Displaying reports 16221-16240 of 88368.Go to page Start 808 809 810 811 812 813 814 815 816 End
Reports until 16:13, Monday 20 November 2023
H1 General
oli.patane@LIGO.ORG - posted 16:13, Monday 20 November 2023 (74322)
Ops DAY Shift End

TITLE: 11/21 Day Shift: 16:00-00:00 UTC (08:00-16:00 PST), all times posted in UTC
STATE of H1: Observing at 157Mpc
INCOMING OPERATOR: Tony
SHIFT SUMMARY: Observing and Locked for 8.5 hours. Been a quiet day.
LOG:

16:00UTC In NOMINAL_LOW_NOISE, I rejected the ASC-X_FIBR_LOCK_BEAT_RFMIN value change of -10 so it is still at -20dB, and put us into Observing.

Start Time System Name Location Lazer_Haz Task Time End
18:51 FAC Randy EX n Delivery 20:21
19:24 FAC Chris, locksmith OSB, EY, EX n Changing external locks 22:28
H1 ISC
oli.patane@LIGO.ORG - posted 10:57, Monday 20 November 2023 - last comment - 15:03, Monday 20 November 2023(74310)
ALS Beatnote Strength Fluctuations Due to VEA Temperature Changes

Building off of 74289, I checked to see how the ALS-X beatnote strength was doing, and then tried to confirm that the issues were actually a result of the EX VEA temperature changes.

Currently, the beatnote strength is still oscillating, and last night it was back to dipping below -10dB when at its minimum (attachment1).

Here is what I noticed when looking to confirm the beatnote-temperature correlations:

ALS-X_FIBR_DEMOD_RFMON is very sensitive to changes in EX VEA temperatures
- Already pretty obvious based on how the temp changes from Friday affected ALS-X_FIBR_DEMOD_RFMON, but even small changes/normal daily fluctuations in EX VEA temperatures can change the beatnote strength by almost 2dB.
- Here are two stretches of time comparing the EX VEA average temperatures and how it affects ALS-X_FIBR_DEMOD_RFMON (attachment2, attachment3).
    - Sometimes the temperature affects the beatnote strength more than other times.


ALS-Y_FIBR_DEMOD_RFMON is NOT very sensitive to changes in EY VEA temperatures
- Comparing ALS-Y_FIBR_A_DEMOD_RFMON and FMC-EY_VEA_AVTEMP_DEGF(attachment4), although VEA temp changes are definitely affecting the beatnote strength (especially during large temperature excursions), the beatnote strength is able to stay very consistant.

It is especially interesting seeing the difference between the amount of variation in ALS-X_FIBR_DEMOD_RFMON as compared to ALS-Y_FIBR_DEMOD_RFMON as temperature changes (attachment5). To be fair, the temperatures at EX do seem to fluctuate more (higher highs and lower lows) than at EY, but the way that the ALS-X beatnote strength reacts still seems to me to be much more exaggerated than it should be (like if ALS-Y was going through the EX temperature changes).

Images attached to this report
Comments related to this report
daniel.sigg@LIGO.ORG - 15:03, Monday 20 November 2023 (74320)

The demod readback goes thru channel 1 of the demodulator concentrator D1100691 before reaching the EtherCAT end station chassis 2 at the L5 terminal. The RF monitor is read using this calibration dB = 16.67 * V - 72.

LHO VE
david.barker@LIGO.ORG - posted 10:28, Monday 20 November 2023 (74308)
Mon CP1 Fill

Mon Nov 20 10:24:12 2023 INFO: Fill completed in 24min 7secs

Gerardo confirmed a good fill curbside.

Images attached to this report
H1 AOS (CAL)
louis.dartez@LIGO.ORG - posted 09:59, Monday 20 November 2023 (74302)
Discrepancy in DARM loop between pyDARM model and measurements taken with IFO
As mentioned in LHO:74264, Sheila and I have been taking a closer look at the DARM loop and our ability to model the PUM crossover in particular in response to LHO:73937. I'm attaching a plot of the DARM OLG with the pyDARM model (blue trace), OLG measurements taken as part of the weekly calibration measurement suite (orange dots), and Sheila's measurements from Friday (LHO:74264) with the Y2L and P2L gains nominal (green circles) and with the Y2L and P2L gains reduced by a factor of two (red crosses).

There's a few interesting things going on that I'll have to elaborate on in a future alog. Briefly, the model clearly doesn't match any of our measurements below about 20Hz (zoomed out version also attached). 

More plots of just the PUM and UIM crossovers are in progress.





Relevant Links:

- Sheila takes OLG measurements with different Y2L and P2L gains (LHO:74264)
- Sheila attempts taking PUM crossover measurements at the L2 LOCK band (LHO:74226)
- Gabriele finds nonlinear high frequency noise coupling in DARM and suggests offloading ETMX L3 (LHO:73937)
Images attached to this report
H1 PSL
ryan.short@LIGO.ORG - posted 09:40, Monday 20 November 2023 (74304)
PSL 10-Day Trends

FAMIS 20003

Chiller alarm 2 days ago, has been addressed (alog 74276).

The FSS TPD signal has continued to drop over the past 10 days; currently is at 0.72V.

Images attached to this report
H1 DetChar (DetChar)
matthew.ball@LIGO.ORG - posted 09:35, Monday 20 November 2023 (74303)
Data Quality Shift Report 2023-11-13 to 2023-11-19

Link to full report

Summary of the report:

H1 General
oli.patane@LIGO.ORG - posted 08:05, Monday 20 November 2023 (74301)
Ops DAY Shift Start

TITLE: 11/20 Day Shift: 16:00-00:00 UTC (08:00-16:00 PST), all times posted in UTC
STATE of H1: Observing at 151Mpc
OUTGOING OPERATOR: Austin
CURRENT ENVIRONMENT:
    SEI_ENV state: CALM
    Wind: 9mph Gusts, 7mph 5min avg
    Primary useism: 0.02 μm/s
    Secondary useism: 0.43 μm/s
QUICK SUMMARY:

Looks like we had reached NOMINAL_LOW_NOISE at 15:42UTC and it wasn't able to go into Observing because of the one EXISC sdf from yesterday (ASC-X_FIBR_LOCK_BEAT_RFMIN being -20 instead of -10dB). I reverted the SDF and put us in Observing.

H1 General
anthony.sanchez@LIGO.ORG - posted 00:09, Monday 20 November 2023 (74300)
Sunday Ops EVE Shift End

TITLE: 11/20 Eve Shift: 00:00-08:00 UTC (16:00-00:00 PST), all times posted in UTC
STATE of H1: Observing at 147Mpc
INCOMING OPERATOR: Austin
SHIFT SUMMARY:

5:59 UTC lockloss  see alog: https://alog.ligo-wa.caltech.edu/aLOG/index.php?callRep=74299

ALSX Camera looks stange. Like there is an extra light from somewhere on ALSX cam.
Gaurdian ALX Node in Error. This is what Oli had warned me about.
I changed H1:ALS-X_FIBR_LOCK_BEAT_RFMIN back down to -20 as Oli has done.

I Started Initial Alinment.
Initial Alignment complete.
Running through SDF revert changes H1:ALS-X_FIBR_LOCK_BEAT_RFMIN back to -10. so I changed it back to -20 again.

Lock had to go through AQUIRE_PRMI right after an initial alignment.

Current H1 Status: Locked for 6 minutes and Observing.


LOG:
no log

 

Images attached to this report
H1 AOS (Lockloss)
anthony.sanchez@LIGO.ORG - posted 22:16, Sunday 19 November 2023 (74299)
Strange looking lockloss 1384452871

Lockloss from Nominal Low noise. Wind is not high, MicroSeism is elevated, but no earthquakes rolled through, PI's looked good.
https://ldas-jobs.ligo-wa.caltech.edu/~lockloss/index.cgi?event=1384452871
The lockloss plots look like they dont have any data, and soem of the data looks glitchy maybe. The ISC_LOCK Log that Lockloss select also brings up looks like it's missing data, the time stamp is several hours ago.

More investigations needed.

 

Images attached to this report
H1 General
anthony.sanchez@LIGO.ORG - posted 20:21, Sunday 19 November 2023 (74298)
Sunday Mid Day Shift report.

TITLE: 11/20 Eve Shift: 00:00-08:00 UTC (16:00-00:00 PST), all times posted in UTC
STATE of H1: Observing at 161Mpc
CURRENT ENVIRONMENT:
    SEI_ENV state: CALM
    Wind: 10mph Gusts, 7mph 5min avg
    Primary useism: 0.03 μm/s
    Secondary useism: 0.52 μm/s
QUICK SUMMARY:

I do not know what exactly fixed diaggui, But while I was tinkering with different settings it started to work again on the CDS machines & it works on the FOM too.
Other than that it's been a really queit night.

H1 General
anthony.sanchez@LIGO.ORG - posted 16:46, Sunday 19 November 2023 (74297)
Sunday Ops Eve Shift Start

TITLE: 11/20 Eve Shift: 00:00-08:00 UTC (16:00-00:00 PST), all times posted in UTC
STATE of H1: Observing at 150Mpc
OUTGOING OPERATOR: Oli
CURRENT ENVIRONMENT:
    SEI_ENV state: CALM
    Wind: 13mph Gusts, 8mph 5min avg
    Primary useism: 0.03 μm/s
    Secondary useism: 0.67 μm/s
QUICK SUMMARY:

H1 had been locked for 22 minutes and OBSERVING when I arrived.
DARM FOM Is still not working properly.

H1:ALS-X_FIBR_LOCK_BEAT_RFMIN is currently set to -20 and I will be keeping an eye on it.

H1 General
oli.patane@LIGO.ORG - posted 16:44, Sunday 19 November 2023 (74296)
Ops DAY Shift End

RELOCKING
16:00UTC Detector relocking and at FIND_IR.

16:09 I started an Initial Alignment because we started getting stuck in the ACQUIRE_PRMI -> ACQUIRE_DRMI_1F -> ACQUIRE_PRMI cycle again
16:26 Initial Alignment completed

16:28 Lockloss at LOCKING_ALS
16:34 Lockloss at FIND_IR
16:46 Lockloss at CARM_OFFSET_REDUCTION
    - Starting at DHARD_WFS EX started saturating every few seconds, through DHARD_WFS, PARK_ALS_VCO, SHUTTER_ALS, and CARM_OFFSET_REDUCTION
    - 16:46:53 EY, IX, IY, PRM all also saturated, DRMI lost lock, and we went down.
16:50 Lockloss at FIND_IR
16:55 I took us to DOWN due to high winds and microseism

17:35 Tried relocking again
17:35 In LOCKING_ARMS_GREEN, Guardian ALS_XARM node in error
    
17:38 ALS_XARM trying to get to INCREASE_FLASHES but keeps jumping to FAULT
    - Errors PLL; PDH and Beat note strength
    - Node keeps falling into error
    - Tried lowering beatnote strength threshold but only worked for a bit before we lost lock
    - Happened June 29th (70951)

18:49 Lockloss at CARM_OFFSET_REDUCTION
19:05 Lockloss at CARM_OFFSET_REDUCTION, putting detector in DOWN

20:02 Trying relocking
20:56 NOMINAL_LOW_NOISE
21:01 Observing

21:33 Earthquake mode activated due to earthquake from Japan

21:51 Lockloss

RELOCKING
21:55 Starting an Initial Alignment
22:28 USERMSG: disabled pump iss after 10 locklosses. Reset SQZ-OPO_ISS_LIMITCOUNT to clear message.
    - Followed 73053, minus touching up the OPO temp, to fix and relock sqz
22:47 Initial Alignment finished, relocking
23:38 Observing

H1 General (Lockloss)
oli.patane@LIGO.ORG - posted 14:04, Sunday 19 November 2023 - last comment - 14:24, Monday 20 November 2023(74294)
Lockloss

Lockloss 11/19 21:51UTC (Lockloss page has been offline for 85 mins as of when I'm writing this so I cannot link the lockloss page for this lockloss)

Not sure of cause. An earthquake was rolling through (crosshairs at lockloss time) but we had gotten through the biggest part fine. Microseism and wind are still high so it's possible that the lockloss was caused by their combination.

Images attached to this report
Comments related to this report
camilla.compton@LIGO.ORG - 14:24, Monday 20 November 2023 (74317)

1384379987 Looks like some glitches in DARM / Length in the 500ms before, as seen in 73818.

H1 General
oli.patane@LIGO.ORG - posted 11:23, Sunday 19 November 2023 - last comment - 15:14, Sunday 19 November 2023(74290)
Ops DAY Midshift Update

Been having trouble relocking all day. Wind and microseism have been high and increasing, and the ALSX beat note strength being below the minimum has been throwing ALS-XARM into error (74289). Lowering the beat note strength threshold has been working for getting us past LOCKING_ARMS_GREEN, but we have now lost lock at CARM_OFFSET_REDUCTION three times, each time having a lot of EX and other suspensions saturating and causing us to lose lock.

Luckily the wind is starting to lower a bit, so I'll let the detector sit in DOWN for a bit and then hopefully relocking will go much better.

Comments related to this report
oli.patane@LIGO.ORG - 15:14, Sunday 19 November 2023 (74291)ISC, OpsInfo

21:01 Observing again at last!

Edit: I accepted the lower value for the ALS X beat note strength in SDFs for now because if the oscillating continues, we would possibly be back down around -10dB again a couple more times before Monday morning and cause the ALS_XARM node to keep erroring when trying to lock. Tagging OpsInfo and ISC because this needs to be reverted to -10 when the issue is solved.

Images attached to this comment
H1 ISC (OpsInfo)
oli.patane@LIGO.ORG - posted 11:10, Sunday 19 November 2023 - last comment - 13:50, Sunday 19 November 2023(74289)
ALSX Beat Note Strength Issues

I've been getting Guardian ALS_XARM node in error when trying to lock today, something that Austin also got overnight. It looks like ALS-X_FIBR_A_DEMOD_RFMON has been worming its way down(attachment1, attachment2) towards the minimum beat note strength of -10dB since Nov 17th 16:30ish UTC. I have temporarily lowered the beat note strength minimum to be -20dB so that we can lock, but going off of the last time this happened, 70951, the ALSX laser temp and current are going to need to be adjusted.

FRS ticket already opened on this issue: FRS#28430

Images attached to this report
Comments related to this report
oli.patane@LIGO.ORG - 13:50, Sunday 19 November 2023 (74293)

Last time this happened (70951), it was determined to not have been caused by EX VEA temps, but this time it definitely looks like the fluctuation in temperature and/or the new temperature might have been what set it oscillating. It's come back up and is now at positive 3, but might go back down in the next 4.5 - 5.5 hours like it has for the past few days. Every 'bump' spans between ~9 to 11.5 hours long.

Images attached to this comment
H1 AOS (DetChar)
sheila.dwyer@LIGO.ORG - posted 15:22, Friday 17 November 2023 - last comment - 16:15, Friday 17 April 2026(74256)
SQZ angle sweeping

Naoki, Camilla, Sheila

We had another look at some signals that we could use to track or servo the SQZ angle.  

We turned the ADF back on, making a line at 1.3kHz in DARM, and tuned the demod phase for the ADF so that the SQZ angle readout was 0 for the angle we've been using for observing in this lock.  Camilla added a bandstop at 1.3kHz in the SQZ BLRMS4, sum and null channels and checked that the ADF isn't dominating those.  We did a sweep of the CLF6 demod phase like this, with the dither amplitude reduced comapred to Wed (0.01 instead of 0.03 CLK gain).  

In the first attachment we are using BLRMS 4 to demodulate for the noise lock signal, the second cursor shows that the zero crossing of the ADF SQZ angle and the noise lock both correspond roughly with the minimum of noise in the BLRM4 (1kHz BLRMs).  This isn't the same phase as the one that minimizes the brown trace, a blrms centered at 350 Hz. This means we have a frequency dependence of the SQZ angle, so we should probably look into things like the SRCL offset that might be causing this. 

Naoki then turned the ADF phase with the SQZ angle set to minize BLRMS3, as shown in the second screenshot from Camilla (at the begining here you can see that Naoki set the ADF phase so SQZ ANG was zero for CLF demod phase of 145).  We repeated the sweep and see that we have lower SNR for the noise lock using this lower frequency BLRMS, and that the ADF as it is now wouldn't make a good error signal for this sqz angle because it doesn't really go negative. 

Detchar:  We are planning to leave the ADF on over the weekend, which will create a line at 1.3kHz.  We are hoping to use this to track changes in the squeezing angle over time.  

 

Images attached to this report
Comments related to this report
camilla.compton@LIGO.ORG - 12:17, Monday 20 November 2023 (74316)

The ADF calculated SQZ angle (H1:SQZ-ADF_OMC_TRANS_SQZ_ANG) seems to follow our SQZ BLRMs over the weekend, plot attached. Unsure why the SQZ is different lock to lock, i.e. sometimes changes over first 6 hours (1 day ago) and is sometimes is stable (-12 hours ago).

Images attached to this comment
sheila.dwyer@LIGO.ORG - 16:15, Friday 17 April 2026 (89940)

The Wensday work that is referenced above is described with helpful detail in 74223

H1 SQZ
camilla.compton@LIGO.ORG - posted 17:39, Wednesday 25 October 2023 - last comment - 15:00, Monday 20 November 2023(73747)
NLG Sweep with IFO

Sheila, Naoki, Camilla

We took FDS, Anti-SQZ and Mean SQZ data at different NLGs: 14, 17, 43, 82 and 123. We think we see evidence of frequency noise. We see less squeezing at high NLGs: around 4.5dB of SQZ at NLG < 43.1 but at NLG 81.9 and 122.8, squeezing at 1kHz reduced to 3.6 and 3.1dB. Attached Plot.

Each time we optimized SQZ angle around 1kHz. Because of this, the low frequency increased noise at high NLGs could be due to incorrect rotation, rather than phase noise. 

 
Time (UTC)
Demod phase
DTT ref#
NLG
SQZ dB at 900Hz
FDS
21:23:40-21:26:00
152.25
3
13.9
-4.3
FDAS
21:27:30- 21:32:30
228.10
6
13.9
15.2
Mean
21:33:47- 21:38:00
-
7
13.9
12.7
No SQZ
21:40:00 -21:44:50
-
0
-
 
FDS (not optimized) 
21:51:20 -21:54:30
164.57
Deleted
43.1
-4
FDAS (Left ASC off) 
21:56:00 -21:59:45
226.2
5
43.1
19.1
FDS (retake)
22:03:00 -22:06:13
167.42
4
43.1
-4.3
Mean
22:06:45 -22:09:52
-
8
43.1
16.9
FDS
22:37:50- 22:41:00
172.16
9
122.8
-3.1
FDAS
22:42:43-22:45:45
217.67
10
122.8
24.7
Mean
22:46:35 - 22:49:55
-
11
122.8
 
FDS
22:57:30- 22:59:40
169.31
12
81.9
-3.6
FDAS
23:01:20 - 23:02:55
219.56
13
81.9
22.7
Mean sqz
23:03:26 - 23:05:45
-
14
81.9
20
FDS (left IFO here)
23:19:38
154.14
15
17.1
-4.5

Data saved in /camilla.compton/Documents/sqz/templates/dtt/20231025_GDS_CALIB_STRAIN.xml

Time (UTC)
Unamplified OPO OPO_IR_PD_LF
(Scan Seed)
 
OPO trans (uW)
OPO Temp (degC)
Amplified (maxium) H1:OPO_IR_PD_LF
(Scan OPO PZT)
NLG 
(Amplified / (Unamplified Peak Max - Unamplified Dark Offset))
Peak Maximum Dark Offset
21:15
786e-6
-16e-6
80.54 uW
31.446
0.0112
13.9
21:44
 
 
100.52
31.414
0.0346
43.1
22:17
 
 
120.56
31.402
0.160
199.5 (SQZ loops went unstable) 
22:30
 
 
110.5
31.407
0.06617
82 (didn't use) 
22:36
 
 
115.46
31.405
0.09853
122.8
22:52
 
 
110.5
31.407
0.0657
81.9
22:09
769e-6
-15e-6
80.5
31.424
0.0137
17.1

Vicky and Naoki did an NLG sweep on the Homodyne in 73562.

Images attached to this report
Comments related to this report
camilla.compton@LIGO.ORG - 10:49, Sunday 29 October 2023 (73801)

Edits to NLG table in above alog: Last measurement was at 22:09 23:09 UTC. For measuring unamplified we scanned OPO PZT and for amplified we scanned SEED PZT. 

Setup for each SQZ Measurement:

For FDS: Took SQZ_MANAGER guardian to FREQ_DEP_SQZ and rotate SQZ angle to give best squeezing at 900Hz. 
For ASQZ: Turn off IFO ASC and FC ASC, rotate SQZ angle to give worst squeezing at 900Hz. 
For Mean SQZ: Took SQZ_MANAGER guardian to FREQ_INDEP_SQZ and SQZ_LO_LR to DOWN. 
 
To measure unampilifed signal for NLG:
  • SQZ_MANAGER to NO_SQUEEZING and SQZ_OPO to DOWN
  • Scan the OPO PZT with PZT_1 Enable, Use trigger: No (via SQZ Overview > HAM7 > OPO IR > OPO PZT1)
  • Should see mode scan on H1: SQZ-OPO_IR_PD_LF_OUT. Measure the highest peak as "Unamplified" and the zero level as "Unamplified Dark Offset".
  • Only need to this once (we did at start and end)
To change to another NLG:
  • SQZ_MANAGER to NO_SQUEEZING
  • Change opo_grTrans_setpoint_uW in sqzparams.py (higher power = higher NLG)
  • Reload SQZ_OPO
  • SQZ_OPO to LOCKED_CLF_DUAL_NO_ISS  then LOCKED_CLF_DUAL
    • Check it locks with correct OPO TRANS power, may need to adjust SHG power with waveplate
  • SQZ_OPO to LOCKED_SEED_NLG
  • SQZ_CLF to DOWN
  • Adjust temperature to maximize SQZ-OPO_IR_PD_LF_OUT (with SEED_PZT scanning enabled)
  • Record SQZ-OPO_IR_PD_LF_OUT maximum it is "Amplified" signal.
    • Can now calculate NLG = Amplified / (Unamplified - Unamplified Dark Offset).
  • SQZ_MANAGER to SQZ_READY_IFO (with swap back from SEED to CLF)
victoriaa.xu@LIGO.ORG - 15:00, Monday 20 November 2023 (74318)

Attached are fits of squeezing loss, phase noise, and technical noise to this NLG sweep on DARM: we can infer 25-30% total SQZ losses, ~25 mrad rms phase noise, with technical noise almost -12 dB below shot noise.

Losses are tracked in the SQZ wiki and gsheet. Of the 30% total inferred loss, we expect 20%: that is 7.5% injection loss, and 13.7% readout loss.

Remaining ~10% mystery loss is compatible with mode-mismatch: in sqz-omc single bounce mode scans, LHO:73696, we estimate 8-15% mismatch, and we observe the frequency-dependence of the mismatch as we vary squeezer beam profile using PSAMS: 73400, 73621.

In the fits, [loss, phase noise, technical noise] are fit to the measured SQZ and Anti-SQZ, given the measured NLG, using equations from the aoki paper. "Loss from mean sqz" is the calculated loss from the measured NLG and measured mean-sqz dB; it is not a fit, but depends on the calibration from NLG to generated SQZ dB.

Some summary slides here show the progression of our loss hunting, which are so far lining up with Sheila's projections from 73395.

Images attached to this comment
Non-image files attached to this comment
H1 SQZ
victoriaa.xu@LIGO.ORG - posted 19:03, Wednesday 18 October 2023 - last comment - 15:29, Monday 20 November 2023(73562)
NLG HD sweep, HD mystery losses fully resolved, ~8dB HD SQZ with new crystal spot

Naoki, Sheila, Camilla, Vicky

Summary: After yesterday's crystal move LHO:73535, we re-aligned SQZT7, and now see 8 dB SQZ on the homodyne, up to measured NLG=114 without a phase noise turnaround! This fully resolves the homodyne loss budget, there is 0 mystery loss remaining on the homodyne, from which we can infer 0 mystery losses in HAM7. Back to the IFO afterwards, after 1 day at this new crystal spot, squeezing in DARM is about 4.5dB - 4.8dB, reaching almost 5dB at the start of lock.

We first re-aligned the homodyne to the IFO SQZ alignment, which reached 4.8dB SQZ in DARM yesterday, so we are more confident the alignment back through the VOPO is not clipping. In yesterday's measurements, we had a sign error in the FC-ASC offloading script, which brought us to a bad alignment with limited homodyne squeezing, despite high 98% fringe visibilities. Attached is a screenshot of homodyne FC/ZM slider values with FC+SQZ ASC's fully offloaded (correctly), to which the on-table SQZT7 homodyne is now well-aligned. After Sheila re-aligned the homodyne to the screenshotted FC/ZM values, fringe visibilities are PD1 = 98.5% (loss 3.1%), PD2 = 97.8% (loss 4.2%).

We then did an NLG sweep on the homodyne, from NLG=2.4 (opo trans 20uW) to NLG=114 (opo trans 120uW). Measurements below and attached as .txt, DTT is attached, plots to follow.

-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

unamplified_ir = 0.0014 (H1:SQZ-OPO_IR_PD_LF_OUT_DQ with pump shuttered)

NLG = amplified / unamplified_ir (opo green pump un-shuttered)

@80uW pump trans, amplified = 0.0198 (at start, 0.0196 at end) --> NLG 0.0198/0.0014 ~ 14

@100uW pump trans, amplified = 0.0046 (at start, 0.0458 at end) --> NLG 0.046/0.0014 = 33

@120uW pump trans, amplified = 0.16 --> NLG 0.16/0.0014 = 114

@60uW pump trans, amplified = 0.011 (at start, 0.0107 at end) --> NLG = 7.86

@40uW pump trans, amplified = 0.0059 (at start, 0.0059 at end) --> NLG = 4.2

@20uW pump trans, amplified = 0.0034 (at start, --- at end) --> NLG = 2.4

trace reference opo_green_trans
(uW)
NLG SQZ dB CLF RF6 demod angles (+)
LO shot noise @ 1.106 mA, -136.3 dB 10 80 14    
Mean SQZ 11     +13  
SQZ 12     -8.0 162.0
ASQZ 13     +16 245.44
NLG=33   100 33    
Mean SQZ 14     +16.7  
SQZ 15     -8.0 170.5
ASQZ 16     +19.9 237.85
NLG = 114   120 114    
Mean SQZ 17     +22.5  
SQZ 19     -8.0 177.98
ASQZ 18     +25.6 230.13
NLG = 7.9   60 7.9    
Mean SQZ 20     +9.7  
SQZ 21     -7.7 154.28
ASQZ 22     +12.4 253.83
NLG = 4.2   40 4.2    
LO SN check 4       ~0.1dB lower?
Mean SQZ 23     +6.8  
SQZ 24     -6.3 140.6
ASQZ 25     +9.6 262.64
NLG = 2.4   20 2.4    
Mean SQZ 26     +3.8  
SQZ 27     -4.8 135.45
ASQZ 28     +6.3 -100.5
LO shot noise @ 1.06 mA, 29        

All measurements had PUMP_ISS engaged throughout; we manually tuned the ISS setpoint for different NLGs. For low NLG (20uW trans) we manually engaged ISS. LO power (shot noise) drifted ~5% over the measurement, see trends.

NLG Sweep Procedure:

DTT saved in $(userapps)/sqz/h1/Templates/dtt/HD_SQZ/HD_SQZ_8dB_101823_NLGsweep.xml

Images attached to this report
Non-image files attached to this report
Comments related to this report
victoriaa.xu@LIGO.ORG - 11:06, Thursday 19 October 2023 (73581)

Using Dhruva's nice plotting code for NLG sweeps from LHO:67242, here are some plots of squeezing vs. NLG, and calibrating the OPO lasing threshold and various green powers at this new crystal spot. Data & updated plotting code attached.

NLG sweep data summary here:

SHG Launched (mW) OPO Green Refl (mW) OPO Green Trans (uW)  NLG 
 
Mean SQZ (dB) SQZ (dB) Anti-SQZ (dB)
10.8 1 80 14 13 -8.0 16
13.3 1.3 100 33 16.7 -8.0 19.9
16 1.5 120 114 22.5 -8.0 25.6
8.4 0.8 60 7.9 9.7 -7.7 12.4
5.9 0.6 40 4.2 6.8 -6.3 9.6
3.5 0.4 20 2.4 3.8 -4.8 6.3

To-do: Look into the fits of loss & technical noise.

  • Technical noise here fits to -8.8dB below LO shot noise @ 1.1mA (-136dB, dark noise @ -159dB) -- check this is correct/real. For reference with the old homodyne PD in Feb 2023, LHO:67223 with 5.5dB sqz, technical noise fit to -9.6dB below LO shot noise @ 1.08mA (calibration suspect, HD_DIFF_DC_OUT was around -134-135dB, dark noise @ - 157dB). Would be interesting if this about -10dB technical noise below shot noise on the homodyne is real technical noise injection from the squeezer, and/or if it can be improved with upcoming squeezer laser PMC installation.
  • Loss estimate from NLG sweep - our homodyne loss budget accounts for 13% losses (after scaling back all over-estimations of losses), but this sweep fits to 7% losses, so we should resolve this over-estimated loss discrepancy.
Non-image files attached to this comment
michael.zucker@LIGO.ORG - 14:58, Thursday 19 October 2023 (73599)

Outstanding work, well done!

victoriaa.xu@LIGO.ORG - 15:29, Monday 20 November 2023 (74319)

Attached here is a re-fitting of this homodyne NLG sweep, which fits [loss, phase noise, technical noise] to measured SQZ+ASQZ, given measured NLG. It also shows the calculated loss from measured mean-sqz and NLG (which relies on accurate calibration of NLG --> generated SQZ dB). The same fitting was done for NLG sweeps on DARM the following week LHO:73747.

The previous anlaysis was fitting [phase noise, technical noise] using the loss calcuated from mean-squeezing. Compared to the earlier analysis, I think these fits here are closer. 

We budgeted 13% HD loss for this homodyne measurement:  1 - [0.985(opo) * 0.96(ham7) * 0.98(sqzt7 optics) * 0.977(HD PD QE) * 0.96((visibility~98%)**2)] = 13%.

This fit to the NLG HD sweep suggests ~11% homodyne loss, 7 mrad rms phase noise, with technical noise about -10 dB below 1.1 mW LO shot noise. Note HD dark noise is -22 dB below shot noise, suggesting mystery technical noise on the homodyne.

Images attached to this comment
Displaying reports 16221-16240 of 88368.Go to page Start 808 809 810 811 812 813 814 815 816 End