Since the Faraday swap on monday, we have more green power.
On the table we measured 50mW going towards the chamber, where previously we had measured 47mW (6% increase). On QPD A we had a 10% increase in the sum, on QPD B a 30% increase in the sum.
Alexa and I measured 19uW arriving on ISCT1, compared to 12 uW measured in alog 9191 ( 60% increase from January)
We also noticed that the beam quality was better than it had been. We measured some profiles, it is a guassian beam, they will be attached to this post later.
When the arm is locked we see a 35% increase in the transmitted power measured by the COMM BBPD, compared to last week.