Thorlabs Beam M² Data Export Date MM-DD-YYYY;07-09-2026 Time HH:MM:SS;08:35:25 Beam Profiler Model;BP209-VIS Serial Number;M00280451 Profiler Azimuth Angle ;[deg]; Wavelength;[nm];1064 Resolution;[µm];1.2 Beam Quality Measurement;M2 Beam Diameter Measure Method;Beam Width Clip (1/e²) * * * * * * * * * * After Lens Measurement Results * * * * * * * * * * Parameter;Unit;Result; M²x'; ;1.35; M²y'; ;1.35; Rayleigh Length X';mm;14.14; Rayleigh Length Y';mm;12.68; Beam Waist Diameter X';µm;160.73; Beam Waist Diameter Y';µm;152.18; Beam Waist Asymmetrie; ;0.95; Beam Waist Position X';mm;131.86; Beam Waist Position Y';mm;130.77; Astigmatism;mm;1.09; Relative Astigmatism; ;0.08; Divergence Angle X';deg;0.651; Divergence Angle Y';deg;0.687; Divergence Asymmetry;%;1.056; * * * * * * * * * * Original Waist Measurement Results * * * * * * * * * * Parameter;Unit;Result; M²x'; ;1.35; M²y'; ;1.35; Rayleigh Length X';mm;1250.76; Rayleigh Length Y';mm;1275.99; Beam Waist Diameter X';µm;1511.61; Beam Waist Diameter Y';µm;1526.37; Beam Waist Asymmetrie; ;1.01; Beam Waist Position X';mm;2314.74; Beam Waist Position Y';mm;2484.63; Astigmatism;mm;169.89; Relative Astigmatism; ;0.13; Divergence Angle X';deg;0.069; Divergence Angle Y';deg;0.069; Divergence Asymmetry;%;0.990; * * * * * * * * * * BEAM DIAMETERS * * * * * * * * * * Z;X';Y'; 200.00;804.17;863.55; 180.00;573.17;610.57; 160.00;358.14;380.30; 140.00;175.78;186.05; 130.05;158.07;144.57; 128.27;162.16;149.74; 126.50;170.73;156.00; 124.72;180.29;166.59; 122.94;192.40;180.48; 121.17;203.50;193.35; 120.00;212.92;204.58; 119.39;219.44;211.70; 100.00;404.05;413.29; 80.00;619.91;644.02; 60.00;842.19;876.35; 40.00;1050.73;1089.22; 20.00;1278.25;1326.63; 0.00;1496.16;1567.04;