Displaying reports 1761-1780 of 77270.Go to page Start 85 86 87 88 89 90 91 92 93 End
Reports until 02:09, Saturday 11 May 2024
H1 PEM
michael.landry@LIGO.ORG - posted 02:09, Saturday 11 May 2024 - last comment - 09:29, Monday 13 May 2024(77777)
Aurora over LHO

The May 10 CME impact on Earth's magnetic field has produced a category G5 geomagnetic storm event, yielding magnificent aurorae here in the Tri-Cities and the world over.  Photos attached.

Images attached to this report
Comments related to this report
jeffrey.kissel@LIGO.ORG - 09:29, Monday 13 May 2024 (77807)EPO
Nice photos! Tagging EPO!
H1 General
oli.patane@LIGO.ORG - posted 01:09, Saturday 11 May 2024 (77776)
Ops EVE Shift End

TITLE: 05/11 Eve Shift: 2300-0800 UTC (1600-0100 PST), all times posted in UTC
STATE of H1: Observing at 155Mpc
INCOMING OPERATOR: Ryan C
SHIFT SUMMARY: Currently Observing and have been Locked for 5 hours. We had had a lockloss when I first came in and then another lockloss a couple of hours later. Relocking was overall pretty hands-off, with the same relocking issues we've been having lately (lockloss from LOCKING_ALS and ENGAGE_DRMI_ASC). We had a superevent S240511i at 03:15 UTC, a little over one minute after getting back into Observing after the second lockloss (phew)!
LOG:

23:01UTC Detector relocking after lockloss
23:12 Lockloss from LOCKING_ALS
23:14 Lockloss from LOCKING_ALS
23:24 Lockloss from PRMI_ASC
00:16 NOMINAL_LOW_NOISE
00:18 Observing

01:14 Lockloss
01:26 Lockloss from ENGAGE_DRMI_ASC
01:29 Started an Initial Alignment
01:54 Initial Alignment done, relocking
03:11 NOMINAL_LOW_NOISE
03:13 Observing
03:15 Superevent S240511i https://gracedb.ligo.org/superevents/S240511i/view/

08:10 Have been locked for 5 hours

H1 General (Lockloss)
oli.patane@LIGO.ORG - posted 18:15, Friday 10 May 2024 - last comment - 21:28, Friday 10 May 2024(77772)
Lockloss

Lockloss @ 05/11/2024 01:14 UTC

Comments related to this report
oli.patane@LIGO.ORG - 20:17, Friday 10 May 2024 (77773)

03:13 UTC Observing

 

oli.patane@LIGO.ORG - 21:28, Friday 10 May 2024 (77774)

test mass ndscope Unsure if LL seen by EX L3 before DARM or not.

Images attached to this comment
LHO General
ryan.short@LIGO.ORG - posted 16:38, Friday 10 May 2024 (77769)
Ops Day Shift Summary

TITLE: 05/10 Day Shift: 1430-2330 UTC (0730-1630 PST), all times posted in UTC
STATE of H1: Lock Acquisition
INCOMING OPERATOR: Oli
SHIFT SUMMARY: SQZ issues were fixed this morning so H1 has recovered some range since last night. Two locklosses today, but otherwise quiet. H1 is currently relocking at ACQUIRE_DRMI.

LOG:

Start Time System Name Location Lazer_Haz Task Time End
16:57 FAC Kim H2 n Technical cleaning 17:14
18:20 AOS Betsy, Mitchell Opt Lab n Checks 18:31
22:00 SQZ Kar Meng Opt Lab n Checking equipment 22:12
H1 General
oli.patane@LIGO.ORG - posted 16:30, Friday 10 May 2024 (77770)
Ops EVE Shift Start

TITLE: 05/10 Eve Shift: 2300-0800 UTC (1600-0100 PST), all times posted in UTC
STATE of H1: Lock Acquisition
OUTGOING OPERATOR: Ryan S
CURRENT ENVIRONMENT:
    SEI_ENV state: CALM
    Wind: 6mph Gusts, 2mph 5min avg
    Primary useism: 0.02 μm/s
    Secondary useism: 0.12 μm/s
QUICK SUMMARY:

Lost lock right before I got in and we've currently had a few locklosses from ALS and DRMI, so if we lose lock again I'll start an initial alignement.

H1 General (Lockloss)
ryan.short@LIGO.ORG - posted 16:08, Friday 10 May 2024 - last comment - 17:21, Friday 10 May 2024(77768)
Lockloss @ 23:01 UTC

Lockloss @ 23:01 UTC - link to online lockloss tool

No immediately obvious cause; DARM1 saw the first motion as usual.

Comments related to this report
oli.patane@LIGO.ORG - 17:21, Friday 10 May 2024 (77771)

00:18 UTC Observing

H1 ISC (OpsInfo)
ryan.short@LIGO.ORG - posted 14:34, Friday 10 May 2024 (77767)
CHECK_VIOLINS State Now Turns on 2W Damping

I've added a parameter into the CHECK_VIOLINS_BEFORE_POWERUP state which engages 2W damping settings if the saturation checker flags that it's not safe to power up. Once someone requests ISC_LOCK to a higher state to continue locking and the saturation check passes, all damping will be turned back off before continuing.

Speaking of 2W damping settings, I've edited the VIOLIN_DAMPING Guardian to now allow 2W damping for all four quads (previously settings were only enabled for ITMX and ETMX). For modes which have dedicated settings for 2W damping, the Guardian will use those, and for those that don't, the otherwise nominal settings will be used. We had the opportunity to test these 2W damping settings earlier in the week when violins were rung up, and all of them appear to work well, motivating these improvements to our automated damping.

H1 CAL (CDS, ISC)
jeffrey.kissel@LIGO.ORG - posted 14:08, Friday 10 May 2024 - last comment - 09:27, Monday 13 May 2024(77735)
Update on OMC DCPD Transimpedance Amplifier Characterization Post OMC Swap: No Forward Progress, Only Sideways Progress and Broken Electronics
J. Kissel [with support from L. Dartez, F. Llamas, and D. Schaetzl]

Executive Summary
In Feb 2024, we identified that both the DCPDA and DCPDB channels of the in-vacuum OMC DCPD transimpedance amplifier (TIA) frequency response dropped by 0.3% below ~25 Hz (LHO:75986) and that the frequency-dependent balance of the DCPDs changed in a different way at the 0.2 dB or 10^(0.2/20) = 2.5% level (LHO:76232). Since then, with what limited person power we had, we've been scrapping together ~2 hours of time during maintenance days to try to re-characterize the response of the TIA in order to fit and re-compensate the response, and then re-re-balance the DCPDs. We've also done a lot of thinking in between. This aLOG covers the saga since then, and executive summary is that we don't have good enough measurements of the TIA response yet, we have broken the functionality of the TIA monitor path in the S2300003 D2200215 whitening chassis, and it needs to be replaced in order to move forward with characterizing the in-vacuum TIA.

Timeline
    Editorial note: You'll find that the story is fraught with missing institutional knowledge due to lack of person power and interferometer time. It would pay for us to have had a spare TIA system *out* of vacuum to test our characterization methods and train new staff.

    . 2024-02-26: In LHO:75986 after measuring twice to ensure we weren't being fooled by vacuum pressure and thermalization, we use the remote, DAC driven characterization setup -- measuring the product of the TIA, the whitening, the whitening compensation, and the TIA compensation -- we conclude that the response of the TIAs has changed, and is no longer as flat as it was the last time we measured it, which was Jul 2023 (LHO:71225).
  
    . 2024-02-26: That same day, we measured the TIA response "alone" using the same remote, DAC driven measurement, but instead turning OFF the analog whitening, and the compensation for both -- as the first attempt to "just get the data we need quickly in order to get the job done." We never processed this data because (a) we didn't have the person power, and (b) we later remember that it's not good enough to use for updating the compensation because this measurement doesn't contain the TIA response "alone," but instead also 
        - several computational delays,
        - the anti-imaging chassis from the DAC,
and it's limited by the 16 kHz sampling frequency of the DAC system so we can't characterize everything we need for the TIA, given that there are 
        - 2x 10kHz poles in the TIA itself, and
        - 1x 44kHz pole in the whitening chassis, even with the z:p = 1:10 Hz whitening filter turned OFF.

    . From 2024-02-26 thru 2024-04-29: I worked offline with Dean to try to understand the electrical grounding paths in the OMC DCPD signal chain, because it's the only think I can imagine that could cause a *common* change in frequency response to both channels. I now have a much better understanding of the electrical grounding in the OMC signal chain, have produced G2400755, and Dean has made an update to the O4 OMC DCPD wiring diagram accordingly D2100716-v3. However, I conclude that any potential change in the in-vacuum electrical grounding during the OMC swap just is not possible, and the DCPDs are tied to earth ground by and at the in-air whitening chassis, so this is likely NOT the source of the frequency dependent change.

    . 2024-04-02: Louis and Francisco finally muster up the person power to go out to the HAM6 ISC-R5 rack to repeat the long, arduous, measurement needed. However, they misinterpret my diagrams from the 2023-03-06 and 2023-03-10 measurements (LHO:67801 and LHO:68167) against the reality of the D2200215-style whitening chassis, and drive 1 [V_pk] source voltage backwards against the monitor path's buffer amplifiers. The first attachment shows
        - page 1: how one should characterize the TIA response, buffer amplifiers, and measurement setup
        - page 2: how one should characterize the measurement setup and buffer amplifiers in order to divide it out of the page 1 measurement, and
        - page 3: how the measurement setup and buffer amplifier characterization was misinterpreted on 2024-04-02.
      Further, as is standard practice, they took SR785 setup notes from "the most recent good measurement available in the SVN," which was the 2023-03-10 data set where Hsiang-yu and I were exploring options. Thus, they used some settings that were meant for testing rather than those which get the best results out of the measurement. More on this later.

    . 2024-04-09: With Louis and Francisco away for the Solar Eclipse, I head out to the floor to take another crack at it. While I get the measurement setup right, I don't know yet, but I still use bad SR785 settings from the 2023-03-10 date. As such, I get results that are too noisy below 50 Hz to make fits. In the second attachment, I show a quad bode-plot of the DCPDA (page 1) and DCPDB (page 2) where I divide 
       - the Last Best Good measurement -- the 2023-03-06 data and
       - the 2024-04-09 data
      against the model fit from that data (see poles and zeros listed in LHO:67809). While one can see hints of the ~0.3% change, it's just too noisy to be confident that we can get a good fit from it.

    . from 2024-04-09 thru 2024-04-23: With review from Louis, we think hard about why the measurement is noisy. I get a hunch about the SR785 parameters' settle time and settle cycles vs. step (or impulse) response of the TIA response itself, insisting that the settle time of 100 [msec] we've been using since 2022 is not long enough for what the impulse response of the TIA circuit is (because we're perpetually trying cut corners and speed up this measurement). I make a bunch of plots that show the step response of the TIA compared with the bad 2023-03-10 settle settings, but while looking back through my notes I also discover that a difference between the 2023-03-06 and 2023-03-10 measurements is that I had decreased the SR785 source voltage from 4 V_pk to 1 V_pk. The third attachment shows plots that show the good vs. bad SR785 settings and why.

       - page 1: this compares the settle parameters, visually showing the crossover frequency where, as the swept sine measurement sweeps down in frequency, the SR785 switches over from using settle time to settle cycles, given that it takes the larger of (settle time) vs. (settle cycles / frequency). The dashed lines are the *bad* too little settle time or cycle parameters, and the solid lines are the good, just long enough, settle parameters. Turns out the cross-over frequency isn't that different, but it's important to understand *where* it is, so that you understand *where* in the frequency response this switch happens so you know which parameter to increase when you see noise in the TIA response at a certain frequency.

       - page 2: Now thinking about the *amplitude* of the excitation vs. the step response of the circuit, I show the magnitude of the modeled response of the TIA with a few points highlighted. Note that, rather than displaying the TIA response magnitude in its "fundamental" units of [Volts output / Amps of current from DCPDs] = [V/A], I display it in the units of this characterization measurement, which is through the 100e3 [V/A] series resistance of the "test" input. Given the intellegent design of the circuit, that means than this measurement of the TIA -- which has transimpedance of 100e3 [V/A] in the gravitational wave band -- has a magnitude of 1 [V/V] in the gravitational wave band. 
            :: the blue "x" and yellow "o" are the cross-over frequencies of the "too little" BAD and the "just enough" GOOD settle params. Good to know that these are *not* in the middle of the 25 Hz resonance feature. But, these cross-over frequency that are the lower bound of a region where we must compare the "desired response" amplitude with the "step response" amplitude after the chosen amount of settle time -- if the "step response" amplitude is still a large fraction of the "desired response" amplitude after (settle cycles/frequency) amount of time, then the data is going to be incoherent (since the circuit is still "ringing" from the incoherent "excitation" that is the step-change in frequency).
            :: the red square, at 102.4 kHz represents the lowest magnitude point in the "settle time" regime. This is where, again we'll need to make sure that amplitude of the "desired response" amplitude at that frequency is larger than the step response amplitude after the total settle time is complete.
            :: the black triangle, at "DC," in this case 0.1 Hz, shows the magnitude of the TIA response at low frequency, at 0.002 [V/V]. This is relevant because the step response will asymptote to this value. Also it's the lowest magnitude point of the whole transfer function, but it still needs to be extremely well resolved -- just as much as the ~1.2 [V/V] region around 25 Hz. So one needs to be conscious of the excitation amplitude (be it 1V, or 4V), filtered through this circuit (thus 0.002 [V_pk] or 0.008 [V_pk]) with respect to the ADC noise of the SR785 at the given input range. Note, the input "range" (which really the amount of attenuation of the input) for the A and B inputs of CH2 has been auto-ranged in the past with 4 [V_pk] source (and locked as static during the measurement once found) to be +16 [dBVpk] == 6.3 [V_pk]. The noise floor at the range setting, at this low of frequency is ~0.1 [mV] (see e.g. figure B3 of Hoffman 2009), so a non-negligible fraction of 0.002 [V_pk] response to 1 [V_pk] source, which is why increasing the source amplitude by a factor of 4 is necessary.

        - Page 3: This shows the full step response of the TIA circuit as a function of time. One can see, as expected, since the maximum gain of the circuit is ~ 1 [V/V], the maximum amplitude of the step response roughly hits the value of the source input within ~100 [usec] of the step. However, we can see that the step response doesn't appreciable attenuate until 0.1 [sec], or 100 [msec]!

        - Page 4: Here, I zoom in on the amplitude and time of the step response, and compare the BAD SR785 settings with the good SR785 settings. Namely, at the two interesting frequency points from page 2 -- the settle time/settle cycle crossover, and the lowest magnitude point in the settle time region. With 1 V input (page 4), at the bad 101 [msec] settle time, and (1/crossover freq) = 101 [msec] crossover frequency cycle duration, the "desired response" amplitude of circuit is comparable to step response at 100 [msec]. This is why the ~25 Hz region of the measurement has been "noisy" all these years we've been measuring the TIA, and why we need to increase the settle time to 250 [msec]. The 102.4 kHz region was still OK, because there's upwards of 10e3 cycles in 100 [msec], oodles of averages to get good SNR. 

        - Page 5: This shows the same plot, but with 4 [V_pk] source voltage step input and settle time of 250 [msec] and 2 settle cycles. Here, we see that it's not an awesome amount of margin between the step response and the desired circuit response at the interesting frequencies, but it's *definitely* better than 100 [msec] and 1 cycle.

     . 2024-04-23: So -- we're good right? We take the hit in patience / IFO time, and increase the settle time from 101 [msec] to 250 [ms], settle cycles from 1 to 2, and increase the excitation amplitude from 1 to 4 [V_pk], and we'll get the bet measurement ever, right?  No. The fourth attachment shows my utter dismay in that -- yes, while the measurement is much less noisy after having improved the SR785 settings -- BUT the magnitude and frequency response of the 2024-04-23 is subtlety different than the 2023-04-09 and 2023-03-06 measurements at the 0.5 [%] / 1 [deg] level. And that one can trace that difference down to that the measurement setup characterization has a non-linear response to excitation voltage.
        - Pages 1 and 3 show the quad bode the same quad bode plot, but now it shows the 2023-03-06, 2024-04-09, and the 2024-04-23 measurement all together. See the ~1% magnitude discrepancy?
        - Pages 2 and 4 show a regular bode plot showing the measurement setup response for all three 2023-03-06, 2024-04-09, and the 2024-04-23 measurements. See the overall magnitude discrepancy, the high-frequency response discrepancy? Maddening.

     As long as we're going insane, Louis and I convince ourselves (via the data sheet) that the AD8672 op amps in this monitor path on the front-interface board, D2200043 of the D2200215 whitening chassis assembly (which are the same used in the main gravitational wave path on the primary functional board, D2200044) that "there's no way this op-amp can drive +/-4V with an input range of +/- 2.5V and an output drive of +/-3.7V!" ... but we were looking at the table for +/-5V operational state, rather than the +/-15 V operational state, which can happily receive +/-12V and drive +/-13V. But that's what drove me to the EE lab...

     . 2024-04-24 and 2024-04-25: Since we have spares of these OMC Whitening Chassis in the EE lab, I repeated this "measurement setup" characterization on both channels of spare chassis serial numbers S2300002 and S2300004. 
         - The fifth attachment shows the results the results of these 2x chassis, each with 2x channels, measurements as a function of input voltage. The spare chassis show no evidence of non-linearity in scale, or frequency-dependent magnitude or phase from 0.5 [V_pk] to 5 [V_pk], to the level of the ratio of the 0.5 [V_pk] to all other amplitude excitations being noisy around the 1.0 +/- 0.0001 (or 0.01 [%] or 1 [HOP]) magnitude and 0 +/- 0.02 [deg] in phase across a 10 Hz to 102.4 kHz span.
         - The sixth attachment compares across all four of these channels at 4.0 [V_pk] source amplitude. All four instantiations of this monitor path in spare chassis show no difference from each other at the same level, using S2300002 CHA as the reference, the ratio of the other channels to it is within 1 +/- 0.0001 in magnitude and 0.1 [deg] in phase.

     . 2024-04-30: To put the nail in the coffin, I go out and measure the S2300003 chassis' monitor path as a function of input voltage. 
         - The seventh attachment shows the damning evidence -- that the frequency response and scale of this monitor path on the S2300003 chassis changes with input voltage, where the other two instantiations do not.
         - The eigth attachment shows that the response to 1 [V_pk] source has changed over time, as I compare the 2023-03-10 measurement setup characterization to the 2024-04-30 results.

And, while I can't definitely say that the incorrect measurement setup from 2024-04-02 borked the op-amps in the monitor path, it's the only smoking gun I have.

"OK, Jeff. You're lost. Why do we care so much about measuring 1.0 from the monitor path of this whitening Chassis?"
I hear you, I feel lost. But, in the world where we're trying to update the TIA compensation at the 0.3% level, and the scale and frequency dependence of the measurement tools you're using are non-linear and frequency dependent at the 0.5 [%] / 1.0 [deg] level in the 1 to 100 kHz region, and you're trying to fit for poles at ~10 kHz, then you're gunna get the wrong answer.

As such, to restate the executive summary we need to replace the S2300003 D2200215 OMC whitening chassis with one of the spares, in order to get a re-characterization of the OMC TIA after the OMC swap.

*sigh*
Non-image files attached to this report
Comments related to this report
jeffrey.kissel@LIGO.ORG - 09:27, Monday 13 May 2024 (77806)CDS
The scripts used to analyze the data and create the plots from the above aLOG live in
    /ligo/svncommon/CalSVN/aligocalibration/trunk/Common/Electronics/H1/
        DCPDTransimpedanceAmp/OMCA/S2100832_SN02/20240423/Data/
            quick_compare_data_20240423.m (rev 12938)

        DCPDWhitening/OMCA/S2300002/20240424/
            quick_plot_data_20240424.m (rev 12940)

Even though the name and location indicates that the script should only be a "quick plot/comparison of the data" from that measurement date, it does in fact contain the measurement-date- and whitening-chassis-serial-number- spanning analysis from this aLOG. 

Similarly, the data used in the above aLOG lives scattered across the folder structure,
    /ligo/svncommon/CalSVN/aligocalibration/trunk/Common/Electronics/H1/
        DCPDTransimpedanceAmp/OMCA/S2100832_SN02/
            20230306/Data/                         # Pre-OMC-swap data that was used for fitting. Does not have "measurement setup" data to divide out of measurement
            20230310/Data/                         # Pre-OMC-swap data where we tried to gather the measurement setup characterization, and also explored the excitation amplitude parameter space.
            20240402/Data/                         # Diagrams of bad measurement setup, and some notes about it (no data was saved)
            20240409/Data/                         # Noisy data and measurement notes with bad SR785 parameters from 20230310, and suspect TIA pickoff monitor path response in whitening chassis
            20240423/Data/                         # Good Data/SR8785 params, measurement notes, analysis scripts and suspect TIA pickoff monitor path response in whitening chassis
        
        DCPDWhitening/
            S2300002/20240424/Data/                # EE Shop, benchtop measurement to show TIA monitor pickoff path response is excitation amplitude independent
            S2300003/20240430/Data/                # With chassis still installed, in-situ measurement to show TIA pickoff path response depends on excitation amplitude
            S2300004/20240425/Data/                # EE Shop, benchtop measurement to show TIA monitor pickoff path response is excitation amplitude independent


Apologies for the mess. 
LHO General
tyler.guidry@LIGO.ORG - posted 13:30, Friday 10 May 2024 (77765)
EX Fan Reapair
Supply Fan 1 at EX had faulty bearing(s) which was causing excessive noise. Macdonald Miller was on site this week to carry out the repairs. The work began Monday (5/6) and was completed Tuesday (5/7). The duration of the work was approximately the same on both days (8am-4:30pm) with brief lunch breaks around noon. All of the work was contained to the mechanical space of the end station. 

T. Guidry
H1 General (Lockloss)
ryan.short@LIGO.ORG - posted 11:52, Friday 10 May 2024 - last comment - 00:23, Saturday 11 May 2024(77764)
Lockloss @ 18:35 UTC

Lockloss @ 18:35 UTC - link to online lockloss tool

No immediately obvious cause; ETMX saw the first erratic movement as usual.

Comments related to this report
ryan.short@LIGO.ORG - 13:48, Friday 10 May 2024 (77766)

H1 back to observing at 19:56 UTC.

oli.patane@LIGO.ORG - 00:23, Saturday 11 May 2024 (77775)
Images attached to this comment
LHO VE
david.barker@LIGO.ORG - posted 10:13, Friday 10 May 2024 (77760)
Fri CP1 Fill

Fri May 10 10:10:49 2024 INFO: Fill completed in 10min 45secs

Travis confirmed a good fill curbside. Note we are now in "summer mode" with TC-mins railing at -200C.

Images attached to this report
H1 General (EPO, ISC, OpsInfo, PEM, SEI)
thomas.shaffer@LIGO.ORG - posted 09:58, Friday 10 May 2024 (77759)
Quick Low Noise % vs Wind Speed for O4a

I made a quick plot of low noise % vs wind speed for O4a and thought it was interesting enough to share. This is by no means an analysis, just a quick look. Someone should probably do something similar to what Laurence and Sheila made for O3 for O4a (alog49682), but that someone is not me at the moment. Unlike on Figure 6 of Sheila's similar ER7 plot (T1500368), there isn't a clear drop off at a particular speed. This could be due to directional impacts and the wind fence, or a plethora of other factors. Figure 19 on this paper (<a href="https://dcc.ligo.org/DocDB/0150/P1800038/005/BRSLIGOCQG.pdf">P1800038</a> is again similar with O1 and O2 data, but in m/s. I'm very curious about looking around times of lock losses or failed acquisitions, but I'm not making promises if I'll ever get to those.

This is minute trends of the max wind, binned in integers, then taking the percent of time we were in ISC_LOCK states of >=600.

Images attached to this report
H1 General (SQZ)
ryan.short@LIGO.ORG - posted 08:49, Friday 10 May 2024 - last comment - 10:17, Friday 10 May 2024(77755)
H1 Out of Observing for SQZ Fixes

H1 was out of observing from 14:53 to 15:36 UTC to fix some issues with the SQZ system as we were seeing last night.

After Sheila adjusted the fiber polarization (alog77754), I reverted Corey's changes to the SQZ Guardians' nominal states (listed in his alog77740), then optimized the SQZ angle by running SQZ_MANAGER through 'SCAN_ALIGNMENT_FDS' and 'SCAN_SQZANG,' increasing BNS range by ~10Mpc.

Comments related to this report
ryan.short@LIGO.ORG - 10:17, Friday 10 May 2024 (77761)

After optimizing the SQZ angle, I accepted SDF diffs for alignment changes of ZM4 and ZM6 (screenshot attached).

Images attached to this comment
H1 ISC
jenne.driggers@LIGO.ORG - posted 13:12, Tuesday 07 May 2024 - last comment - 10:51, Friday 10 May 2024(77690)
SR2 alignment when beam centered on AS_C, before vs after SR3 shift

Sheila asked a good question the other day of, Did SR2 alignment change between the beginning of O4b (when things were still good) and when we had the bad losses through the OFI (when things were bad, before the big shift).  The answer: no, I don't think SR2 moved very much (according to its top mass osems) when the the losses through the OFI showed up. It did move about 10 urad in yaw (see table below), which I plan to look into further.

I looked at several times throughout the last few weeks when ALIGN_IFO guardian had just finished up state 58, SR2_ALIGN at 10 W, which it now does every time initial alignment is automatically run.  These should all be single bounce off of ITMY, with the beam centered on AS_C by adjusting the SR2 sliders, for some given SR3 slider position (nothing automatic touches the SR3 sliders). 

In the table, I summarize the SR3 and SR2 top mass osems.  I've got 3 categories of times for the IFO situation:

Note that this table is not chronological, since I've grouped rows by IFO situation rather than time.  The SR2 and SR3 osem values that are in bold are the ones to compare amongst each other.  There does seem to be a 10 urad shift in SR2 yaw between the April 21st and April 23rd times.  There are no other run-throughs of the SR2_ALIGN state of ALIGN_IFO between these times to check.  This SR2 yaw shift (which is consistent even when we revert sliders to the 'pre shift' values and run SR2_ALIGN) is notable, but not nearly as large as what we ended up using for steering around the spot in the OFI.

IFO 'situation' Date / time [UTC] AS_C NSUM value SR3 Pit [M1_DAMP_P_INMON] SR3 Yaw [M1_DAMP_Y_INMON] SR2 Pit [M1_DAMP_P_INMON] SR2 Yaw [M1_DAMP_Y_INMON]
(1) before EQ, before loss, before alignment shift 17 Apr 2024 00:19:00 0.0227 -281.5 -612.2 569.8 35.3
(1) after EQ, before loss, before alignment shift 21 Apr 2024 20:08:30 0.0227 -281.5 -611.9 571.9 35.3
(2) after EQ, after loss, before alignment shift 23 Apr 2024 23:10:00 0.0193 -281.9 -612.2 572.7 26.7
(2) after EQ, after loss, shift temporarily reverted to check 7 May 2024 18:10:00 0.0187 -282.3 -616.0 558.5 23.2
(3) after EQ, after loss, after alignment shift 25 Apr 2024 12:18:20 0.0226 -291.7 -411.1 599.6 1150.0
(3) after EQ, after loss, after alignment shift 7 May 2024 19:11:15 0.0226 -292.4 -408.8 597.6 1149.9

 

Comments related to this report
jenne.driggers@LIGO.ORG - 14:08, Tuesday 07 May 2024 (77693)

After a quick re-look, that 10 urad move in SR2 yaw seems to have come during maintenance, or sometime later than the time that the loss showed up. 

In the attachment, the vertical t-cursors are at the times from the table in the parent comment on April 21st and 23rd.  The top row is SR2 pitch and yaw, and the bottom row is SR3.  The middle row shows our guardian state (i.e. when we were locked), and kappa_c which is indiciative of when we started to see loss.  In particular, there are 3 locks right after the first t-cursor, and they all have quite similar OSEM values for SR2 yaw (also the times between locks are similar-ish).  Those three locks are the last one with no loss, one with middling-bad loss, and one with the full loss.  So, it wasn't until after we had our full amount of loss that SR2 moved in yaw.  I haven't double-checked sliders yet, but probably this is a move that happened during maintenance day.

Images attached to this comment
sheila.dwyer@LIGO.ORG - 10:51, Friday 10 May 2024 (77762)

I'm using Jenne's times above to do a similar check, but looking at times when ALIGN_IFO was in state 65 (SRY align) because in that state the AS WFS centering servos are on.  This state is run shortly after state 58, so I'll reuse Jenne's numbers to refer to times and the IFO situation.

This table indicates that changes in AS power are consistent between AS_C and the AS WFS, so the beam transmitted by OM1 and reflected by OM1 see similar losses.  This makes it seem less likely that a bad spot on OM1 is the problem (and points to probably being an issue with the OFI), although it's not impossible that a loss on OM1 is seen in the same way for transmission and reflection.

    AS_C sum AS_C normalized to first row AS_A sum AS_A normalized to first row AS_B sum AS_B normalized to first row
1 April 17 00:20:15 UTC 0.0626 1 5264   5104  
1 April 21 20:09:43 UTC 0.0629 1.005 5283 1.004 5114 1.002
2 April 23 23:34:15 UTC 0.0534 0.853 4595 0.873 4359 0.854
2 AS centering was not run this time            
3 April 25 12:19:36 UTC 0.0622 0.993 5209 0.989 5083 0.996
3 May 7 19:12:30 UTC 0.0624 0.997 5241 0.996 5118 1.003

 

H1 ISC
sheila.dwyer@LIGO.ORG - posted 09:20, Monday 06 May 2024 - last comment - 09:10, Monday 13 May 2024(77641)
DRMI PRCL gain change

Sheila, Jenne, Tony, Camilla

We've had locklosses in DRMI because the PRCL gain has been to high when locked on REFL1F.  Tony looked and thinks that this started on 77583, the day of our big shift in the output alignment.

Today we acquired DRMI with half the gain in the PRCL input matrix for 1F, this one acquisition was fast.  I've attached the OLG measurements for PRCL and MICH after the change. 

Tony is working on making histograms of the DRMI acquisition times, before the 23rd, from the 23rd to today, and eventually a histogram from today for the next few weeks to evaluate if this change has an impact on the DRMI acquisition times.

Jenne also found that it seems out POP18 build up seems higher in DRMI since the 23rd.

Images attached to this report
Comments related to this report
jenne.driggers@LIGO.ORG - 11:45, Friday 10 May 2024 (77763)

I'm no longer quite so sure about the conclusion that Pop18 is higher, or at least enough to really matter.

Here are 2 screenshots that I made extremely quickly, so they are not very awesome, but they can be a placeholder until Tony's much more awesome version arrives.  They both have the same data, displayed 2 ways.

The first plot is pop18 and kappaC versus time.  The x-axis is gpstime, but that's hard to interpret, so I made a note on the screenshot that it ranges from about April 20th (before The Big Shift) to today.  Certainly during times when the optical gain was low, Pop18 was also low.  But, Pop18 is sometimes high even before the drop in optical gain.  So, probably it's unrelated to The Big Shift.  That means that the big shift in the output arm is not responsible for the change in PRCL gain (which makes sense, since they should be largely separate).

The second plot is just one value versus the other, to see that there does seem to be a bit of a trend that if kappaC is low, then definitely Pop18 is low.  But the opposite is not true - if pop18 is low kappaC isn't necessarily low.

The last attachment is the jupyter notebook (you'd have to download it and fix up the suffix to remove .txt and make it again a .ipynb), with my hand-typed data and the plots.

Images attached to this comment
Non-image files attached to this comment
sheila.dwyer@LIGO.ORG - 09:10, Monday 13 May 2024 (77805)

I actually didn't load the guardian at the time of this change, so it didn't take effect until today.

So, we'd like histograms of DRMI acquitisiton times from before April 23rd, from April 23rd until today, and for a few weeks from today.

Displaying reports 1761-1780 of 77270.Go to page Start 85 86 87 88 89 90 91 92 93 End