H1 SUS ETMY PUM Driver S1102654 /ligo/gitcommon/suspensions/ligo/electronics/fits/ fit_ITMY_PUM_driver_20220117.py on 2021-01-17 ~13:00 PST. The Answer: STATE 1 (ACQ OFF, LP OFF) SUMMARY UL -- [z]:[p] = [12.1803]:[113.6837] LL -- [z]:[p] = [12.3093]:[114.7967] UR -- [z]:[p] = [12.3441]:[115.0118] LR -- [z]:[p] = [12.0803]:[112.4055] STATE 2 (ACQ ON, LP OFF) SUMMARY UL -- [z]:[p] = [ 1.3617 3.5809 144.8768 41543.1133]:[ 3.576 76.4377 166.1999 32658.7107] LL -- [z]:[p] = [ 1.3749 77.0526 145.2179 13137.7848]:[ 76.9103-0.0061j 76.9103+0.0061j 166.635 -0.j 12732.3392-0.j ] UR -- [z]:[p] = [ 1.3616 2.45 146.0432 42866.3372]:[ 2.4426 76.4422 166.7998 33494.721 ] LR -- [z]:[p] = [ 1.3603 1.8388 143.3509 41010.595 ]:[ 1.8202 76.1815 165.2128 32255.937 ] STATE 3 (ACQ OFF, LP ON) SUMMARY UL -- [z]:[p] = [ 12.1710 27.3049 107.1420 ]:[ 2.4296 112.8847 1205.6673 ] LL -- [z]:[p] = [ 12.3310 27.5800 109.2670 ]:[ 2.4669 114.6605 1218.8928 ] UR -- [z]:[p] = [ 12.3413 27.6735 107.3107 ]:[ 2.4654 114.0302 1208.4609 ] LR -- [z]:[p] = [ 12.0733 28.0830 105.9011 ]:[ 2.4977 110.9624 1199.6369 ] STATE3 / STATE 1 SUMMARY UL -- [z]:[p] = [ 3.3488 27.2313 108.1098 ]:[ 2.4192 3.3597 1207.2547 ] LL -- [z]:[p] = [ 2.2592 27.5952 109.4840 ]:[ 2.2436 2.4807 1220.5785 ] UR -- [z]:[p] = [ 3.4359 27.6321 108.3918 ]:[ 2.4543 3.4491 1209.9594 ] LR -- [z]:[p] = [ 3.0491 27.9953 107.5277 ]:[ 2.4829 3.0633 1201.2130 ] AOSEM IMPEDANCE SUMMARY UL -- [z]:[p] = [ 125.6849 834.3799 1751.1282]:[ 131.0161 1175.8141 17809.2085] LL -- [z]:[p] = [ 116.6856 723.2315 1341.4576 2736.4936]:[ 121.2433 869.6445 2361.2862 18917.8614] UR -- [z]:[p] = [ 123.929 834.9612 1825.0572]:[ 129.1906 1192.9006 17749.4195] LR -- [z]:[p] = [ 122.4376 837.1883 1845.9303]:[ 127.5633 1198.3934 17672.9644]