Reports until 20:05, Friday 26 April 2024
X1 SUS (SUS)
oli.patane@LIGO.ORG - posted 20:05, Friday 26 April 2024 (77455)
Searching for the perfect position for the secondary prism on the BBSS

Ibrahim A, Oli P, Mark B, Betsy W

Context:

M3 on the beam splitter has secondary prisms that are placed along the barrel below the primary prisms. These secondary prisms are placed at a location below the primary prism, and if placed in the correct spot, act as a clamp, damping the oscillations moving down the wire loop into the optic. We've been working on finding the optimal secondary prism position for the BBSS.

Experiment:

The experiment for determining this position was developed for the BSFM in 2011(T1100086), and we used a similar setup for our three (successful) runs for the BBSS. This is basically how the experiment is set up to work:

  1. A biased solenoid is placed a few millimeters away from one of the M2-M3 loop wires. The solenoid is driven at the wire’s predetermined resonance frequency using a waveform generator.
  2. The wire goes from M2 down through M3’s fixed primary prism and (the reason for this experiment) moveable secondary prism.
  3. Phonograph needle is rested on wire below the secondary prism and the other end is connected to a spectrum analyzer. Five FFT averages are taken to give us an attenuated peak at the resonant frequency. This is a reading. We took 4/5 measurements for each location, where each measurement consisted of five 8-second averages.
  4. We move the prism down by a millimeter or two and take more readings.
  5. From the results, we determine the lowest amplitude peak/highest attenuation and note the prism position.
  6. This becomes the new BBSS secondary (glued on) prism position.

Results + Commentary:

Note: Mark has an upcoming alog about the run of data and issues that we took when he was here, but in LESS IMPORTANT Auxiliary Investigations we will be focusing on the data from the second and third runs, but here in the results will still be showing the data we got with him.

The plot attached shows the attenuation measured versus the position (positive distance relative to the mass’ centerline) of the secondary prism. Each color is a separate run, and the more negative the attenuation, the quieter the reading was, so we are looking for a location with the lowest values where the secondary prism would do the most damping.

We found that there is a region between ~29.00mm and ~31.5mm below the horizontal center of the mass where the amplitude is minimized. There are approximately three anomalous data points in between, with one confirmed being due to cross-coupling of above wire. As of now, we’re still working on understanding this data and have spoken to Mark B and Jeff K about how we can go about accurately interpreting the raw data (including essential questions about the calibration and setup. See below.)

LESS IMPORTANT Auxiliary Investigations - Details, Details, Details

    Read for Clues/Hints (things we noticed, changed etc.):

Images attached to this report