Displaying reports 14421-14440 of 89077.Go to page Start 718 719 720 721 722 723 724 725 726 End
Reports until 16:06, Tuesday 02 April 2024
LHO General
corey.gray@LIGO.ORG - posted 16:06, Tuesday 02 April 2024 (76870)
Tues DAY Ops Summary

TITLE: 04/02 Day Shift: 15:00-23:00 UTC (08:00-16:00 PST), all times posted in UTC
STATE of H1: Preventive Maintenance
INCOMING OPERATOR: Oli
SHIFT SUMMARY:

LOG:

H1 ISC (IOO)
jennifer.wright@LIGO.ORG - posted 15:04, Tuesday 02 April 2024 - last comment - 15:28, Tuesday 02 April 2024(76883)
Investigating if we have clipping through Faraday Isolator between IM2 and IM3

Jennie W, Camilla

 

We set up this measurement by mis-aligning ITMX, ITMY, BS, PRM and SRM during the maintenance period (after the PSL and IM suspensions came back up).

NB. PSL was at 2W but ISS was off.

Then we adjusted IM3 till we were centred in pitch and yaw on IM4 TRANS.

Then we searched each direction of IM1 and 2 that gave us clipping on IM4 TRAns - IE. THE nsum DROPPEED.

Then we went back to a point near clipping on IM 2 and mpved IM1 in that DOF to try and bring back the NSUM since we managed to do this for all DOFs (up and down in pitch and yaw)we assume no translation oft he beam improves NSUM so we are probably not clipping through OFI.

First image is starting sliders.

Second is sliders after we  aligned IM3.

This is ndscope trends throughout our tests. By far the largest improvement to IM4 Trans was moving IM3 and nothing we did with IM1 or 2 really improved this.

Images attached to this report
Comments related to this report
camilla.compton@LIGO.ORG - 15:28, Tuesday 02 April 2024 (76891)

Attached is how much we had to move IM2 from it's nominal (20257,-4177) before the IM4 TRANS NSUM started drop, at which piont we're asumming we are clipping the faraday. It's closet (1000-200 counts) from clipping in one direction in Yaw, but centered in Pit.

Translating the beam through the faraday in Pit and Yaw didn't give us any better NSUM values. As Jennie explains we did this by moving IM2 (Tried IM2 Pitch to 19057 and  22057 (nominal 20257) and Yaw to -5577 and -3477 (nominal -4177) and using IM1 to correcting back towards the center of the IM4_TRANS QPD.

We conulde clipping the faraday is not an issue. 

Images attached to this comment
H1 PSL
ryan.short@LIGO.ORG - posted 15:03, Tuesday 02 April 2024 (76890)
PSL PMC Alignment and Rotation Stage Calibration

This morning I tweaked up the PSL PMC alignment remotely using the picomotor-controlled mirrors with the ISS off. I was only able to get about 60mW of power increase out of the PMC, but I decreased reflected power by about 200mW, so perhaps there's some mode-matching that could be optimized in the future.

After that, I recalibrated the rotation stage using the following process:

  Power in (W) D B (Minimum power angle) C (Minimum power)
Old Values 103.151 1.990 -24.823 0.000
New Values 102.287 1.990 -24.825

0.000

Images attached to this report
Non-image files attached to this report
H1 AOS
jason.oberling@LIGO.ORG - posted 15:00, Tuesday 02 April 2024 - last comment - 10:54, Thursday 04 April 2024(76889)
FARO Progress Update, 4/2/2024

J. Oberling, R. Crouch, T. Guidry

Since the last update we have tried a couple of things.

First, we attempted to align the FARO to the BSC2 chamber directly.  We placed the FARO in the biergarten and shot the chamber-side door flanges.  We used the SMR to sweep along the outside of the +X side of the +Y door flange, and the +Y side of the +X door flange.  The shots were done as circles, which we could then place a point at circle center to represent the center of the door flange.  The hope was we could then project lines to represent the X and Y axes, and use that to align to our origin using a Plane/Axis/Center Point alignment routine.  I won't go into much detail, as it didn't work.  We used this alignment to look at PSI-6 in the biergarten, and it was off by over 1cm; we then looked at LV-35 and that one was off by even more.  I don't remember exact values as we didn't save screenshots since the devaiations were so large.  Let's say that if these deviations were true, I'm not sure how we would have a functioning, aligned, and alignable IFO.  So, that experiment didn't work, moving on.

Today we decided to try using height marks 901, 902, and 903 to set our Z axis alignment.  We looked at marks 901 and 902 with an autolevel and compared them to our BSC2 door flange scribe average we had used the water tube level to measure (alog 75974) and the listed local Z coordinate from T1100187.  Results:

Not bad, and definitely not the almost +4mm the FARO was measuring for these marks w.r.t. BTVE-1.  If we use our differential height survey from alog 75771 we can also calculate the local Z axis coordinate for mark 903, which comes out to -79.8 mm (versus the listed coordinate of -79.9 mm from T1100187), and again not the almost +4mm reported by the FARO when measured w.r.t. BTVE-1.  It's looking more and more like there was some error in the Z axis position of BTVE-1; whether that error was in setting the monument itself or in setting BSC2 w.r.t. to BTVE-1 we cannot say.  The fact of the matter is that when compared to BSC2 as it sits right now based on our water level survey, the height marks we've looked at are pretty close to their listed coordinates and BTVE-1 is definitely not.  We also happened to do the same differential height survey with BTVE-1 (also in alog 75771), so we can calculate a local Z coordinate for BTVE-1 based on our height mark 903 (which we have now tied to BSC2 Z=0).  Doing so gives a local Z axis coordinate for BTVE-1 of -1060.3 mm (Z903 - deltaZBTVE-1/903 = -79.8 - 980.5); this becomes -1060.9 mm once we convert to our global coordinate frame (which is decidedly not the -1057.2 mm all of the old documentation lists it as).

Now that we have local Z coordinates for 901, 902, 903, and BTVE-1 that have all been tied to BSC2 Z=0, we used these to set a new alignment.  First, we need global Z coordiantes for our 3 height marks.  We got these by using a autolevel to set a magnetic nest inline with the height mark to be measured.  The FARO was put into an intial alignment using BTVE-1, PSI-1, PSI-2, and PSI-6 (X and Y are generally pretty good, Z is off but we don't care yet), and we then placed the SMR on the nest we placed inline with 903.  From this we got an X/Y coordinate for the nest that we could use to calculate our global Z.  Once this was done we added 903 into the alignment routine, and removed PSI-1, PSI-2, and PSI-6 for the Z axis fit.  we then moved the FARO to a position where we could see 901 and 902 and shot them in the same way we did 903.  Ultimately, we ended up with an alignment using BTVE-1, PSI-1, PSI-2, and PSI-6 for X and Y and 901, 902. and 903 for Z.  The coordinates used for the height marks were (format is [X,Y,Z]):

Tyler has a screenshot of the results of the alignment that he'll post as a comment to this alog.  One thing to note here: We were not using the new BTVE-1 Z as part of the Z axis alignment, only marks 901, 902, and 903.  It's somewhat comforting to see the Z axis coordinate of BTVE-1 as calculated by the alignment routine as close as it is to what we think the nominal shoud be based on our water level survey of BSC2.  The PSI monuments are all out, especially PSI-2.  I'm not sure how concerning this is, as we don't trust the Z coordinates of the PSI monuments anyway.

One caveat here, all of the height marks we used are almost in a line down the Y axis, with very little deviation in the X axis coordinate (a deltaX of less than 7mm across ~25m of the Y axis).  I'm not entirely confident we're picking up the global tilt of our X axis with this configuration.  We want to take this alignment and test and refine it; first thing we can do is get some more height marks spread along the X axis to hopefully catch the global X axis tilt.  More to come!

Comments related to this report
tyler.guidry@LIGO.ORG - 10:54, Thursday 04 April 2024 (76951)


		
		
Images attached to this comment
H1 SEI
jim.warner@LIGO.ORG - posted 14:29, Tuesday 02 April 2024 (76888)
BRSX remote mass adjuster connected to X PICO A driver

The BRSX remote mass adjust pico motor was finally connected to the EX pico controller this morning. This was done with the pigtail D2400020 that Fil made, to allow the BRS to use channel 5 on the controller. Channel 5 is then sent to a DB9 that goes into the box, and connects to a DB9 soldered to the pins on a feedthru on BRSX vacuum vessel. 

I tested that we are able to run the motor from the MEDM, and made an adjustment to the mass on the BRSX to try to get us better centered. I can now recenter the BRS from my office, nice.

The pico controller screen is accessed through LSC->Picmotors->End X (X PICO A) screen. To recenter push "Enabled", push the  wait for the err codes to clear, set the step size and speed. "Sprint" is fine for the speed, for the BRS at least. The motor is, I think, 10k steps per revolution, when UW was here, we used 1250 steps per move. Today, I went +2250 steps, saw that was the wrong direction, then -5000 steps from and then a final +1000 steps to disengage the mass adjuster drive yoke. The net move was -1750 steps. This is recorded in epics as H1:SYS-MOTION_X_PICO_A_CURRENT_X_POSITION, but that seems to depend on the channels selected on the chassis.

 The buttons on the pico motor screen are kind of confusing " < " makes positive steps, and " > " is negative. This moved the BRSX driftmon channel about -20k counts, and it just barely on the bottom edge of it's readout range. I've turned the voltage on the heater down to bring it back to center, but that will take a day or two to settle out, so I've taken the BRS out of loop for now.

H1 DetChar (DetChar)
kaylahbreanne.mcgowan@LIGO.ORG - posted 14:12, Tuesday 02 April 2024 (76887)
DQ Shift Report for March 18-24, 2024
Summary (highlights) of findings during the DQ Shift

-Engineering run with ongoing commissioning
-Numerous lock-losses due to earthquakes/seismic activity
-BNS range averages between 145 and 150 Mpc. Sometimes there are fluctuations to 130 Mpc.
-Glitchgrams show a reoccuring glitch around 40 Hz that occurs a few times a day.
-The h(t) spectrogram showed sporatatic increases of amplitude relative to the median in low-frequency range. Anomalies can be associated with drops in BNS range.
-Fscans are unavailable at this time.
-A few pyCBC triggers with new SNR > 10 (short) occured 2 out of 7 days.
-PEM, ASC and LSC type channels showed often as round winners in Hveto associated with low-frequency noise around 40 Hz and 50 Hz respectively.
-New channels of interest include IMC and HPI. Channel activity can be linked to sustained winds. Also could be round winners because of shorter observation runs.



DQ Shift Report: 

https://wiki.ligo.org/DetChar/DataQuality/DQShiftLHO20240318
H1 SQZ (ISC)
victoriaa.xu@LIGO.ORG - posted 13:34, Tuesday 02 April 2024 - last comment - 10:25, Friday 05 April 2024(76885)
Unsqueezed DARM excess noise getting lower, more similar to O4a

Attachment 1 - Unsqueezed (left) vs. squeezed (right) DARMs in ER16, with respect to O4a. Like Gabriele's plots, the ratios on the bottom show a given trace / O4a reference, so ratio > 1 means more noise now than O4a.

Colors: Rainbow colors are ordered past to present: e.g. Red = O4a, ... Purple = now.

Attachment 2 - SQZ dBs after subtracting non-quantum noise, with various recent PSAMS settings. Upper subplot: Anti-SQZ, Lower subplot: SQZ. Lines are a moving average of the data. Some caveats for the subtraction: here I am using the same H1 unsqueezed quantum noise model for all subtractions; this may not be the most accurate, but I think it should be fairly close.

 

Reference alogs wth times that I used to grab GDS_STRAIN_CLEAN:

Images attached to this report
Comments related to this report
victoriaa.xu@LIGO.ORG - 10:25, Friday 05 April 2024 (76978)

Adding FDS times from the DARM comparison in 76955, and only including times with sqz/no-sqz comparable to those recent darm comparisons.

Attachment 1 - Unsqueezed (left) vs. squeezed (right) DARMs in ER16, with respect to O4a. Like Gabriele's plots, the ratios are trace / O4a reference, so ratio > 1 means more noise now than O4a.

Attachment 2 - SQZ dBs after subtracting non-quantum noise. Analysis with non-quantum noise subtraction for this week's PSAMS data in 76925 and 76949 ongoing.

Attachment 3 - Comparison to a very basic gwinc noise model (pulled from H1 NB) for squeezed darm. For example the dashed purple "Unsqueezed QN model" is (almost) what I am using for subtracting non-quantum noises.

Colors are rainbow-ordered time-wise.  Red = O4a reference.  Purple = 4/4 times from 76955 for FDS DARM comparisons.

 

Note: here the PSAMS settings are referenced in terms of strain voltages. Just pointing out that LLO has a guardian to servo PSAMS voltage to meet the "_TARGET" strain gauge voltage LLO:63403.

 

Reference alogs wth times used to grab GDS_STRAIN_CLEAN:

  • O4a references - sqz data set 74935, and Gabriele's times from e.g. 76825
  • 76434 - 3/17, PSAMS 120/120
  • 76540 - 3/20-3/25, PSAMS 150/90
          --> corresponding SQZ data set saw ~10-15% higher frequency-dependent losses at 100 Hz vs. kHz unlike earlier Mar16, possibly due to PSAMS settings following 76507
  • 7675776777 - 3/27-29, PSAMS 100/100, since SQZ ZM alignment guardian
  • 76872 - 4/2, PSAMS 100/100 - opportunistic NO SQZ ~1hour
  • 76955 - 4/4, DARM comparison
Images attached to this comment
H1 ISC (ISC)
marc.pirello@LIGO.ORG - posted 12:59, Tuesday 02 April 2024 - last comment - 09:36, Tuesday 09 April 2024(76884)
H1:IMC-REFL_A_DEMOD_LOMON out of range

This could be a bad readback, or real, we are not sure.  The signal should be 23dB, but the readback is showing 16.8dB on NDSCOPE and has been trending down for a while, it is throwing an error flag in beckhoff.  This signal is located in ISC-R1 slot U16 on the left side.  The LO signal is 9MHz and is fed through a delay line chassis, there are also baluns, cabling, etc...  We were not able to determine the source of the readback error before noon, so this will have to wait until a maintenance window.

M. Pirello, F. Mera, D. Sigg

Comments related to this report
daniel.sigg@LIGO.ORG - 09:36, Tuesday 09 April 2024 (77050)

Measured RF power at the

  • Delay line input = 14.4 dBm
  • Demod input = 11.2 dBm

The analog readback of the power detector was 5.89V at the front panel. This has been confirmed to be the same as the slow controls readback, which corresponds to 23.3dBm.

Seems like disconnecting and connecting the LO cable" fixed" the problem. Flaky cable and/or connector?

Images attached to this comment
H1 SUS (SUS)
rahul.kumar@LIGO.ORG - posted 12:53, Tuesday 02 April 2024 (76881)
ETMX ESD bias voltage sign flipped for 3hrs this morning

This morning I flipped the sign (after talking to the Operator) of the ETMX ESD lock bias voltage from -450V to +450V for 3hrs to reduce the charge build up in the optic.

After 3hrs I reversed the sign and set it to nominal.

ndscope trends are attached below.

Images attached to this report
H1 PEM (PEM)
marc.pirello@LIGO.ORG - posted 12:51, Tuesday 02 April 2024 (76882)
EX Accelerometer Testing

Per WP11795 I looked at the EX Accelerometer Power Conditioner based on what Robert said about one of the channels not functioning.  Visually, there is no power on channel 8, the power good light is off.  We moved the signal to channel 9 and will go back to fix channel 8 when time permits.  This channel was working when the chassis was installed.

Images attached to this report
H1 General
camilla.compton@LIGO.ORG - posted 12:42, Tuesday 02 April 2024 (76880)
LVEA Swept

Lights off, paging system unplugged, WAP stayed off. Some unused extension cords unplugged, genie lift charging block unplugged. Turned off SQZ laptop and monitor. 

Faro is still plugged in (expect it is staying warmed up to be used next week) 
HAM3 dust monitor has a fairly long extension cable.
Some of Robert's PEM equipment is still plugged in, photo attached.

All else looked good, followed T1500386

Images attached to this report
H1 CDS (CDS, PSL)
patrick.thomas@LIGO.ORG - posted 12:09, Tuesday 02 April 2024 - last comment - 16:26, Tuesday 02 April 2024(76879)
Updated PSL Beckhoff PLC
WP 11799.

Patrick T., Ryan S.

The PSL Beckhoff PLC has been updated to match the changes made at LLO to add a watchdog channel. This is the code from 'TwinCAT Projekt_LIGO_240124.zip' in https://dcc.ligo.org/E2100235. The now running version is in git at https://git.ligo.org/cds/ifo/beckhoff/lho-psl and commit 9b6ad86cb26660b136242b334bd4831a683f1c5b. It appears there was also a change to fix a bug where toggling the noise eater tripped the NPRO. Ryan verified that this now appears to be fixed. I also removed the ${IFO}.PSL.LASER.MAIN.SHUTTER alias of the bEnable_Shutter2 variable, which according to LLO is not being used. The channel was not being exported to EPICS.

There were two minor issues. The first was that the version of a couple of the terminals found from a scan of the EtherCAT bus did not match what was in the IO configuration, and this appeared to be the cause of an INIT error for a couple of terminals (presumably those) when the system was put into run mode. This was fixed by changing the IO configured versions to match the scanned versions. The other was that I forgot to comment out the inversion of the signal from laser PD1, which was not realized until we started everything and were wondering why the power was reading as negative for that PD. I fixed that in a new commit and restarted the code again.
Comments related to this report
ryan.short@LIGO.ORG - 16:26, Tuesday 02 April 2024 (76896)

The new pump diode watchdog will trip off the system in the event any of the diodes drops below a set percentage, much like the already existing watchdogs on the output of the NPRO and both amplifiers. I've set this trip level at 80% and enabled the WD. I also added an indicator to the PSL overview medm screen showing the WD is enabled (H1:PSL-LASER_PDWD).

Images attached to this comment
LHO VE
david.barker@LIGO.ORG - posted 11:00, Tuesday 02 April 2024 (76878)
Tue CP1 Fill

Tue Apr 02 10:06:37 2024 INFO: Fill completed in 6min 34secs

Jordan confirmed a good fill curbside

Images attached to this report
H1 SEI (CDS)
erik.vonreis@LIGO.ORG - posted 10:15, Tuesday 02 April 2024 (76877)
Picket Fence updated

Picket Fence was updated.  The only major change is that one channel "HLID" was dropped for another, "HWUT".

H1 PSL
ryan.short@LIGO.ORG - posted 09:08, Tuesday 02 April 2024 (76873)
PSL Beckhoff Settings Table

We have an upgrade slated today for the PSL Beckhoff computer; I captured a screenshot of the settings table in case it gets overwritten and am posting it here for posterity.

Images attached to this report
H1 SQZ (OpsInfo)
victoriaa.xu@LIGO.ORG - posted 08:39, Tuesday 02 April 2024 - last comment - 10:12, Tuesday 02 April 2024(76868)
Resolving SQZ locking issues last night/today

Camilla, Vicky

Screenshot of how we found the SQZ_FC and SQZ_OPO this AM. Two separate issues.

Only to-do might be fixing the SQZ_FC issue in FIND_IR. I hope the SQZ_OPO_LR issue is solved, but we will watch out for it.

More description of the issues from this morning/last night and how to fix it:

Images attached to this report
Comments related to this report
victoriaa.xu@LIGO.ORG - 09:06, Tuesday 02 April 2024 (76872)

Conveniently there are some nicer glitch-free NO SQZ and FDS times:

NO SQZ: 1396084368 - 1396087119 (~46 min)

FDS: 1396072807 - 1396075570 (~46 min)

Images attached to this comment
camilla.compton@LIGO.ORG - 10:12, Tuesday 02 April 2024 (76874)

Eric, Camila

The inital issue that stopped the IFO observing at 2024/04/02 08:05 UTC is that the OPO unlocked. It couldn't relock as the OPO PZT1 wasn't scanning the correct range, Vicky has now fixed this (above).
When OPO unlocked, SQZ_MANAGER jumped to LOCK_OPO (#28).
During the time the OPO was trying to relock, the SQZ_FC was still requested to IR_LOCKED. It's not possible for FC to lock without OPO locked, it tried for ~30minutes before stopping/getting caught on cdu.getdata.
 
Eric ad I added a  turn_off_fc() function in all SQZ_MANAGER states below LOCK_FC, this will take SQZ_FC to DOWN if it's not already in DOWN on FC_ MISALIGNED.
 
We also went ahead and used the from timeout_utils import call_with_timeout fucntion. We will watch on relocking that there is no issues with this. 
Images attached to this comment
H1 SUS (CDS, ISC, SQZ, SUS)
jeffrey.kissel@LIGO.ORG - posted 15:29, Monday 01 April 2024 - last comment - 10:04, Tuesday 02 April 2024(76850)
Model Prep for HAM Doubles and Singles Watchdog Upgrade Complete
J. Kissel
ECR E1700387
IIET 9392
WP 11743

Here, this week we tackle the last of the suspension types that need suspension watchdog upgrades after having finished the End Station, Corner BSC, and HAM Triple suspensions this past month (LHO:76269, LHO:76545, and LHO:76712, respectively).

WP 11797
While doing so, I'm also going to address the OMC ASC routing bugs found in the h1susifoout top-level model and its OMCS_MASTER library part (which used to live in h1susomc.mdl), and fix those too.

As such, I'm doing a linear combination of the following changes:
    (1) In library part, removed sending watchdog out to top level
    (2) In library part, removed ODC parts from DAMP and LOCK banks
    (3) In library part, removed never-usered software shutter system
    (4) In library part, upgraded watchdog trigger generator to actual functional RMS
    (5) At top level, removed USER DACKILL
    (6) At top level, removed wires to ground (i.e. things never used) at top-level for WD, SHUTTER, and ODC outputs and SHUTTER input
    (7) At top level, cleaned up receiving of ISC ASC signals (see discussion in LHO:76842)

Below is the list of models, their respective library parts, and the optics in that model that this impacts, followed by a call out of which of the above changes happened
h1susim       HSSS_MASTER    IM1, IM2, IM3, IM4      (1), (2), (4), (5), (6)
h1sushtts     HSSS_FF_MASTER RM1, RM2                (1), (2), (3), (4), (6)
h1susifoout   OMCS_MASTER    OMC                     (1), (2), (4), (5), (6), (7)
              OFIS_MASTER    OFI                     (1), (2), (4), (6)
              HSSS_MASTER    OM1, OM2, OM3           (1), (2), (4), (5), (6)
h1sussqzout   HSSS_MASTER    ZM4, ZM5, ZM6           (1), (2), (4), (5), (6)
h1sussqzin    HDDS_MASTER    ZM1, ZM3                (1), (4)
              OPOS_MASTER    OPO                     (1), (2), (4), (6)
              HSSS_MASTER    ZM2                     (1), (2), (4), (5), (6)

I've test-compiled all of the above listed models, so they're ready for re-compile, install, and restart of the front-end process tomorrow morning.
All model changes have been committed to the userapps repo, either under
    /opt/rtcds/userapps/release/sus/h1/models/
    /opt/rtcds/userapps/release/sus/common/models/
where the above mentioned models live.

Screenshots describing the changes will be posted in the comments.
Comments related to this report
jeffrey.kissel@LIGO.ORG - 16:19, Monday 01 April 2024 (76856)CDS, ISC, SUS
To address the removal of the transverse, vertical, and roll possibility and removing the "bad route" of the L, P, and Y OMC ASC control, I've done the following:
    (1) At the top level of the h1susifoout model, I've removed the ingestion of the T, V, and R IPCs. These still get send out into the dolphin IPC fabric from the h1omc model, because I didn't want to mess with that model. This means that if someone ever tries to use the OMC-ASC_DOF2TT matrix elements for "SUS V, SUS R or SUS T", the output of those elements of that matrix will go into nothing, essentially terminated.

        See before vs. after screenshots of the bottom left corner of the h1susifoout.mdl top level IPC area.
 
    (2) I've modified the OMCS_MASTER model to completely delete the "OMC_M1_ASC" bank, and removed all simulink inputs that landed in that bank from the top level h1susifoout model, to the top level of the OMCS_MASTER model, to the M1 stage of the OMCS_MASTER.

        See before vs. after area surrounding the call to the OMCS_MASTER.mdl in the h1susifoout.mdl top level model, 

        And then drilling down into the OMCS_MASTER library part, see
            (i) before and after, the top level of the OMCS_MASTER library part.

            (ii) before vs. after, the M1 stage of the OMCS_MASTER library part.

    (3) Because the OMC ASC alignment input channels are stored inthe frames at 256, I had to update the DAQ channel list to switch to using the M1_LOCK banks rather than the M1_ASC banks. Note, there are therefore 3 less channels stored in the frames.

         This is seen in (2)(i) before vs. after cited above.

    (4) While there, I also deleted other never-used / impossible to use infrastructure that I installed early on in the 2010s because it was easier to copy and past than to customize. Now that we've gone through 4 observing runs, and there's no hardware to support it, I'm deleting the infrastructure to clean things up (lest it gets copied forward again by some future me). This includes
        (a) The "LKIN" parts, which are originally designed for demodulating an angular dither signal in order to diagonalize angular drive. But, there are only two OSEMs for both pitch and yaw on the OMCS's OSEM configuration, so the "coil balancing" methods, using the lock-in just will not work, since we can't drive "pringle." DELETED.
        (b) Many other SUS that have right-cylinders for optics, which have a center of rotational actuation that we're trying to align to a geometric rotational center in order to minimize angle to length coupling by "dithering." Sometimes you might hear the gains applied to the actuation basis called the "A2L" gains. For the OMCS, the beam is coming in at the edge of a square breadboard that's suspended beneath the actuators, of which again, there're only two. This, like the LKIN system, the DITHER system just doesn't make any physical sense.
     Because of the reasons outlined in (a) and (b), the inputs to the M1 stage of the OMCS_MASTER library part have always been terminated. That means that the inputs from ISC don't even appear at the top level. So, I've ripped both of these out.
   
         This is seen in (2)(ii) before vs. after cited above.
    
Images attached to this comment
jeffrey.kissel@LIGO.ORG - 10:04, Tuesday 02 April 2024 (76876)
J. Kissel, D. Barker, P. Fritschel

Model changes described above in LHO:76856 now covered under (fast-track approved) ECR E2400116 and IIET Ticket 30863, escalated to such formality because the SUS-OMC_M1_ASC_[LTVRPY]_IN1 channels *were* in the GDS broadcaster frames which is under strict ECR control. They're now replaced by OMC_M1_LOCK_[LPY]_IN1 (i.e. three less channels because TVR don't exist).

After making these changes, the impacted model and library have been committed to the userapps SVN under
    /opt/rtcds/userapps/release/sus/h1/models/h1susifoout.mdl            rev 27374
    /opt/rtcds/userapps/release/sus/common/models/OMCS_MASTER.mdl        rev 27375
H1 SUS
jeffrey.kissel@LIGO.ORG - posted 12:37, Monday 01 April 2024 - last comment - 10:02, Tuesday 02 April 2024(76842)
OMC ASC Inputs to H1 SUS OMC Drive Path Has Been Doing it Wrong For Years. We Can Fix it Tomorrow.
J. Kissel, [Corroborated by J. Wright and J. Driggers]
WP 11797

While upgrading the OMCS_MASTER.mdl library part for the OMC Suspension (which now lives in the h1susifoout.mdl top-level model), I found two distressing things:
    (1) The OMC ASC signals that are sent from the h1omc.mdl model -- running at 2kHz -- are not anti-imaged at all before being sent to the OMC SUS DAC -- running at 16 kHz. So, we're sending unnecessary imaging noise to the SUS DAC.
    (2) The OMC ASC L, P, amd Y signals are being sent through a filter bank that does NOT past through the DRIVEALIN matrix, rendering 2019's attempt at diagonalization (LHO:47488) unused.

These bugs have been long standing.
I see it now as I touch the h1susifoout.mdl and the OMCS_MASTER library part for watchdog upgrades, but I'd previously identified the issue when 
 - I was upgrading the front-end models for O4 (i.e. when the OMC SUS transitioned from the h1susomc.mdl to the h1susifoout.mdl) in 2021 (LHO:59652), and
 - When we were identifying which SUS models used the broken "TrueRMS" bloick -- there's a preceding note in the simulink model that suggests this has been around since 2015 or earlier.

We can fix this tomorrow, with *only* a change to the h1susifoout.mdl and OMCS_MASTER.mdl models, and we're already doing that anyways for the SUS watchdog upgrade.


Attached are several series of screenshots showing:
    (1) The signal path from the h1omc.mdl model where the OMC ASC control signals begin, to where they land in the OMC SUS driving in the EULER basis (again WITHOUT going through the DRIVEALIGN MATRIX)
    (2) MEDM Screens showing that in practice we only drive L, P, and Y (as one should for an ASC signal) and we don't use T, V, or R drive and likely never will.
    (3) The existing filters that will need to be copied over, and the DRIVEALIGN gain values that will likely need to be re-addressed -- since effectively we've NOT been using them.

Note -- I checked on how this is done at LLO, and they *do* drive through the LOCK banks and DRIVALIGN banks (BUT they don't have any anti-imaging filters in pace either, even though their ISCINF FM1 banks are ON... whoops)...
So, when we change the OMCS_MASTER library part, this won't affect them.
Images attached to this report
Comments related to this report
jeffrey.kissel@LIGO.ORG - 15:05, Monday 01 April 2024 (76845)ISC
The ASC P and Y filters in play during nominal low noise are as follows:
     Bank            Module     Name         Design String
     OMC_M1_ASC_P    FM0        SUSinvert    zpk([0.12+i*3.85;0.12-i*3.85;0.12+i*2.18;0.12-i*2.18],[0.3+i*1.75;0.3-i*1.75],1,"n")
                     FM10       LP8          zpk([],[2.58028+i*3.68676;2.58028-i*3.68676;1+i*7.93725;1-i*7.93725],1,"n")

     OMC_M1_ASC_Y    FM0        FM1          zpk([0.033+i*0.54;0.033-i*0.54;0.25+i*3.95;0.25-i*3.95],[1.6;1.6],1,"n")
                     FM10       LP8          zpk([],[2.58028+i*3.68676;2.58028-i*3.68676;1+i*7.93725;1-i*7.93725],1,"n")

These might be from 2015 -- the only aLOG I can find that even vaguely discusses these filters is LHO:16402.
(Note, these are distinctly separate from the primary OMC ASC control filters that were simiplified by Dan Brown and Evan Hall in 2022; LHO:65861).

Attached are bode plots of these two filters and their products for PITCH (in RED, on the left) and for YAW (in BLUE, on the right).
Note, the FM0 plant inversion filters resolve to have notches at
    Notch Freq     1st (Hz)  2nd (Hz)
       P DOF          2.21     3.84
       Y DOF          0.54     3.94


For comparison, I plotted the M1 [PIT,YAW] to M2 [PIT,YAW] transfer functions from the latest dynamical model from 2021 when I improved the damping loops (LHO:60049) and I get resonances at 
    Notch Freq     1st (Hz)  2nd (Hz)
       P DOF          1.92     3.98
       Y DOF          0.46     4.13


So ... maybe when we fix the infrastructrure, we should also update the plant inversion...
Images attached to this comment
Non-image files attached to this comment
jeffrey.kissel@LIGO.ORG - 16:20, Monday 01 April 2024 (76858)ISC, SYS
LHO aLOG 76856 describes the model prep for fixes made to the h1susifoout.mdl top level part and OMCS_MASTER.mdl library part in order to fix this bug.
jeffrey.kissel@LIGO.ORG - 10:02, Tuesday 02 April 2024 (76875)
J. Kissel, D. Barker, P. Fritschel

Model changes described above in LHO:76856 now covered under (fast-track approved) ECR E2400116 and IIET Ticket 30863, escalated to such formality because the SUS-OMC_M1_ASC_[LTVRPY]_IN1 channels *were* in the GDS broadcaster frames which is under strict ECR control. They're now replaced by OMC_M1_LOCK_[LPY]_IN1 (i.e. three less channels because TVR don't exist).

After making these changes, the impacted model and library have been committed to the userapps SVN under
    /opt/rtcds/userapps/release/sus/h1/models/h1susifoout.mdl            rev 27374
    /opt/rtcds/userapps/release/sus/common/models/OMCS_MASTER.mdl        rev 27375
H1 PSL
ryan.short@LIGO.ORG - posted 10:07, Monday 01 April 2024 - last comment - 15:33, Tuesday 02 April 2024(76839)
PSL 10-Day Trends

FAMIS 20022

The main chiller alarm was active last night, so this morning I added 100mL of water to stope the low-level alarm.

We plan on doing a remote PMC alignment tomorrow during maintenance to address the recent rise in PMC reflected power.

Images attached to this report
Comments related to this report
ryan.short@LIGO.ORG - 15:33, Tuesday 02 April 2024 (76892)PSL

Here are the actual trends from this week; seems I posted a repeat of last weeks'.

Images attached to this comment
Displaying reports 14421-14440 of 89077.Go to page Start 718 719 720 721 722 723 724 725 726 End