List of lines marked for H1-CAL-DELTAL-EXT in data set 17may2015 Frequencies are interpolated, but amplitudes are max values within +/-1 mHz Particular comb types: b = Bounce Mode r = Roll mode C = Calibration line L = Lock-in oscillator S = 64-Hz s = 16-Hz F = Four-Hz (almost) M = Power Mains N = Nearly-Power Mains Comb harmonics found below 2000 Hz ---------------------------------- Comb 1 (type b, offset=0.000000): Frequency (offset + harmonic x fund freq) Ampl (m/rtHz) Bar (logarithmic) b 9.7767 ( 1 X 9.7767) 3.274751e-14 ******************** b 19.5534 ( 2 X 9.7767) 8.674675e-18 ********* Comb 2 (type b, offset=0.000000): Frequency (offset + harmonic x fund freq) Ampl (m/rtHz) Bar (logarithmic) b 9.8316 ( 1 X 9.8316) 3.396338e-13 *********************** b 19.6632 ( 2 X 9.8316) 1.688156e-17 ********** Comb 3 (type r, offset=0.000000): Frequency (offset + harmonic x fund freq) Ampl (m/rtHz) Bar (logarithmic) r 13.8190 ( 1 X 13.8190) 1.518910e-16 ************* Comb 4 (type r, offset=0.000000): Frequency (offset + harmonic x fund freq) Ampl (m/rtHz) Bar (logarithmic) r 13.8900 ( 1 X 13.8900) 5.442048e-15 ****************** Comb 5 (type r, offset=0.000000): Frequency (offset + harmonic x fund freq) Ampl (m/rtHz) Bar (logarithmic) r 13.9303 ( 1 X 13.9303) 6.291841e-14 ********************* Comb 6 (type r, offset=0.000000): Frequency (offset + harmonic x fund freq) Ampl (m/rtHz) Bar (logarithmic) r 13.9790 ( 1 X 13.9790) 6.944433e-16 *************** Comb 7 (type C, offset=0.000000): Frequency (offset + harmonic x fund freq) Ampl (m/rtHz) Bar (logarithmic) C 33.1000 ( 1 X 33.1000) 2.097711e-17 ********** Comb 8 (type C, offset=0.000000): Frequency (offset + harmonic x fund freq) Ampl (m/rtHz) Bar (logarithmic) C 34.7000 ( 1 X 34.7000) 9.501481e-17 ************ Comb 9 (type C, offset=0.000000): Frequency (offset + harmonic x fund freq) Ampl (m/rtHz) Bar (logarithmic) C 35.3000 ( 1 X 35.3000) 4.429150e-19 ***** Comb 10 (type C, offset=0.000000): Frequency (offset + harmonic x fund freq) Ampl (m/rtHz) Bar (logarithmic) C 36.7000 ( 1 X 36.7000) 3.376571e-17 *********** Comb 11 (type C, offset=0.000000): Frequency (offset + harmonic x fund freq) Ampl (m/rtHz) Bar (logarithmic) C 310.0000 ( 1 X 310.0000) 1.875620e-17 ********** Comb 12 (type L, offset=0.000000): Frequency (offset + harmonic x fund freq) Ampl (m/rtHz) Bar (logarithmic) L 322.1000 ( 1 X 322.1000) 4.174604e-17 *********** Comb 13 (type C, offset=0.000000): Frequency (offset + harmonic x fund freq) Ampl (m/rtHz) Bar (logarithmic) C 534.7000 ( 1 X 534.7000) 6.835166e-17 ************ Comb 14 (type C, offset=0.000000): Frequency (offset + harmonic x fund freq) Ampl (m/rtHz) Bar (logarithmic) C 538.1000 ( 1 X 538.1000) 2.064659e-17 ********** Comb 15 (type C, offset=0.000000): Frequency (offset + harmonic x fund freq) Ampl (m/rtHz) Bar (logarithmic) C 539.3000 ( 1 X 539.3000) 4.959605e-20 *** Comb 16 (type C, offset=0.000000): Frequency (offset + harmonic x fund freq) Ampl (m/rtHz) Bar (logarithmic) C 540.7000 ( 1 X 540.7000) 1.056945e-16 ************* Comb 17 (type M, offset=0.000000): Frequency (offset + harmonic x fund freq) Ampl (m/rtHz) Bar (logarithmic) M 60.0000 ( 1 X 60.0000) 1.041910e-17 ********** M 120.0000 ( 2 X 60.0000) 8.540242e-20 *** M 180.0000 ( 3 X 60.0000) 1.308655e-19 **** M 240.0000 ( 4 X 60.0000) 1.612209e-17 ********** M 300.0000 ( 5 X 60.0000) 1.518401e-19 **** M 360.0000 ( 6 X 60.0000) 4.381369e-20 ** M 420.0000 ( 7 X 60.0000) 1.092931e-19 **** M 480.0000 ( 8 X 60.0000) 1.402602e-19 **** M 540.0000 ( 9 X 60.0000) 4.626199e-20 ** Comb 18 (type s, offset=0.000000): Frequency (offset + harmonic x fund freq) Ampl (m/rtHz) Bar (logarithmic) s 32.0000 ( 2 X 16.0000) 7.899031e-19 ****** s 48.0000 ( 3 X 16.0000) 1.652752e-19 **** s 64.0000 ( 4 X 16.0000) 5.293597e-19 ****** s 80.0000 ( 5 X 16.0000) 1.229020e-19 **** s 96.0000 ( 6 X 16.0000) 1.403687e-19 **** s 112.0000 ( 7 X 16.0000) 7.192878e-20 *** s 128.0000 ( 8 X 16.0000) 3.655696e-19 ***** s 144.0000 ( 9 X 16.0000) 9.752989e-20 *** s 160.0000 ( 10 X 16.0000) 1.702255e-19 **** s 176.0000 ( 11 X 16.0000) 9.254924e-20 *** s 192.0000 ( 12 X 16.0000) 1.726100e-19 **** s 208.0000 ( 13 X 16.0000) 1.030167e-19 **** s 224.0000 ( 14 X 16.0000) 7.487349e-20 *** s 240.0000 ( 15 X 16.0000) 1.612209e-17 ********** s 256.0000 ( 16 X 16.0000) 5.975004e-19 ****** s 272.0000 ( 17 X 16.0000) 1.420393e-19 **** s 288.0000 ( 18 X 16.0000) 2.914390e-19 ***** s 304.0000 ( 19 X 16.0000) 3.544260e-20 ** s 320.0000 ( 20 X 16.0000) 1.492698e-19 **** s 336.0000 ( 21 X 16.0000) 5.213425e-20 *** s 352.0000 ( 22 X 16.0000) 4.606560e-19 ***** s 368.0000 ( 23 X 16.0000) 2.624298e-19 ***** s 384.0000 ( 24 X 16.0000) 2.209552e-19 ***** s 400.0000 ( 25 X 16.0000) 1.633990e-19 **** s 416.0000 ( 26 X 16.0000) 1.537112e-19 **** s 432.0000 ( 27 X 16.0000) 2.547921e-19 ***** s 448.0000 ( 28 X 16.0000) 1.307537e-19 **** s 464.0000 ( 29 X 16.0000) 8.890587e-20 *** s 480.0000 ( 30 X 16.0000) 1.402602e-19 **** s 496.0000 ( 31 X 16.0000) 1.490235e-19 **** s 512.0000 ( 32 X 16.0000) 9.446757e-20 *** s 528.0000 ( 33 X 16.0000) 2.319366e-19 ***** s 544.0000 ( 34 X 16.0000) 3.013668e-19 ***** s 560.0000 ( 35 X 16.0000) 1.520849e-19 **** s 576.0000 ( 36 X 16.0000) 2.913213e-19 ***** s 592.0000 ( 37 X 16.0000) 1.153138e-19 **** s 608.0000 ( 38 X 16.0000) 1.257051e-19 **** s 624.0000 ( 39 X 16.0000) 5.824669e-20 *** s 640.0000 ( 40 X 16.0000) 1.256302e-19 **** s 656.0000 ( 41 X 16.0000) 7.980882e-20 *** s 672.0000 ( 42 X 16.0000) 3.348011e-19 ***** s 688.0000 ( 43 X 16.0000) 2.468348e-19 ***** s 704.0000 ( 44 X 16.0000) 5.106877e-20 *** s 720.0000 ( 45 X 16.0000) 8.010980e-20 *** s 736.0000 ( 46 X 16.0000) 1.991772e-19 **** s 752.0000 ( 47 X 16.0000) 2.605353e-19 ***** s 768.0000 ( 48 X 16.0000) 1.931609e-19 **** s 784.0000 ( 49 X 16.0000) 2.869886e-19 ***** s 800.0000 ( 50 X 16.0000) 3.485110e-19 ***** s 816.0000 ( 51 X 16.0000) 2.795304e-19 ***** s 832.0000 ( 52 X 16.0000) 1.276569e-19 **** s 848.0000 ( 53 X 16.0000) 9.055072e-20 *** s 864.0000 ( 54 X 16.0000) 1.321787e-19 **** s 880.0000 ( 55 X 16.0000) 1.004739e-19 **** s 896.0000 ( 56 X 16.0000) 2.218541e-19 ***** s 912.0000 ( 57 X 16.0000) 1.064239e-19 **** s 928.0000 ( 58 X 16.0000) 2.842821e-19 ***** s 944.0000 ( 59 X 16.0000) 2.195213e-19 ***** s 960.0000 ( 60 X 16.0000) 1.667878e-19 **** s 976.0000 ( 61 X 16.0000) 2.625834e-19 ***** s 992.0000 ( 62 X 16.0000) 3.238651e-19 ***** s 1008.0000 ( 63 X 16.0000) 1.733031e-19 **** s 1024.0000 ( 64 X 16.0000) 1.914195e-19 **** s 1040.0000 ( 65 X 16.0000) 4.403794e-19 ***** s 1056.0000 ( 66 X 16.0000) 1.401894e-19 **** s 1072.0000 ( 67 X 16.0000) 2.077247e-19 **** s 1088.0000 ( 68 X 16.0000) 1.322874e-19 **** s 1104.0000 ( 69 X 16.0000) 1.722345e-19 **** s 1120.0000 ( 70 X 16.0000) 3.649303e-19 ***** s 1136.0000 ( 71 X 16.0000) 1.825761e-19 **** s 1152.0000 ( 72 X 16.0000) 1.052383e-19 **** s 1168.0000 ( 73 X 16.0000) 2.819866e-19 ***** s 1184.0000 ( 74 X 16.0000) 3.725837e-19 ***** s 1200.0000 ( 75 X 16.0000) 1.127150e-19 **** s 1216.0000 ( 76 X 16.0000) 9.811566e-20 *** s 1232.0000 ( 77 X 16.0000) 1.971585e-19 **** s 1248.0000 ( 78 X 16.0000) 1.316440e-19 **** s 1264.0000 ( 79 X 16.0000) 1.787153e-19 **** s 1280.0000 ( 80 X 16.0000) 2.404328e-19 ***** s 1296.0000 ( 81 X 16.0000) 4.414466e-19 ***** s 1312.0000 ( 82 X 16.0000) 1.188290e-19 **** s 1328.0000 ( 83 X 16.0000) 2.438473e-19 ***** s 1344.0000 ( 84 X 16.0000) 2.511370e-19 ***** s 1360.0000 ( 85 X 16.0000) 2.748823e-19 ***** s 1376.0000 ( 86 X 16.0000) 2.363566e-19 ***** s 1392.0000 ( 87 X 16.0000) 1.360338e-19 **** s 1408.0000 ( 88 X 16.0000) 2.558889e-19 ***** s 1424.0000 ( 89 X 16.0000) 2.685299e-19 ***** s 1440.0000 ( 90 X 16.0000) 1.210149e-19 **** s 1456.0000 ( 91 X 16.0000) 5.764311e-19 ****** s 1472.0000 ( 92 X 16.0000) 3.051366e-19 ***** s 1488.0000 ( 93 X 16.0000) 3.573614e-19 ***** s 1504.0000 ( 94 X 16.0000) 1.516638e-19 **** s 1520.0000 ( 95 X 16.0000) 3.282876e-19 ***** s 1536.0000 ( 96 X 16.0000) 3.099572e-19 ***** s 1552.0000 ( 97 X 16.0000) 2.092912e-19 **** s 1568.0000 ( 98 X 16.0000) 1.627724e-19 **** s 1584.0000 ( 99 X 16.0000) 3.828801e-19 ***** s 1600.0000 ( 100 X 16.0000) 2.151944e-19 **** s 1616.0000 ( 101 X 16.0000) 4.210007e-19 ***** s 1632.0000 ( 102 X 16.0000) 1.521758e-19 **** s 1648.0000 ( 103 X 16.0000) 3.012875e-19 ***** s 1664.0000 ( 104 X 16.0000) 1.452470e-19 **** s 1680.0000 ( 105 X 16.0000) 2.389904e-19 ***** s 1696.0000 ( 106 X 16.0000) 3.555633e-19 ***** s 1712.0000 ( 107 X 16.0000) 1.852697e-19 **** s 1728.0000 ( 108 X 16.0000) 4.739292e-19 ****** s 1744.0000 ( 109 X 16.0000) 3.333513e-19 ***** s 1760.0000 ( 110 X 16.0000) 2.977242e-19 ***** s 1776.0000 ( 111 X 16.0000) 2.196622e-19 ***** s 1792.0000 ( 112 X 16.0000) 1.955510e-19 **** s 1808.0000 ( 113 X 16.0000) 5.292044e-19 ****** s 1824.0000 ( 114 X 16.0000) 4.711269e-19 ****** s 1840.0000 ( 115 X 16.0000) 5.371570e-19 ****** s 1856.0000 ( 116 X 16.0000) 5.767506e-19 ****** s 1872.0000 ( 117 X 16.0000) 3.495521e-19 ***** s 1888.0000 ( 118 X 16.0000) 4.104073e-19 ***** s 1904.0000 ( 119 X 16.0000) 6.299388e-19 ****** s 1920.0000 ( 120 X 16.0000) 1.797518e-19 **** s 1936.0000 ( 121 X 16.0000) 2.052885e-19 **** s 1952.0000 ( 122 X 16.0000) 1.827042e-19 **** s 1968.0000 ( 123 X 16.0000) 2.903300e-19 ***** s 1984.0000 ( 124 X 16.0000) 3.298533e-19 ***** Comb 19 (type F, offset=13.997700): Frequency (offset + harmonic x fund freq) Ampl (m/rtHz) Bar (logarithmic) F 13.9977 ( 0 X 3.9994) 8.459745e-17 ************ F 17.9971 ( 1 X 3.9994) 3.083661e-17 *********** F 21.9965 ( 2 X 3.9994) 1.609684e-17 ********** F 25.9959 ( 3 X 3.9994) 9.520783e-18 ********* F 29.9953 ( 4 X 3.9994) 5.899419e-18 ********* F 33.9947 ( 5 X 3.9994) 3.897166e-18 ******** F 37.9941 ( 6 X 3.9994) 2.648160e-18 ******** F 41.9935 ( 7 X 3.9994) 1.854741e-18 ******* F 45.9929 ( 8 X 3.9994) 1.328512e-18 ******* F 49.9923 ( 9 X 3.9994) 9.303675e-19 ****** F 53.9917 ( 10 X 3.9994) 6.735205e-19 ****** F 57.9911 ( 11 X 3.9994) 5.474742e-19 ****** F 61.9905 ( 12 X 3.9994) 4.524984e-19 ***** F 65.9899 ( 13 X 3.9994) 3.647248e-19 ***** F 69.9893 ( 14 X 3.9994) 2.864954e-19 ***** F 73.9887 ( 15 X 3.9994) 2.388288e-19 ***** F 77.9881 ( 16 X 3.9994) 1.952002e-19 **** F 81.9875 ( 17 X 3.9994) 1.603236e-19 **** F 85.9869 ( 18 X 3.9994) 1.288193e-19 **** F 89.9863 ( 19 X 3.9994) 1.034059e-19 **** F 93.9857 ( 20 X 3.9994) 8.144616e-20 *** F 97.9851 ( 21 X 3.9994) 6.799753e-20 *** F 101.9845 ( 22 X 3.9994) 5.692240e-20 *** F 105.9839 ( 23 X 3.9994) 5.269174e-20 *** F 109.9833 ( 24 X 3.9994) 4.988944e-20 *** F 113.9827 ( 25 X 3.9994) 4.089195e-20 ** F 117.9821 ( 26 X 3.9994) 3.847340e-20 ** F 121.9815 ( 27 X 3.9994) 3.908532e-20 ** F 125.9809 ( 28 X 3.9994) 3.537298e-20 ** F 129.9803 ( 29 X 3.9994) 3.156804e-20 ** F 133.9797 ( 30 X 3.9994) 3.050299e-20 ** F 137.9791 ( 31 X 3.9994) 2.712008e-20 ** F 141.9785 ( 32 X 3.9994) 2.577463e-20 ** F 145.9779 ( 33 X 3.9994) 2.738702e-20 ** Comb 20 (type S, offset=0.000000): Frequency (offset + harmonic x fund freq) Ampl (m/rtHz) Bar (logarithmic) S 64.0000 ( 1 X 64.0000) 5.293597e-19 ****** S 128.0000 ( 2 X 64.0000) 3.655696e-19 ***** S 192.0000 ( 3 X 64.0000) 1.726100e-19 **** S 256.0000 ( 4 X 64.0000) 5.975004e-19 ****** S 320.0000 ( 5 X 64.0000) 1.492698e-19 **** S 384.0000 ( 6 X 64.0000) 2.209552e-19 ***** S 448.0000 ( 7 X 64.0000) 1.307537e-19 **** S 512.0000 ( 8 X 64.0000) 9.446757e-20 *** S 576.0000 ( 9 X 64.0000) 2.913213e-19 ***** S 640.0000 ( 10 X 64.0000) 1.256302e-19 **** S 704.0000 ( 11 X 64.0000) 5.106877e-20 *** S 768.0000 ( 12 X 64.0000) 1.931609e-19 **** S 832.0000 ( 13 X 64.0000) 1.276569e-19 **** S 896.0000 ( 14 X 64.0000) 2.218541e-19 ***** S 960.0000 ( 15 X 64.0000) 1.667878e-19 **** S 1024.0000 ( 16 X 64.0000) 1.914195e-19 **** S 1088.0000 ( 17 X 64.0000) 1.322874e-19 **** S 1152.0000 ( 18 X 64.0000) 1.052383e-19 **** S 1216.0000 ( 19 X 64.0000) 9.811566e-20 *** S 1280.0000 ( 20 X 64.0000) 2.404328e-19 ***** S 1344.0000 ( 21 X 64.0000) 2.511370e-19 ***** S 1408.0000 ( 22 X 64.0000) 2.558889e-19 ***** S 1472.0000 ( 23 X 64.0000) 3.051366e-19 ***** S 1536.0000 ( 24 X 64.0000) 3.099572e-19 ***** S 1600.0000 ( 25 X 64.0000) 2.151944e-19 **** S 1664.0000 ( 26 X 64.0000) 1.452470e-19 **** S 1728.0000 ( 27 X 64.0000) 4.739292e-19 ****** S 1792.0000 ( 28 X 64.0000) 1.955510e-19 **** S 1856.0000 ( 29 X 64.0000) 5.767506e-19 ****** S 1920.0000 ( 30 X 64.0000) 1.797518e-19 **** S 1984.0000 ( 31 X 64.0000) 3.298533e-19 ***** Comb 21 (type D, offset=0.000000): Frequency (offset + harmonic x fund freq) Ampl (m/rtHz) Bar (logarithmic) D 59.3155 ( 1 X 59.3155) 9.321867e-19 ****** D 118.6310 ( 2 X 59.3155) 8.834137e-20 *** Comb 22 (type E, offset=0.000000): Frequency (offset + harmonic x fund freq) Ampl (m/rtHz) Bar (logarithmic) E 36.9733 ( 1 X 36.9733) 6.027555e-19 ****** E 73.9466 ( 2 X 36.9733) 5.331451e-19 ****** E 110.9199 ( 3 X 36.9733) 1.358187e-19 **** E 147.8932 ( 4 X 36.9733) 6.588426e-19 ****** E 184.8665 ( 5 X 36.9733) 3.178739e-19 ***** E 221.8398 ( 6 X 36.9733) 3.143686e-19 ***** E 258.8131 ( 7 X 36.9733) 1.497924e-19 **** E 295.7864 ( 8 X 36.9733) 7.740361e-19 ****** E 332.7597 ( 9 X 36.9733) 7.345576e-20 *** E 369.7330 ( 10 X 36.9733) 1.139548e-18 ******* E 406.7063 ( 11 X 36.9733) 2.286807e-19 ***** E 443.6796 ( 12 X 36.9733) 6.675017e-19 ****** E 480.6529 ( 13 X 36.9733) 1.053286e-19 **** E 517.6262 ( 14 X 36.9733) 7.943448e-19 ****** E 554.5995 ( 15 X 36.9733) 1.428754e-19 **** E 591.5728 ( 16 X 36.9733) 6.682199e-19 ****** E 628.5461 ( 17 X 36.9733) 1.552704e-19 **** E 665.5194 ( 18 X 36.9733) 5.514371e-19 ****** E 702.4927 ( 19 X 36.9733) 1.170144e-19 **** E 739.4660 ( 20 X 36.9733) 7.235271e-19 ****** E 776.4393 ( 21 X 36.9733) 3.236870e-19 ***** E 813.4126 ( 22 X 36.9733) 6.871208e-19 ****** E 850.3859 ( 23 X 36.9733) 1.221418e-19 **** E 887.3592 ( 24 X 36.9733) 7.230846e-19 ****** E 924.3325 ( 25 X 36.9733) 1.217429e-19 **** E 961.3058 ( 26 X 36.9733) 6.817552e-19 ****** E 998.2791 ( 27 X 36.9733) 9.417681e-19 ****** E 1035.2524 ( 28 X 36.9733) 2.958061e-19 ***** E 1072.2257 ( 29 X 36.9733) 1.290729e-19 **** E 1109.1990 ( 30 X 36.9733) 5.697759e-19 ****** E 1146.1723 ( 31 X 36.9733) 1.894004e-19 **** E 1183.1456 ( 32 X 36.9733) 4.570131e-19 ***** E 1220.1189 ( 33 X 36.9733) 1.159998e-19 **** E 1257.0922 ( 34 X 36.9733) 7.456581e-19 ****** E 1294.0655 ( 35 X 36.9733) 1.878796e-19 **** E 1331.0388 ( 36 X 36.9733) 2.353345e-19 ***** E 1368.0121 ( 37 X 36.9733) 3.421471e-19 ***** E 1404.9854 ( 38 X 36.9733) 9.029505e-19 ****** E 1441.9587 ( 39 X 36.9733) 1.817377e-19 **** E 1478.9320 ( 40 X 36.9733) 5.408922e-19 ****** E 1515.9053 ( 41 X 36.9733) 2.081987e-19 **** E 1552.8786 ( 42 X 36.9733) 5.663798e-19 ****** E 1589.8519 ( 43 X 36.9733) 1.955283e-19 **** E 1626.8252 ( 44 X 36.9733) 5.386580e-19 ****** E 1663.7985 ( 45 X 36.9733) 1.815808e-19 **** E 1700.7718 ( 46 X 36.9733) 6.755061e-19 ****** E 1737.7451 ( 47 X 36.9733) 2.875197e-19 ***** E 1774.7184 ( 48 X 36.9733) 9.101411e-19 ****** E 1811.6917 ( 49 X 36.9733) 1.793191e-19 **** E 1848.6650 ( 50 X 36.9733) 6.064694e-19 ****** E 1885.6383 ( 51 X 36.9733) 2.315700e-19 ***** E 1922.6116 ( 52 X 36.9733) 5.657082e-19 ****** E 1959.5849 ( 53 X 36.9733) 1.808916e-19 **** E 1996.5582 ( 54 X 36.9733) 5.993952e-19 ****** Comb 23 (type G, offset=0.000000): Frequency (offset + harmonic x fund freq) Ampl (m/rtHz) Bar (logarithmic) G 75.2300 ( 1 X 75.2300) 5.650544e-19 ****** G 150.4600 ( 2 X 75.2300) 4.786849e-20 *** Comb 24 (type H, offset=0.000000): Frequency (offset + harmonic x fund freq) Ampl (m/rtHz) Bar (logarithmic) H 99.9989 ( 1 X 99.9989) 1.151681e-19 **** H 199.9978 ( 2 X 99.9989) 1.537022e-19 **** H 299.9967 ( 3 X 99.9989) 1.837480e-19 **** H 399.9956 ( 4 X 99.9989) 1.808672e-19 **** H 499.9945 ( 5 X 99.9989) 4.663949e-19 ****** H 599.9934 ( 6 X 99.9989) 1.727749e-19 **** H 699.9923 ( 7 X 99.9989) 1.632279e-19 **** H 799.9912 ( 8 X 99.9989) 1.590969e-19 **** H 899.9901 ( 9 X 99.9989) 1.661713e-19 **** H 999.9890 ( 10 X 99.9989) 6.638706e-19 ****** H 1099.9879 ( 11 X 99.9989) 1.430272e-19 **** H 1199.9868 ( 12 X 99.9989) 1.250370e-19 **** H 1299.9857 ( 13 X 99.9989) 1.386724e-19 **** Comb 25 (type N, offset=0.000000): Frequency (offset + harmonic x fund freq) Ampl (m/rtHz) Bar (logarithmic) N 59.9954 ( 1 X 59.9954) 8.914744e-18 ********* N 119.9908 ( 2 X 59.9954) 1.734097e-19 **** N 179.9862 ( 3 X 59.9954) 1.194757e-19 **** N 239.9816 ( 4 X 59.9954) 4.568680e-20 ** N 299.9770 ( 5 X 59.9954) 1.321572e-19 **** N 359.9724 ( 6 X 59.9954) 7.677886e-20 *** Quad suspension violin modes found below 2000 Hz -------------------------------------------------- Frequency Ampl (m/rtHz) Bar (logarithmic) Q 500.0540 1.875717e-19 **** Q 500.2115 2.111922e-19 **** Q 501.0911 6.837285e-19 ****** Q 501.2075 1.914244e-19 **** Q 501.2544 1.125682e-19 **** Q 501.4511 4.291879e-19 ***** Q 501.6072 4.471661e-16 ************** Q 501.6828 3.379318e-19 ***** Q 501.7492 1.893888e-18 ******* Q 501.8120 3.540886e-18 ******** Q 502.6216 1.759788e-18 ******* Q 502.7450 3.434911e-19 ***** Q 503.0067 1.648757e-18 ******* Q 503.1192 1.353738e-15 **************** Q 504.8049 8.421481e-15 ****************** Q 504.8714 1.553473e-18 ******* Q 505.5870 3.255312e-15 ***************** Q 505.7072 3.768438e-16 ************** Q 505.7100 4.594068e-16 ************** Q 505.8059 1.512066e-15 **************** Q 506.9220 3.591584e-16 ************** Q 507.1589 1.479609e-15 **************** Q 507.1941 3.954132e-16 ************** Q 507.3916 9.818416e-16 *************** Q 507.9916 1.893329e-18 ******* Q 508.0098 1.676502e-18 ******* Q 508.1462 6.080023e-19 ****** Q 508.2061 7.226787e-19 ****** Q 508.2206 1.847623e-18 ******* Q 508.2892 4.440535e-17 *********** Q 508.5871 3.216031e-18 ******** Q 508.6614 1.799532e-16 ************* Q 991.7478 1.388656e-16 ************* Q 991.9345 1.042019e-16 ************* Q 992.4256 2.695087e-15 ***************** Q 992.7944 9.191528e-16 *************** Q 994.2767 2.796105e-15 ***************** Q 994.6456 5.489510e-15 ****************** Q 994.7331 1.411453e-16 ************* Q 994.8973 8.186311e-16 *************** Q 995.3650 6.224299e-17 ************ Q 995.6447 1.359711e-15 **************** Q 996.2517 1.786856e-17 ********** Q 996.5296 1.357216e-15 **************** Q 997.7169 1.485482e-15 **************** Q 997.8868 7.275091e-15 ****************** Q 998.6645 1.247629e-16 ************* Q 998.8050 1.110122e-14 ******************* Q 1003.6673 6.524590e-17 ************ Q 1003.7788 1.648572e-16 ************* Q 1003.9071 4.262611e-17 *********** Q 1004.0782 8.984392e-17 ************ Q 1004.5370 1.892429e-16 ************* Q 1005.1694 2.213103e-16 ************** Q 1005.9378 2.242579e-16 ************** Q 1006.5031 7.930463e-16 *************** Q 1008.4502 5.017011e-18 ********* Q 1008.4938 4.259899e-16 ************** Q 1009.0273 3.999826e-16 ************** Q 1009.2089 3.466584e-18 ******** Q 1009.4402 3.314371e-16 ************** Q 1009.4863 3.600768e-16 ************** Q 1009.6234 4.241757e-18 ******** Q 1009.6825 2.312520e-18 ******** Q 1456.1793 2.918963e-16 ************** Q 1456.8448 2.242283e-16 ************** Q 1461.4125 3.835923e-16 ************** Q 1461.7318 2.589147e-17 *********** Q 1461.8627 6.243383e-17 ************ Q 1462.0311 8.789925e-16 *************** Q 1462.3129 4.051772e-16 ************** Q 1462.5991 3.739838e-17 *********** Q 1463.0994 3.470524e-16 ************** Q 1467.4759 2.262205e-17 *********** Q 1467.9648 4.470766e-16 ************** Q 1470.3809 1.755062e-15 **************** Q 1470.8263 1.322916e-16 ************* Q 1471.9279 9.294864e-17 ************ Q 1472.4505 1.336832e-16 ************* Q 1472.5268 3.205782e-17 *********** Q 1474.0800 6.057009e-17 ************ Q 1475.0976 2.313791e-16 ************** Q 1475.2510 2.150263e-17 ********** Q 1476.3779 1.450781e-17 ********** Q 1478.1701 6.111175e-17 ************ Q 1478.6459 2.023265e-17 ********** Q 1482.5840 8.466869e-17 ************ Q 1484.0765 5.843903e-17 ************ Q 1484.4293 1.522690e-18 ******* Q 1484.5241 8.156747e-17 ************ Q 1484.5731 1.364639e-18 ******* Q 1484.6685 7.455261e-17 ************ Q 1922.9259 2.338904e-16 ************** Q 1923.6124 2.665879e-17 *********** Q 1923.8582 1.964312e-18 ******* Q 1923.8644 1.449818e-16 ************* Q 1924.6770 8.434641e-17 ************ Q 1924.9178 3.852445e-17 *********** Q 1926.2402 8.747033e-18 ********* Q 1927.4652 4.264302e-19 ***** Q 1927.4662 6.137983e-18 ********* Q 1928.4664 8.919755e-17 ************ Q 1929.3172 1.585377e-17 ********** Q 1931.5738 6.803297e-19 ****** Q 1932.1391 2.376757e-17 *********** Q 1932.3359 1.774950e-17 ********** Q 1932.6117 2.384644e-16 ************** Q 1940.3232 2.038662e-18 ******* Q 1940.6643 3.864447e-16 ************** Q 1941.3501 1.275518e-17 ********** Q 1942.1270 5.441152e-18 ********* Q 1942.1751 1.958547e-17 ********** Q 1942.3900 4.032629e-18 ******** Q 1943.7780 3.960320e-17 *********** Q 1944.1332 3.215838e-18 ******** Q 1946.7318 2.485923e-17 *********** Q 1947.7089 1.481911e-17 ********** Q 1954.4574 1.493959e-17 ********** Q 1955.9208 1.432710e-17 ********** Q 1957.3347 3.912228e-18 ******** Q 1959.0215 6.527076e-18 ********* Beam splitter violin modes found below 2000 Hz ------------------------------------------------ Frequency Ampl (m/rtHz) Bar (logarithmic) B 299.4100 1.373450e-19 **** B 299.5400 9.552369e-19 ****** B 299.5783 2.007264e-19 **** B 299.7033 9.892853e-19 ****** B 302.2221 9.142438e-19 ****** B 303.3058 1.001566e-18 ******* B 599.1365 8.948132e-19 ****** B 599.4233 1.206580e-18 ******* B 604.4871 8.453061e-19 ****** B 606.6669 1.025896e-18 ******* B 898.7788 1.285682e-19 **** B 899.2300 1.292322e-19 **** B 906.8300 1.287981e-19 **** B 910.1085 1.341089e-19 **** Single lines found below 1000 Hz ---------------------------------- Frequency Ampl (m/rtHz) Bar (logarithmic) X 5.4060 2.612161e-14 ******************** X 6.0600 9.821926e-15 ****************** X 19.6080 1.510055e-17 ********** X 27.4170 2.830231e-17 *********** X 27.5930 3.438610e-18 ******** X 27.8203 2.292371e-18 ******** X 27.8606 3.667243e-18 ******** X 28.2137 3.860399e-17 *********** X 29.9695 2.555521e-18 ******** X 30.2164 1.993863e-18 ******* X 35.7049 1.321442e-18 ******* X 40.9365 5.247770e-17 ************ X 42.9150 7.327316e-19 ****** X 46.0890 4.613957e-18 ******** X 51.3150 1.056687e-19 **** X 64.3724 2.154358e-19 **** X 75.2300 5.650544e-19 ****** X 81.0000 7.717217e-18 ********* X 83.3180 2.730312e-20 ** X 99.9740 1.143574e-19 **** X 106.1900 2.108981e-20 * X 138.6100 5.968052e-20 *** X 138.6600 4.420359e-20 ** X 170.7500 3.124725e-20 ** X 359.9275 4.395967e-20 ** X 369.7305 8.815782e-19 ****** X 452.8830 2.545964e-19 ***** X 453.3980 1.418349e-19 **** X 1150.0000 1.612473e-19 **** X 1672.0000 2.464303e-19 ***** X 1704.0000 2.299713e-19 ***** X 1880.0000 1.901050e-19 **** X 1896.0000 2.747790e-19 ***** X 1999.9200 3.550047e-19 *****