J. Kissel, for the Calibration Team I've updated the results from LHO aLOG 21825 and G1501223 with an ASD from the current lock stretch, such that I could display the computed time dependent correction factors, which have recently been cleared of systematics (LHO aLOG 22056), sign errors (LHO aLOG 21601), and bugs yesterday (22090). I'm happy to say, that not only does the ASD *without* time dependent corrections still fall happily within the required 10%, but if one eye-balls the time-dependent corrections and how they would be applied at each of the respective calibration line frequencies, they make sense. To look at all relevant plots (probably only interesting to calibrators and their reviewers), look at the first pdf attachment. The second and third .pdfs are the money plots, and the text files are a raw ascii dump of respective curves so you can plot them however or whereever you like. All of these files are identical to what is in G1501223. This analysis and plots have been made by /ligo/svncommon/CalSVN/aligocalibration/trunk/Runs/O1/H1/Scripts/produceofficialstrainasds_O1.m which has been committed to the svn.
Apparently, this script has been moved to a slightly different location. The script can be found at
/ligo/svncommon/CalSVN/aligocalibration/trunk/Runs/O1/H1/Scripts/DARMASDs/produceofficialstrainasds_O1.m