ibrahim.abouelfettouh@LIGO.ORG - posted 08:17, Wednesday 13 December 2023 (74778)
OPS Day Shift Start

TITLE: 12/13 Day Shift: 16:00-00:00 UTC (08:00-16:00 PST), all times posted in UTC
STATE of H1: Corrective Maintenance
OUTGOING OPERATOR: TJ
CURRENT ENVIRONMENT:
    SEI_ENV state: CALM
    Wind: 8mph Gusts, 6mph 5min avg
    Primary useism: 0.02 μm/s
    Secondary useism: 0.25 μm/s
QUICK SUMMARY:

IFO is attempting to relock after ADC card failure in h1iopseib3. Erik, Richard and TJ worked on this about an hour before shift and got everything up and running again. See alog 74774 for details. EX seems misaligned (due to a trip) so may go through initial alignment if this lock attempt doesn't work.

Currently locking ALS though TJ mentioned lower ALSX/Y arm transmition than usual (by about 0.1/0.2 counts) - a weeklong issue apparently.

H1 CDS
thomas.shaffer@LIGO.ORG - posted 06:49, Wednesday 13 December 2023 - last comment - 11:10, Wednesday 13 December 2023(74774)
h1iopseib3 ADC error

Looks like the lock loss was caused by an ADC error on h1iopseib3. Erik is currently working on the issue.

Comments related to this report
erik.vonreis@LIGO.ORG - 07:23, Wednesday 13 December 2023 (74775)

[ TJ, Erik, Richard ]

Failure happend at 13:16:22 UTC. 

The second ADC showed timeout.  On reboot of the front end, the Linux kernel crashed.

Richard cycled power on the IO Chassis and we rebooted the front end again.  This time it did not crash, but could not see the second DAC.

This sounds similar to the events a year ago described here: https://alog.ligo-wa.caltech.edu/aLOG/index.php?callRep=66570

Richard is going to re-seat the second DAC.

erik.vonreis@LIGO.ORG - 07:28, Wednesday 13 December 2023 (74776)

Richard found that the LEDs on the second DAC did not come up as they did with the first DAC.  The second DAC again did not appear on the PCI bus.

Richard is going to replace it.

erik.vonreis@LIGO.ORG - 08:03, Wednesday 13 December 2023 (74777)

We're back up and running with the new DAC.

filiberto.clara@LIGO.ORG - 11:10, Wednesday 13 December 2023 (74783)

16 Bit DAC Installed: SN 180904-122
16 Bit DAC Removed: SN 100922-07

H1 CAL (AOS)
louis.dartez@LIGO.ORG - posted 01:05, Wednesday 13 December 2023 - last comment - 09:04, Wednesday 13 December 2023(74771)
Follow up to LHO DARM modeling efforts
Gabriele, Sheila, Artem, Louis


Since Gabriele found low frequency noise (~4 Hz) non-linearly coupling into DARM at higher frequencies (~15-25 Hz) (see LHO:73937), Sheila & I have been working to prop up a dependable model of the DARM loop that can be used to estimate stability margins and offload actuation from ETMX L3 to L2 and L1. N.B. we're not 100% certain that the noise we're chasing is due to the ETMX ESD.


1. The DARM Model

To estimate DARM loop stability, we've been primarily looking at modeling and directly measuring IN1/IN2 at the input to each stage's LOCK bank filter. IN1/IN2 at the input to the L3 LOCK bank filter (often measured as (IN1/EXC)/(IN2/EXC) so as to not be biased towards 1) is the open loop gain, G, of the DARM loop. We've been calling IN1/IN2 measured at the input to the L2 and L3 LOCK banks Gp and Gu, respectively. This is because they each behave somewhat like an effective open loop gain as measured at those points in the offloaded DARM loop. Similar to G, Gp, and Gu (the "OLGs"), we calculate the loop suppression, 1/(1-G), and the closed loop gain, G/(1-G), for each our three test points (G, Gp, and Gu) at the input of each ETMX LOCK bank filter.
Figure 1: LHO_DARM_with_measurements_for_G_and_Gu.png
Figure 1 above is a Bode plot of G, Gp, and Gu for the LHO DARM loop overlaid with measurements taken for G and Gu. The OLG, G, measurement was taken from a recent calibration swept sine and the measurement of Gu is courtesy of Sheila from LHO:74226. The two measurements shown for G and Gu match the model pretty well. There is some noise in the Gu measurement but the structure is in pretty good agreement between the model and the data. The measurement of G also matches pretty well but begins to deviate below 10 Hz. This is likely in large part due to error in fitting the sensing function, C, at low frequencies. It's worth noting that the UIM stage has multiple UGFs just below 2 Hz and 4 Hz; this is okay but not ideal. We tried taking a measurement of Gp but we've been unable to get good agreement between the current DARM loop model and these measurements (this is mentioned in LHO:74264). We've only tried relatively low coherence measurements so far in this current sprint so we plan to get some better measurements of this soon to compare against the model. However, it's worth noting that the PUM has been a bear to model properly for a while (see for example the PUM discussion in LHO:63450). A few years ago Jenne successfully modeled length-to-pitch-to-length cross-coupling that matched data taken at the PUM stage (LHO:48738). In light of the recent modeling push, Sheila and I are have begun working to follow up on this work to include yaw in the hopes of modeling and suppressing this effect from coupling to DARM. More on that later as our current major goal is to reduce ETMX actuation as introduced at the top. The Gu points in Figure 1 were calculated from a measurement of IN1/EXC at the input of the L2 stage LOCK bank filter. Direct measurement of Gu = IN1/IN2 was difficult due to low coherence. The relationship between Gu and IN1/IN2 is Gu = (IN1/EXC)/(1+(IN1/EXC)). See Section 4 of DCC:LIGO-T2300436 for a discussion on that relationship.

2. Comparison of LHO and LLO DARM Models

Here is a quick comparison of recent DARM loop models for LHO and LLO. For searching purposes I'm writing their modeled OLG UGFs: LHO UGFs: 1: 70.45 Hz, 26.97 degrees LLO UGFs: 1: 78.75 Hz, 35.89 degrees
Figure 2: Zoomed out, LHO_LLO_comparison_OLG_zoom_out.png
Figure 3: Zoomed in low band, LHO_LLO_comparison_OLG_zoom_low_freq.png
The DARM OLG magnitudes between LLO and LHO differ at low frequencies by roughly 3 orders of magnitude near 1Hz. Here's an even higher zoom near 1Hz: LHO_LLO_comparison_OLG_super_zoom_low_freq.png. As an independent check into whether that makes sense, Artem pulled up the external disturbance sent to each actuation stage at both sites using CAL-DELTAL_CTRL_{UIM,PUM,TST}_DBL_DQ and saw that they're similar between LLO and LHO (direct link to this plot). Then he looked at the LHO and LLO error signals and found a discrepancy of 2 orders of magnitude in the expected direction (direct link to this plot). So that all seems to roughly track, which is great news. See LHO:74744 for Artem's discussion.
Figure 4: Zoomed in bubble, LHO_LLO_comparison_OLG_zoom_bubble.png

3. Instability of 4.5Hz boost filter not predicted by DARM model

In LHO:74226, Sheila tried engaging a L2 LOCKL FM2, "boost4.5", which caused a lockloss. She successfully engaged the boost without a lockloss the day before just before the Tuesday maintenance period. Here is a comparison of the LHO model with and without this boost engaged.
Figure 5: Zoomed in bubble, LHO_compare_toggle_4p5khz_L2_LOCK_boost.png
The hope was that we would look at the model with the boost included and be able to confirm that the resultant loop configuration is unstable. However, we aren't able to point to anything that obviously indicates that that's the case. The corresponding loop suppression plot shows some peaking near 10Hz but nothing that would have us consider it a smoking gun. ===== So the DARM modeling efforts are progressing but there is still much to do..
Images attached to this report
Comments related to this report
louis.dartez@LIGO.ORG - 09:04, Wednesday 13 December 2023 (74780)
after discussing a bit with Gabriele this morning and taking a closer look at the scenario with the 4.5Hz boost applied, I think I was too quick to suggest that the instability disagrees with the model.

1. There is some minor gain peaking the TST stage near 10Hz with the boost applied compared to the nominal state.
2. Gp inches closer to unity at 10Hz. Given that we already don't trust the Gp model right now, this could put us close enough to unity to cause issues at 10Hz, which is also where the phase goes to 0.
3. OLG also starts to dip to unity near 10Hz. 

all this together may actually suggest that the loop with the 4.5Hz boost engaged, or the transition from the nominal state to the new state with the boost engaged, could be unstable. So perhaps Sheila's experience with having the boost be fine one day and cause a lockloss the next day is consistent with the model afterall. 
H1 General
anthony.sanchez@LIGO.ORG - posted 00:05, Wednesday 13 December 2023 (74773)
Tuesday EVE End Shift

TITLE: 12/13 Eve Shift: 00:00-08:00 UTC (16:00-00:00 PST), all times posted in UTC
STATE of H1: Observing at 158Mpc
INCOMING OPERATOR: TJ
SHIFT SUMMARY:
00:27 UTC HWS was restarted and took us out of Observing by making SF changes for a minute. The HWS Engineer making the changes was still in the control room and managed to handle the SDF changes which allowed us to get back to Observing.  See Alog : https://alog.ligo-wa.caltech.edu/aLOG/index.php?callRep=74770

00:29 UTC OBSERVING was reached again.

After this a small earthquake rolled through which we survived, but otherwise the night was quite.
Everthing went smooth like butter.

LOG:
No log

H1 General
anthony.sanchez@LIGO.ORG - posted 16:11, Tuesday 12 December 2023 (74768)
Tuesday EVE Shift Start

TITLE: 12/13 Day Shift: 16:00-00:00 UTC (08:00-16:00 PST), all times posted in UTC
STATE of H1: Observing at 156Mpc
OUTGOING OPERATOR: Corey
CURRENT ENVIRONMENT:
    SEI_ENV state: CALM
    Wind: 5mph Gusts, 3mph 5min avg
    Primary useism: 0.02 μm/s
    Secondary useism: 0.33 μm/s
QUICK SUMMARY:
H1 Is locked at NOMINAL_LOW_NOISE and OBSERVING for the past hour.
Violins lookin great!

 

LHO General
ryan.short@LIGO.ORG - posted 16:04, Tuesday 12 December 2023 (74765)
Ops Day Shift Summary

TITLE: 12/12 Day Shift: 16:00-00:00 UTC (08:00-16:00 PST), all times posted in UTC
STATE of H1: Preventive Maintenance
INCOMING OPERATOR: Tony
SHIFT SUMMARY: Relatively light maintenance day followed by some trouble getting locked, but after adjusting thresholds in ALS Guardians, was able to lock without much issue.

LOG:

Start Time System Name Location Lazer_Haz Task Time End
16:06 FAC Karen EY - Technical cleaning 17:16
16:06 FAC Kim EX - Technical cleaning 17:29
16:07 FAC Randy EY -   17:05
16:08 VAC Janos, Jordan MX, EX - Turbopump testing 19:35
16:09 TCS Camilla LVEA YES Turning off HWS power supplies 16:17
16:24 VAC Gerardo CS, FCES, EX YES Purge air system testing 20:20
16:35 VAC Travis LVEA YES Turbopump testing 19:50
16:36 FAC Chris +1 LVEA, Outbuildings YES Fire extinguisher inspections 17:50
16:51 CDS Fil EX - PEM cable termination 18:20
17:05 FAC Randy LVEA YES Moving scissor lift to HAM3 17:55
17:08 AOS Betsy, Ali LVEA YES Walkabout 18:25
17:08 TCS Camilla Optics Lab LOCAL CO2 profiler testing 18:25
17:16 FAC Karen, Kim FCES - Technical cleaning 18:10
17:30 FAC Tyler LVEA YES 3IFO 17:41
17:40 TCS TJ Optics Lab LOCAL CO2 profiler testing 18:25
17:52 SQZ Daniel, Nutsinee LVEA YES   19:10
18:13 SEI Jim LVEA YES Move CPS rack (HAM7 ISI down) 18:24
18:21 CDS Fil LVEA YES BSC2 temp sensor 20:15
18:25 SEI Jim CR - HAM6 transfer functions 19:11
18:28 TCS TJ, Camilla, Ali LVEA YES CO2X beam profiling, HWS table checks 20:15
18:45 FAC Karen, Kim LVEA YES Technical cleaning 20:00
19:18 SEI Jim CR - ITMY transfer functions 20:00
21:34 TCS Camilla LVEA - Turning on HWS cameras 21:41
21:45 FAC Richard LVEA - Moving scissor lift 21:50
22:16 VAC Gerardo, Fil MX - Compressor troubleshooting Ongoing
H1 ISC (OpsInfo)
thomas.shaffer@LIGO.ORG - posted 15:34, Tuesday 12 December 2023 (74766)
ALS green transmission threshold lowered

Ryan S, Jenne D, Sheila

The ALS_COMM and DIFF nodes will check on the green arm transmission when finding IR to ensure that the arms are aligned well enough to move on. This threshold has been set to 0.75 but I lowered it to 0.65 since we've ran into this threshold as few times in the last week and have been unable to improve the arm transmission.

Last week this happened on Thursday (alog74655) and I just manualed around the state to continue locking. It worked for that lock to get us back up, and then it wasn't a problem with future locks until today. Today, the arm transmission was low for both arms so we went to confirm that there was no PR3 movement, and then we restored optics back to the start of the previous lock. This also didn't help. Looking at the arm power over the last few weeks (attachment 1), after the PR3 and table work from last Tuesday (alog74618) it has been lower than before then but as low as we were seeing today. The current thinking is that the PR3 move last week brought our powers lower, and while the table work improved the COMM beatnote, it did nothing for the arm powers. Once the IFO gets to a better full lock alignment, this alignment tends to give us better ALS power next lock acquisition, so lowering this threshold should help us get to that point and hopefully help sequential locks.

The long term solution would be to go back on table to realign the rest of the COMM path to camera, then renormalize the arm powers so that 1 is actually our max power, then bring this threshold back to the 0.75 that it was*. Going on that table always carries rick of making things worse, so we definitely don't want to be doing this right before the holidays. We will reconvene in 2024 on this issue if it's still an issue.

*There's a comment in the guardian code that it was 0.85 from Aug 12, 2020.

Images attached to this report
LHO VE
jordan.vanosky@LIGO.ORG - posted 13:36, Tuesday 12 December 2023 (74763)
Functionality Test Performed on EX/MX Turbo Pumps

Jordan, Janos

We ran the functionality test on the main turbopumps in MX and EX during Tuesday Maintenance (12/12/23). The scroll pump is started to take pressure down to low 10^-02 Torr, at which time the turbo pump is started, the system reaches low 10^-08 Torr after a few minutes, then the turbo pump system is left ON for about 1 hour, after the hour the system goes through a shut down sequence.
 

MX Turbo:

Bearing Life:100%

Turbo Hours: 120

Scroll Pump Hours: 211

EX Turbo:

Bearing Life:100%

Turbo Hours: 277

Scroll Pump Hours: 6317

Closing WP 11573 and FAMIS 24846 FAMIS 24870

H1 TCS
camilla.compton@LIGO.ORG - posted 13:08, Tuesday 12 December 2023 (74760)
CO2X Beam Scan Started

TJ, Camilla. WP11574

We borrowed the Ophir PH00235 USB MSP-NS-Pyro 9/5 bean scanner from CIT. It has a PH00092 removable 7.5" FL lens on the front and is on a 500mm rail. WE checked the scanner and rail worked in the lab using our Nanoscan laptop and the Nanoscan v1 software. 

On CO2X (layout T1200007) we added three Gold coated steering mirrors just after the power control waveplate. Added a beam block panel on the back of the table. While aligning we kept ther CO2 power requested at 0.1W.  Started aligning the beam to the beamscan but didn't finish before the end of maintenance. 

We left the beanscan and the three mirrors on the table, we moved the mirror that was in the path ("A" on photo) out of the path. Plan to finish aligning and take data next Tuesday. The cables and driver boxes are in a tote labeled "nanoscan" in the TCS cabinet and the laptop is back in the optics lab cabinet.  Photos of table and beampath (nominal lighter red) attached. 

Images attached to this report
H1 General
thomas.shaffer@LIGO.ORG - posted 13:05, Tuesday 12 December 2023 (74761)
LVEA swept

Genie lift found plugged into outlet near TCSY table, unplugged.

Robert's shaker is still connected to near HAM2, it is plugged in but powered off.

All else looked good, followed T1500386

 

H1 CDS (DAQ)
david.barker@LIGO.ORG - posted 11:01, Tuesday 12 December 2023 - last comment - 13:17, Wednesday 13 December 2023(74755)
WP11568 TW0 raw minute trends offload

As the first part of the TW0 raw minute trend file offload, tw0 is now writing to a new area freeing up the old files for transfer. 

nds0 was restarted at 10:44 PST to serve the past 6 months of data from their temporary location as the files are being transferred to h1daqframes-0. The file copy takes about 30 hours.

Comments related to this report
david.barker@LIGO.ORG - 16:22, Tuesday 12 December 2023 (74769)

File copy was started 11:50 Tue. As of 16:05 43 of 256 dirs had been copied. ETA 13:10 Wed.

david.barker@LIGO.ORG - 13:17, Wednesday 13 December 2023 (74786)

Copy completed at 13:15:44 PST. I will do the nds0 change and file deletion tomorrow morning.

H1 TCS
camilla.compton@LIGO.ORG - posted 08:26, Tuesday 12 December 2023 - last comment - 13:12, Wednesday 13 December 2023(74750)
ITM HWS Camera's Powered off at 16:15UTC, search for cause of DARM Comb

This morning at 8:15am (16:15UTC) I turned off the external power supply that powered both ITM HWS CCD cameras (Dalsa 1M60), located on the floor under the HWS table.  We'll plan to check the magnetometer data to see if the 74738 comb is still present.

After we stopped HWS code and remotely turned off camera's in 74738, the DARM comb still remained. Dan, Daniel, Nutsinee expect this could be the framegrabber in the HWS computer still sending a 7Hz signal. We can troubleshoot this and how the camera's are grounded to the optics table later. 

Comments related to this report
ansel.neunzert@LIGO.ORG - 09:29, Tuesday 12 December 2023 (74752)

Looks like the comb is gone in the magnetometer channel. Figure 1 pre, figure 2 post.

Images attached to this comment
ansel.neunzert@LIGO.ORG - 10:02, Tuesday 12 December 2023 (74753)

Minor update: appending an hour-long "post" spectrum just to confirm & provide better comparison with previous 1-hour plots.

Images attached to this comment
ansel.neunzert@LIGO.ORG - 11:20, Tuesday 12 December 2023 (74756)

Camilla says the camera was turned on but not initialized at 10:03 Pacific / 18:03 UTC. Took a 30m spectrum starting 18:05 utc, attached. There is a strong near-1Hz comb and a strong near-57 Hz comb.

Images attached to this comment
camilla.compton@LIGO.ORG - 12:44, Tuesday 12 December 2023 (74759)

At 20:40UTC (12:40PT) we powered off both ITM computers h1hwsmsr and h1hwsmsr1. Dave has been notified. In 30 minutes we can try power cycling the cameras to see if the computers are off whether they re-initalize causing a comb. 

ansel.neunzert@LIGO.ORG - 13:34, Tuesday 12 December 2023 (74762)

Combs are still there with ITM computers off. Looks pretty much identical to the previous spectrum (very tiny frequency shift in the near-1Hz comb, though).

Images attached to this comment
camilla.compton@LIGO.ORG - 13:44, Tuesday 12 December 2023 (74764)

At 21:30UTC Erik unplugged the fiber connections that run from the back of the computers to the HWS cameras. Computers still off.

At 21:37UTC I power cycled the external supply to the cameras, to test if the cameras still turn on with a comb with no computer/frame grabber connection.

camilla.compton@LIGO.ORG - 16:37, Tuesday 12 December 2023 (74770)

When the camera's restanted with the fiber link to the computers disconneted, we again saw the 1Hz and 57Hz combs, plot attached. It appears default for ITMX is 57Hz and ITMY is 1Hz. We should check if the camera software allows us to to put these to zero/ off.

For now we've replugged in the fiber connections, turned on the computers and restarted the hws code, both at 7Hz (this created some sdf diffs from the H1:TCS-ITMY_HWS_{}_POS_{X,Y} channels restarting with default values that kicked us out of observing, sorry)

Lasers still off, will turn lasers on tomorrow during commissioning to avoid sdf diffs.

Images attached to this comment
camilla.compton@LIGO.ORG - 13:12, Wednesday 13 December 2023 (74785)

Turned on both ITM SLEDs at 21:08UTC. These H1:TCS-ITM{X,Y}_HWS_SLEDSHUTDOWN channels should be added to sdf.

H1 AOS
sheila.dwyer@LIGO.ORG - posted 15:44, Wednesday 15 November 2023 - last comment - 12:11, Friday 15 December 2023(74226)
DARM offloading

Louis and I have been looking at increasing the offloading of the signal from the ESD to the PUM and UIM, in light of 73913

I have attempted a few times to make measurements by exciting at the L2 LOCK L filter, this results in poor coherence even for amplitudes that are quite large in DARM, and turning up the amplitude slightly has caused a lockloss by saturating the ESD. 

On Tuesday morning I engaged the PUM boost (L2 LOCKL FM2, boost4.5 which is a boost with a little resG aaround 4.5Hz).  Some history (with links to old alogs about this filter is here: 48767) we used to run with this boost but turned it off in early O3 because angle to length cross couplings were causing instabilities of the DARM loop.  On Tuesday morning I engaged this boost, and we stayed locked.  The first attachment shows a comparison of our ETMX drives to LLO's (LLO also uses ETMY L1 for DARM control, but I haven't plotted that here), H1's usual configuration is in gold and the time with the PUM boost engaged is in red. The ESD drive RMS is reduced by more than a fator of 2 with the PUM boost engaged.  I quickly tried a swept sine injection, and saw that the coherence was somewhat better than earlier measurements with a very small amplitude injection, so it seemed promising that we might be able to get a better measurement of the cross over with the PUM boost engaged.  

Today I engaged this boost again during commisoning time and we immediately lost lock.  The third attachment shows that turning on the boost certainly seems to be the cause of the lockloss. 

We can probably rely on the pyDARM model for the PUM crossover, since the calibration measurements validate the model above 10Hz. Today I was able to make a measurement of the UIM crossover, which Louis can use to compare to pyDARM from 1-10Hz. 

 

Images attached to this report
Non-image files attached to this report
Comments related to this report
louis.dartez@LIGO.ORG - 08:57, Wednesday 13 December 2023 (74779)
this is discussed in LHO:74771
louis.dartez@LIGO.ORG - 12:11, Friday 15 December 2023 (74828)
Gabriele, Louis

The comparison between LLO's and LHO's ETMX L3 drive above (fist attachment in Sheila's post) confused us. The difference between the two LHO traces makes sense. But there are several things wrong with the channel from LLO (blue trace in top right hand plot). First of all, we do not have 2 orders of magnitude in difference between the L3 MASTER OUT drives at LLO vs LHO. Also, the calibration lines show up at the wrong frequencies in the LLO trace.

It turns out that if you use NDS2 and leave the Epoch Start date at the default 6/1/1980, DTT sees two of some channels with the same name (but different sampling frequencies!), see DTT_NDS_issue.png. The result in this case was confusion of what DTT was plotting in Sheila's alog. We think that DTT got confused by having multiple channels with the same name but differing sampling frequencies in its lookup list, leading to DTT plotting the 16kHz channel but somehow applying the wrong sampling frequency. To avoid this, we had to re-remember to be careful when copying and pasting similar channel names from one text box to another within DTT and to set the Epoch Start field to something more reasonable than 1980.
Images attached to this comment
H1 CAL
vladimir.bossilkov@LIGO.ORG - posted 08:29, Friday 28 July 2023 - last comment - 12:31, Tuesday 12 December 2023(71787)
H1 Systematic Uncertainty Patch due to misapplication of calibration model in GDS

First observed as a persistent mis-calibration in systematic error monitoring Pcal lines which measure PCAL / GDS-CALIB_STRAIN affecting both LLO and LHO, [LLO Link] [LHO Link], characterised by these measurements consistently disagreeing with the uncertainty envelope.
It us presently understood that this arises from bugs in the code producing the GDS FIR filters there exists a sizeable discrepancy, which Joseph Betzwieser is spear-heading a thorough investigation to correct,

I make a direct measurement of this systematic error by dividing CAL-DARM_ERR_DBL_DQ / GDS-CALIB_STRAIN , where the numerator is further corrected for kappa values of the sensing, cavity pole, and the 3 actuation stages (GDS does the same corrections internally). This gives a transfer function of the difference induced from errors in the GDS filters.

Attached in this aLog, and its sibling aLog in LLO, is this measurement in blue, the PCAL / GDS-CALIB_STRAIN measurement in orange, and the smoothed uncertainty correction vector in red. Attached also is a text file of this uncertainty correction for application in pyDARM to produce the final uncertainty, in the format of [Frequency, Real, Imaginary].

Images attached to this report
Non-image files attached to this report
Comments related to this report
ling.sun@LIGO.ORG - 15:33, Friday 28 July 2023 (71798)

After applying this error TF, the uncertainty budget seems to agree with monitoring results (attached).

Images attached to this comment
ling.sun@LIGO.ORG - 13:02, Thursday 17 August 2023 (72299)

After running the command documented in alog 70666, I've plotted the monitoring results on top of the manually corrected uncertainty estimate (see attached). They agree quite well.

The command is:

python ~cal/src/CalMonitor/bin/calunc_consistency_monitor --scald-config  ~cal/src/CalMonitor/config/scald_config.yml --cal-consistency-config  ~cal/src/CalMonitor/config/calunc_consistency_configs_H1.ini --start-time 1374612632 --end-time 1374616232 --uncertainty-file /home/ling.sun/public_html/calibration_uncertainty_H1_1374612632.txt --output-dir /home/ling.sun/public_html/

The uncertainty is estimated at 1374612632 (span 2 min around this time). The monitoring data are collected from 1374612632 to 1374616232 (span an hour).

 

Images attached to this comment
jeffrey.kissel@LIGO.ORG - 17:01, Wednesday 13 September 2023 (72871)
J. Kissel, J. Betzwieser

FYI: The time at which Vlad used to gather TDCFs to update the *modeled* response function at the reference time (R, in the numerator of the plots) is 
    2023-07-27 05:03:20 UTC
    2023-07-26 22:03:20 PDT
    GPS 1374469418

This is a time when the IFO was well thermalized.

The values used for the TDCFs at this time were
    \kappa_C  = 0.97764456
    f_CC      = 444.32712 Hz
    \kappa_U  = 1.0043616 
    \kappa_P  = 0.9995768
    \kappa_T  = 1.0401824

The *measured* response function (GDS/DARM_ERR, the denominator in the plots) is from data with the same start time, 2023-07-27 05:03:20 UTC, over a duration of 384 seconds (8 averages of 48 second FFTs).

Note these TDCF values list above are the CAL-CS computed TDCFs, not the GDS computed TDCFs. They're the value exactly at 2023-07-27 05:03:20 UTC, with no attempt to average further over the duration of the *measurement*. See attached .pdf which shows the previous 5 minutes and the next 20 minutes. From this you can see that GDS was computing essentially the same thing as CALCS -- except for \kappa_U, which we know
 - is bad during that time (LHO:72812), and
 - unimpactful w.r.t. the overall calibration.
So the fact that 
    :: the GDS calculation is frozen and
    :: the CALCS calculation is noisy, but is quite close to the frozen GDS value is coincidental, even though
    :: the ~25 minute mean of the CALCS is actually around ~0.98 rather than the instantaneous value of 1.019
is inconsequential to Vlad's conclusions.

Non-image files attached to this comment
louis.dartez@LIGO.ORG - 00:54, Tuesday 12 December 2023 (74747)
I'm adding the modeled correction due to the missing 3.2 kHz pole here as a text file. I plotted a comparison showing Vlad's fit (green), the modeled correction evaluated on the same frequency vector as Vlad (orange), and the modeled correction evaluated using a dense frequency spacing (blue), see eta_3p2khz_correction.png. The denser frequency spacing recovers error of about 2% between 400 Hz and 600 Hz. Otherwise, the coarsely evaluated modeled correction seems to do quite well. 
Images attached to this comment
Non-image files attached to this comment
ling.sun@LIGO.ORG - 12:31, Tuesday 12 December 2023 (74758)

The above error was fixed in the model at GPS time 1375488918 (Tue Aug 08 00:15:00 UTC 2023) (see LHO:72135)