TITLE: 12/23 Day Shift: 16:00-00:00 UTC (08:00-16:00 PST), all times posted in UTC
STATE of H1: Observing at 160Mpc
OUTGOING OPERATOR: Tony
CURRENT ENVIRONMENT:
SEI_ENV state: CALM
Wind: 5mph Gusts, 4mph 5min avg
Primary useism: 0.04 μm/s
Secondary useism: 0.42 μm/s
QUICK SUMMARY:
TITLE: 12/23 Eve Shift: 00:00-08:00 UTC (16:00-00:00 PST), all times posted in UTC
STATE of H1: Observing at 156Mpc
INCOMING OPERATOR: Tony
SHIFT SUMMARY:
H1 was locked and observing the entire shift; current lock stretch coming up on 20 hours. One event candidate, S231223j, but nothing else of note on this very quiet evening.
State of H1: Observing at 161Mpc
H1 has been locked for almost 16 hours. Very quiet evening so far.
TITLE: 12/22 Eve Shift: 00:00-08:00 UTC (16:00-00:00 PST), all times posted in UTC
STATE of H1: Observing at 160Mpc
OUTGOING OPERATOR: Ryan C
CURRENT ENVIRONMENT:
SEI_ENV state: SEISMON_ALERT
Wind: 8mph Gusts, 5mph 5min avg
Primary useism: 0.12 μm/s
Secondary useism: 0.39 μm/s
QUICK SUMMARY:
H1 has been locked and observing for almost 12 hours. All other systems look good.
The first and second attached plots show the BNS range with and without ADF servo. It takes 3 hours to reach the maximum BNS range without ADF servo, but it takes only 1 hour with ADF servo. During thermalization, the optimal squeezing angle can drift due to the drift of SRC detuning and SQZ-IFO mode matching. Since the ADF at 1.3 kHz is within the SRC bandwidth 80 kHz, the ADF can track the squeezing angle drift by SRC detuning and SQZ-IFO mode matching which CLF cannot track. The ADF servo uses the ADF squeezing angle (SQZ-ADF_OMC_TRANS_SQZ_ANG) as error signal and feeds it back to the CLF demod phase (SQZ-CLF_REFL_RF6_PHASE_PHASEDEG) to keep the optimal squeezing angle during thermalization.
TITLE: 12/22 Day Shift: 16:00-00:00 UTC (08:00-16:00 PST), all times posted in UTC
STATE of H1: Observing at 158Mpc
INCOMING OPERATOR: Ryan S
SHIFT SUMMARY:
5.7 from the Dominican Republic came through around 18:55, EQ mode at 18:56UTC, lots of SRM saturations
I noticed that ITMX mode12 was slowly ringing up and turned off the gain and applied settings that worked previously, alog74430.
Theres been some extra noise from the MSR today, Tony went in and found H1FS0 to have an error which was found be a faulty backup power supply which was swapped alog74998
The SQZer lost lock and we went into comissioning at 2:28UTC, it relocked and we went back into Observing at 22:30UTC.
Microseism continues to rise
LOG:
| Start Time | System | Name | Location | Lazer_Haz | Task | Time End |
|---|---|---|---|---|---|---|
| 21:33 | EE | Fil | MidY | N | Drop of parts | 22:14 |
| 22:23 | CDS | Tony | H2 ecl | N | Look for spare power supplies | 22:27 |
(Travis S., Gerardo M.)
Late entry.
Last Tuesday Travis and I troubleshooted the lost signal from HAM2 annulus ion pump. Upon getting to HAM1/HAM2 the controllers for both AIPs were noted good, see photos, meaning that they both had a couple of LEDs on, this is good, then the connection for the signal cable on the back of the controller for HAM2 AIP was verified. Voltage read zero when measured at the rack in the mechanical room, then the signal was measured at the controller but on the front panel, we got some sensible voltage ~1.4 VDC, so based on the results we replaced the controller, however that did not fix the issue with the lost signal.
After a brief break (went to X-End to install a fan motor on aftercooler for purge air system) and with a laptop on hand, the signal cable from the controller for HAM2 AIP was removed, I then noticed that the signal for HAM1 AIP went red, reconnected the cable for HAM2 AIP and signal for HAM1 AIP was back. Long story short the signal cables never got switched, and with this new information I reseated the signal cable for HAM1 AIP, then the signal for HAM2 AIP was good, this was the loose connection, we still don't know why it lost connection, perhaps someone was up on top of HAM1 or in the vicinity shaking things around.
A FRS ticket is open to switch the signal cables at the rack 30026, since this is the easy way to fix this.
A long sustained error sound was coming from the MSR.
Upon inspection, the H1FS0 server had one of it's power supplies fail.
Erik and I have replaced the power supply around 22:30 UTC.
The replacement PSU came from a server from the far left side of the test stand rack in the H2 building. The PSU was not being used in the test stand.
Jonathan, Erik, EJ, Dave:
A follow up on the unexpected IPC errors we saw sitewide after the power cycle of h1seib1 Tue 17.47 19th December 2023 PST (cycled because of an ADC issue).
Details can be found in FRS30025
When h1seib1 rejoined the dolphin network (i.e. when it re-enabled its IX dolphin switch port during its boot cycle) most, but not all, frontends with Dolphin and/or long-range Dolphin IPC receivers saw Rx errors (see attachment).
This error condition persisted for 64 seconds.
For some front ends the error rate was small (e.g. a dozen single errors spread over 64 seconds) and for others it was larger (over 100 errors, multiple per second (up to 8), continuous stretch of errors for 15 seconds)
Unfortunately the HAM ISI models were in the list of models which saw 15 seconds of continuous errors. This caused their Payload Watchdogs to trip (they trip on 10 seconds or more continuous IPC Rx errors) (see attachment)
Currently the only event we have detected which happened at the time the IPC errors stopped was when the h1isiitmy model got going on h1seib1, at which point it would have started writing to its dolphin card.
STATE of H1: Observing at 155Mpc
We've been locked and observing the whole shift, we survived a 5.3 from the Dominican Republic earlier.
In advance of the Holiday break, I topped off the water level for the PSL chiller. I added ~150mL of water to hit the MAX line.
(Filiberto C., Gerardo M.)
The motor for the aftercooler was removed and replaced (see aLOG 74833), the fan motor for the aftercooler was tested and is fully functional. Next maintenance period the purge air system will be tested.
Closed FRS ticket 29978.
Fri Dec 22 10:13:19 2023 INFO: Fill completed in 13min 15secs
Gerardo confirmed a good fill curbside. TC mins -99C,-92C, trip is -70C
All looks good
Closes FAMIS 26272
The only interesting item was MY_FAN1_270_{1,2} is seeing some big spikes 1.0< yesterday and today (only really fan2 today).
Closes FAMIS 26223
Laser Status:
NPRO output power is 1.829W (nominal ~2W)
AMP1 output power is 67.86W (nominal ~70W)
AMP2 output power is 136.9W (nominal 135-140W)
NPRO watchdog is GREEN
AMP1 watchdog is GREEN
AMP2 watchdog is GREEN
PMC:
It has been locked 15 days, 20 hr 0 minutes
Reflected power = 16.6W
Transmitted power = 109.8W
PowerSum = 126.4W
FSS:
It has been locked for 0 days 5 hr and 2 min
TPD[V] = 0.9381V
ISS:
The diffracted power is around 2.0%
Last saturation event was 0 days 5 hours and 3 minutes ago
Possible Issues: None
No clear reason, https://ldas-jobs.ligo-wa.caltech.edu/~lockloss/index.cgi?event=1387278013
TITLE: 12/22 Day Shift: 16:00-00:00 UTC (08:00-16:00 PST), all times posted in UTC
STATE of H1: Observing at 159Mpc
OUTGOING OPERATOR: Tony
CURRENT ENVIRONMENT:
SEI_ENV state: CALM
Wind: 9mph Gusts, 8mph 5min avg
Primary useism: 0.03 μm/s
Secondary useism: 0.26 μm/s
QUICK SUMMARY:
Jenne, Naoki, Louis, Camilla, Sheila
Here is comparison of the DARM CLEAN spectrum with OM2 hot vs cold. The second screenshot shows a time series of OM2 cooling off. The optical gain increased by 2%, as was seen in the past (for example 71087). Thermistor 1 shows that the thermal transient takes much longer (12 + hours) than what thermistor 2 says (2 hours).
Louis posted a comparison of the calibration between the two states, there are small differences in calibration ~1% (74913). While the DARM spectrum is worse below 25Hz, it is similar at 70 Hz where we in the past thought that the sensitivity was worse with OM2 cold. From 100-200 Hz the sensitivity seems slightly better with OM2 cold, some of the peaks are removed by Jenne's jitter subtraction (74879) but there also seems to be a lower level of noise between the peaks (which could be small enough to be a calibration issue). At high frequency the cold OM2 noise seems worse, this could be because of the squeezing. We plan to take data with some different squeezing angles tomorow and will check the squeezing angle as part of that.
So, it seems that this test gives us a different conculsion than the one we did in the spring/summer, and that now it seems that we should be able to run with OM2 cold to have better mode matching from the interferometer to the OMC. We may have not had our feedforwards well tuned in the previous test, or perhaps some other changes in the noise mean that the result is different now.
Is this additonal nosie at low frequency due to the same non-stationarity we oberved before and we believe is related to the ESD upconversion? Probably not, here's why.
First plot compares the strain spectrum from two times with cold and hot OM2. This confirms Sheila's observation.
The second and third plots are spectrograms of GDS-CALIB_STRAIN during the two periods. Both show non-stationry noise at low frequency. The third plot shows the strain spectrogram normalized to the median of the hot OM2 data: beside the non-stationariity, it looks like the background noise is higher below 30 Hz.
This is confirmed by looking at the BLRMS in the 16-60 Hz region for the two times, as shown in the fourth plot: its higher with cold OM2
Finally, the last plot shows the correlation between the ESD RMS and the strain BLRMS, normalized to the hot OM2 state. There is still a correlation, but it appear again that the cold OM2 state has an additional background noise: when the ESD RMS is att the lower end, the strain BLRMS setlles to higher values
Here is the same comparison, without squeezing. Using times from 74935 and 74834
This suggests that where cold OM2 seems better than hot OM2 above that is due to the squeezing (and the jitter subtraction Jenne added, which is also on in this plot for cold OM2 but not for hot OM2). And the additional noise with cold OM2 reaches up to about 45Hz.
After we optimized ADF demod phase in 74972, the BNS range seems better and consistently 160-165Mpc. The attached plot shows the comparison of OM2 cold/hot with/without SQZ. The OM2 cold with SQZ is measured after optimization of ADF demod phase and other measurements are same as Sheila's previous plots.
This plot supports what Sheila says in the previous alogs.
Echo-ing the above, and summarizing a look at OM2 with sqz in both Sept 2023 and Dec 2023 (running gps times dictionary is attached here).
If we compare the effect of squeezing -- there is higher kHz squeezing efficiency with hot OM2. We can look at either just the darm residuals dB[sqz/unsqz] (top), or do subtraction of non-quantum noise (bottom) which shows that hot OM2 improved the kHz squeezing level by ~0.5 dB at 1.7 kHz (the blue sqz blrms 5). This is consistent with summary pages: SQZ has not reached 4.5 dB since cooling OM2 74861. Possibly suggests better SQZ-OMC mode-matching with hot OM2.
Without squeezing, cold om2 has more optical gain and more low-freq non-quantum noise. Better IFO-OMC mode-matching with cold OM2.
In total, it's almost a wash for kHz sensitivity: heating OM2 loses a few % optical gain, but recovers 0.2-0.5 dB of shot noise squeezing.
It's worth noting the consistent range increases with SQZ tuning + improvements: even in FDS, there is a non-zero contribution of quantum noise down to almost 50 Hz. For example Naoki's adjustment of sqz angle setpoint on 12/21 74972 improved range, same for Camilla's Jan sqz tuning 75151. Looking at DARM (bottom green/purple traces), these sqz angle tunings reproducibly improved quantum noise between about 60-450 Hz.
Here are some more plots of the times that Vicky plotted above.
The first attachment is just a DARM comparison with all 4 no sqz times, OM2 cold vs hot in December vs September.
Comparing OM2 hot September vs December shows that our sensitivity at from 20-40 Hz has gotten worse since September, the MICH coherence seems lower while the jitter and SRCL coherence seem similar. The same comparison for OM2 cold shows that with OM2 cold our sensitivity has also gotten worse from 15-30 Hz.
Comparing cold vs hot, in September the MICH coherence did get worse from 60-80 Hz for cold OM2, which might explain the worse sensitivity in that region. The MICH coherence got better from 20-30 Hz where the sensitivty was better for cold OM2. The December test had better tuned MICH FF for both hot and cold OM2, so this is the better test of the impact of the curvature change.
As Gabriele pointed out with his BRUCO, 74886 there is extra coherence with DHARD Y for cold OM2 at the right frequencies to help explain the extra noise. There isn't much change in the HARD pitch coherence between these December times, but the last attachment here shows a comparison of the HARD Y coherences for hot and cold OM2 in December.
Peter asked if the difference in coherence with the HARD Yaw ASC was due to a change in the coupling or the control signal.
Here is a comparison of the control signals with OM2 hot and cold, they look very similar at the frequencies of the coherence.
Late entry, work done last tuesday. The corner station, filter cavity end station and the X-End station purge air systems were operated and tested, this work was done in preparation for the upcoming vents.
The corner station system was turned on and ran for 3 hours, then the dry air was sampled, the output at the drying towers produced a dew point of -41.3 deg C, a good reading. No other issues were noted with the compressor or the drying towers. The compressor (Kobelco) unit is to have a yearly maintenance done in the next couple of weeks.
The filter cavity end station system also ran for 3 hours, the dry air was tested, this system produced a dew point of -43.6 deg C. No real issues with the system were encountered, compressor and drying towers. However, the system requires power to run the drying tower, one automatic drain for the receiver tank, and a second automatic drain for the compressor all 115 AC power, we have an outlet with 4 dedicated plugs for such process, however the day of the test, 3 out of the 4 outlets were being used by other non vacuum systems.
The test at the X-End was terminated when the aftercooler fan failed to start, after 3 different tries. The system was powered off and troubleshooted with help from Filiberto, the fan motor was dead. A new fan motor for the aftercooler was ordered. FRS ticket opened, see 29978.
X-End issue resolved, see here 74989.