Configuration of SR785 options used to capture OMCA DCPD Channels A&B Voltage ASD measurements. # Configuration of SR785 ## Display Setup: Measurement group FFT A Trace: FFT Chan 1, LogMag, Units: dB units off, pk units: RMS, PSD units on. B Trace: FFT Chan 2, View: LogMag, Units: dB units off, pk units: RMS, PSD units: on ## Source: off ## input - source: analog - analyzer config: dual chan - config menu (channel 1 & channel 2 settings) - input mode: A-B - coupling: AC - Channel Range: - Aa filter: on - anti whitening filter: off - auto range: up only - auto offset: OFF ## Window Menu: Window default: BFH -> switched to Hanning ## Averaging Menu: - Compute Averages: yes - Averaging Type: linear/fixed length - number of averages: 5 - display average: RMS - Average Preview: Off - Preview Time: 2s ### More - Overload Reject: Off - Trigger Average Mode: Time Record ## Output Menu: - Hard Copy: ASCII Dump Frequency Span: 10Hz 12.9kHz Frequency lines: 800 16Hz RBW # Data Exports ## Export 1 Span: 6.4kHz, 800 lines, Low Freq: 10Hz, RBW: 8Hz -6dB Vpk range (both channels) ### Files ``` 2022-07-13_H1OMC_OMCA_DCPDA_800lines_10-6410Hz_inprange_m6dBVpk.TXT 2022-07-13_H1OMC_OMCA_DCPDB_800lines_10-6410Hz_inprange_m6dBVpk.TXT ``` ## Export 2 Span: 6.4kHz, 800 lines, Low Freq: 10Hz, RBW: 8Hz ADC Range: 4dB Vpk range, ### Files ``` 2022-07-13_H1OMC_OMCA_DCPDA_800lines_10-6410Hz_inprange_p4dBVpk.TXT 2022-07-13_H1OMC_OMCA_DCPDB_800lines_10-6410Hz_inprange_p4dBVpk.TXT ``` ## Export 3 span: 51.2kHz, start: 1khz, 800 lines, rbw: 64Hz adc range: -4dB Vpk ### Files ``` 2022-07-13_H1OMC_OMCA_DCPDA_800lines_1k-52p2kHz_inprange_m4dBVpk.TXT 2022-07-13_H1OMC_OMCA_DCPDB_800lines_1k-52p2kHz_inprange_m4dBVpk.TXT ``` ## Export 4 span: 51.2kHz, start: 1khz, 800 lines, rbw: 64Hz adc range: -6dB Vpk ### Files ``` 2022-07-13_H1OMC_OMCA_DCPDA_800lines_1k-52p2kHz_inprange_m6dBVpk.TXT 2022-07-13_H1OMC_OMCA_DCPDB_800lines_1k-52p2kHz_inprange_m6dBVpk.TXT ``` ## Export 5 span: 51.2kHz, start: 50kHz, 800 lines, RBW: 64Hz Number of averages: 100 adc range: -6dB Vpk ### Files ``` 2022-07-13_H1OMC_OMCA_DCPDA_800lines_50-101p2kHz_inprange_m6dBVpk.TXT 2022-07-13_H1OMC_OMCA_DCPDB_800lines_50-101p2kHz_inprange_m6dBVpk.TXT ```