krishna.venkateswara@LIGO.ORG - posted 14:57, Wednesday 05 November 2014 - last comment - 10:32, Friday 07 November 2014(14872)
BRS and T240 temperature sensor noise

K. Venkateswara

I had installed temperature sensors on BRS and GND_T240 yesterday as described in 14825. The first plot shows the trend over a day along with the PEM_VEA temperature sensor. The count to Kelvin conversion was expected to be 1.56e-3 K/count. This seems roughly consistent with the temperature of the T240. The BRS temperature sensor shows a lower magintude and a phase offset due to it's extra thermal shielding and larger mass.

The attached pdf shows the ASD of the temperature sensors and the coherence between them and their respective instruments. The temperature sensors are mostly limited by ADC noise. An op-amp based pre-amp of 50-100 gain would be useful. BRS_RY_Out shows a little bit of coherence near few mHz while T240 X, Y and Z show no coherence with the temperature sensor.

Images attached to this report
Non-image files attached to this report
Comments related to this report
richard.mittleman@LIGO.ORG - 07:16, Thursday 06 November 2014 (14884)

You should add the ADC noise to this plot

krishna.venkateswara@LIGO.ORG - 10:32, Friday 07 November 2014 (14910)

I've added a comment about it in 14909. I'm not sure how to display the ADC noise in DTT. In any case, it shouldn't be limited by ADC noise any more.