Beam Parameters Report				
Timestamp	06/08/26 at 15:59:34			
User				
SID4 S/N	SID4-1657			
Version	PhaseStudio v2402.0.4			
Wavelength (nm)	1000.00			
Magnification	1.00			
Parameters	Unit	R	Min	Max
M	--	1.22	1.15	1.30
2W0	mm	0.5156	0.4878	0.5413
z0	mm	-378.3795	-384.4954	-372.5148
zR0	mm	170.2205	162.7818	176.8621
zR	mm	1017.6942	1070.9716	961.4691
2W	mm	1.2566	1.2549	1.2584
Divergence	mrad	3.0288	2.9969	3.0604
Azimuth NF		35.6741		
Azimuth FF		3.9033		
Near Field				
2W	mm	1.2567		
2Wmin	mm	1.2620		
2Wmax	mm	1.2513		
Azimuth		35.6741		
2Wx	mm	1.2584		
2Wy	mm	1.2549		
Radii of curvature				
Rx	mm	-453.6234		
Ry	mm	-456.2705		
Rmin	mm	-453.4115		
Rmax	mm	-456.4852		
Thetamin		-15.2741		
Rfoc	mm	-454.9431		
Rast	mm	-134675.6719		
ThetaAst		-15.2741		