Reports until 00:57, Wednesday 13 January 2016
H1 CAL (SUS)
jeffrey.kissel@LIGO.ORG - posted 00:57, Wednesday 13 January 2016 - last comment - 19:17, Wednesday 06 July 2016(24917)
H1SUSETMY L1 to L3 (UIM to TST) High Frequency Actuation Characterization
J. Kissel, E. Goetz

Having a few new breakthrough ideas on the UIM actuation system (see LHO aLOG 24914), we explored whether we are modeling the [m/N] L1/UIM force to L3/TST displacement transfer function incorrectly. This was done by driving the UIM out to 600 [Hz] and measuring the response in DARM. Not only did we find the expected-but-not-yet-modeled wire violin modes at ~330 [Hz], 420 [Hz], and some at ~500 [Hz], but we found several bending-mode resonances at 111 [Hz] and 166 [Hz]. Indeed, upon first glance, we think the 111 [Hz] resonance is the remaining missing frequency dependence that explains the turn-up seen at 100 [Hz] in all previous measurements of the UIM to TST transfer function.

We'll process in more detail some time in the future, but check out the attached screen shots and be amazed at how not 1/f^6 the L1 to L3 transfer function is.

--------
We'd started by exciting the L1 stage via awggui in a broad-band fashion such that we could catch all of the wire violin mode frequencies watching DARM. As Evan mentions, we had suspected that these wire resonances -- documented in T1300876 -- were the source of the deviation from 1/f^6, and they'd just not been included in the SUS dynamical model -- [[EDIT -- Brett has now included them in the model, and they are a non-negligible effect; see LHO aLOG 24915]]. This broadband TF is shown in the first attachment. Black is the with excitation ON, and red is ambient (to distinguish the ~505 [Hz] fiber violin modes from the ~495 [Hz] TOP to UIM wire violin modes, and the Beam Splitter violin modes & 331.7 [Hz] calibration line from the Sus. Point to TOP wire violin modes). Only the peaks of the wire violin mode resonances are visible above the DARM noise; driving them any higher breaks the IFO's lock.

Just in case, we drove down to the same ~80 [Hz] region, and BINGO! We also found new, unexpected resonances at 166 [Hz] and even as low as 111 [Hz]! (see second and third attachments) Our best guess for the source of these resonances are imperfect actuators. Perhaps the bending of the L-bracket that mounts the OSEM coil to the reaction chain's UIM (see D060375, see page 1 for the total assembly, lock at pg 14 for the L-bracket). Though, if it were these L brackets, I'd suspect there would be 4 independent resonances... also it doesn't look like enough moving mass to have resonance frequencies as low as ~100 [Hz]... dunno, will think more.

Finding something at 111 [Hz], we then took a careful swept sign measurement covering it, and indeed, it looks strikingly like another piece of the UIM puzzle. (see 4th attachment). We also grabbed a PCAL to DARM transfer function over this frequency vector, so we can turn the TF into an absolute calibration later.

For future reference, the templates for the swpet sine TFs are here:
/ligo/svncommon/CalSVN/aligocalibration/trunk/Runs/PostO1/H1/Measurements/FullIFOActuatorTFs/
2016-01-12_L1toDARM_FullLock.xml
2016-01-12_PCALYtoDARM_FullLock.xml

and I attach screenshots of the awggui sessions used to excite L1 and PCAL in a broad-band fashion (DARM_IN1 ASDs during the broad-band excitations were taken using the standard wall FOM for DARM).
Images attached to this report
Comments related to this report
brett.shapiro@LIGO.ORG - 11:16, Wednesday 13 January 2016 (24925)SUS

Weird about the 111 Hz and 160 Hz modes. If it is a mechanical mode of the UIM OSEMs, it is probably more likely to be the magnet-flag assembly, which makes a nice long cantilever, and is attached to the main chain itself. See page page 16 of D060375. Additionally, if for some reason the set screw holding the flag assembly in place is loose, you met get lower frequency modes.

The L bracket is on the reaction chain, so if the mode was in that it have to couple through to the main chain via the magnetic field gradient inside the coil; it's possible but one more step removed.

You could try exciting these modes one OSEM at a time, to see if it is coming from one in particular. If we get lucky, maybe we'll find there is a simple fix, like tightening a set screw.

I also wonder if these features exist on other test mass suspensions.

norna.robertson@LIGO.ORG - 14:14, Friday 15 January 2016 (24972)SUS, SYS
The 111Hz feature is very likely from the first internal mode of the UIM blades which is not surprising sicne it will ebt there at some level due to cross-coupling.  The frequency is very close - see for example

https://alog.ligo-la.caltech.edu/aLOG/index.php?callRep=16740

where the frequencies were measured at LLO for their quads to be in the range ~111 to 112 Hz.


As for the 166 Hz, I don't have a good idea. It is not the second resonance of these blades. Lab measurements of such a blade here at Caltech give the second mode at around 325 Hz. Funnily enough this is ~twice the observed feature, but I can't think why we would see something at half the frequency of a blade mode.
kiwamu.izumi@LIGO.ORG - 19:17, Wednesday 06 July 2016 (28206)

I was asked by Jeff to fit the UIM data so that we can include the peaks at 111 and 167 Hz in our calibration model. After some struggle, I ended up doing an emperical zpk model which gave me the following parameters:

=========================
gain = 2.270401e-09
f:pole0 = 1.113398e+02
Q:pole0 = 5.596904e+14
f:pole1 = 1.950899e+02
Q:pole2 = 4.743158e+00
f:zero0 = -1.133450e+02
Q:zero0 = 6.220362e+01
=========================

In addition to these fitted parameters, I had fixed zpks which are 6 poles at 1 Hz and 1 complex pole at 166.7 Hz with a Q of 200. The attached shows a comparison of the fitting and data. I have used fminsearch to minimize a weighted residual. I didn't have an energy to compute the uncertainties in the estimated parameters.

One thing I don't like with this fitting is that the fitted model falls faster than the nominal 1/f^6 slope above approximately 160 Hz due to the extra poles that I put in to make the fitting better at frequencies below.

 

The code and resultant figure can be found at:

  • /ligo/svncommon/CalSVN/aligocalibration/trunk/Runs/PostO1/H1/Scripts/FullIFOActuatorTFs/uim_fitting_adventure.m
  • /ligo/svncommon/CalSVN/aligocalibration/trunk/Runs/PostO1/H1/Results/FullIFOActuatorTFs/2016-01-12_UIM_tf_fitting.pdf
Non-image files attached to this comment