Displaying reports 1081-1100 of 82965.Go to page Start 51 52 53 54 55 56 57 58 59 End
Reports until 16:31, Wednesday 07 May 2025
H1 ISC
betsy.weaver@LIGO.ORG - posted 16:31, Wednesday 07 May 2025 (84312)
HAM1 PSL ALS beam in chamber now
Gearing up to align the 1 PSL ALS beam mirror in HAM1, we opened the ALS shutter. Camilla immediately found the beam and dumped it into a v-dump to align tomorrow. This shutter will stay open for a couple days while we finish alignment of this path alongside the other paths.
Images attached to this report
LHO General
ryan.short@LIGO.ORG - posted 16:31, Wednesday 07 May 2025 (84311)
Ops Day Shift Summary

TITLE: 05/07 Day Shift: 1430-2330 UTC (0730-1630 PST), all times posted in UTC
STATE of H1: Planned Engineering
INCOMING OPERATOR: None
SHIFT SUMMARY: Lots more progress today; PM1 is in place and EX wind fence work is going well.
LOG:

Start Time System Name Location Lazer_Haz Task Time End
15:07 SAF LVEA is Laser HAZARD LVEA YES LVEA is Laser HAZARD Ongoing
14:50 FAC Kim, Nelly LVEA - Technical cleaning 15:15
14:51 FAC Ken LVEA - HAM4/5 cable trays 22:51
15:55 AOS Betsy LVEA - Talking to Ken 16:13
15:56 FAC Kim H2 N Technical cleaning 16:11
16:14 VAC Gerardo LVEA - Talking to Ken 16:30
16:32 ISC Keita, Jennie LVEA YES HAM1 alignment (Jennie out @ 18:59) 19:31
17:01 EE Fil LVEA - HAM1 racks 18:38
17:05 SEI Jim, Mitchell, Randy, Tony EX N Wind fence work 19:03
17:11 EE Jackie FCES N Measurements 19:11
17:45 VAC Travis, Tyler MY N Moving equipment 19:20
18:14 FAC Nellie EX N Tech clean 18:46
18:34 VAC Gerardo LVEA - Isolating OMC turbo 18:42
19:08 ISC Daniel LVEA - ISC rack cable istallation 20:21
19:13 SUS Rahul LVEA - PM1 install 20:46
19:13 ISC Camilla LVEA YES HAM1 POP path alignment Ongoing
19:31 CDS Dave MY N PEM sensor electronics 20:01
19:35 ISC Sheila LVEA YES HAM1 POP path alignment 22:20
19:55 ISC Elenna LVEA YES HAM1 POP path alignment 20:56
20:05 SEI Jim, Randy, Mitchell EX N Wind fence work 21:46
20:14 FIT Ibrahim Y-arm N On a walk 20:54
20:29 ISC Keita OptLab N Looking at spare Siskyou mount 20:58
20:32 CAL Tony PCalLab Local Start measurement 20:38
20:42 ISC Oli LVEA - Unlocking rotation stage 20:51
21:26 VAC Travis MY N Roughing pump work 22:19
21:37 PCAL Tony PCAL Lab Y(local) Getting stuff to ship to France 21:37
21:39   Betsy LVEA YES Seeing if HAM1 team needs help 22:03
21:50 SUS Rahul LVEA - PM1 adjustments 23:03
22:13 ISC Elenna LVEA YES HAM1 POP path alignment Ongoing
23:03 AOS Betsy LVEA - Checking on HAM1 team Ongoing
H1 ISC
sheila.dwyer@LIGO.ORG - posted 16:27, Wednesday 07 May 2025 - last comment - 15:21, Monday 19 May 2025(84307)
HAM1 POP path beam profile and distance measurements

location number on drawing

distance horizontal 13.5% diameter [um] vertical 13.5% diameter [um] photo of profiler location photo of profiles photo of beam scan measurements
1 52 mm from dichroic M10 6243 64040 here and here 9022 9023
2 147 mm from dichroic M10 6202 6365 9028 9026 9027
3 119 mm from HR of 50/50 BS M15, also 295mm from center of lens L2 870 874 9031 9030 9029
4 153 mm from HR of 50/50 BS, approximate location of LSC diode 279 284 9035 9032 9033
5 128 mm from HR of 50/50 BS 721 726 9036 9037 9038
6 139 mm from HR of 50/50 483 490 9041 9040 9039
7 353 mm from HR of 50/50 BS 3280 3372 9042 9043 9044

Camilla made these measurements with 20W input power into the IMC, PRM and ITMY misaligned single bounce beam.  We didn't place the 90/10 BS M12 in the pop path yet so that we would have about 35uW to measure beam profiles. There's a rough pen sketch of where these locations are in this photo.

Camilla also made ruler measurements of some distances:

Images attached to this report
Comments related to this report
elenna.capote@LIGO.ORG - 19:48, Wednesday 07 May 2025 (84313)

I think these measurements look pretty good, and the result is not that much different than the model (at least in terms of where to put the diodes). The biggest surprise is that Keita and I were pretty sure the beam coming into HAM1 is just over 2 mm in radius, when in reality it is more like 3 mm in radius.

I have attached plots of both the horizontal and vertical propagations, but you can see that the results for each direction are very similar. The black points are the measurements Sheila and Camilla took today, and I fit them with a la mode and the 2 inch POP lens (f = 334 mm), using the distances from the dichroic mirror that Sheila and Camilla measured.

If we place the POP LSC diode about 140 mm from M15 and the POP WFS about 200 mm from M15, we should get the beam sizes we want on each. I think these positions are a little closer to M15 than they are in the drawing, but I don't think that's a problem.

I also attached my a la mode script.

Images attached to this comment
Non-image files attached to this comment
sheila.dwyer@LIGO.ORG - 15:21, Monday 19 May 2025 (84471)

Linking here that Keita added distances to the refl air and popair periscopes here: https://alog.ligo-wa.caltech.edu/aLOG/index.php?callRep=84345

H1 ISC (ISC, SYS)
keita.kawabe@LIGO.ORG - posted 15:47, Wednesday 07 May 2025 - last comment - 16:01, Wednesday 07 May 2025(84308)
HAM1 Wednesday

Morning (JennieW, Rahul, Keita)

We used the PRX flashes to align the POP path.

POP periscope location is good but the drawing is not.

The POP periscope position, which was set yesterday by Camilla and others, was right. That means that the drawing on D1000313-v19 is wrong. The periscope in reality is about an inch toward -Y direction relative to D1000313-v19. See the first picture, which was shot with a cellphone inserted under the top periscope mirror and looking straight down the bottom mirror. This means that the dichroic (M12) needed to be shifted by the same amount too.

Since the distance betwen the IFO and the lens for POP WFS doesn't matter that much, everything downstream (i.e. 90:10, PM1 tip-tilt, a lens, 50:50 and POP LSC as well as POP WFS) will be installed using the drawing.

We mainly rotated the periscope mirror clamps around the post for rough alignment, but we might have changed the mirror height by a millimeter or two in the process. 

PM1, which is calld that because it's the 1st (and the last) suspended Mirror for POP, is somehow called RM3 in D1000313. Systems please fix it.

Set the IR beam spot height/position on the periscope as well as the dichroic

The IR beam is supposed to be about 6mm or 1/4" lower than the center line of both of the periscope mirrors as well as the dichroic. This is because the green ALS beams are supposed to be ~13mm higher than IR. See L1200282 “CPy-X, CPx-Y” case on Table 1.

Top Periscope Mirror

It was almost impossible to see how much the beam is lower than the center of the top and bottom periscope mirror. Using the IR viewer card, I and Jennie agreed that the beam is lower than the center, but we could not quantitatively say how much especially on the top. We'll leave it as is, and if the green beam from the end station is too high we will have to use pico because we periscope is already as high as possible.

Bottom peri mirror

If everything is as intended, the bottom periscope mirror is 4" high from the ISI surface and the POP beam is 1/4" lower than that, therefore the POP beam is (1-0.25*sqrt(2)) = 0.646" = 16.4mm away from the bottom edge of the mirror.

Using a ruler in chamber (and measuring the dimensions of a spare Siskiyou mount using caliper), the height of the bottom periscope mirror center was calculated to be ~4.07" from the ISI surface, i.e. 0.7" too high. This means that, when the beam height measured from the ISI is as designed (i.e. 4"-1/4"), the POP beam is (1-(0.25+0.07)*sqrt(2))=0.547"=13.9mm away from the bottom edge of the mirror.

If you have difficulty understanding this, see the cartoon.

POP beam radius is ~2mm, so 13.9mm (or even 13mm for that matter) looks like a safe distance to me. I don't see the need to readjust the height of the bottom periscope mirror.

I adjusted the top periscope mirror to set the beam height right after the bottom peri mirror to be ~3.75" using the IR viewer card and a ruler.

Dichroic

I placed the dichroic about 1" into -Y direction relative to the drawing (because I had to), and used the bottom periscope mirror to set the beam height close to the dichroic to be ~3.75".

Then I used the dichroic to steer the beam into the direction of the location for PM1 without placing 90:10.

For the beam profile measurement, the downstream alignment is done without 90:10. Later we will install 90:10 back in place and do the final alignment.

Images attached to this report
Comments related to this report
keita.kawabe@LIGO.ORG - 16:01, Wednesday 07 May 2025 (84310)

And here's a memo of how we "measured" the height of the center of the bottom periscope mirror.

Images attached to this comment
H1 TCS
matthewrichard.todd@LIGO.ORG - posted 14:26, Wednesday 07 May 2025 (84306)
Estimating surface defocus of test masses and absorption from HOM spacing

M. Todd, S. Dwyer


As derived in previous alogs, we are able to relate the HOM spacing observed in each arm to the surface defocus of the test masses -- which is a combination of self-heating and ring heater power (ignoring CO2 affects on the ITM RoC). From the fits we've made of the HOM spacing / surface defocus change as a function of ring heater power we can get a value for the ring heater to surface defocus coupling factor.

Theoretically from this we should be able to solve for the self heating contribution in the test masses as well -- allowing us to constrain things like the coupling of absorbed power to surface defocus at the ITMs if we assume to know the arm power and absorption values (from HWS).


Upper Limits

If we assume no absorption in ETMs (obviously not physical), and we assume the HWS values for the ITM absorptions are correct, then with a HOM spacing measurement from each arm we can get an upper limit of the coupling factor of self-heating to surface defocus for each ITM (they shouldn't be different but this is a good exercise).

Assuming alpha is the absorption coefficient, i subscript is for the ITM, and x/y is which arm. P_y,i_rh is the itmy ring-heater. G-factors are the product of ITM and ETM g-factors. Then from the formula in section 1.2 of the notes file : Gy = Gyc - B*L*gyic*(Pyerh+Pyirh) - L*(Ai*alpha_yi*Pyarm*gyec + beta*Ai*alpha_e*Pyarm*gyic), we can solve for Ai which is the coupling factor of self-heating to surface defocus.

Parameter Value Notes
alpha_x,i 430 ppm from alog 76937
alpha_y,i 375 ppm from alog 76937
alpha_x,e 0 ppm  
alpha_y,e 0 ppm  
P_y,i_rh 0.000 W T0 = 1417899757
P_x,i_rh 0.850 W  
P_x,e_rh 1.950 W  
P_y,e_rh 2.146 W  
P_yarm
385159 W
T0 = 1417899757
P_xarm 385159 T0 = 1417899757
Gx 0.8149 T0 = 1417899757
Gy 0.8198

                    TMS * pi

G    =   cos2 (   ----------------   )

                      FSR

Ai_y -26 uD/W  
Ai_x -39 uD/W  

Middling Values

If we assume quoted absorption in ETMs (measured by LIGO, on galaxy), and we assume the HWS values for the ITM absorptions are correct, then with a HOM spacing measurement from each arm we can get a more realistic value of the coupling factor of self-heating to surface defocus for each ITM (they shouldn't be different but this is a good exercise).

Assuming alpha is the absorption coefficient, i subscript is for the ITM, and x/y is which arm. P_y,i_rh is the itmy ring-heater. G-factors are the product of ITM and ETM g-factors. Then from the formula in the notes file : Gy = Gyc - B*L*gyic*(Pyerh+Pyirh) - L*(Ai*alpha_yi*Pyarm*gyec + beta*Ai*alpha_e*Pyarm*gyic), we can solve for Ai which is the coupling factor of self-heating to surface defocus.

Parameter Value Notes
alpha_x,i 430 ppm from alog 76937
alpha_y,i 375 ppm from alog 76937
alpha_x,e 200 ppm  
alpha_y,e 210 ppm  
P_y,i_rh 0.000 W T0 = 1417899757
P_x,i_rh 0.850 W  
P_x,e_rh 1.950 W  
P_y,e_rh 2.146 W  
P_yarm
385159 W
T0 = 1417899757
P_xarm 385159 T0 = 1417899757
Gx 0.8149 T0 = 1417899757
Gy 0.8198

                    TMS * pi

G    =   cos2 (   ----------------   )

                      FSR

Ai_y -16 uD/W  
Ai_x -26 uD/W  

Summary

Both of these values indicate there is certainly an overestimation of the self-heating impact on surface defocus.

For reference, the current TCS-SIM values for this coupling factor are Ai_y = Ai_x = -36.5 uD/W.   More examination is required into this.


Links to previous alogs:

  1. Estimating HOM spacing shift from self-heating alone:  https://alog.ligo-wa.caltech.edu/aLOG/index.php?callRep=84172
Non-image files attached to this report
LHO VE (VE)
gerardo.moreno@LIGO.ORG - posted 10:55, Wednesday 07 May 2025 (84303)
Morning Purge Air Check

Dry air skid checks, water pump, Kobelco, all nominal.  Noticed extra noise coming from the left dryer tower (rattling), maybe a one of the left side one-way valves is showing its age.

Dew point measurement at HAM1 -45.7 °C.

.

Images attached to this report
LHO VE
david.barker@LIGO.ORG - posted 10:28, Wednesday 07 May 2025 (84302)
Wed CP1 Fill

Wed May 07 10:12:33 2025 INFO: Fill completed in 12min 29secs

Gerardo confirmed a good fill curbside.

Images attached to this report
H1 CDS
david.barker@LIGO.ORG - posted 10:18, Wednesday 07 May 2025 - last comment - 15:56, Wednesday 07 May 2025(84300)
Geist environmental monitors installed in out-buildings

WP12510 FRS34002

I have installed the four Geist Watchdog 1250 rack mount units in the CDS computer racks at MX, MY, EX and EY.

At the end stations these units are mounted close to the top of the CDS front-end computer racks, forward facing (see attached photo of EY). The end station computer racks are located in the first room you enter from outside.

At the mid stations these units are mounted in the VEA computer rack, located roughly in the center of the VEA. All the internal roll-up doors are open, essentially making the whole building one very large room.

EPICS IOCs were created to read out these units, and their channels were added to the DAQ yesterday.

I have added an environment section to the CDS overview (see attachment) which provides visual alerts if any out-building sensor is out-of-bounds.

The LEDs show light_level (L), sound_level (S), temperature (T) and humidity (H).

In the example shown, the EX lights are on.

Images attached to this report
Comments related to this report
david.barker@LIGO.ORG - 10:26, Wednesday 07 May 2025 (84301)

Providing the Geist units with a network connection to the ADMIN-LAN:

At the midstations, the ADMIN-LAN is provided by a fiber-to-ethernet converter box, originally directly connected to the PEM front-end computer IPMI port. This is now sent to a 4-port ethernet hub, which feeds the IPMI port and the Geist unit.

At EY it was possible to reconfigure sw-ey-stk(2) to give ADMIN 4 more ports, and remove 2 each from SLOW and FE. This required moving all four SLOW cables 4 ports to the right, which was done quickly with no impact to HWS, BRS, HEPI-PUMP.

At EX it was not possible to do the same, so here two unused SLOW ports (15,16) were reallocated the ADMIN.

david.barker@LIGO.ORG - 15:56, Wednesday 07 May 2025 (84309)

To complete the install I removed the MY rack doors (front and rear) to make them the same as MX and improve front-end computer cooling (see trend).

Images attached to this comment
LHO General
ryan.short@LIGO.ORG - posted 07:50, Wednesday 07 May 2025 (84299)
Ops Day Shift Start

TITLE: 05/07 Day Shift: 1430-2330 UTC (0730-1630 PST), all times posted in UTC
STATE of H1: Planned Engineering
OUTGOING OPERATOR: None
CURRENT ENVIRONMENT:
    SEI_ENV state: LIGHT_MAINTENANCE_WINDY
    Wind: 1mph Gusts, 0mph 3min avg
    Primary useism: 0.02 μm/s
    Secondary useism: 0.14 μm/s
QUICK SUMMARY: More alignment in HAM1 and wind fence work at EX lined up for today.

H1 ISC
camilla.compton@LIGO.ORG - posted 17:00, Tuesday 06 May 2025 (84292)
ISC Alignment Work, Day 5 in HAM1: DC centering loops working, REFL alignment finished (apart from BDs), 2W now into IMC

Oli, Keita, Elenna, Jennie, Sheila, Ryan S, Ryan, Jeff, Camilla

Day 1: 84193, Day 2: 84228, Day 3: 84230 and 84239, Day 4: 84274

Realsied that yesterday we had the POP aligned causing flashes and need to have ITMX misaligned to work on the REFL path best. Elenna and Keita re-took the final two REFL path beam profile measurements, Travis and I turned down the purge air for these. Turned back up to maxium afterwards.

Rahul installed PM1, slightly out of the way of the POP path so it will need to be moved ~12" to be in it's final postion. It's beam dump will later be added.

Oli, Shiela and I recentered the ASC REFL B diode as yesterday we left it with the beam off mostly off the diode. We swapped the pico cables to the two on sled picos so now the match the medm labels and tested them both. Sheila then centered ASC REFL B using picos. Around 400 NSUM counts on each diode with 200mW of PSL beam in.

Elenna, Sheila and Jeff did some troubleshooting of the DC centering loops (use ASC REFL A and B signals and feedback to RM1 and RM2), there was two issues: one was that the RM1 and RM2 sign flip was no effecting the LOCK filters, Jeff chnegd this (see 84289), secondly it seemed that what was the DC signal of ASC REFL A is now ASC REFL B, Ryan is doing some alog digging to check if this was known in the last HAM1 vent. It's possible the DC signals have just been the wrong way around and this would make some sense why the pico labeled were swapped, but if the RF signals are swapped it will cause lots of issues locking the IFO. After these things were fixed, the DC centering loops were working. 

Oli and I found the position the LSC sensors should be to have the beam radius on the diodes to be ~0.25mm and placed them there, aiming from 0.2 - 0.3mm. Beam profile attached at 165mm from M6 BS HR surface to front of BS, adding 5mm for diode body placement to account for the distance in front of body the diode is and that the measurement plane of scan is 0.74mm behind the surface. 
We placed both LSC REFL A and B 170mm from the HR surface of the M6 BS. We aligned these by eye and got 0.4 on each H1:LSC-REFL_{A,B}_LF_OUT16, we dumped the reflected beams. Once the DC centering loops were on,  Elenna and myself then checked that the beams were centered by using the steering mirrors and ndscope channels. Also checked the pico motor in the REFL path worked in PIT and YAW.

Oli turned up the PSL input to 2W and then de-energized and locked out.

Elenna and I could then see POP flashes, we started to align the periscope but ran out of time before finishing. 

ITMX left misaligned so there3's no POP beam.

Images attached to this report
LHO General
thomas.shaffer@LIGO.ORG - posted 16:36, Tuesday 06 May 2025 (84297)
Ops Day Shift End

TITLE: 05/06 Day Shift: 1430-2330 UTC (0730-1630 PST), all times posted in UTC
STATE of H1: Planned Engineering
INCOMING OPERATOR: None
SHIFT SUMMARY: EX wind fence work and more HAM1 beam measurements were the main activites today. PM1 is now on table and hooked up. Sensor correction has been turned on for the corner station by taking SEI_ENV to LIGHT_MAINTENANCE_WINDY.
LOG:

Start Time System Name Location Lazer_Haz Task Time End
15:07 SAF LVEA is Laser HAZARD LVEA YES LVEA is Laser HAZARD 07:07
14:35 FAC Chris EX, MX, EY, MY N Searching for lift 16:01
14:57 FAC Kim, Nellie LVEA - Technical cleaning 17:21
15:35 VAC Jordan LVEA - Purge air and pump checks 17:17
15:49 SEI Jim, Mitchell EX N Wind fence work 18:49
16:01 FAC Chris LVEA - FAMIS checks 16:20
16:09 EE Jackie FCES N Looking for oscilloscope 18:09
16:20 FAC Chris, Pest Ctrl. LVEA, EX, MX, EY, MY - Checking traps 17:27
16:23 FAC Ken LVEA - HAM4/5 cable trays 23:00
16:24 CDS Fil LVEA - Electronics install in HAM6 racks 18:34
17:12 ISC Camilla, Oli LVEA YES HAM1 beam alignment 19:44
17:21 FAC Kim, Nellie FCES N Technical cleaning 18:20
17:28 OPS RyanC OptLab N Checking dust monitor 17:49
17:33 FAC Tyler LVEA - HEPI elbow removal 17:47
17:39 ISC Sheila LVEA YES Joining HAM1 crew 18:51
18:03 ISC Keita LVEA YES Joining HAM1 crew 18:33
18:05 VAC Travis LVEA - Dropping off parts 18:13
18:06 CAL Tony PCal Lab Local Starting measurement 18:22
18:20 ISC Betsy, Elenna LVEA - Checking in with HAM1 crew 18:33
18:48 FAC Nellie MY N Technical cleaning 19:19
18:50 OPS RyanC CER, LVEA - Looking for box to restart 19:38
19:57 VAC Travis MX N Work on roughing pump 21:12
19:59 CDS Dave CER - Power cycling dust monitor box 20:08
20:07 TCS TJ LVEA - Picture of TCS table 20:22
20:15 SUS Rahul LVEA - PM1 install 21:49
20:16 SEI Jim, Mitchell, Corey, Randy EX N Wind fence work 22:14
20:16 CDS Dave EY N PEM sensor install 22:16
20:48 OPS RyanC CER - Restart dust monitor contoller box 21:05
21:10 ISC Elenna, Keita LVEA YES HAM1 beam profiling (Keita out 2242) ongoing
21:14 VAC Travis, Camilla LVEA - Turning down purge air (Travis out 21:18) 21:21
21:30 - Matt, grandparents LVEA yes Tour 22:05
21:31 SUS Oli LVEA yes Checking on RM offset directions 21:51
21:47 PCAL Tony PCAL lab yes Swap spheres 21:57
21:54 CDS Marc LVEA yes Checking connector fitment 22:14
22:22 ISC Camilla LVEA YES HAM1 work ongoing
22:52 ISC Oli LVEA YES Energizing rotation stage 23:12
H1 SUS (SEI, SUS)
edgard.bonilla@LIGO.ORG - posted 16:26, Tuesday 06 May 2025 - last comment - 15:58, Monday 12 May 2025(84296)
HAM5 drive template for SR3 OSEM calibration

Follow up to 83605.

I created a few templates to do a SR3 OSEM calibration. I called them

sus/trunk/HLTS/H1/SR3/Common/Data/
yyyy-mm-dd_hhmm_H1ISIHAM5_ST1_WhiteNoise_ISO_{X,Y,Z}_0p05to40Hz_calibration.xml

The measurement should be run with all gains on SR3_M1_DAMP_{L,T,V,R,P,Y} set to - 0.1 . Just to make sure the loop gain does not mess with the OSEM measurement of the cage above resonance.

The idea is that we ought to be able to only use the cartesian measurements of the ISI to calibrate all the OSEMs in the suspensions attached by using linear algebra. This is the first pilot for that idea. We want about 100 points between 5-10Hz to make sure we can make a "decent" fit.

___

When Oli runs these templates, the plan is to use them to calibrate the SR3 OSEMs for the estimator work (see 84171 for the latest news as of the making of this post).
We will use Oli's feedback to work out the kinks of the calibration procedure and to export it to other ISIs at a later time.

 

Comments related to this report
oli.patane@LIGO.ORG - 17:41, Tuesday 06 May 2025 (84298)

I've run these three transfer function measurements with settings as follows:

  • HAM5 in ISOLATED, although with HEPI locked but the HEPI ISO gain set to 0 to bypass the locked HEPI
  • SR3 in ALIGNED
  • SR3 DAMP gains all set to -0.1

The measurements can be found in /ligo/svncommon/SusSVN/sus/trunk/HLTS/H1/SR3/Common/Data/as 2025-05-07_0000_H1ISIHAM5_ST1_WhiteNoise_ISO_{X,Y,Z}_0p05to40Hz_calibration.xml. They have been committed to the svn as r12295.

Checking the coherence between each excitation channel (H1:ISI-HAM5_ISO_{X,Y,Z}_EXCMON) and SR3's euler basis DOFs, the coherence between 5 and 10 Hz was good between the following channels:

ISO_X -> DAMP_T

ISO_Y -> DAMP_L

ISO_Z -> DAMP_V, but also pretty good coherence with DAMP_L and DAMP_T
 

edgard.bonilla@LIGO.ORG - 15:45, Monday 12 May 2025 (84365)SEI, SUS

I have run the calibration script to get the calibration factors for the SR3 OSEMs. The (multiplicative) factors are:

    LF: 1.3206
    RT: 1.3705
    SD: 1.3806
    T1: 2.4539
    T2: 1.4819
    T3: 1.4125

The factors are calculated by fitting a frequency-independent real factor to the measured uncalibrated transfer functions to fit the modeled one. The fit range is from 5 to 15 Hz. The attached plots show the before and after of the process.

 

A few observations:

1) As we had seen before, the Left and Right (LF and RT) OSEMs are pretty well matched to one another, which bodes well for our PR3 Yaw OSEM estimator plans.
2) The T1|T2|T3 OSEMs that measure the vertical direction have their orientation flipped respect to the model. I don't think there's a problem, but it is worth pointing out.
3) The T1 OSEM, which is used to distinguish roll is 70% off from T1 and T2. The large factor feels concerning, but it is consistent with other measurements of SR3. See for example 83963, where in the attached .pdf the Vertical to roll coupling is very clean (page 12). And the difference of T1, T2, and T3 can be seen very directly in page 24.

- I think it might be worth revisiting some of these transfer functions after recalibrating just to confirm that we're getting the numbers right.

Images attached to this comment
edgard.bonilla@LIGO.ORG - 15:58, Monday 12 May 2025 (84367)

If we wanted the OSEM measurements to match with the GS13 calibrations between 5-15 Hz, the factors that should be in the OSEM input filters should be:

    T1: 3.627
    T2: 1.396
    T3: 1.345
    LF: 1.719
    RT: 1.490
    SD: 1.781

Since this change will increase the effective gain of the damping loops by a factor of 30% at a minimum, we must compensate by changing the gains for the filter modules in the M1 DAMP filter bank.

H1 SUS (SUS)
rahul.kumar@LIGO.ORG - posted 15:43, Tuesday 06 May 2025 - last comment - 15:26, Thursday 08 May 2025(84293)
HAM1 vent work - PM1 (Tip Tilt suspension) installed in the chamber

This afternoon I went into HAM1 chamber and installed PM1 which is a Tip Tilt suspension (ICS link - D1001396-07, previously ZM1 in HAM5). PM1 is currently sitting at the edge of the table (as shown in the picture here) and will be later moved to its final position sometime this week. The suspension was clamped down with four dog clamps. The other information related to PM1 is listed below,

- The BOSEMs have the following serial number - D060106-C s/n 583, s/n 562, s/n 263, s/n 201

- The quadrupus cable for the BOSEMs have the following S/N - D1000234 - S1000695 (length 55in). 

I have applied the following offsets and gains in the OSEM input filter of PM1 and accepted it in the SDF, see here.

OLC OFFSET GAIN
26440 - 13220 1.13
28894 - 14447 1.04
24529 - 12264.5 1.22
29276 - 14638 1.02

 

 

 

 

I switched on the coil drivers (Rack C4 in the CER, page 14) and satellite amp (floor SUS R1, page 13), as per the systems wiring diagram for HAM1 - D0902810.

I am currently testing the electronics chain and will post the results (osem spectra and transfer function measurements) asap.

Images attached to this report
Comments related to this report
rahul.kumar@LIGO.ORG - 15:22, Thursday 08 May 2025 (84327)SUS

Late entry:- Yesterday, we moved SUS PM1 to its desired location in HAM1 chamber. Later we were able to fine tune the position mechanically (including the pitch and yaw of the mirror) with the IR laser beam. The pictures are attached below.

Images attached to this comment
rahul.kumar@LIGO.ORG - 15:26, Thursday 08 May 2025 (84328)

This morning I found that the Voltmon for PM1 was not working, Daniel fixed it by switching on the filter modules associated with it. Earlier, I crosschecked the wiring and chassis in the CER to confirm they are correctly attached and switched on.

H1 PEM (CDS)
ryan.crouch@LIGO.ORG - posted 15:39, Tuesday 06 May 2025 - last comment - 11:53, Wednesday 07 May 2025(84295)
Diode room dust monitor issues today

The DM first starting having issues at ~11am 2 weeks ago on April 22nd (Tues) for seemingly no reason, there wasn't any work or alogs that day that would made us suspicious of causing this.

Following my alog comments on alog84282, Dave showed me where the physical comtrol box is in the CER, restarting it brought the connection back but I then found that the dust monitor had no flow :(. This DM hasn't been used since we got it calibrated in April of 2024, so I swapped it for the last pumpless spare we had which has good flow but now it's not connecting again. I've tried powercycling the DM itself, the ioc, and the comtrol box then the ioc which is what got the other DM to reconnect, and it still getting "No reply from device within 1000 ms" for all its PVs. The DM it self is working and reading counts properly.

Doing a telnet network status (comand "ss -e") on h0epics for the diode room port yielded:

timer:(keepalive,38min,0) uid:1001 ino:20939307 sk:ffff8800b67bd500
ESTAB       0      0                                                                                        10.105.0.80:37347                                                                                    10.105.0.100:8000 

 

Comments related to this report
ryan.crouch@LIGO.ORG - 11:53, Wednesday 07 May 2025 (84304)

I opened up the DM that had no flow to find that the internal tubing was not even connected, I swapped this one back this morning and it came right back no problem and we can see it on epics. The one that I took off has some kind of network issue.

Images attached to this comment
H1 ISC
elenna.capote@LIGO.ORG - posted 16:55, Thursday 01 May 2025 - last comment - 09:17, Thursday 08 May 2025(84230)
REFL Beam Profile Measurements

Keita, Elenna, Jennie W.

We have taken three beam profile measurements along the REFL path on HAM1: one in the location where REFL WFS B will go, one in the location where REFL WFS A will go, and one measurement further "downstream" where we placed a steering mirror after where WFS B will go and steered back towards the edge of the table. We will post further details later, more measurements to be made along this path tomorrow.

Comments related to this report
keita.kawabe@LIGO.ORG - 16:32, Friday 02 May 2025 (84241)

Note that the same glitches we had in the original installation (alog 8934) were still there. Quoting my alog from 2013,

it was still difficult to obtain good data because of some kind of glitches. It's not clear if it was due to NanoScan or the beam, the beam was well damped and was not moving on the viewer card, there was no noticable intensity glitch either. But the symptom was that the statistics window shows nice steady data for anywhere from one second to 30 seconds, then there's some kind of glitch and the scan/fit image looked noticably different (not necessarily ugly), the diameter mean becomes larger and the stddev jumps to a big number (like 10% or more of the mean, VS up to a couple % when it's behaving nice), and the goodness of fit also becomes large. Somehow no glitch made the beam diameter number smaller. I just kept waiting for a good period and cherry-picked.

keita.kawabe@LIGO.ORG - 16:30, Friday 02 May 2025 (84240)

We measured the beam radius using NanoScan at four points around the WFS sled (roughly WFSA position, roughly WFSB position, far field 1, far field 2). We used D4sigma numbers instead of 1/e**2 numbers. NanoScan outputs diameter not the radius, and the table below shows the raw number.

We assumed that the WFS position would be ~0.5" from the +Y edge of the WFS sled for both A and B. Distances were measured using stainless steel rulers and are relative to the 50:50 splitter on the WFS sled that also acts as the steering mirror for WFSa.

position distance [mm] 2*avg(wx) [um] 2*std(wx) 2*avg(wy) 2*std(wy)
WFSA 94 670.26 2.34 778.95 2.82
WFSB 466.5 793.73 6.38 711.29 11.95
downstream 2 788.5 1484.15 12.46 1387.24 58.32
downstream 1 1092.5 2253.78 50.67 2119.24 68.30

In all of the above measurements, "Profile averages" was 10, "Rolling profile Averages" was 3.

We also measured between M5 and 50:50 splitter for ASC-LSC split as well as between M2 and RM1. Numbers will be added to this alog.

We'll also measure the beam size at LSC REFL_B location on Monday before proceeding to POP path.

elenna.capote@LIGO.ORG - 16:41, Friday 02 May 2025 (84242)

Here are some comments about the measurement process:

The beam profiler is difficult to use because the profiler head easily swivels once it is place. The swivel seems to be driven by the fact that the cable is very stiff and made stiffer by the addition of the foil so it is cleanroom safe. Several times today, I would pick up or set down the profiler and the head would swivel. I tried tightening the screw holding the post to the base, and I tried tightening the screw that holds the post to the head, but it is not tight enough to prevent swiveling. I found the best method was to line up the profiler in the designed location, and hold the head and cable in place while someone else ran the measurement. That makes this a minimum two person job, but there was enough juggling that having a third person was sometimes helpful.

When we went in around 3 pm to do the final measurement of the day I measured the particle count: 0.3u was 10 and 0.5u was 0. I used the standing particle counter on the +Y side of the HAM1 chamber- briefly unplugged to carry it over to the -Y side for the measurement. I didn't measure when closing up because Keita is heading out to do a few more tasks on HAM1. The handhelp particle counter isn't working, so we have to carry this large one on a stand around to use.

keita.kawabe@LIGO.ORG - 11:10, Monday 05 May 2025 (84252)

WFS sled is still excellent, 84 to 85 deg Gouy phase separation.

In the attached, four measurement points have error bars both in the position and the beam size but it looks negligible. There's no concern for WFS, it's good to go as is. 

However, just for the record, the astigmatism is bigger now (which is inconsequential in that ASC DOF separation is determined by the Gouy phase even if there's an astigmatism). The waist location difference is ~49mm now VS ~14mm or so before (just eyeballing the old plot from alog 8932) for a beam with the Rayleigh range of ~200mm. Not sure if this is the result of the AOI change or beam position change on curved mirrors and lenses, but I won't fix/correct this.

Images attached to this comment
elenna.capote@LIGO.ORG - 11:40, Monday 05 May 2025 (84254)

This morning we entered to do one more beam profile measurement. First Jennie and I refoiled the cable of the nanoscan profiler, since it was very stiff from multiple layers of foil. Then, before opening the cover to the table, I measured the dust counts by carrying the stand particle counter over to our working side like I did on Friday. The read was 0 and 0 for 0.3um and 0.5 um particles. I know it was working however, because as I carried the counter over outside the cleanroom it counted 19 each of 0.3 and 0.5 um particles.

Then, Jennie and I took one more beam profile measurement, this time on the LSC REFL path, after the final beamsplitter (M18). LSC REFL A (on transmission of M18) is placed on the table as in the drawing, but the LSC REFL B sensor (reflection of M18) was further away relative to the splitter. My quick rough measurement showed that LSC REFL B was about 160 mm away from M18.

I measured the distance of LSC REFL A to the front surface of M18 to be 128 mm. Then, I set LSC REFL B off to the side, and placed the profiler about 128 mm away from M18 on reflection of the splitter. We measured the beam profile, and then I re-placed LSC REFL B, this time at a distance of 128 mm to M18.

elenna.capote@LIGO.ORG - 12:26, Monday 05 May 2025 (84256)

I have attached a very rough drawing of the REFL path and the locations where we made beam profile measurements. Each X on this drawing marks a beam profile measurement location. I also marked the Xs with letters A-G.

The measurements Keita reports above correspond the measurements C, D, E and F on this drawing. The difference between E and F, which is not depicted in my drawing, is a different placement of the temporary steering mirror relative to the sled.

 We still need to report details on the measurements for locations A, B, and G.

Images attached to this comment
keita.kawabe@LIGO.ORG - 15:31, Monday 05 May 2025 (84257)

Beam size upstream of the WFS sled

Unfortunately this is preliminary.

We measured the beam size at 4 different location upstream of the WFS sled marked as A, B, C and D. D data cannot be used as there's no data/picture of D locaiton but that's fine as far as position A data is good. Unfortunately, though, the position A horizontal width looks narrower than it really is (2nd attachment). The beam might be clipping in the nanoscan aperture or there might be a ghost beam or bright background light in the Region Of Interest (ROI), or ROI is defined poorly, effectively clipping the beam. Must remeasure.

LSC REFL_B (and therefore REFL_A) beam radius is ~0.1mm, which is tinier than my preference, the diode is 3mm (in diameter) so the beam could be larger.  The diodes are placed close to the focus of the lens upstream (number 18 in a circle in the first attachment) so the beam won't move when the beam position moves on that lens. Moving away from that position will be fine as far as the deviation is much smaller than the focal length (~200mm). Rayleigh range is like 3cm or maybe smaller (0.1mm waist -> RR=10*pi mm),  it should be easy to double the beam size by moving the sensors away from the lens by a couple inches. We'll do this after POP alignment.

Location Distance from the closest component wx [um] std(wx) [um] wy [um] std(wy) [um]
A

225mm downstream of M2, hard to measure the position accurately.

Nanoscan wx*2 number looks narrower than it really is. Must remeasure.

2683.6/2
14.2/2
3562.9/2 4.4/2
B 303mm downstream of M5. 3936.9/2 64.5/2 3960.8/2 83.1/2
C

128mm downstream of the last 50:50 for LSC REFL_A/B. LSC-REFL_B location (tentative).

211.6/2 12.7/2 247.3/2 4.5/2
D Exact position unknown, between RM1 and M2, less than 1400 downstream of M2. Beam size numbers look good. 3703.6/2 3.0/2 4332.8/2 4.5/2

After everything is done we'll make a good measurement of distances between everything by either using a long/short ruler (preferred) or counting bolt holes or both.

Images attached to this comment
camilla.compton@LIGO.ORG - 09:04, Tuesday 06 May 2025 (84263)

Yesterday Betsy and I measured the distances between these optics:

  • 856mm from RM1 to RM2, measured 827mm between the front D1000767 plate structure of each SUS and Don and Rahul measured on D1001396 the horizontal distance between plate and optic is 14.3mm for each RM1 and RM2.
  • 895mm from RM2 to M5: Front of RM2 mirror to HR of M5 mirror
  • 758mm from M5 to M6 (BS): HR of M5 mirror to front HR of M6 BS (-X side of M6)
  • 468mm from M6 (BS) to 2" Lens on SLED: front HR of M6 BS (-X side of M6) to front side of 2" lens (0.5" thick lens holder, from siskiyou FOH 2", so add 0.25" if you want center of lens).
elenna.capote@LIGO.ORG - 17:24, Tuesday 06 May 2025 (84266)

Camilla and I went back out today to redo the measurements at the locations labeled "A" and "D" in Keita's diagram. This table reports the D4sigma values, like Keita's tables above.

We forgot that we had left ITMX aligned, so the original measurements in this alog are no good. Keita and I remeasured these again today (May 6) and I am updating the table below with the new data. We also got two more measurements in new locations that are not indicated in Keita's diagram.

Location Distance from closest component wx [um] std wx [um] wy [um] std wy [um]
A 238 mm (+- 3 mm) downstream of M2 (nanoscan image) 4038.9/2 1.4/2 4206.2/2 3.3/2
D 314 mm (+- 3 mm) upstream of RM1 (measured from nanoscan front to metal ring around the RM, the mirror surface may be set back from the ring by another 1mm or so, hard to tell) (nanoscan image) 3950.6/2 2.8/2 4315.0/2 2.6/2
New location, after RM2 374 mm upstream of M5 (nanoscan image) 2304.8/2 36.3/2 2335.9/2 37.1/2
New location, between RM1 and RM2 345 mm upstream of RM2 (measured from nanoscan front to metal ring around RM) (nanoscan image) 1650.9/2 2.3/2 1805.1/2 3.2/2

Leaving this older comment: It is difficult to measure these distances well with the ruler, so I would guesstimate error bars of a few mm on each distance measurement reported here.

Some new notes: when we reduce teh purge air flow, the measurements become much more stable and there is no need to "cherry pick" data as Keita discussed in earlier comments. Also, I think we have finally managed to tighten the screws on the nanoscan posts enough that it doesn't slide around anymore.

Images attached to this comment
H1 DetChar (DetChar)
ansel.neunzert@LIGO.ORG - posted 15:08, Thursday 23 January 2025 - last comment - 15:07, Wednesday 07 May 2025(82320)
Relationship between violin mode height and narrow spectral artifact contamination, revisited

Summary

Q: What is the relationship between the strength of violin mode ring-ups and the number of narrow spectral artifacts around the violin modes? Is there a clear cut-off at which the contamination begins?

A: The answer depends on the time period analyzed. There was an unusual time period spanning from mid-June 2023 through (very approximately) August 2023. During this time period, the number lines during ring-ups was much greater than in the rest of O4, and the appearance of the contamination may have begun at lower violin mode amplitudes.

What to keep in mind when looking at the plots.

1. These plots use the Fscan line count in a 200-Hz band around each violin mode region, which is a pretty rough metric, and not good for picking up small variations in the line count. It's the best we've got at the moment, and it can show big-picture changes. But on some days, contamination is present, but only in the form of ~10 narrow lines symmetrically arranged around a high violin mode peak. (Example in the last figure, fig 7) This small jump in the line count may not show up above the usual fluctuations. However, in aggregate (over all of O4) this phenomenon does become an issue for CW data quality. These "slight contamination" cases are also particularly important for answering the question "at what violin mode amplitude does the contamination just start to emerge?" In short, we shouldn't put too much faith in this method for locating a cut-off problematic violin mode height.

2. The violin modes may not be the only factor in play, so we shouldn't necessarily expect a very clear trend. For example, consider alog 79825 . This alog showed that at least some of the contamination lines are violin mode + calibration line intermodulations. Some of them (the weaker ones) disappeared below the rest of the noise when the violin mode amplitude decreased. Others (the stronger ones) remained visible at reduced amplitude. Both clusters vanished when the temporary calibration lines were off. If we asked the question "How high do the violin modes need to be...?" using just these two clusters, we'd get different apparent answers depending on (a) which cluster we chose to track (weak or strong), and (b) which time period we selected (calibration lines on or off). This is because at least some of the contamination is dependent on the presence & strength of a second line, not a violin mode.

Looking at the data

First, let's take a look at a simple scatter plot of the violin mode height vs the number of lines identified. This is figure 1. It's essentially an updated version of the scatter plots in alog 71501. It looks like there's a change around 1e-39 on the horizontal axis (which corresponds to peak violin mode height).

However, when we add color-coding by date (figure 2), new features can be seen. There's a shift at the left side of the plot, and an unusual group of high-line-count points in early O4.

The shift at the left side of the plot is likely due to an unrelated data quality issue: combs in the band of interest. In particular, the 9.5 Hz comb, which was identified and removed mid O4, contributes to the line count. Once we subtract out the number of lines which were identified as being part of a comb, this shift disappears (figure 3).

With the distracting factor of comb counts removed, we still need to understand the high-line-count time period. This is more interesting. I've broken the data down into three epochs: start of O4 - June 21, 2023 (figure 4); June 21, 2023 - Sept 1 2023 (figure 5); and Sept 1 2023 - present (figure 6). As shown in the plots, the middle epoch seems notably different from the others.

These dates are highly approximate. The violin mode ring-ups are intermittent, so it's not possible to pinpoint the changes sharply. The Sept 1 date is just the month boundary that seemed to best differentiate between the unusual time period and the rest of O4. The June 21 date is somewhat less arbitrary; it's the date on which the input power was brought back to 60W (alog 70648), which seems a bit suspicious. Note that, with this data set, I can't actually differentiate between a change on June 21 and a change (say) on June 15th, so please don't be misled by the specificity of the selected boundary.

Images attached to this report
Comments related to this report
kiet.pham@LIGO.ORG - 13:53, Friday 18 April 2025 (83997)DetChar

Kiet, Sheila

We recently started looking into the whether nonlinearity of the ADC can contribute to this by looking at the ADC range that we were using in O4a. 

They are showed in the H1:OMC-DCPD_A_WINDOW_{MAX,MIN} that sum the 4 DC photodiodes (DCPD). They are 18 bits DCPD, so that channel should saturate at 4* 2^17 ~520,000 counts. 

Now there are instances that agree with Ansel report when there are violin mode ring up that we can see a shift in the count baseline.

Jun 29 - Jun 30, 2023 when the baseline seems to shift up and stay there for >1 months, Detchar summary page show significant higher violin mode ring up in the usual 500-520Hz region as well as the nearby region (480-500 Hz) 

Oct 9, 2023 is when the temporary calibration lines are turned off 72096, the down shift happened right after the lines are off (after 16:40 UTC)

During this period, we were using a~5% of the ADC range (difference between max and min channel divided by the total range - 500,000 to 500,000 counts), and it went down to ~2.5 % once the shift happenned on Oct 9, 2023. We want to do something similar with Livingston, using the L1:IOP-LSC0_SAT_CHECK_DCPD_{A,B}_{MAX,MIN} channels to see the ADC range and the typical count values of those channels.

Another thing for us to maybe take a closer look is the baseline count value increase around May 03 2023. There was a change to the DCPC total photocurrent during that time (69358). Maybe worth checking if there is violin mode contaimination during the period before that. 


 

Images attached to this comment
kiet.pham@LIGO.ORG - 10:28, Tuesday 29 April 2025 (84136)DetChar

Kiet, Sheila

More updates related to the ADC range investigation: 

  • ADC ranges comparison between Hanford and Livingston: 
    • At Livingston we are using the L1:IOP-LSC0_SAT_CHECK_DCPD_{A,B}_{MAX,MIN} channels, which were not turned on until 1398035799; these channels saturate at 2^17 counts. 
    • At Hanford we are using the H1:OMC-DCPD_A_WINDOW_{MAX, MIN}, these channels saturate at 4 * 2^17 counts as they are sum over 4 DCPD
    • We looked through the data in Feb 2025 when the violin modes at Livingston were somewhat higher than usual. 
    • The range being used in compatible with Hanford (2-4%), and the count values are also similar as LHO counts/4 and LLO counts are in +- 5000 counts of each others.
  • Comparing the comtaimination before the DARM offset change in ER15:
    • We saw hint of contaimination even before the change (Using Fscan spectrum of May 2nd) 

Further points + investigations:

  • Ansel pointed out that it was odd to have a significant shift in the baseline count value + range when the temporary calibration lines turned off as these calibration lines were not that different in height than other calibration lines (see plot in the Alog 83997)
  • Joseph from LLO gave us a spectrum comparison between H1, L1 raw ADC count, there is a notable difference in the higher order violin modes
    • To do: looking to periods of when the both (1st and 2nd) violin modes ring up and periods of only the first violin mode ring up to see the contaimination caused by the 2nd mode or higher down mixing
  • Evan pointed out that during the period of high contaimination (June 30th - Aug 9th; 2023), the range stayed between 7 - 20%; and LHO in general seemed to have higher rate of saturation + intermitten increase of the ADC range than LLO. 
    • To do: Selecting the periods of stable ADC range in LHO data, and run the average spectrum over those periods to see the level of contaimination, assessing the contribution of the periods with increase ADC range. 

       
Images attached to this comment
kiet.pham@LIGO.ORG - 15:07, Wednesday 07 May 2025 (84305)DetChar

Kiet, Sheila

Following up on the investigation into potential intermixing between higher-order violin modes down to the ~500 Hz region:

The Fscan team compiled a detailed summary of the daily maximum peak height (log10 of peak height above noise in the first violin mode region) for the violin modes near 500 Hz (v1) and 1000 Hz (v2). They also tracked line counts in the corresponding frequency bands: 400–600 Hz for v1 and 900–1000 Hz for v2. This data is available in the Google spreadsheet (LIGO credentials required).

  • We identified dates when both violin modes were elevated (n1_height > 7; n2_height > 8) and when only the fundamental mode was elevated (n1_height > 7; n2_height < 8). For each case, we computed average PSDs using an FFT length of 1800 s. The study period spans from August 10, 2023, to January 14, 2025, starting when ADC counts stabilized after the temporary calibration lines at 24.4 and 24.5 Hz were turned off (see alog  72096)
    • The psds comparisons is shown in vmodes_psds_comparison.png.
      • Note that the number of averages differs between the cases; there are significantly fewer days with only v1 elevated, which explains why the [v1 high, v2 low] spectrum appears noisier in some regions. However, similar features are still present in the [v1 high, v2 high] case.
      • Notably, there appears to be more spectral content in the 450–550 Hz range when both modes are elevated, with certain lines showing significant power (highlighted in green).

 

  • Daily Fscan data around the violin modes is summarized in Pairwise_scatter_plots.png , where n1_height and n2_height are the max peak heights of v1 and v2, and n1_count and n2_count are the corresponding line counts. There appears to be a threshold in violin mode amplitude beyond which line counts increase (based on {n1_height, n2_height} vs. {n1_count, n2_count} trends).
  • We also ploted how n1_count varies with n2_height when n1_height is high in n1_count_vs_n2_height_when_v1_high.png

Next: We plan to further investigate the lines that appear when both modes are high, the goal is to identify possible intermodulation products using the recorded peak frequencies of the violin modes.

Images attached to this comment
Displaying reports 1081-1100 of 82965.Go to page Start 51 52 53 54 55 56 57 58 59 End