#OFI transmission measurements import numpy as np mW = 10**-3 uW = 10**-6 nW = 10**-9 A=np.array([7.589, 7.708, 7.742])*mW A_std=np.array([96.83, 41.13, 47.93])*uW B = np.array([1.025, 1.056, 1.015])*uW B_std = np.array([25.55, 12.11, 12.0])*nW B_bg = np.array([809.0, 845.0, 822.0])*nW B_bg_std = np.array([2.227, 8.0, 4.0])*nW Isolation = [] for meas in [0,1,2]: isolation=(2*(B[meas]-B_bg[meas]))/ A[meas] Isolation.append(isolation) np.array(Isolation) Isolation_mean = np.mean(Isolation) print('The three measurements are: {}'.format(10*np.log10(Isolation))) Isolation_dB = 10*np.log10(Isolation_mean) print('The mean isolation measurede is: {}'.format(Isolation_dB))