, and then land in the ISC racks at PSL-R2. Hence the oddly long 50 ft. number for that journey. Jason also reports that he rounded up to the nearest foot for his measurement of the 9ft run from where the future fiber collimator will go to the PSL enclosure "feed through."
Upon discussion with the SPI team, we want to minimize the number of "patch panel" "fiber feedthrough" connections in order to minimize loss and polarization distortion. As such, we prefer to go directly from the "SPI pick-off in the PSL" fiber collimator directly to the Laser Prep Chassis in SUS-R2. That being said purchase all of the above fiber lengths, such that we can re-create a "fiber feedthrough patch panel full" system as contingency plan. So, for the baseline plan, we'll take the "original, now contingency plan" PSL-R2 to SUS-R2, 100 ft fiber run and use that to directly connect the "SPI pick-off in the PSL" fiber collimator directly to the Laser Prep Chassis in SUS-R2. I spoke with Fil and confirmed that 100 ft is plenty enough to make that run (from SPI pick-off in PSL to SUS-R2).