Today I took "unbiased" OLGs of MICH P and MICH Y (see 67187 for a discussion of unbiased measurements and methods). These loops have not been measured since Gabriele and I updated the loop design in May (69370).
The templates for the these measurements are saved in [userapps]/asc/h1/templates/MICH as 'MICH_{P,Y}_olg_broadband_shaped.xml'. I obtained 40 averages at a 0.015 Hz bandwidth, so it took about 20 minutes to run each measurement.
I have plotted the open loop gain with error shading in the attached plots. You can find the same measurement templates, exported data, and processing code in '/ligo/home/elenna.capote/DRMI_ASC/MICH'.
MICH P appears to have a UGF of 1 Hz with a phase margin of 46 deg. MICH Y appears to have a UGF of 0.55 Hz with a phase margin of 35 deg.
I believe Gabriele and I sought to reduce the UGF of MICH Y more than MICH P because at the time, MICH Y contributed more to the ASC subbudget from 10-30 Hz. However, we are now seeing significant upconversion of low frequency motion in DARM that limits the sensitivity from 20-40 Hz. I will revisit this loop design and prioritize more low frequency suppression to see if we can reduce the DARM RMS further.
Based on this measurement of MICH Y, it appeared the loop is stable for a 13 dB increase in gain, which would put the UGF closer to 1 Hz like MICH P. I raised the gain by a factor of 4, and the loop is stable and there doesn't appear to be excess noise in DARM. I ran a quick injection to check the level of MICH Y relative to DARM.
MICH Y gain is now -2.4 (was -0.6). This is updated in the guardian (lownoise ASC) and SDFed.
A few minutes after posting, Sheila and I noticed CSOFT Y motion increased significantly and the noise in DARM between 20-30 Hz worsened. This extra noise and motion reduced as I reduced the MICH Y gain back to -0.6. Looking at the spectra and RMS of both MICH Y IN1 and LSC DARM IN1, it appears that although the overall RMS of both decreases with the higher gain, the higher gain also increases a 1.3 Hz peak in both MICH Y and DARM. I am undoing these changes and keeping the MICH Y gain nominal (-0.6).
The changes in environmental coupling with the change from 75W to 60W ( https://alog.ligo-wa.caltech.edu/aLOG/index.php?callRep=70808 ) require that we re-measure some of the coupling functions that were measured in the pre-run set of PEM injections in May. Today I began those injections by injecting acoustically in the LVEA, as well as shaking HAM6 and the ITMY cryopump. I also set up for acoustic injections at EX.
In addition Sheila and I did some bias manipulation for ETMX. I injected currents onto the EX building ground and swept the ETMX bias as I had in January (https://alog.ligo-wa.caltech.edu/aLOG/index.php?callRep=67075). The figure shows that there has not been much change in the bias setting that minimizes grounding noise.
TITLE: 08/09 Day Shift: 15:00-23:00 UTC (08:00-16:00 PST), all times posted in UTC
STATE of H1: Commissioning
OUTGOING OPERATOR: Tony
INCOMING OPERATOR: Ryan C
CURRENT ENVIRONMENT:
SEI_ENV state: CALM
Wind: 17mph Gusts, 13mph 5min avg
Primary useism: 0.02 μm/s
Secondary useism: 0.08 μm/s
QUICK SUMMARY:
Comissioners started running test and injections early this morning.
RO Alarm from H0:FMC-CS_WS_RO_ALARM 16:26 UTC
FIRE DRILL Lockloss 17:00 UTC
https://alog.ligo-wa.caltech.edu/aLOG/index.php?callRep=72097
Relocking started while everyone was out side 17:02 UTC
NOMINAL_LOW_NOISE reached at 17:58 UTC
Camera servos finished at 18:11 UTC
Comissioners spent most of the day doing PEM Injections and ASC Open loop measurements.
22:51 UTC Lockloss Unknown cause 10 minutes before Hand off to Ryan C.
https://alog.ligo-wa.caltech.edu/aLOG/index.php?callRep=72112
Current IFO Status relocking in AQUIRE_PRMI
| Start Time | System | Name | Location | Lazer_Haz | Task | Time End |
|---|---|---|---|---|---|---|
| 15:52 | PEM | Robert | Ctrl Rm | N | PEM Injections | 17:54 |
| 15:56 | PEM | Robert | LVEA | N | PEM Injection setup | 16:33 |
| 16:41 | Electric | Ken | Mid Y | n | Replacing lights | 22:02 |
| 16:42 | FAC | Cindy | MEchanical Room | N | technical cleaning. | 17:54 |
| 19:07 | FAC | Cindi | H2 | N | Technical cleaning | 19:20 |
| 19:20 | PEM | Robert | Ctrl Rm | N | PEM Injections | 21:20 |
| 19:55 | ASC | Elena | Ctrl Rm | N | ASC Open Loop Gain Measurements | 20:27 |
| 20:11 | EE | Fil | PCAL Lab | Local | Installing a Laser stop button | 22:11 |
| 21:09 | Tour | jenne & Co. | Roof | n | Tour which includes a View from the roof. | 21:39 |
| 21:53 | PEM | Robert | EX | N | Setting up PEM Injections | 22:53 |
TITLE: 08/09 Day Shift: 15:00-23:00 UTC (08:00-16:00 PST), all times posted in UTC
STATE of H1: Commissioning
OUTGOING OPERATOR: Tony
CURRENT ENVIRONMENT:
SEI_ENV state: CALM
Wind: 18mph Gusts, 10mph 5min avg
Primary useism: 0.03 μm/s
Secondary useism: 0.08 μm/s
QUICK SUMMARY:
I've used the attached "script", which is also in userapps/lsc/h1/scripts/DARM_transitions, to transition DARM control from the ETMX ESD to ITMX, leaving the upper stage control on EX.
The attached screenshot shows the DARM OLG in the normal configuration (EX bias 130V, drivealign gain 184.65) and in the current configuration (IX bias -428V drivealign gain -108.4V).
Robert did a ground injection while sweeping the EX bias.
I then switched the ESD bias back to -125V, and started to switch the DARM control back to EX starting with the settings in 66751. This matched the DARM gain pretty well, with a discrepancy in the OLG measured at 200 Hz of .15%. I didn't have time to do a broadband pcal measurement before we lost lock, but this should have kept the calibration OK. Interestingly, the increase in DARM noise seen in 66751 and 66814, as shown in the attached screenshot.
We were in this configuration starting at around 22:45:30 UTC August 9th, until a lockloss at 22:49:24 UTC. I'm not sure why we lost lock, we were sitting at -125V for 5 minutes for some quiet data when it happened.
22:51 UTC Lockloss from NOMINAL_LOW_NOISE from Unknown cause.
https://ldas-jobs.ligo-wa.caltech.edu/~lockloss/index.cgi?event=1375656582
Reducing the SR2 M1 DAMP P gain from -0.5 to -0.2 and the SRM M1 DAMP L gain from -0.5 to -0.25 did reduce the noise in SRCL and DARM between 3 and 8 Hz.
The main improvement was obtained reducing SR2 damping gain, When reducing the SRM L gain, the 1.3 Hz line appears to get larger. It's worth taking a look at the SRM damping loop and its interaction with the ASC and LSC loops to ttry to track donw the 1.3 Hz peak.
So SRCL and DARM are limited by SR2 P damping noise between 3 and 8 Hz. This change reduced DARM RMS by 20%.
I SDFed the SR2 P gain to -0.2 in observe.snap. I tried to update this in the safe.snap file by changing to the "SDF to safe" guardian state, however, it did not show an SDF diff. This is something to keep an eye out for just in case it gets reverted in safe. Tagging OpsInfo so operators are aware.
Confirmed during this relock that the SR2 M1 DAMP P gain was set to -0.2 in both safe and observe, and SRM M1 DAMP L gain was set to -0.5 in both safe and observe.
The SR2 SAFE and OBSERVE tables are the same, as seen in this table of SDF_REVERT models; alog 69140. As far as I know, no changes have been made since this table was created, except for Daniel's comment.
Tagging OpsInfo to remind people of its existence.
After the SR2_P gain reduction, it looks like the SRCL RMS between 2 and 8 Hz is now equally limited by SR2_M1_DAMP_P and SR3_M1_DAMP_L. Below 2 Hz and in particular at the 1.3 Hz peak SRCL is coherrent wiith SRM_M1_DAMP_L, however that could be recoil from the SRCL lock to M1.
J. Kissel, A. Neunzert, E. Goetz, V. Bossilkov As we continue the investigation in understanding why the noise the region around 102.13 Hz gets SUPER loud at the beginning of nominal low noise segments, and the calibration line seems to be reporting a huge amount of systematic error (see investigations in LHO:72064), Ansel has found that some new electronics noise has appeared in the end station as of Saturday Aug 5 2023 around 05:30a PT at frequency extremely, and unluckily close to the 102.13000 Hz calibration line -- something at 102.12833 Hz; see LHO:72105. While we haven't yet ID'd the cause, and thus no have a solution -- we can still change the calibration frequency to move it away from this feature in hopes that there're not beating together terribly like that are now. I've changed the calibration frequency line frequency to 104.23 Hz as of 21:13 UTC on Aug 09 2023. This avoids (a) LLO's similar frequency at 101.63 Hz, and (b) because the former frequency, 102.13 Hz was near the upper edge of the 9.33 Hz wide [92.88, 102.21) Hz pulsar spin down, "non-vetoed" band, this new frequency 104.23 Hz skips up to the next 18.55 Hz wide "non-veto" band between [104.22, 122.77) Hz according to LHO:68139. Stay tuned - to see if this band-aid fix actually helps, or just spreads out the spacing between the comb, and - as we continue to investigate the issue of from where this thing came. Other things of note: Since - this feature is *not* related to the calibration line itself, - this calibration line is NOT used to generate any time-dependent correction factors and thus the calibration pipeline itself, nor the data it produces is affected - this calibration line is used only to *monitor* the calibration systematic error, - this feature is clearly identified in an auxiliary PEM channel -- and that same channel *doesn't* see the calibration line we conclude that there *isn't* some large systematic error that is occuring, it's just the calculation that's getting spoiled and misreporting large systematic error. Thus, we make NO plan to do anything on further with the calibration or systematic error estimate side of things from this. We anticipate that this now falls squarely into the noise subtraction pipeline's shoulders. Given that this 102.12833 Hz noise has a clear witness channel, and the noise creates non-linear nastiness, I expect this will be an excellent candidate for offline non-linear / NONSENS cleaning. Here's the latest list of calibration lines: Freq (Hz) Actuator Purpose Channel that defines Freq Changes Since Last Update (LHO:69736) 15.6 ETMX UIM (L1) SUS \kappa_UIM excitation H1:SUS-ETMY_L1_CAL_LINE_FREQ No change 16.4 ETMX PUM (L2) SUS \kappa_PUM excitation H1:SUS-ETMY_L2_CAL_LINE_FREQ No change 17.1 PCALY actuator kappa reference H1:CAL-PCALY_PCALOSC1_OSC_FREQ No change 17.6 ETMX TST (L3) SUS \kappa_TST excitation H1:SUS-ETMY_L3_CAL_LINE_FREQ No change 33.43 PCALX Systematic error lines H1:CAL-PCALX_PCALOSC4_OSC_FREQ No change 53.67 | | H1:CAL-PCALX_PCALOSC5_OSC_FREQ No change 77.73 | | H1:CAL-PCALX_PCALOSC6_OSC_FREQ No change 104.23 | | H1:CAL-PCALX_PCALOSC7_OSC_FREQ FREQUENCY CHANGE; THIS ALOG 283.91 V V H1:CAL-PCALX_PCALOSC8_OSC_FREQ No change 284.01 PCALY PCALXY comparison H1:CAL-PCALY_PCALOSC4_OSC_FREQ No change 410.3 PCALY f_cc and kappa_C H1:CAL-PCALY_PCALOSC2_OSC_FREQ No change 1083.7 PCALY f_cc and kappa_C monitor H1:CAL-PCALY_PCALOSC3_OSC_FREQ No change n*500+1.3 PCALX Systematic error lines H1:CAL-PCALX_PCALOSC1_OSC_FREQ No change (n=[2,3,4,5,6,7,8])
Just a post-facto proof that this calibration line frequency change from 102.13 to 104.23 Hz drastically improved the symptom that the response function systematic error, as computed by this ~100 Hz line, was huge for hours while the actual 102.128333 Hz line was loud. The attached screenshot shows a 2 days before and two days after the change (again on 2023-08-09 at 21:13 UTC). The green trace, which shows that there is no longer erroneously reported large error as computed by the 102.13 and then 104.23 Hz lines at the beginning of nominal low noise segments.
This is a quick investigation about 1.6611Hz comb in h(t) which might be related to OM2 heater (alog 71108 and 71801 ).
During Tuesday maintenance, I measured the OM2 heater driver at the driver chassis (https://dcc.ligo.org/LIGO-D2000211) using a breakout board while both the Beckhoff-side cable and vacuum-side cable were attached to the chassis and the driver was heating the OM2 (H1:AWC-OM2_TSAMS_DRV_VSET=21.448W). How the chassis, Beckhoff module and in-vac OM2 are connected are in https://dcc.ligo.org/LIGO-E2100049.
Voltages I looked at are:
I used SR785 (single BNC input, floating mode, AC coupled) for all of the above, and in addition used a scope to look at 2. and 3.
1., 4., 5. and 6. didn't look suspicious at all at low frequency, but 2. and 3. showed a forest of peaks comprising ~1.66Hz and its odd-order harmonics. See the first picture attached, 1.66Hz was derived from 18.25/11, not from the fundamental peak as the fft resolusion was not fine enough for that. (FWIW, frequency of the peaks read off of SR785 are 1.625, 5.0, 8.3125, 11.625, 14.9375 and 18.25Hz). Note that the amplitude is not huge (see the y-axis scale).
For 2. and 3., if you look at wider frequency range, it was apparent that the forest extended to much higher frequency (at least ~100Hz before it starts to get smaller), and something was glitching at O(1Hz) (attached video). No, the glitch is not SR785 auto-adjusting the frontend gain, it's in the signal.
Each glitch was pretty fast (2nd picture, see t~ negative 24ms from the center where there was a glitch of ~2ms or so width) and it's not clear if it's related to the comb. I could not see 1.66Hz periodicity using the scope but, as I already wrote, the comb is not huge in amplitude.
Anyway, these all suggest that the comb is between Beckhoff ground and the driver ground (i.e. it's common mode for Beckhoff positive and negative output). Differential receiver inside the driver seems to have rejected that, we're not pounding the heater element of OM2 with this comb (3rd picture, this corresponds to the measurement 4.).
I cannot say at this point if this is the cause of the comb in h(t) but it certainly looks as if Beckhoff is somehow related.
I haven't repeated this with zero drive voltage from Beckhoff though probably I should have. If the comb is not present in 2. and 3. with zero drive voltage from Beckhoff (I doubt that but it's still possible), we could use an adjustable power supply as an alternative.
The heater driver chassis is mounted on the same rack as the DCPD interface. Maybe we could relocate the driver just in case Beckhoff ground is somehow coupling into the rest of the chassis on the same rack.
It would also be useful to know if something happened to the Beckhoff world at the same time the comb appeared in h(t) apart from that we started heating OM2.
If you're wondering about the DC voltage between Beckhoff and driver, pin 6 and 19 on the back DB25 (Beckhoff P and N output) relative to pin 13 (the driver ground) were about +3.2V and -3.61V, respectively. If there was no gnd difference this should have been 3.4V and -3.4V.
J. Kissel
As a result of recent findings that the collection of lines used to probe the low frequency sensing function at
PCALY_EXC 8.925, 11.575, 15.275, 24.5 [Hz]
DARM_EXC 8.825, 11.475, 15.175, 24.4 [Hz]
are causing excess non-linearities found when looking over long time scales --- see LHO:71964 --- I've turned OFF these calibration lines as of 2023-08-09 16:40 UTC. We were already out of observing for PEM injections anyways, and Robert relinquished a moment for me to quickly turn these off.
I've accepted the changes in both the PCALY (h1caley) and OMC (h1omc) front-end model's SDFs against both the safe and OBSERVE.snaps such that this sticks.
Here's the latest list of calibration lines:
Freq (Hz) Actuator Purpose Channel that defines Freq Changes Since Last Update (LHO:69736)
8.825 DARM (via ETMX L1,L2,L3) Live DARM OLGTFs H1:LSC-DARMOSC1_OSC_FREQ Turned OFF; THIS aLOG
8.925 PCALY Live Sensing Function H1:CAL-PCALY_PCALOSC5_OSC_FREQ Turned OFF; THIS aLOG
11.475 DARM (via ETMX L1,L2,L3) Live DARM OLGTFs H1:LSC-DARMOSC2_OSC_FREQ Turned OFF; THIS aLOG
11.575 PCALY Live Sensing Function H1:CAL-PCALY_PCALOSC6_OSC_FREQ Turned OFF; THIS aLOG
15.175 DARM (via ETMX L1,L2,L3) Live DARM OLGTFs H1:LSC-DARMOSC3_OSC_FREQ Turned OFF; THIS aLOG
15.275 PCALY Live Sensing Function H1:CAL-PCALY_PCALOSC7_OSC_FREQ Turned OFF; THIS aLOG
24.400 DARM (via ETMX L1,L2,L3) Live DARM OLGTFs H1:LSC-DARMOSC4_OSC_FREQ Turned OFF; THIS aLOG
24.500 PCALY Live Sensing Function H1:CAL-PCALY_PCALOSC8_OSC_FREQ Turned OFF; THIS aLOG
15.6 ETMX UIM (L1) SUS \kappa_UIM excitation H1:SUS-ETMY_L1_CAL_LINE_FREQ No change
16.4 ETMX PUM (L2) SUS \kappa_PUM excitation H1:SUS-ETMY_L2_CAL_LINE_FREQ No change
17.1 PCALY actuator kappa reference H1:CAL-PCALY_PCALOSC1_OSC_FREQ No change
17.6 ETMX TST (L3) SUS \kappa_TST excitation H1:SUS-ETMY_L3_CAL_LINE_FREQ No change
33.43 PCALX Systematic error lines H1:CAL-PCALX_PCALOSC4_OSC_FREQ No change
53.67 | | H1:CAL-PCALX_PCALOSC5_OSC_FREQ No change
77.73 | | H1:CAL-PCALX_PCALOSC6_OSC_FREQ No change
102.13 | | H1:CAL-PCALX_PCALOSC7_OSC_FREQ No change
283.91 V V H1:CAL-PCALX_PCALOSC8_OSC_FREQ No change
284.01 PCALY PCALXY comparison H1:CAL-PCALY_PCALOSC4_OSC_FREQ Off briefly between 2023-08-01 22:02 - 22:11 UTC, back on as of 22:16 UTC
410.3 PCALY f_cc and kappa_C H1:CAL-PCALY_PCALOSC2_OSC_FREQ No Change
1083.7 PCALY f_cc and kappa_C monitor H1:CAL-PCALY_PCALOSC3_OSC_FREQ No Change
n*500+1.3 PCALX Systematic error lines H1:CAL-PCALX_PCALOSC1_OSC_FREQ No Change (n=[2,3,4,5,6,7,8])
Wed Aug 09 10:21:40 2023 INFO: Fill completed in 21min 35secs
Gerardo confirmed a good fill curbside.
Benoit, Ansel, Derek
Benoit noticed that for recent locks, the 102.13 Hz calibration line is much louder than typical for the first few hours of the lock. An example of this behavior is shown in the attached spectrogram of H1 strain data on August 5 - this is the first day this behavior appeared. Ansel noted that this feature includes a comb-like structure around the line that is only present in the H1:GDS-CALIB_STRAIN_NOLINES channel and not H1:GDS-CALIB_STRAIN (see spectra for CALIB_STRAIN and CALIB_STRAIN_NOLINES on Aug 5). This issue also visible in the PCAL trends for the 102.13 Hz line.
We are not sure if the excess noise near 102.13 Hz is from the calibration line itself or another noise source that is near the line. However, the behavior has been present for every lock since 12:30 UTC on August 5 2023.
FYI,
$ gpstime Aug 05 2023 12:30 UTC
PDT: 2023-08-05 05:30:00.000000 PDT
UTC: 2023-08-05 12:30:00.000000 UTC
GPS: 1375273818.000000
so... this behavior seems to have started at 5:30a local time on a Saturday. Therefore *very* unlikely that the start of this issue is intentional / human change driven.
The investigation continues....
making sure to tag CAL.
Other facts and recent events:
- Attached are 2 screenshots that show the actual *digital* excitation is not changing with time in anyway.
:: 2023-08-08_H1PCALEX_OSC7_102p13Hz_Line_3mo_trend.png shows the specific oscillator, --- PCALX's OSC7 which drives the 102.13 Hz line's EPICs channel version of its output. The minute trend shows the max, min, and mean of the output, and there's no change in amplitude.
:: 2023-08-08_H1PCALEX_EXC_SUM_3mo_trend.png shows a trend of the total excitation sum from PCAL X. This also shows *no* change in time in amplitude.
Both trends show the Aug 02 2023 change in amplitude kerfuffle I caused that Corey found and a bit later rectified -- see LHO:71894 and subsequent comments, but that was done, over with an solved, definitely by Aug 03 2023 UTC and unrelated to the start up of this problem.
It's also well after I installed new oscillators and rebooted the PCALX, PCALY, and OMC models on Aug 01 2023 (see LHO:71881).
The front-end version of the calibration's systematic error at 102.13 Hz also shows the long, time-dependent issue -- this will allow us to trend the issue against other channels
Folks in the calibration group have found that the online monitoring system for the
- overall DARM response function systematic error
- (absolute reference) / (Calibrated Data Product) [m/m]
- ( \eta_R ) ^ (-1)
- (C / 1+G)_pcal / (C / 1+G)_strain
- CAL-DELTAL_REF_PCAL_DQ / GDS-CALIB_STRAIN
(all different ways of saying the same thing; see T1900169) in calibration at each PCAL calibration line frequency -- the "grafana" pages -- are showing *huge* amounts of systematic error during these times when the amplitude of the line is super loud.
Though this metric is super useful because it's dreadfully obvious that things are going wrong -- this metric is not in any normal frame structure, so you can't compare it against other channels to find out what's causing the systematic error.
However -- remember -- we commissioned a front-end version of this monitoring during ER15 -- see LHO:69285.
That means the channels
H1:CAL-CS_TDEP_PCAL_LINE8_COMPARISON_OSC_FREQ << the frequency of the monitor
H1:CAL-CS_TDEP_PCAL_LINE8_SYSERROR_MAG_MPM << the magnitude of the systematic error
H1:CAL-CS_TDEP_PCAL_LINE8_SYSERROR_PHA_DEG << the phase of the systematic error
tell you (what's supposed to be***) equivalent information.
*** One might say that "what's suppose to be" is the same as "roughly equivalent" due to the following reasons:
(1) because we're human, the one system is displaying the systematic error \eta_R, and the other is displaying the inverse ( \eta_R ) ^ (-1)
(2) Because this is early-days in the front-end system, it uses the "less complete" calibrated channel CAL-DELTAL_EXTERNAL_DQ rather than the "fully correct" channel GDS-CALIB_STRAIN
But because the problem is so dreadfully obvious in these metrics, even though they're only *roughly* equivalent, you can see the same thing.
In the attached screenshot, I show both metrics for the most recent observation stretch, between 10:15 and 14:00 UTC on 2023-Aug-09.
Let's use this front-end metric to narrow down the problem via trending.
There appears to be no change in the PCALX analog excitation monitors either. Attached is a trend of some key channels in the optical follower servo -- the analog feedback system that serves as intensity stabilization and excitation power linearization for the PCAL's laser light that gets transmitted to the test mass -- the actuator of which is an acousto-optic modulator (an AOM). There seems to be no major differences in the max, min, and mean of these signals before vs. after these problems started on Aug 05 2023. H1:CAL-PCALX_OFS_PD_OUT_DQ H1:CAL-PCALX_OFS_AOM_DRIVE_MON_OUT_DQ
I believe this is caused by the presence of another line very close to the 102.13 Hz pcal line. This second line is present at the start of a lock stretch but seems to go away as the lock stretch continues. I have attached a plot showing a zoom-in on an ASD around 102.1-102.2 Hz right after a lock stretch (orange), where the second peak is evident, and well into a lock stretch (blue) where the PCAL line is still present, but the second peak right below it in frequency is gone. This ASD is computed using an hour of data for each curve, so we can get the needed resolution for these two peaks.
I don't know the origin of this second line. However, a quick fix to the issue could be moving the PCAL line over by about a Hz. The second attached plot shows that the spectrum looks pretty clean from 101-102 Hz, so somewhere in there would be probably be okay for a new location of the PCAL line.
Since it looks like the additional noise is at 102.12833 Hz, I did a quick check in Fscan data from Aug 5 for channels where there is high coherence with DELTAL_EXTERNAL at 102.12833 but *not* at 102.13000 Hz. This narrows down to just a few channels:
(lines git issue opened as we work on this.)
As a result of Ansel's discovery, and conversation on the CAL call today -- I've moved the calibration line frequency from 102.13 to 104.23 Hz. See LHO:72108.
This line may have appeared in the previous lock the day before (Aug 4). The daily spectrogram for Aug 4 shows a line near 100 Hz starting at 21:00 UTC.
Looking at alogs leading up to the time Derek notes above, I noticed that Gabriele retuned and tested new LSC FF. This change may be related to this new peak. Remembering some issues we had recently where DHARD filter impulses were ringing up violin modes, I checked the new LSC FF filters and how they are engaged in the guardian. Some of them have no ramp time, and the filter bank is turned on immediately along with the filters in the guardian. I have no idea why that would cause a peak at 102 Hz, but I updated those filters to have a 3 second ramp.
Reloaded the H1LSC model to load in Elenna's filter changes
Now that the calibration line has been moved, the comb-like structure at the calibration line frequency is no longer present (checked in the CLEAN channel).
We can also see the shape of the 102.12833 Hz line much more clearly without the overlapping calibration line. I have attached a plot for reference on the width and shape.
As discussed in todays commissioning meeting, I checked TMSX and ETMX movement for a kick during locking and couldn't see anything suspicious. I did find some increase motion/noise every 8Hz in TMSX 1s into ENGAGE_SOFT_LOOPS when ISC_LOCK isn't explicitly doing anything, plot attached. However this noise was present prior to Aug 4th, (July 30th attached).
TMS is suspicious as Betsy found that TMS's have violin modes ~103-104Hz.
Jeff draws attendtion to 38295, showing modes of quad blade springs above 110Hz, and 24917 showing quad top wire modes above 300Hz.
Elenna's notes with calibration lines off (as we are experimenting with for current lock) we can see this 102Hz peak at ISC_LOCK state ENGAGE_ASC_FOR_FULL_IFO. We were mistaken.
To preserve documentation, this problem has now been solved, with more details in 72537, 72319, and 72262.
The cause of this peak was a spurious, narrow, 102 Hz feature in the SRCL feedforward that we didn't catch when the filter was made. This has been been fixed, and the cause of the mistake has been documented in the first alog listed above so we hopefully don't repeat this error.
C.Swain, L.Dartez, J.Kissel
Finished characterizing the spare whitening chassis, OMC DCPD S2300004.
The last two files are text documents containing updated notes on gathering the transfer function measurement data and noise measurement data of the whitening chassis, respectively.
Adding more details to OMC DCPD S2300004 Characterization for better accessibility:
Whitening ON Fit:
| Channel | Fit Zeros | Fit Poles | Fit Gain |
| DCPDA | [0.997] Hz | [9.904e+00, 4.401e+04] Hz | 435752.409 |
| DCPDB | [1.006] Hz | [9.994e+00, 4.377e+04] Hz | 433177.144 |
These results align very well with the goal of a [1:10] whitening chassis.
A difference of [0.3, 0.6] % from the model for the DCPDA and DCPDB fit zeros, respectively, is present. A difference of [0.96, 0.06] % from the model for the DCPDA and DCPDB fit poles, respectively, is present.
As all the differences are below 1%, the precision of the results is satisfactory.
Noise Measurements:
Average ASD for Whitening ON: 300 nV/rtHz
Average ASD for Whitening OFF: 30 nV/rtHz
Average ASD SR785 Noise Floor Level: 20 nV/rtHz
Whitening ON and Whitening OFF noise is consistent with OMC DCPD S2300003 noise measurements. Noise Floor is slightly higher than the S2300003 Noise Floor measurements (~20nV/rtHz compared to ~6nV/rtHz). This difference in noise floor measurements could be caused by a difference in measurement setup when gathering S2300003 data compared to S2300004 data.
Overall:
The measurments of the OMC DCPD S2300004 whitening chassis align well with what is to be expected from a z:p = [1 : (10, 44e3)] Hz whitening filter, with an average ASD noise floor of 300nV/rtHz in the Whitening ON state.
SR785 noise floor level is slightly higher than previously recorded from the OMC DCPD S2300003 whitening chassis but does not contribute to either the Whitening OFF or Whitening ON states, so this difference may effectively be ignored.