Wed Jan 15 10:06:29 2025 ALERT: Fill done (errors) in 6min 25sec
TCs started very high, around +40C, and therefore TCmins [-28C, -27C] did not exceed the trip temp (-60C). After the TCs had flatlined I manually cancelled the fill.
OAT (-1C, 31F) and foggy.
Diag Main has now got a PSL_Chiller message telling me to Check the PSL Chiller.
I ran the PSL Status script to hopefully get some insight.
Laser Status:
NPRO output power is 1.842W
AMP1 output power is 70.15W
AMP2 output power is 137.0W
NPRO watchdog is GREEN
AMP1 watchdog is GREEN
AMP2 watchdog is GREEN
PDWD watchdog is GREEN
PMC:
It has been locked 28 days, 23 hr 15 minutes
Reflected power = 25.55W
Transmitted power = 102.4W
PowerSum = 127.9W
FSS:
It has been locked for 0 days 3 hr and 32 min
TPD[V] = 0.781V
ISS:
The diffracted power is around 4.2%
Last saturation event was 0 days 3 hours and 32 minutes ago
Possible Issues:
PMC reflected power is high
Check chiller (probably low water)
PSL probably wants some water and someone to chill with.
The plots below seem like it might be a little lonely.
I just got the verbal alarms alert to Check the PSL Chiller.
I topped off the PSL chiller with ~100mL of water.
TITLE: 01/15 Day Shift: 1530-0030 UTC (0730-1630 PST), all times posted in UTC
STATE of H1: Observing at 155Mpc
OUTGOING OPERATOR: Ryan S
CURRENT ENVIRONMENT:
SEI_ENV state: CALM
Wind: 3mph Gusts, 1mph 3min avg
Primary useism: 0.03 μm/s
Secondary useism: 0.35 μm/s
QUICK SUMMARY:
When I walked in H1 had been locked and Observing for 33 minutes!
Unknown lockloss this morning at 13:17 UTC
H1 Manny relocked without assistance this morning and no one was even woken up.
PS: It's a little icey out in the parking lot so be prepared stepping out of your car.
ETMY mode1s damping wasn't going well this morning and it was slowly rising (it did this a few times over the previous weekend as well), I went from +60 of phase to +30 and flipped the sign of the gain +0.1 -> -0.1, it has been damping for the past hour and has damped past where it was turning around with the previous settings.
Camilla, TJ, Marc, Fil. WP#12281
We attached the AOM drive cable from the back of the D1300649 chassis to the lowest TEST point in the PEM feed through photo on the CO2X table. We used two barrel connectors (photos attached) to do this as it looks like there used to be an AOM driver on the table that the signal went into before going to the AOM.
We thought that we could use the digital filters in the h1tcscs model to create a loop with this output and feed to the Synrad UC-2000 PWM controller (needs 0-10VDC). The max of CTRL2 was capped at +/-2 (unsure why) and this was actually +/- 0.6V on the BNC via an oscilloscope. We'll need to increase this by ~x10 to get PWM to work. Reverted changed sdfs.
There was an unknown cable also labeled AOM drive coming into the table, not connected to anything photo.
Matt Todd and I checked that neither of these BNC barrels were grounded to the CO2X table.
TITLE: 01/15 Eve Shift: 0030-0600 UTC (1630-2200 PST), all times posted in UTC
STATE of H1: Observing at 152Mpc
INCOMING OPERATOR: Ryan S
SHIFT SUMMARY:
IFO is in NLN and OBSERVING as of 05:40 UTC (20 min lock!)
Overall, bad recovery from maintenance. Here’s what happened in 7 short stories:
Big thanks to Erik and Sheila who helped a lot with troubleshooting.
LOG:
None
The changes here 82263 were reverted.
H1OMC models were reverted to version 5.3.0 of the RCG that uses the linear ramp. The IFO was consistently losing lock after a filter ramp down.
WP12272 h1omc0 new RCG, quadratic filter ramping
Erik, Dave:
h1omc0 models were built against RCG 5.31 which introduces quadratic smoothing to ramped filter switching. All the models running on this frontend have the new rcg (h1iopomc0, h1omc, h1omcpi).
The overview was modified to show that h1omc0 has a different rgc than h1susex by colour coding the RCG: dark_blue=5.31(quadratic filter ramp and variable duotone frequence), light-blue-5.30 (LIGO DAC) and green = 5.14 (standard)
WP12274 h1guardian1 reboot
TJ, Erik:
TJ rebooted h1guardian1 to reload all the nodes. The hope is that this will eliminate the leap-second warnings we have been seeing on certain nodes.
Tue14Jan2025
LOC TIME HOSTNAME MODEL/REBOOT
08:06:21 h1omc0 h1iopomc0 <<< Install RCG5.31
08:06:35 h1omc0 h1omc
08:06:49 h1omc0 h1omcpi
17:30:10 h1omc0 h1iopomc0 <<< Revert back to RCG5.30
17:30:24 h1omc0 h1omc
17:30:38 h1omc0 h1omcpi
TJ, Camilla WP12277.
Started the work done in 81734 on the EX HWS at EY. Swapped the fiber collimator to a CFCS11-A adjustable SMA fiber collimator. Still need to swap to a 50um fiber, remove HWS-L3 and realign. SLED left off.
Repeated 82151, with H1:IOP-OAF_L0_MADC{2,3}_TP_CH{10-13} 65kHz channels on CO2Y. WP# 12261.
Plots attached of the DC and AC out channels. These signals are straight from the PD in counts, before the filtering to undo the D1201111 pre-amp listed in 81868. PWM is at 5kHz, as can be seen in the spectrum.
I misread the graph, for CW 100%
TITLE: 01/14 Day Shift: 1530-0030 UTC (0730-1630 PST), all times posted in UTC
STATE of H1: Lock Acquisition
INCOMING OPERATOR: Ibrahim
SHIFT SUMMARY: Currently relocking, we just lost lock at LOWNOISE_ESD_ETMX.
LOG:
| Start Time | System | Name | Location | Lazer_Haz | Task | Time End |
|---|---|---|---|---|---|---|
| 22:08 | OPS | LVEA | LVEA | N | LASER SAFE | 16:01 |
| 15:48 | FAC | Kim, Nelly | EX | n | Tech clean | 16:46 |
| 16:01 | FAC | Chris | XARM | n | Big Green versus tumbleweeds, finished at 23:00 | 17:02 |
| 16:16 | CAL | Sheila | CR | n | IM4 trans cal check | 17:01 |
| 16:17 | Camilla, Mitchell | LVEA (WB) | n | Looking for parts | 16:24 | |
| 16:31 | VAC | Janos, Jordan, Travis, JC | MX, EX | n | Air supply replacement | 22:36 |
| 16:39 | PSL | Mayank, Sivananda, Rick, Rahul, Keita | LVEA (H2 PSL+tour) | n | Grabbing parts (Keita out 18:21) | 18:48 |
| 16:40 | PCAL | Tony | PCAL Lab | y(local) | Preparing stuff to ship | 17:31 |
| 16:47 | FAC | Kim, Nelly | EY | n | Tech clean | 18:09 |
| 16:48 | EE | Fil | CER | n | Checking for OMC0 necessities | 17:05 |
| 16:52 | Christina | OSB Receiving | n | Forklifting stuff into the bins | 18:01 | |
| 17:04 | FAC | Chris, pest control | LVEA, EX, MX, EY, MY, FCES | n | Pest control | 20:11 |
| 17:05 | EE | Fil | EX, EY | YES, n | Checking all racks | 19:56 |
| 17:26 | FAC | Eric, contractor | LVEA | n | Patching wall holes | 19:28 |
| 17:36 | PCAL | Francisco | PCAL Lab | y(local) | Grabbing stuff for PCAL meas | 17:52 |
| 17:45 | PEM | RyanC | EX, FCES | n | Checking dust monitors | 18:31 |
| 17:49 | PCAL | Francisco | EX | YES | PCAL measurements | 20:02 |
| 18:09 | FAC | Kim, Nelly | LVEA | n | Tech clean | 19:08 |
| 18:11 | TCS | Camilla | LVEA | n | Adjusting CO2 power | 18:22 |
| 18:30 | SEI | Jim | CR | n | Testing filters on BSCs | 20:26 |
| 18:34 | IAS | RyanC | LVEA | n | Setting up FARO for next week | 18:47 |
| 18:37 | TCS | Camilla | LVEA | n | CO2 laser work | 19:10 |
| 18:52 | PCAL | Rick, Mayank, Sivananda | EX | YES | Tour | 20:01 |
| 19:11 | HWS | TJ, Camilla | EY | n | Adding new collimator to HWS | 20:20 |
| 19:41 | VAC | Janos, JC | EY, MY, LVEA | YES | Fitting new exhaust filters for roughing pipes | 23:53 |
| 20:02 | PCAL | Francisco | PCAL Lab | y(local) | Dropping stuff off | 20:32 |
| 20:07 | PCAL | Rick, Sivananda, Mayank | PCAL Lab | y(local) | tour | 20:59 |
| 20:37 | EE | Fil | LVEA | n | Rack checks | 22:29 |
| 20:47 | SEI | Jim | CR | n | Tests on ETMX | 22:27 |
| 21:06 | OPS | Camilla | LVEA | YES | Transitioning LVEA to HAZARD | 21:27 |
| 21:27 | TCS | Camilla, TJ | LVEA | Y | HWS table work | 22:19 |
| 22:19 | OPS | Camilla | LVEA | Y -> N | SAFE transition | 22:26 |
| 22:46 | TCS | Camilla | LVEA | N | TCSY adjustment | 22:53 |
| 23:10 | EE | Daniel | LVEA | N | HAM1 investigation | 23:32 |
During one of the large eqs over the weekend all of the BSC-ISI tripped. I checked one of the chambers and the trip was due to large low frequency drive railing the stage 2 horizontal actuators. The earthquake was big enough SEI_ENV went LARGE_EQ, well after the IFO lost lock, but also after the ISIs started tripping. In SEI_ENV I've reduced the threshold on the peak mon to 6000, about 1.5x the largest eq we've ever ridden out, down from 10000. I've also changed the stage 2 blends used in this state to some 1.5hz blends which roll-off the gs13s at low frequency much more aggressively. I don't think these changes would have been enough to prevent the ISI trips for this particular earthquake, the ISIs started tripping while peak mon was around 4000, so there must have been a large amount of ground motion that was out of band for that channel.
TITLE: 01/15 Eve Shift: 0030-0600 UTC (1630-2200 PST), all times posted in UTC
STATE of H1: Lock Acquisition
OUTGOING OPERATOR: Ryan C
CURRENT ENVIRONMENT:
SEI_ENV state: CALM
Wind: 2mph Gusts, 0mph 3min avg
Primary useism: 0.03 μm/s
Secondary useism: 0.39 μm/s
QUICK SUMMARY:
IFO is LOCKING at MOVE_SPOTS
The LN2 truck is yet to arrive and so it may cause a lockloss.
In 81638 we removed a DARM1 FM1 boost because of a glitch when ramping it off causing locklosses during ESD transitions in preparation to switch back to ETMX. Today the CDS team updated H1OMC0 models with an improved filter ramping: 82263. We hope this will allow us to keep the boost us which gives us more range against high microseism while relocking (it's always off by NLN).
Uncommented line 3058 from ISC_LOCK.py and reloaded: ezca.get_LIGOFilter('LSC-DARM1').turn_on('FM1'). If we have locklossses at ISC_LOCK state 557 or 558, at the operator can re-comment this line out. Tagging OpsInfo.
Sheila, Camilla, Erik
We lost lock 12s after this DARM1 FM1 filter was turned off, we're not sure if the filter changes are the cause. Are trying to relock again.
We think we were a little confused and have been turning on FM1 the whole time, as it was still turning on in PREP_DC_READOUT_TRANSITION. Unsure if it was just luck that the glitch disappeared when we made the change Dec 5th. Will look into more...
Re-commented out line 3058 today at 1:25 UTC after losing lock at the same state LOWNOISE_ESD_ETMX (558).
J. Kissel, T. Sanchez, E. Goetz, L. Dartez *EDIT* This limitation is only pulling out data with test points from the A0/B0 filter outputs due to them being recorded in single precision, not double precision. The actual data for all internal calculations is double precisions, and in fact for the final calibrated gravitational wave strain is both calculated in, and then stored in frames as, double precision. Back in July 2024 when I started to characterize the super-16 kHz-Nyquist frequency data off the OMC DCPDs with the live 524 kHz channels. See LHO:78516 for the whole story, but we got stalled when we ran into what we believed was some sort of single-precision, numerical precision noise, limited at the equivalent of 1e12 [A/rtHz] DCPD current or 1e-6 [V/rtHz] ADC voltage. In LHO:78559, we ruled out single-point precision calculation of the ASD when we ran the same DCPD signal through a special version of DTT which uses double-precision to calculate the pwelch algorithm. That version of the data proved that the high frequency limitation is still there, and NOT the precision of DTT. In that same data set, we also showed that if you ask DTT to remove the mean, i.e. large DC component of the signal, it also did NOT have any impact on this limit. And it's in removing the mean that we reveal / confirm that it is "single-point precision" limit in the test point readbacks. Check out 1st attachment which is the data from LHO:78559, but with no DTT calibration applied. That means the channels are calibrated into the units of whatever they are coming off of the front-end -- in this case milliamps, or [mA]. The DC value of the test point channel during the time of measurement was ~20 [mA]. The front-end computes all its filtering in double precision, but the readbacks of the products of those calculations are single-point precision. An IEEE 754 32-bit base-2 floating-point, single precision channel has 24 bits of significance (excluding the sign and exponent) to hold the entire frequency dependent content of the time-series that has a DC value of ~20 [mA]. The (front end filtering algorithm?) rounds the 20.43 DC component to the nearest 2^n value, i.e. 2^5 = 32 ["mA"]. Eq. 2.2 of Liquid instruments article on quantization noise suggests that the amplitude spectral density of 1 bit spread across the 0 to 2^18 Hz (f_Nyquist) frequency range over which we care is n_{ASD,RMS} = sqrt( DELTA^2 / (12 * f_Nyquist) ) = DELTA * sqrt( 1 / 12 * 1/f_Nyquist ) where - DELTA is the minimum step resolution (i.e. the peak value / number of significant bits), - the factor or 1/12 comes from the expectation value of the noise power derived from the integral of the product of instantaneous noise power and the probability that that power is distributed across one, specifically the least significant, bit (and we use the square root because we want the amplitude not the power) - the factor of 1/f_Nyquist comes from spreading out the (presumably frequency independent) power over the entire frequency range (and again, we use square root because we want the amplitude not the power) In line-by-line math, that's [[ 32 ["mA"_DC] peak value * (1/2^24) significant bits in single precision) ]] [[ 1 bit * (1/sqrt(12)) expectation value quantization noise amplitude spread across 1 bit * (1/sqrt(2^18 Hz)) quantization noise spread across 0 to Nyquist frequency range ]] = [ 32 / (2^24) ] * sqrt[ 1 / (12 * 2^18) ] = 1.07539868e-9 ["mA"/rtHz] This number *exactly* the high-frequency asymptote we see. BINGO! The next step was then to create pick-off paths of the ADC channels and high-pass -- i.e. remove the large DC component of the signal -- in the front-end. Importantly, this has to be the *first* filter, so the DC component is removed before any other calculation is done. We added the infrastructure and installed the filtering (LHO:78956, LHO:78975), but have not come back to the data until today. Today, we're finally looking at the front-end OMC DCPD data that's been high-pass filtered with a 5th order, 1 Hz high-pass, with 40 dB stop-band attenuation and a 1 dB ripple ellip("HighPass",5,1,40,1) The odd (5th) order means that DC component is completely suppressed, leaving only the remaining frequency-dependent accumulated RMS, which is 5.8901e-4 [mA_RMS] to upper limit of the dynamic range and define the precision limit. SIDE QUEST -- Fractional numbers are much less intuitive to "just round up to the nearest power of 2^n," so the equivalent of "converting 20.43 [mA] to 32 ["mA"]" for 5.8901e-4 is instead a process of, Converting 5.8901e-4 to floating point binary, # sign exponent fraction 0 01110100 00110100110110111110111 # round up the fraction part 0 01110100 01000000000000000000000 # convert back to decimal 0.00061035156 That takes the quantization limit down to = [ 6.1035156e-4 / (2^24) ] * sqrt[ 1 / (12 * 2^18) ] = 2.05e-14 ["mA"/rtHz] Take a look at 2nd attachment, which compares the normal nominal OMC DCPD data to this 1 Hz high passed data. One can see all of the AA filtered data all the way out to the 232 kHz Nyquist frequency, because that data only goes as low as 1e-13 [mA]. Finally in the last attachment, we re-cast this into ADC noise units, to show where the data against the trace we'd had as a bench mark before. Note *this* comparison is a false comparison because the digital AA filtering is applied after the ADC noise is added. So -- don't read anything into the fact that the resolved noise goes below the ADC noise -- it's just a guide to the eye and a bench mark to remind folks that the numerical precision limit is *not* ADC noise. In conclusion, if we want to investigate the OMC DCPD data above 7 kHz with test points, we need to make sure to use a version where the data is high-passed significantly. So, now we can actually begin doing that...
FAMIS26026
Last week's report - alog82184. All spectra look good to me and agree with last week's report.
FAMIS31405
I replaced both sock filters for fresh ones and inspected the radiator air filters.