TITLE: 09/25 Day Shift: 1430-2330 UTC (0730-1630 PST), all times posted in UTC
STATE of H1: Observing at 154Mpc
OUTGOING OPERATOR: Tony
CURRENT ENVIRONMENT:
SEI_ENV state: CALM
Wind: 5mph Gusts, 3mph 3min avg
Primary useism: 0.02 μm/s
Secondary useism: 0.16 μm/s
QUICK SUMMARY: H1 has been locked for 2.5 hours. Looks like the lockloss last night was from yet another earthquake, this time likely a M5.7 from Mexico. Calibration and commissioning time scheduled today from 15:30 to 20:00 UTC.
TITLE: 09/25 Eve Shift: 2330-0500 UTC (1630-2200 PST), all times posted in UTC
STATE of H1: Earthquake
INCOMING OPERATOR: Tony
SHIFT SUMMARY:
IFO is in LOCKING in LOCKING_ARMS_GREEN
Very calm shift with a fully auto Lock acquisition from the M6.1 Venezuela EQ this afternoon (alog 87128).
We had one Lockloss due to another earthquake from Venezuela, M6.3. Lockloss alog 87134. I turned on the ASC High Gain before it happened, but we understandably still lost lock.
I've just set IFO to auto-lock, which it will do once the EQ rings down (I expect it to take another 30 mins to get to a lockable state).
LOG:
| Start Time | System | Name | Location | Lazer_Haz | Task | Time End |
|---|---|---|---|---|---|---|
| 16:48 | VAC | Pump | LVEA | N | AIP pumping on HAM6 | 23:43 |
| 22:10 | ISC | Jennie | Opt Lab | Local | ISS array work | 23:31 |
| 22:54 | VAC | Gerardo | LVEA | N | Checking on AIP | 23:16 |
Lockloss due to another EQ from Venezuela - 6.3 Magnitude
I previously reported results from DARM offset step tests in October and last week. The overall goal here is to figure out what overall effect the power outage has had on the IFO, especially since we have lost 1% optical gain. For example, see this alog about comparing modulation depth tests.
However, after looking closer at the DARM offset measurement results, and especially investigating the effect of whether leaving the OMC ASC on during the measurement matters, I would like to revise the previously reported measurement results.
Some background: this test aims to measure the contrast defect light by changing the DARM offset in mA while injecting strong PCAL lines to capture the DARM optical gain. We expect the relationship between optical gain (mW/pm) and DARM offset power (mW) to be quadratic. We assume by fitting the data to a parabola, the resulting y-intercept can tell us the contrast defect light on the DCPDs, if we had zero DARM offset.
Some considerations:
First question, does leaving the OMC ASC on matter? Jennie has been doing a lot of these measurements, and her experience has been, yes, because the OMC ASC offsets are set at one DARM offset and are not reset at each different offset. We do not expect the OMC alignment to drift significantly during the 15 minutes of this measurement, so the best practice has been to turn the OMC ASC OFF for the measurement. However, I didn't do it the first time I measured last week, but I got significantly better results (using Craig's code) with the ASC ON than I did with the ASC OFF.
Second question, what is the best way to fit this data? This apparently has been a question for a long time, see Gabriele's comment to 30573 in 2016. I believe the model I described above suggests that the vertex of the resulting parabola should be centered around x=0, that is zero optical gain. However, I discovered this week that Craig's code, which I used to report the results in October and last week (see first two links), fits a nonzero vertex (i.e. y = a(x-x0)^2 + b instead of y = ax^2 + b). Of course, that will change the answer significantly! I don't understand the mechanism that would cause us to move the center of this vertex away from x=0, so I refitting the data to follow y = ax^2 + b.
Results:
I am revising the fitted contrast defects to be:
| Frequency | October 2024 (OMC ASC on) [mW] | September 2025 (OMC ASC on) [mW] | September 2025 (OMC ASC off) [mW] |
| 255.0 Hz | 1.088 +- 0.033 | 1.219 +- 0.032 | 1.256 +- 0.005 |
| 410.3 Hz | 1.086 +- 0.030 | 1.214 +- 0.042 | 1.240 +- 0.008 |
Conclusions:
This measurement shows that the contrast defect is higher now than it was in October. I don't know if we can attribute all of this to the power outage; we lost 1% optical gain comparing kappa c from before and after the vent, giving us at least 2% less optical gain between October and now. It also shows that there is perhaps a small effect related to the OMC ASC being off: slightly higher contrast and a better overall fit.
This also increases our upper limit on the possible homodyne angle- in October I stated that the homodyne angle upper limit would be about 7 degrees, but with this revised value it should have been more like 8.8 degrees. Now it is 9.3 degrees.
TITLE: 09/24 Day Shift: 1430-2330 UTC (0730-1630 PST), all times posted in UTC
STATE of H1: Earthquake
INCOMING OPERATOR: Ibrahim
SHIFT SUMMARY: Very quiet day of H1 observing until an earthquake caused a lockloss in the late afternoon. Once the ground motion rings down, Ibrahim will start relocking.
LOG:
| Start Time | System | Name | Location | Lazer_Haz | Task | Time End |
|---|---|---|---|---|---|---|
| 16:48 | VAC | Pump | LVEA | N | AIP pumping on HAM6 | 23:43 |
| 16:10 | FAC | Kim | Opt Lab | N | Technical cleaning | 16:26 |
| 17:02 | SEI | Jim, Mitchell | Opt Lab | N | Taking parts out of clean bags | 17:30 |
| 17:58 | CAL | Tony | PCal Lab | N | Looking for a cable | 18:34 |
| 20:08 | ISC | Jennie | Opt Lab | Local | ISS array work | 21:00 |
| 20:41 | CAL | Tony | PCAL Lab | N | Dropping off supplies | 21:25 |
| 20:49 | AOS | Betsy | Opt Lab | N | Checking bag contents | 21:16 |
| 21:26 | SPI | Jeff, RyanS | Opt Lab | N | Looking at SPI optics | 21:41 |
| 22:10 | ISC | Jennie | Opt Lab | Local | ISS array work | 00:09 |
| 22:54 | VAC | Gerardo | LVEA | N | Checking on AIP | 23:16 |
TITLE: 09/24 Eve Shift: 2330-0500 UTC (1630-2200 PST), all times posted in UTC
STATE of H1: Earthquake
OUTGOING OPERATOR: Ryan S
CURRENT ENVIRONMENT:
SEI_ENV state: LARGE_EQ
Wind: 6mph Gusts, 3mph 3min avg
Primary useism: 0.69 μm/s
Secondary useism: 0.18 μm/s
QUICK SUMMARY:
IFO is DOWN due to EARTHQUAKE
Just waiting for this 6.2 EQ from Venezuela to pass through and then back to LOCKING
Lockloss @ 22:31 UTC after 11.5 hours locked - link to lockloss tool
Caused by S-waves from a M6.1 EQ out of Venezuela. Happened less than a minute after the start of the transition to EQ mode, so I didn't get a chance to try the ASC Hi-Gain feature to ride through it (although, in this case I don't suspect it would have helped). Holding H1 in DOWN as it's still 10 minutes before the R-waves are supposed to hit.
Wed Sep 24 10:08:21 2025 INFO: Fill completed in 8min 18secs
Gerardo confirmed a good fill curbside.
TITLE: 09/24 Day Shift: 1430-2330 UTC (0730-1630 PST), all times posted in UTC
STATE of H1: Observing at 153Mpc
OUTGOING OPERATOR: Tony
CURRENT ENVIRONMENT:
SEI_ENV state: CALM
Wind: 2mph Gusts, 0mph 3min avg
Primary useism: 0.01 μm/s
Secondary useism: 0.14 μm/s
QUICK SUMMARY: H1 has been locked for 3.5 hours. One lockloss overnight with an unknown cause, but relocking seems to have gone without issue. We will have a short planned commissioning window from 16:00 to 17:00 UTC.
The hour of commissioning time did not happen today due to calibration report issues at LLO. This has been added onto tomorrow's regularly scheduled commissioning time, so we will have a total of 4.5 hours (with the first 30 minutes dedicated to calibration sweeps, as usual).
Back in March, I ran a modulation depth test, with several goals in mind related to providing useful calibration information for modeling. That alog is still unfortunately sitting in my drafts. However, I was able to use the results to make a side-by-side comparison with a modulation depth test Sheila ran last Thursday, after the power outage. We are still trying to understand what the overall effect of the power outage was on the IFO. Namely, we have lost 1% of optical gain, 86964.
Some background: the modulation depth test aims to measure the fraction of carrier, 9 MHz and 45 MHz power at each port. This is done by measuring the powers at the nominal settings, and then individiually stepping the 9 and 45 MHz modulation up or down by a known value. Using a measured calibration of V/dBm and rad/V (see alog 62883), and using the bessel functions, the power fraction of each field can be measured at each diode based on how much the total diode power changes at each step. (note: I still have a to do list item to better calibrate the modulation depth in radians using OMC scan data).
Procedure: I stepped both down and up in modulation depth, resulting in 5 different measurements (nominal, 9 down, 9 up, 45 down, 45 up). In March, I stepped down by 3 dBm and up by 2 dBm, but in September we were able to step both down and up by 3 dBm. I measured for 3 minutes at each step in March, and 1 minute at each step in September.
EDIT: I realized I made an error, and the first results I report below are actually from Feb 2025. The significant difference here is that in my February measurement I only stepped down by 3 dBm for 9 and 45 MHz each, so there is less data to fit. In my March measurement, I stepped up 2 dBm and down 3 dBm, getting 5 total different measurement times. There should be very little difference in the interferometer between February and March 2025, so the differences in the results I believe are due to the fact that more points (5 versus 3) gives you a much better fit to the data. I would compare March and now for a more accurate understanding of the differences.
February results, well before power outage, only fit from 3 data points:
| Field | Input | POP | REFL | AS |
| carrier | 0.9771 | 0.9842 | 0.9419 | 0.3386 |
| 9 MHz | 0.01278 | 0.01527 | 0.02950 | 0.1787 |
| 45 MHz | 0.01003 | 0.000474 | 0.02860 | 0.4827 |
March results, before power outage, fit from 5 data points:
| Field | Input | POP | REFL | AS |
| carrier | 0.9779 | 0.9831 | 0.9164 | 0.2134 |
| 9 MHz | 0.01212 | 0.01332 | 0.04059 | 0.2761 |
| 45 MHz | 0.009687 | 0.002302 | 0.04431 | 0.5413 |
September results, after power outage (and reduction of PSL power, attenuation at IMC REFL), fit from 5 data points:
| Field | Input | POP | REFL | AS |
| carrier | 0.9783 | 0.9833 | 0.9333 | 0.3556 |
| 9 MHz | 0.01195 | 0.01325 | 0.02533 | 0.1852 |
| 45 MHz | 0.009486 | 0.001909 | 0.04199 | 0.4641 |
EDIT: Including the March results (and trusting them more than the February results) changes the conclusion. The input ratios are very similar between all three measurements. This is also true for POP where carrier and 9 ratios are concerned. 45 MHz is the hardest field to measure because the 9 and carrier are so strong at POP. There may be less 45 MHz at POP, or this is just the measurement uncertainty. At REFL, there may be an increase in carrier light now after the power outage, and there may be half as much 9 MHz as in March.
The most dramatic differences are at the AS port. Just comparing the February and March measurements, there may be considerable uncertainty in how much of each field is at AS in general. However, if we choose to believe the March results, this would suggest a significant increase in carrier at AS, and a significant decrease in 9 and 45 MHz. However, comparing February and now, the 9 MHz and carrier are nearly the same, and the 45 seems to have decreased.
Rereading this alog, I see that I should say specifically which diodes measure these powers:
Input == IM4 trans
POP == POP A LF
REFL == REFL A LF
AS = AS_C DC NSUM
Here are plots of various channels during the March and September mod depth tests. The shaded region indicates which step was being taken at the time, and the dotted line matching each shading color indicates the median of the channel at that time.
WP 12802
ECR E2400330
Modified List T2500232
The following SUS SAT Amps were upgraded per ECR E2400330. Modification improves the whitening stage to reduce ADC noise from 0.05 to 10 Hz.
| Suspension | Old | New | OSEM |
| PRM M2 | S1100080 | S1100064 | ULLLURLR |
| PRM M3 | S1100106 | S1000285 | ULLLURLR |
| SRM M2 | S1100078 | S1100095 | ULLLURLR |
| SRM M3 | S1000274 | S1100091 | ULLLURLR |
| ITMY L1 UIM | S1100070 | S1100129 | ULLLURLR |
| ITMX L1 UIM | S1100141 | S1000287 | ULLLURLR |
| BS M2 | S1000295 | S1100117 | ULLLURLR |
| ETMY UIM | S1100159 | S1100088 | ULLLURLR |
F. Clara, J. Kissel, O. Patane
Here's the characterization data and fit results for S1100064, assigned to PRM M2's ULLLURLR OSEMs. This sat amp is a UK 4CH sat amp, D0900900 / D0901284. The data was taken per methods described in T080062-v3, using the diagrammatic setup shown on PAGE 1 of the Measurement Diagrams from LHO:86807. The data was processed and fit using ${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Scripts/ plotresponse_S1100064_PRM_M2_ULLLURLR_20250917.m Explicitly, the fit to the whitening stage zero and pole, the transimpedance feedback resistor, and foton design string are Optic Stage Serial_Number Channel_Number OSEM_Name Zero_Pole_Hz R_TIA_kOhm Foton_Design PRM M2 S1100064 CH1 UL 0.0961:5.24 120.25 zpk([5.24],[0.0961],1,"n") CH2 LL 0.0950:5.18 120.25 zpk([5.18],[0.0950],1,"n") CH3 UR 0.0961:5.24 120.25 zpk([5.24],[0.0961],1,"n") CH4 LR 0.0962:5.25 120.00 zpk([5.25],[0.0962],1,"n") The attached plot and machine readable .txt file version of the above table are also found in ${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Results/ Per usual, R_TIA_kOhm is not used in the compensation filter -- but after ruling out an adjustment in the zero frequency (by zeroing the phase residual at the lowest few frequency points), I nudged the transimpedance a bit to get the magnitude scale within the ~0.25%. shown in the attached results. Any scaling like this will be accounted for instead with the absolute calibration step, i.e. Side Quest 4 from G2501621, a la what was done for PR3 and SR3 top masses in LHO:86222 and LHO:84531 respectively.
Here's the characterization data and fit results for S1000285 , assigned to PRM M3's ULLLURLR OSEMs. This sat amp is a UK 4CH sat amp, D0900900 / D0901284. The data was taken per methods described in T080062-v3, using the diagrammatic setup shown on PAGE 1 of the Measurement Diagrams from LHO:86807. The data was processed and fit using ${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Scripts/ plotresponse_S1000285_PRM_M3_ULLLURLR_20250917.m Explicitly, the fit to the whitening stage zero and pole, the transimpedance feedback resistor, and foton design string are Optic Stage Serial_Number Channel_Number OSEM_Name Zero_Pole_Hz R_TIA_kOhm Foton_Design PRM M3 S1000285 CH1 UL 0.0956:5.22 120.250 zpk([5.22],[0.0956],1,"n") CH2 LL 0.0967:5.27 120.250 zpk([5.27],[0.0967],1,"n") CH3 UR 0.0955:5.21 120.375 zpk([5.21],[0.0955],1,"n") CH4 LR 0.0950:5.18 120.375 zpk([5.18],[0.0950],1,"n") The attached plot and machine readable .txt file version of the above table are also found in ${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Results/ Per usual, R_TIA_kOhm is not used in the compensation filter -- but after ruling out an adjustment in the zero frequency (by zeroing the phase residual at the lowest few frequency points), I nudged the transimpedance a bit to get the magnitude scale within the ~0.25%. shown in the attached results. Any scaling like this will be accounted for instead with the absolute calibration step, i.e. Side Quest 4 from G2501621, a la what was done for PR3 and SR3 top masses in LHO:86222 and LHO:84531 respectively.
Here's the characterization data and fit results for S1100095 , assigned to SRM M2's ULLLURLR OSEMs. This sat amp is a UK 4CH sat amp, D0900900 / D0901284. The data was taken per methods described in T080062-v3, using the diagrammatic setup shown on PAGE 1 of the Measurement Diagrams from LHO:86807. The data was processed and fit using ${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Scripts/ plotresponse_S1100095_SRM_M2_ULLLURLR_20250917.m Explicitly, the fit to the whitening stage zero and pole, the transimpedance feedback resistor, and foton design string are Optic Stage Serial_Number Channel_Number OSEM_Name Zero_Pole_Hz R_TIA_kOhm Foton_Design SRM M2 S1100095 CH1 UL 0.0955:5.22 120 zpk([5.22],[0.0955],1,"n") CH2 LL 0.0975:5.33 120 zpk([5.33],[0.0975],1,"n") CH3 UR 0.0951:5.19 120.25 zpk([5.19],[0.0951],1,"n") CH4 LR 0.0955:5.20 120.25 zpk([5.20],[0.0955],1,"n") The attached plot and machine readable .txt file version of the above table are also found in ${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Results/ Per usual, R_TIA_kOhm is not used in the compensation filter -- but after ruling out an adjustment in the zero frequency (by zeroing the phase residual at the lowest few frequency points), I nudged the transimpedance a bit to get the magnitude scale within the ~0.25%. shown in the attached results. Any scaling like this will be accounted for instead with the absolute calibration step, i.e. Side Quest 4 from G2501621, a la what was done for PR3 and SR3 top masses in LHO:86222 and LHO:84531 respectively.
Here's the characterization data and fit results for S1100091 , assigned to SRM M3's ULLLURLR OSEMs. This sat amp is a UK 4CH sat amp, D0900900 / D0901284. The data was taken per methods described in T080062-v3, using the diagrammatic setup shown on PAGE 1 of the Measurement Diagrams from LHO:86807. The data was processed and fit using ${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Scripts/ plotresponse_S1100091_SRM_M3_ULLLURLR_20250917.m Explicitly, the fit to the whitening stage zero and pole, the transimpedance feedback resistor, and foton design string are Optic Stage Serial_Number Channel_Number OSEM_Name Zero_Pole_Hz R_TIA_kOhm Foton_Design SRM M3 S1100091 CH1 UL 0.0983:5.37 120.25 zpk([5.37],[0.0983],1,"n") CH2 LL 0.0959:5.23 120.25 zpk([5.23],[0.0959],1,"n") CH3 UR 0.0955:5.23 120.00 zpk([5.23],[0.0955],1,"n") CH4 LR 0.0957:5.24 120.00 zpk([5.24],[0.0957],1,"n") The attached plot and machine readable .txt file version of the above table are also found in ${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Results/ Per usual, R_TIA_kOhm is not used in the compensation filter -- but after ruling out an adjustment in the zero frequency (by zeroing the phase residual at the lowest few frequency points), I nudged the transimpedance a bit to get the magnitude scale within the ~0.25%. shown in the attached results. Any scaling like this will be accounted for instead with the absolute calibration step, i.e. Side Quest 4 from G2501621, a la what was done for PR3 and SR3 top masses in LHO:86222 and LHO:84531 respectively.
Here's the characterization data and fit results for S1100129 , assigned to ITMY L1 (UIM) ULLLURLR OSEMs. This sat amp is a UK 4CH sat amp, D0900900 / D0901284. The data was taken per methods described in T080062-v3, using the diagrammatic setup shown on PAGE 1 of the Measurement Diagrams from LHO:86807. The data was processed and fit using ${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Scripts/ plotresponse_S1100129_ITMY_L1_ULLLURLR_20250916.m Explicitly, the fit to the whitening stage zero and pole, the transimpedance feedback resistor, and foton design string are Optic Stage Serial_Number Channel_Number OSEM_Name Zero_Pole_Hz R_TIA_kOhm Foton_Design ITMY L1 S1100129 CH1 UL 0.0956:5.23 121.75 zpk([5.23],[0.0956],1,"n") CH2 LL 0.0966:5.28 120.00 zpk([5.28],[0.0966],1,"n") CH3 UR 0.0978:5.34 120.00 zpk([5.34],[0.0978],1,"n") CH4 LR 0.0966:5.27 120.00 zpk([5.27],[0.0966],1,"n") The attached plot and machine readable .txt file version of the above table are also found in ${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Results/ Per usual, R_TIA_kOhm is not used in the compensation filter -- but after ruling out an adjustment in the zero frequency (by zeroing the phase residual at the lowest few frequency points), I nudged the transimpedance a bit to get the magnitude scale within the ~0.25%. shown in the attached results. Any scaling like this will be accounted for instead with the absolute calibration step, i.e. Side Quest 4 from G2501621, a la what was done for PR3 and SR3 top masses in LHO:86222 and LHO:84531 respectively.
Here's the characterization data and fit results for S1100287 , assigned to ITMX L1 (UIM) ULLLURLR OSEMs. (Note the typo in Fil's main entry -- he quotes S1000287, but it's S1100287. This sat amp is a UK 4CH sat amp, D0900900 / D0901284. The data was taken per methods described in T080062-v3, using the diagrammatic setup shown on PAGE 1 of the Measurement Diagrams from LHO:86807. The data was processed and fit using ${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Scripts/ plotresponse_S1100287_ITMX_L1_ULLLURLR_20250916.m Explicitly, the fit to the whitening stage zero and pole, the transimpedance feedback resistor, and foton design string are Optic Stage Serial_Number Channel_Number OSEM_Name Zero_Pole_Hz R_TIA_kOhm Foton_Design ITMX L1 S1100287 CH1 UL 0.0958:5.23 120 zpk([5.23],[0.0958],1,"n") CH2 LL 0.0966:5.28 120 zpk([5.28],[0.0966],1,"n") CH3 UR 0.0941:5.14 120 zpk([5.14],[0.0941],1,"n") CH4 LR 0.0955:5.23 120 zpk([5.23],[0.0955],1,"n") The attached plot and machine readable .txt file version of the above table are also found in ${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Results/ Per usual, R_TIA_kOhm is not used in the compensation filter -- but after ruling out an adjustment in the zero frequency (by zeroing the phase residual at the lowest few frequency points), I nudged the transimpedance a bit to get the magnitude scale within the ~0.25%. shown in the attached results. Any scaling like this will be accounted for instead with the absolute calibration step, i.e. Side Quest 4 from G2501621, a la what was done for PR3 and SR3 top masses in LHO:86222 and LHO:84531 respectively.
Here's the characterization data and fit results for S1100117 , assigned to BS M2 ULLLURLR OSEMs. This sat amp is a UK 4CH sat amp, D0900900 / D0901284. The data was taken per methods described in T080062-v3, using the diagrammatic setup shown on PAGE 1 of the Measurement Diagrams from LHO:86807. The data was processed and fit using ${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Scripts/ plotresponse_S1100117_BS_M2_ULLLURLR_20250917.m Explicitly, the fit to the whitening stage zero and pole, the transimpedance feedback resistor, and foton design string are Optic Stage Serial_Number Channel_Number OSEM_Name Zero_Pole_Hz R_TIA_kOhm Foton_Design BS M2 S1100117 CH1 UL 0.0970:5.31 120 zpk([5.31],[0.0970],1,"n") CH2 LL 0.0975:5.33 120 zpk([5.33],[0.0975],1,"n") CH3 UR 0.0967:5.30 120 zpk([5.30],[0.0967],1,"n") CH4 LR 0.0955:5.22 120.375 zpk([5.22],[0.0955],1,"n") The attached plot and machine readable .txt file version of the above table are also found in ${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Results/ Per usual, R_TIA_kOhm is not used in the compensation filter -- but after ruling out an adjustment in the zero frequency (by zeroing the phase residual at the lowest few frequency points), I nudged the transimpedance a bit to get the magnitude scale within the ~0.25%. shown in the attached results. Any scaling like this will be accounted for instead with the absolute calibration step, i.e. Side Quest 4 from G2501621, a la what was done for PR3 and SR3 top masses in LHO:86222 and LHO:84531 respectively.
Here's the characterization data and fit results for S1100088 , assigned to ETMY L1 (UIM) ULLLURLR OSEMs. This sat amp is a UK 4CH sat amp, D0900900 / D0901284. The data was taken per methods described in T080062-v3, using the diagrammatic setup shown on PAGE 1 of the Measurement Diagrams from LHO:86807. The data was processed and fit using ${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Scripts/ plotresponse_S1100088_ETMY_L1_ULLLURLR_20250916.m Explicitly, the fit to the whitening stage zero and pole, the transimpedance feedback resistor, and foton design string are Optic Stage Serial_Number Channel_Number OSEM_Name Zero_Pole_Hz R_TIA_kOhm Foton_Design ETMY L1 S1100088 CH1 UL 0.0968:5.29 120 zpk([5.29],[0.0968],1,"n") CH2 LL 0.0956:5.23 120 zpk([5.23],[0.0956],1,"n") CH3 UR 0.0955:5.22 120 zpk([5.22],[0.0955],1,"n") CH4 LR 0.0959:5.24 120 zpk([5.24],[0.0959],1,"n") The attached plot and machine readable .txt file version of the above table are also found in ${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Results/ Per usual, R_TIA_kOhm is not used in the compensation filter -- but after ruling out an adjustment in the zero frequency (by zeroing the phase residual at the lowest few frequency points), I nudged the transimpedance a bit to get the magnitude scale within the ~0.25%. shown in the attached results. Any scaling like this will be accounted for instead with the absolute calibration step, i.e. Side Quest 4 from G2501621, a la what was done for PR3 and SR3 top masses in LHO:86222 and LHO:84531 respectively.
During Tuesday maintenance, we swapped the HAM6 AIP (Starcell). Note this annulus system is connected to HAM5 via the septum plate. We vented the lines with dry nitrogen and left a continuous nitrogen purge(~.3 psi) of the line during the pump swap. Nitrogen attached to HAM5 pump out port while HAM6 pump out port was left open to atmosphere.
No issues during the swap, annulus system is now pumping at both HAM5 & 6 ports with an aux cart and turbo pair. As of end of maintenance, the HAM6 cart was at ~3E-5 Torr, HAM5 cart at ~1E-4 Torr. These pumps will continue running until pressure is <1E-5 Torr at which point the ion pumps will be powered on.
Carts are placed on foam for isolation, and a piece of foam between the flex hose running up to HAM6 pump out port and HAM6 chamber. See attached pictures.
Work permit will be closed once pumps are disconnected from chambers.
Update.
IFO was out of lock due to an earthquake, I went in to the LVEA to check on the aux-carts pumping down on the annuli for HAM5 and HAM6. HAM5 aux-cart was good and pumping down on the annulus, however HAM6 aux-cart safety valve somehow managed to trip between yesterday and today, time is unknown as of now, I restored aux cart, and opened the valve. Aux-cart for HAM6 was reporting a dubious pressure number of 1.26 x 10-07 Torr.
After restoring pumping to HAM6 annulus, both aux carts are reporting more believable numbers.
(Jordan, TJ, Gerardo)
Late entry.
TJ powered ON the ion pumps over the weekend, that allowed for the pumps to reach very good vacuum pressure on the shared annuli system, then on Tuesday morning, Jordan isolated the annuli system for HAM5 and HAM6 from the mechanical pumps and turned off the aux carts.
A couple of hours later we removed the small can turbos, flex hoses and aux carts from the HAM5/6 area, to conclude the replacement of the HAM5 annulus ion pump body.