Both tests: 25 Hz excitation for 240 seconds Channel 1 Test Start time: 1367347574 Excited Channel: H1:PEM-CS_ACC_HAM6VAC_SEPTUM_Y_DQ Ratio of accelerometer ASD during excitation to excited accelerometer ASD: Channel Frequency (Hz) - Ratio (counts/counts) H1:PEM-CS_ACC_PSL_PERISCOPE_X_DQ 25 - 2.928035428340081e-05 50 - 0.00038980800309218466 75 - 0.001905537093989551 100 - 0.025286663323640823 H1:PEM-CS_ACC_PSL_PERISCOPE_Y_DQ 25 - 5.4996326070977375e-05 50 - 0.0005994198727421463 75 - 0.0027604594361037016 100 - 0.03899331018328667 H1:PEM-CS_ACC_PSL_TABLE1_X_DQ 25 - 2.0073284758836962e-05 50 - 0.00021208969701547176 75 - 0.0008946711895987391 100 - 0.01016039215028286 H1:PEM-CS_ACC_PSL_TABLE1_Y_DQ 25 - 3.644914613687433e-05 50 - 0.00035056404885835946 75 - 0.0014438519719988108 100 - 0.017143158242106438 H1:PEM-CS_ACC_PSL_TABLE1_Z_DQ 25 - 5.021219840273261e-05 50 - 0.0004129698208998889 75 - 0.00173992442432791 100 - 0.012193546630442142 H1:PEM-CS_ACC_PSL_TABLE2_Z_DQ 25 - 5.936908200965263e-05 50 - 0.00028819666476920247 75 - 0.0012952868128195405 100 - 0.012602935545146465 H1:PEM-CS_ACC_ISCT1_REFL_Y_DQ 25 - 4.789167724084109e-05 50 - 0.00026065559359267354 75 - 0.00018125087080989033 100 - 0.0011250507086515427 H1:PEM-CS_ACC_SQZT6_HOMODYNE_X_DQ 25 - 3.9790040773368673e-07 50 - 2.242629761894932e-06 75 - 9.499200132268015e-06 100 - 0.00043681554961949587 H1:PEM-CS_ACC_IOT2_IMC_Y_DQ 25 - 0.00012056363630108535 50 - 0.0011140212882310152 75 - 0.0023326773662120104 100 - 0.09929320216178894 H1:PEM-CS_ACC_LVEAFLOOR_HAM6_Z_DQ 25 - 5.92994547332637e-05 50 - 0.0006934930570423603 75 - 0.00038522749673575163 100 - 0.002293755766004324 H1:PEM-CS_ACC_HAM4_SR2_X_DQ 25 - 1.0349498552386649e-05 50 - 0.00015397997049149126 75 - 0.0003584604128263891 100 - 0.001145710120908916 H1:PEM-CS_ACC_EBAY_FLOOR_Z_DQ 25 - 0.0001398378808517009 50 - 0.0010734731331467628 75 - 0.0029689576476812363 100 - 0.008483119308948517 H1:PEM-CS_ACC_ISCT6_SQZLASER_X_DQ 25 - 7.280699355760589e-05 50 - 0.000635225442238152 75 - 0.002018025843426585 100 - 0.0034471233375370502 Channel 16 Test Start time: 1367347851 Excited Channel: H1:PEM-CS_ACC_PSL_TABLE2_Z_DQ Ratio of accelerometer ASD during excitation to excited accelerometer ASD: Channel Frequency (Hz) - Ratio (counts/counts) H1:PEM-CS_ACC_PSL_PERISCOPE_X_DQ 25 - 3.452219971222803e-05 50 - 0.0003267276333644986 75 - 0.0017197505803778768 100 - 0.02678137645125389 H1:PEM-CS_ACC_PSL_PERISCOPE_Y_DQ 25 - 4.0488503145752475e-05 50 - 0.0004522359522525221 75 - 0.0022657932713627815 100 - 0.0505606010556221 H1:PEM-CS_ACC_PSL_TABLE1_X_DQ 25 - 2.249857789138332e-05 50 - 0.00020146660972386599 75 - 0.0008005812996998429 100 - 0.011113373562693596 H1:PEM-CS_ACC_PSL_TABLE1_Y_DQ 25 - 3.0556100682588294e-05 50 - 0.000279019121080637 75 - 0.0011613329406827688 100 - 0.02129446342587471 H1:PEM-CS_ACC_PSL_TABLE1_Z_DQ 25 - 4.053770680911839e-05 50 - 0.0003352837229613215 75 - 0.0013892995193600655 100 - 0.012590707279741764 H1:PEM-CS_ACC_ISCT1_REFL_Y_DQ 25 - 4.537958011496812e-05 50 - 0.0002024308341788128 75 - 0.00021175495930947363 100 - 0.0010948141571134329 H1:PEM-CS_ACC_SQZT6_HOMODYNE_X_DQ 25 - 5.605581350209832e-07 50 - 7.898095645941794e-06 75 - 1.5055758922244422e-05 100 - 0.0003843605809379369 H1:PEM-CS_ACC_IOT2_IMC_Y_DQ 25 - 0.00013724557356908917 50 - 0.0010309992358088493 75 - 0.002245798707008362 100 - 0.08699029684066772 H1:PEM-CS_ACC_LVEAFLOOR_HAM6_Z_DQ 25 - 5.570419307332486e-05 50 - 0.0006299180095084012 75 - 0.00036953791277483106 100 - 0.002074664691463113 H1:PEM-CS_ACC_HAM4_SR2_X_DQ 25 - 1.0829509847098961e-05 50 - 0.00017505453433841467 75 - 0.0003116489970125258 100 - 0.0011309105902910233 H1:PEM-CS_ACC_EBAY_FLOOR_Z_DQ 25 - 0.00011985340825049207 50 - 0.0009439214481972158 75 - 0.0025017305742949247 100 - 0.006966133136302233 H1:PEM-CS_ACC_ISCT6_SQZLASER_X_DQ 25 - 6.93980255164206e-05 50 - 0.00048796291230246425 75 - 0.0016080839559435844 100 - 0.00361593347042799 H1:PEM-CS_ACC_HAM6VAC_SEPTUM_Y_DQ 25 - 1.1462258044048212e-05 50 - 0.00020961553673259914 75 - 0.00019388871442060918 100 - 0.0012029058998450637