Attached below are the first set of transfer function measurement results for the Tip Tilt suspension PM1 in HAM1 chamber. The purge air flow in chamber was reduced to enable good coherence value and minimum cross coupling for this in-air measurements, and the results are very much visible when compared to earlier data (not posted here) which was very noisy. The plots attached below are for Length, Pitch and Yaw degree of freedom. Yaw dof. still shows some cross coupling/unwanted peak at 1.28Hz approximately, however it should go away in vacuum.
The templates for this measurement are stored at the following location,
/ligo/svncommon/SusSVN/sus/trunk/HTTS/H1/PM1/SAGM1/Data
2025-05-14_2100_H1SUSPM1_M1_WhiteNoise_L_0p01to50Hz.xml
2025-05-14_2100_H1SUSPM1_M1_WhiteNoise_P_0p01to50Hz.xml
2025-05-14_2100_H1SUSPM1_M1_WhiteNoise_Y_0p01to50Hz.xml
Hence, I can confirm that the suspension and the electronics chain is healthy.
Chiller pump 1 at Mid X was overhauled; bearings and seal were replaced, stator was cleaned, and impeller was inspected. Current wear ring is in decent shape and was left in place. Pump was reinstalled this morning.
I branched and added some functionality to my misaligned GUI located on ASC IFO_ALIGN_COMPACTEST. I wrote another script to take a lock start and end and run the GUI every 5 minutes, take a screenshot then stitch them all together to make a gif of the lock so that we can see the changing alignments throughout lock stretches. It's a little slow, it was taking 20 minutes per hour till I realized I was redoing a large calculation each iteration unnecessarily, changing that brought it down to ~10 mins/hour of lock.
Here's an example gif of a short 2 hour lock starting at 03/30/25 18:13
I've added PM1 to HAM1 since posting this.
Wed May 14 10:11:12 2025 INFO: Fill completed in 11min 8secs
Gerardo confirmed a good fill curbside.
Dry air skid checks, water pump, kobelco, drying towers all nominal.
Dew point measurement at HAM1 -43.8 °C
.
TITLE: 05/14 Day Shift: 1430-2330 UTC (0730-1630 PST), all times posted in UTC
STATE of H1: Planned Engineering
OUTGOING OPERATOR: None
CURRENT ENVIRONMENT:
SEI_ENV state: MAINTENANCE
Wind: 15mph Gusts, 8mph 3min avg
Primary useism: 0.04 μm/s
Secondary useism: 0.16 μm/s
QUICK SUMMARY:
Expected work today:
CS:
HAM1 ISI Balancing & testing
HAM1 SUS testing.
END X: - Note Bee aware of Bees:
Wind fence work
END Y:
SEI BRS Pump down
Here we study the effect of thermal deformation on the SRCL detuning as we increase the input power. This work shows a more signifcant impact of the spherical aberrations and HOM thermal distortions on the SRCL detuning than the mode mismatch, or misalignment.
Building on the previous study of how mode mismatch affects SRCL detuning, I now turn to exploring the thermal effects on SRCL detuning. When laser light passes through optics, a small portion of it is absorbed, leading to some heating. This thermal absorption causes the optic to either physically deform or develop a refractive index gradient both of which can distort the wavefront. More background on this process is available.
A thermally distorted thin lens can be thought of as a combination of a quadratic lens term and a residual distortion that includes spherical aberrations and other higher-order modes. In FINESSE, we simulate this by generating an optical path difference (OPD) map based on the absorbed power. We then fit and subtract the quadratic component, which corresponds to the ideal thin lens (the effect explored in the earlier study). The remaining residuals represent the non-quadratic distortions we’re now interested in analyzing for their impact on SRCL detuning.
It’s important to note that in this model, the cavity geometry like, the radius of curvature remains unchanged. As a result, mode mismatch is not a factor in this part of the analysis.
These residual distortions have a stronger effect on detuning than the lensing term alone, as shown in the plots below. This model assumes 0.13 ppm absorption in the ITM coatings. The model fails to lock with higher absorption values. No thermal effects are applied to the ETMs in this analysis.
TITLE: 05/13 Day Shift: 1430-2330 UTC (0730-1630 PST), all times posted in UTC
STATE of H1: Planned Engineering
INCOMING OPERATOR: None
SHIFT SUMMARY:
GRB-Short E571183 @ 15:36:28
Test T571184 @ 16:00:09
EX N2 fill completed.
Water tank taken to EX to wate the sand.
SEI BRS upgrade work continues , possibly pump down tomorrow.
LVEA has been transitioned to LASER SAFE
LOG:
Start Time | System | Name | Location | Lazer_Haz | Task | Time End |
---|---|---|---|---|---|---|
15:07 | SAF | LVEA is Laser HAZARD | LVEA | YES | LVEA is Laser HAZARD | 19:20 |
14:46 | FAC | Kim & Nellie | LVEA | - | Technical Cleaning | 15:50 |
15:19 | VAC | Jordan | LVEA | - | Purge air checks | 15:35 |
15:20 | ISC | Camilla | LVEA HAM1 | yes | Beam alignment inside HAM1 | 19:10 |
15:23 | SEI | Corey, Mitchell, Randy, jim | EX | N | Building a wind fence (Mitchell out 17:29) | 19:14 |
15:49 | SEI | Micheal R. & Shoshana | EY | N | SEI BRS Work | 18:56 |
15:50 | FAC | Tyler | EX | n | Checking on Bees | 16:48 |
15:57 | SUS | Rahul | LVEA HAM1 | N | Installing a beam dump on PM1 | 16:32 |
16:09 | ISC | Betsy | LVEA HAM1 | Yes | Beam Alignment on HAM1 | 18:57 |
16:11 | PSL | Jason | LVEA | - | Walking the PSL Chiller lines | 16:22 |
16:22 | CDS | Dave | Remote | N | restarting EDC | 16:25 |
16:22 | CDS | Erik | Remote | N | Restarting EXT alerts | 16:37 |
16:37 | ISC | Ibrihim | LVEA HAM1 | Yes | Beam Alignment with HAM1 | 17:16 |
17:09 | FAC | Nellie & Kim | MY & MX | n | Tecnical cleaning | 18:26 |
17:30 | SEI | Chris | EX | N | Wind fence | 19:26 |
17:39 | ISC | Jenne | LVEA | YES | Going in to check on HAM1 peeps | 18:57 |
17:46 | FAC | Christina | Recycling | n | Fork lifting heavy materials into recycling | 18:34 |
18:33 | VAC | Jordan | LVEA | - | Setting up Vac equipment near +Y arm | 18:48 |
19:03 | FAC | Nellie | HAM Shaq | N | Technical Cleaning | 20:48 |
19:07 | Tour | Rick + 4 | LVEA | yes | Giving a tour of the LVEA to a small group | 20:20 |
19:10 | SAFETY | TJ, Ollie | LVEA | - | Transitrioning to LASER SAFE | 19:26 |
19:31 | Jog | Ryan C | Y-arm | N | Running Y arm | 20:19 |
19:38 | SEI | Micheal R & Shoshana | ey | N | SEI BRS balancing | 21:47 |
20:32 | ISI | Jim & TJ | LVEA HAM1 | N | Balancing ISI | 23:42 |
22:07 | SEI | Betsy | LVEA HAM1 | N | Inspecting clamps and misc | 22:25 |
22:31 | VAC | Janos | LVEA | N | Checking out general things ∠(‘-‘ ) | 22:42 |
23:45 | TCS | Camilla | Optics lab | N | Inspecting ZnSe optics | 01:45 |
Closes FAMIS37252, last checked in alog83817
The diode room DM was having some issues that I fixed the other week, lab2 is also a known issue. No other issues of note.
Betsy, Ibrahim
Today, Betsy and I replaced the non-plated BBSS bottom wire loop with the gold plated wire loop, which was done successfully and took roughly two hours. We've confirmed the BBSS is centered and aligned. We will monitor and inspect the BBSS in the next few days and coming week. Stay tuned for new transfer functions etc.
Gallery of the wire loop below!
Went in to monitor the BBSS alignment after pulling this new wire. All alignments were the same as yesterday. I have since covered the BBSS to take TFs, which have been saved as of writing this. Next, the plan is to compare the gold plated wire loop TFs with the non-plated wire loop TFs.
- Cable table bracket for the bosem cables have been added to RM1 suspension in HAM1 - see here.
- Beam dump for catching the transmitted light through PM1 optic has also been attached as shown in several pictures attached below. Camilla and I adjusted/fine-tuned (angle etc.) it as well.
Also, I thoroughly inspected PM1 (BOSEM flags and mirror holder) for any rubbing or touching, didn't find anything. I adjusted the BOSEM bracket for my own satisfaction, since one or two of the BOSEM flags were not exactly looking centered between the PD/LED, however the adjustment was minimal (the bracket could hardly move) and didn't create much of a difference. I will be posting the measurement results asap.
Betsy, Jenne, Camilla
Day 1: 84193, Day 2: 84228, Day 3: 84230 and 84239, Day 4: 84274, Day 5: 84292, Day 6: 84314, Day 7: 84334, Day 8: 84361
This morning we finished our last laser hazard tasks in HAM1 mainly at 20W in, did some final checks and went back to laser safe.
POP LSC Pico: This pico was too close ot the edge of its range so I turned the mirror mount so there was range and realigned onto the center of the POP LSC diode.
Final Checks: We checked ndscope for all LSC and ASC diodes while blocking and unblocking the beam to check there was signal on all. Yesterday Elenna and I removed all of the temporary metal name tags and this morning I checked that there were no thumb knobs left on any optics. As it doesn't make sense for PM1 to be misaligned, we suggest having misaligned offsets of (0,0) for this, although if we really wanted, a setting of (1500,1500) would ge the beam off the diodes.
Have lots of component photos that will be attached later.
[Shoshana, Michael] We've started preparing to install the BRS mass adjuster to the End-Y BRS. The plan is to follow the same procedure/same parts as the End-X install outlined in LIGO-T2400043 and SEI log 1886 (https://alog.ligo-la.caltech.edu/SEI/index.php?callRep=1886). We've taken apart the BRS and discovered that doesn't match the designs on the DCC. How the parts are arranged blocks access to where the pico-motor mount should attach, but we think we have a work around that should work. The electronics/wiring of the BRS is as expected and we've finished all the wiring for the pico-motor so that it should attach to the feed through. Assuming all goes well we plan on installing the mass adjuster parts and begin testin.
[Shoshana, Michael] We've managed to get all of the hardware/parts installed in and we've closed up the BRS chamber. We had to add an inch of shims beneath to motor mount in order to get it to fit/align properly, but otherwise there were only minor complications during installation. We've tested the pico-motor and the mass adjuster using the pico-motor driver that we brought from UW and both seem to work fine. For in air balancing, we left the new mass adjuster centered to increase range for future adjustments for when the BRS is pumped down and running, and tried to just stick to moving masses that are inaccessible when the BRS chamber is pumped down. We unfortunately reached the maximum range with the internal masses and had to slightly move the manual mass adjustment system (what is currently used to adjust the center of mass) from center, but that might be returned to center after we re-balance it when it's pumped down. Right now the resonance frequency looks to be around ~7mHz (around 130 second period) which is about the same as it was before Mass Adjuster installation, but we'll check again after the chamber has been pumped down. Tomorrow we'll finish all of the wiring and electronics to hook up the pico-motor to the LIGO system. The plan is to pump down the BRS chamber tomorrow and re-balance and test the pico-motor some more. The reference pattern has a higher intensity than expected and we aren't sure why. Right now our best guess is that the light source drifted slightly, and we'll look into it more tomorrow.
EPO tag for BRS pics
[Shoshana, Michael] Pumped down the BRS chamber overnight and started the ion pump this morning and got it down to 1.9e-6 Torr before we left end End-X. We also wired up the picomotor to the LIGO picomotor controller system. It is hooked up to the 7th channel X-direction and we tested it out and were able to hear it spinning for both directions. The BRS's thermal insulation was reapplied the box closed and the temperature sensors and heating plate were all re-attached and plugged back in. The reference beam's intensity has gone down to be closer to normal somehow, so it doesn't seem to be anything to worry about We might go back to End-Y one more time tomorrow to clean up the wiring and do a final check of the vacuum pressure. We waited for the temperature to equilibrate a bit before balancing because we were hoping that as the temperature rises it would drift back to center, but we ended up using the mass adjuster to try and balance it. It looks like the + - wires for the damping were switched, meaning when the damping was on it would ring up the BRS. Fixed by changing a line in the BRS code[IF H1_ISI_GND_BRS_ETMY_CAPDRIVE>=0] by switching the [>=0] to [<=0] and switching [H1_ISI_GND_BRS_ETMY_CAPOUTL] and [H1_ISI_GND_BRS_ETMY_CAPOUTR], and [H1_ISI_GND_BRS_ETMY_RELAYL] and [H1_ISI_GND_BRS_ETMY_RELAYR] FROM THE END-X INSTALL: Coupling/decoupling move: 1.25k steps Maximum: +-140k steps Be careful: +-100k steps NOTE: MOVING PICOMOTOR +(POSITIVE) DIRECTION TRANSLATES TO MOVING THE BRS UP TOTAL MOVEMENT TODAY:+21k steps
Centered both ETMY and ETMX BRSs. For ETMX net movement was +2200 steps, for ETMY net movement was -3200 steps. For ETMX we saw that the DRIFTMON was moved by about ~3.27 counts per step, and ~2.3 counts per step for ETMY
Cleaned everything up for the ETMY BRS and relabeled all the wires. The final reading that we saw for the ion pump was 9.9e-7 Torr (186uA, 6950V) which seems about right. We left some extra mass adjuster parts with Jim just in case.
Shoshana, Michael
We took tilt subtraction spectra as a final life check of the BRSs. Both BRSs appear to be in good working order and doing their jobs well.
Sheila, Betsy, Tyler, Oli, Jennie, Camilla,
Day 1: 84193, Day 2: 84228, Day 3: 84230 and 84239, Day 4: 84274, Day 5: 84292, Day 6: 84314, Day 7: 84334
Photos attached of: | REFL (BDV1) | POP (BDV2) |
Location | photo | |
Open Position | photo | photo |
Closed Position | photo | photo |
Beamdump Location | photo | photo |
Beamdump Name | BD5 | BD4 |
Beam dumped | photo | photo |
PSL ALS: we decided that the PSL ALS path was too close to the edge to the VP (84334) we tried moving M13 so the beam traveled in-between M11 and L2 but this didn't give enough separation at the bottom pericope mirror of ISCT1. We then put M1 back to a similar position to wear we started an decided that it was good enough away form clipping the VP. Photos attached of M15 location, beam on M15, beam next to L2, VP simulator, UPM, LPM.
The logic of the REFL diverter has been changed to alternate beam. This required a restart of the slow controls software.
Before lunch, Tyler, Oli and Betsy removed ISCT1 and the VP simulator.
POP Beam to diodes:
Jennie and I aligned the POP beam onto the POP X ASC diode (using flashes then single bounce, with 20W in). We needed to adjust the dark offsets (they are about the same size as the 1.5uW single bounced beam), sdfs accepted in safe attached, properly need to be redone and saved once we get to observe. All segments responded to the beam and a flashlight and we centered the beam in pitch and yaw using PM1 (alignment sliders P:100, Y:50).
Elenna and I aligned the POP beam onto the POP A LSC diode (using flashes then single bounce). We checked the POP LSC picomotor was working as expected. Checked beam was centered on the diode by it was centered by going to the middle of the edges of the diode. All segments responded to the beam and a flashlight and we centered the beam in pitch and yaw using PM1, ndscope. Note that we'll want to check this again tommor0w as we need to offload the picomotor in yaw as it doesn't have much range left.
We couldn't see the didoe reflected beams (using POP flashes, sensitive IR card and IR viewer) so I dead rekon'ed the positions of beam dumps BD1 and BD2, see attached photo.
Leaving IMC locked with 2W in.
Still to do: offload POP LSC pico and check alignment, place BD6a, 6b, 7, 9 (couldn't see beams easily tpday at 2W), final checks (see other alog).
Betsy, Camilla, Keita
Name | Path | Beam being dumped | Checked? |
BD1
|
POP
|
LSC POP_A Reflection
|
|
BD2
|
POP
|
ASC POP_A WFS Reflection
|
|
BD3
|
POP
|
RM3 Transmission
|
|
BD4
|
POP
|
BDV2 Reflection
|
|
BD5
|
REFL
|
BDV1 Reflection
|
|
BD6a
|
REFL
|
M7 Ghost
|
|
BD6b
|
REFL
|
M7 Ghost
|
|
BD7
|
REFL
|
Unknown, beam from upstream of M2
|
|
BD8
|
REFL
|
M2 Transmission
|
YES: photo
|
BD9
|
REFL
|
Unknown, beam from upstream of M2
|
|
BD10
|
REFL
|
ASC REFL_A Reflection
|
YES: photo
|
BD11
|
REFL
|
ASC REFL_B Reflection
|
YES: photo
|
BD12
|
REFL
|
RM2 Transmission
|
YES: photo
|
BD13
|
REFL
|
LSC REFL_A Reflection
|
YES: photo
|
BD14
|
REFL
|
LSC REFL_B Reflection
|
YES: photo
|
BD15
|
REFL
|
M2 Ghost (photo showing distance from main beam)
|
YES: photo
|
BD16
|
REFL
|
RM1 Transmission
|
YES: photo
|
BD17
|
PSL
|
Nozzle Baffle on PSL-HAM1 VP
|
YES: photo
|
BD18
|
PSL
|
Nozzle Baffle on PSL-HAM1 VP
|
YES: photo
|
Name
|
Path
|
Beam being dumped
|
Checked?
|
BD1
|
POP
|
LSC POP_A Reflection
|
YES: photo
|
BD2
|
POP
|
ASC POP_A WFS Reflection, beam clears by ~15mm, photo
|
YES: photo
|
BD3
|
POP
|
RM3 Transmission (Beam too weak, estimated placement)
|
Estimated: photo
|
BD4
|
POP
|
BDV2 Reflection MOVED LOCATIONS
|
YES: photo
|
BD5
|
REFL
|
BDV1 Reflection, alos catching 2 ghost beams, photo
|
YES: photo
|
BD6a
|
REFL
|
M14 Ghost MOVED LOCATIONS
|
YES, see BD6b.
|
BD6b
|
REFL
|
M14 Ghost
|
YES: photo
|
BD7
|
REFL
|
Unknown, beam from upstream of M2, MOVED LOCATIONS
|
YES: photo
|
BD8
|
REFL
|
M2 Transmission
|
YES: photo
|
|
|
|
Not using, dumped by BD8
|
NewBD
(BD9)
|
REFL
|
REFL ASC A low reflection, beam bounces off ISI, photo
Call this new beamdump BD9 as now not using BD9. |
YES: photo
|
BD10
|
REFL
|
ASC REFL_A Reflection
|
YES: photo
|
BD11
|
REFL
|
ASC REFL_B Reflection
|
YES: photo
|
BD12
|
REFL
|
RM2 Transmission
|
YES: photo
|
BD13
|
REFL
|
LSC REFL_A Reflection
|
YES: photo
|
BD14
|
REFL
|
LSC REFL_B Reflection
|
YES: photo
|
BD15
|
REFL
|
M2 Ghost (photo showing distance from main beam)
|
YES: photo
|
BD16
|
REFL
|
RM1 Transmission
|
YES: photo
|
BD17
|
PSL
|
Nozzle Baffle on PSL-HAM1 VP
|
YES: photo
|
BD18
|
PSL
|
Nozzle Baffle on PSL-HAM1 VP
|
YES: photo
|
HiPWR
|
REFL
|
M14 and M17 Reflection
|
YES: photo
|
EPO tag for pics of laser safety in practice.
We ended up using pico in HAM3 to bring down green ALS beam as well as IR POP rather than raising the top periscope mirror after confirming that it was safe to do so.
We first confirmed that the X arm and corner alignment were still the same as yesterday. X arm was still locked to 00 green, and when I misaligned PRM the IR flash in H1:LSC-TR_X_IN1 was reaching 0.8 or more.
We convinced ourselves that the alignmernt was reasonable and probably similar to the alignment before the vent, that the ALS beams were pretty high but was still hitting the top periscope mirror before just like ALSX beam is as of now. We discussed the use of the pico in HAM3 (instead of moving the top periscope mirror up) and decided that it's OK to do so as far as the IR beam is far from clipping somewhere in HAM2.
We moved the HAM3 pico for PIT in "Down" direction by 700 counts and the IR beam position looked about right. We moved the pico further down by 1000 counts (that's the total of 1700 counts down from the initial state) and the IR beam missed the bottom of the top periscope mirror but no sign of clipping in HAM2, so we were comfortable to go 1000 counts "UP" and search for an optimum positision. In the end we ended up going back to 700 counts "Down" relative to where we started.
Sheila and Camilla adjusted top and bottom periscope mirror. M10 (dichroich) was also touched up. Steering mirrors for in-air beams were also roughly adjusted, These are M7 (for REFL air), M11 (for green ALS), M13 (for IR ALS beam from the PSL) and M16 (for POP air).
Now we're ready to move ISCT1L in position to confirm that all beams hit the periscope mirrors on the table.
Photos attached of the Green X-arm and POP beams on the HAM1 top periscope mirror before the HAM2 pico-ing and after the HAM2 pico-ing in the final position.
Tyler fork-lifted ISCT1 into position and we correctly set the height. There was some confusion due to a typo by me on the -X-Y corner height measurement that we've now corrected.
PSL ALS and Green ALS Beams:
To get the PSL ALS and green ALS beams through the VP simulator and onto the top periscope mirror, Sheila needed to move both final steering mirrors (M11 and M13) to have the beams more centered on the VP simulator:
Photos of the final positions attached, on the VP simulator (inner red circle is the location of the actual window), on the upper periscope mirror, on the lower periscope mirror and on the ALS shutter, we expect to need to do some steering of the UPM to get the final beams better on the LPMs, this will be done once ISCT1 is installed and realigned.
POP air and REFL air alignment
(Note: At this point, HAM1 drawing https://dcc.ligo.org/D1000313-v19 was not useful for exact placement of the mirrors because, as was reported earlier, POP and REFL beam position coming from HAM2 into HAM1 are different from what the drawing suggests by an inch or so. )
POP air steering mirror (M16) had to be relocated closer to the periscope, otherwise the POP air beam hit the -X edge of the viewport simulator when we tried to center POP air on the periscope mirror on ISCT1. But then the REFL air beam became close to the M16 mirror mount, so we ended up moving M16 MUCH closer to the periscope and the REFL air steering mirror (M7) got further down in -X direction. As a result, two beams now crosses with each other inside the chamber. See the first attachment where the new POP air beam is in orange and the new REFL air path is in cyan. This is for illustration purpose only.
2nd attachment: Overview of the POP and REFL air path in chamber.
3rd and 4th attachment: Detailed location of M16 and M7, respectively.
5th attachment: REFL air beam is comfortably away from the M16 mirror holder.
6th attachment: Beam position on the viewport simulator. Dimmer beam is POP air.
7th and 8th: REFL and POP beam on the top periscope mirror of ISCT1.
This means that the beam diverters WILL have to be relocated and reconfigured.
Just swapping the positions might not work.
I wonder if using one diverter at the edge of the table will catch both beams. (That's not a good idea as we might want to use in-air POP WFS if in-vac one doesn't work well.)
Distance from POP air steering mirror (M16) to the top POP periscope mirror in ISCT1 was: 34 1/4 + 30 1/4 = 64.5"
Distance from REFL air steering mirror (M7) to the top REFL periscope mirror in ISCT1 was: 13 3/4 + 30 + 31.5 = 75.25"
Measurement was done using long-ish ruler, but it wasn't long enough so we had to measure the distance in multiple segments. I'd say that the error bar of this is roughly +-3/4".
Betsy, Camilla, Keita, Jason This morning, we took a closer look at eh small PSL-ALS beam path hitting the 1 and only mirror in HAM1 before heading to IOT1. It looked a bit clipped/elongated and was coming out of the light pipe toward the -Y side and veering off course onto the table (when compared to the new layout map). Given work on the path on the PSL table (for SPI), Jason and Keita went in to the PSL and revisited how the beam launches at the periscope. They could see scatter somewhere in the light pipe so it was probably clipping in there. Using the top ALS periscope mirror, Jason was able to yaw the beam back to the designed position of the mirror in HAM1 (Keita on the phone with Jason in the box, relaying to Betsy who moved the mirror and beam dump on HAM1). Now the beam looks nicely round, more centered in the entry viewport window, and runs along the table per the map we are aligning to D1000313-v19. We pointed the steering mirror so the beam leaves the table to also match the map. Next up is the X-arm peek and tweaking of alignments to get onto the IOT1 periscope mirrors.
First attachment shows the beam right after the viewport, shot by Betsy's cellphone. The beam looked much worse to my eyes on the viewer card.
Anyway, Jason and I found that
3rd attachment shows ALS-L2 where the beam was offset by about half of the open aperture radius of the lens mount (the lens itself is 1").
We used ALS-M2 and ALS-M3 to recenter the beam on two reference irises. 4th attachment shows ALS-L2 after this adjustment where the beam is more centered. Not perfect but much better.
On the 5th picture the beam shows no sign of clipping but the beam is close to the -Y edge of the baffle hole for the viewport.
Next Jason moved the top periscope mirror to steer the beam to where it's supposed to be in HAM1. The beam looked more centered on the baffle (6th picture).
7th picture: The location of the steering mirror for this beam on HAM1, which was set according to the HAM1 layout drawing.
8th and 9th show the PSL-ALS beam right after it is reflected by the steering mirror, and right before it leaves HAM1. We might have to fine-adjust this.
As Keita notes, before yawing the top periscope mirror he and I checked the 2 ALS alignment irises that sit under the periscope between mirrors IO_ALS_M3 and IO_ALS_M4 (the bottom ALS periscope mirror). The beam was off on both alignment irises to the -Y side, suggesting a beam shift had occured sometime between the end of the PSL laser upgrade in Jan 2022 and now (the last time I recall looking at this alignment was at the end of the laser upgrade); using an IR viewer Keita looked at the beam on IO_ALS_L2 (the 2nd ALS mode matching lens) and the beam was off-center on the lens as well. The beam looked decently well aligned (not perfect) along the IO_ALS_M1 to IO_ALS_M2 path, but beyond that it was well off. It's unclear what casued this shift. There was the recent SPI pickoff install last month that potentially could have bumped IO_ALS_M2, since they were working in that area, but I don't think that alone would be enough to cause the shift we saw. The beam needed to be walked in the +Y direction using both IO_ALS_M2 and IO_ALS_M3 to re-center on the alignment irises, but bumping only a single mirror would more than likely have caused an angular shift in the beam and be easily corrected using only the mirror that was bumped. This defintely was not the case, as we needed both IO_ALS_M2 and IO_ALS_M3 to walk the beam back to the irises. So as of right now I cannot say what caused this misalignment, it's entirely possible the ALS beam has been misaligned for a while.
As noted above, we used IO_ALS_M2 and IO_ALS_M3 to re-center the ALS beam on the alignment irises, and at this point the beam was no longer clipping when traversing the ALS light pipe. Back on the HAM1 table, the mirror that directs the ALS beam onto ISCT1 was still too far in the -Y direction for comfort (getting a little to close to the future home of the JAC and its associated optics). I then went into the enclosure and yawed IO_ALS_M5 (the top ALS periscope mirror) until the mirror in HAM1 was in its as-designed location. I've attached a picture I took of how the ALS beam now looks when entering the ALS light pipe; it's now a little closer to the +Y edge of the light pipe. Before adjustment the outside edge of the beam was very roughly 1" or so away from the light pipe edge (did not get a picture of this), now it's very roughly 1/2" or so away from the edge.