C. Cahillane
I have updated the LHO uncertainty budget, fixed some bugs, and generated fewer, more relevant plots.
All calibration versions are included in every budget.  I will review the details each version below:
C00: No kappas, no static systematics applied
C01: No kappas, static systematics applied
C02: kappa_tst and kappa_C applied, kappa_pu and cavity pole not applied, static systematics applied
C03: "Perfect", so all kappas and cavity pole applied, static systematics applied
Whenever I do ratio comparisons between model response functions, the "perfect" C03 model is in the denominator.
***** PDFs 1, 2, and 3 *****
These contain the C01, C02, and C03 uncertainty budgets at GPS Time = 1126259461.  (Kappas provided by Darkhan in aLOG 11499.  Thanks Darkhan!)
Plot 1 is the calibration version response function over C03's response function. 
Plot 2 is four plots.  The mag and phase squared uncertainty components plots and the total mag and phase strain uncertainty by itself.
Plots 3-6 are just enlarged versions of Plot 2.
***** PDF 4 and 5 *****
PDF 4 is the Sensing fits and statistics
PDF 5 are the Actuation stages fits and statistics
***** PDF 6 *****
This 4 way subplot of all the calibration versions together on one plot.  I plot the mag and phase response functions and their ratios.
For the C01 calibration version, we have under 7% and 4 degrees of uncertainty (See PDF 1, Plot 2)
For the C02 calibration version, we have under 6% and 3.5 degrees of uncertainty (See PDF 2, Plot 2)