After discovering that the MPC polarization controller creates a peak at 27kHz in the EX PLL noise spectrum, I decided to take some more measurements at the end station with the controller off.
PLL Servo Board
PDH Servo Board
PLL error signal measured out of PFD IMON, and PDH error signal measured out of demod IMON.
The lowest RMS comes from the PLL Boost 1: OFF, and the PDH Boost 2: ON. With the PLL Boost 1 off there is no gain peaking and the the OLTF of the PLL is stable with a UGF of 22kHz and a phase margin of 50 deg. I need to take an OLTF with the PDH Boost 2 on.
Adjusting EX PLL boost 1 ON/OFF and EX PDH boost 2 ON/OFF as above, the various Comm PLL Error signals (PFD IMON) were measured.