After decoupling the pumping components used during the recent bake out of the Y-end RGA, I exposed the RGA to the Y-end vacuum volume, energized the filament and let it come into equilibrium for an hour or more. I then let the RGA scan continuously with the multiplier (SEM) on for an additional hour or so while I gathered up my mess(es). I periodically checked the scanning as I walked past the screen. At one point, I noticed that the spectrum was changing rapidly towards the "dirty". I monitored the scanning and noted that after reaching a temporary maximum, the amus which had increased then returned to near their original values. After consulting with the Jeff B. (the operator on shift), I feel that the observed changes in partial pressures were likely the result of IFO locking attempts as they coincide closely in time. Perhaps something gets hot when the IFO is locked or when mirrors are steered? See attached scans
If true that could be kind of scary (!) Can we set an RGA in MID (stripchart) mode and run time series following the main peaks through a locking attempt?
I could imagine baking the adsorbed water off the ETM and perhaps nearby baffles. But this should not persist (or repeat) after the first good cavity buildup.
Mike - Chandra's stated goal is to eventually continuously trend 7 AMUs (max allowed by software) at each building. The observance cited in this aLOG obviously would have been missed while in Faraday. Too bad that the RGAs don't live long with their SEMs on 24/7. As we install/commission the RGAs and as she works out the issues with the CDS and/or GC folks this trending will eventually be happening. J. Smith - The partial pressures that are changing are too small and not expected to show up on the total pressure gauges. From the graphic scans and knowing that the total pressure at the Y-end is 2 x 10-9 torr, we see that the partial pressures that changed are small (10-12 torr) - but still interesting because they are measurable and even more interesting if the changes can be shown to be tied to some IFO locking activity. (Science interesting? Who knew?)
Doh!!! Here are the .txt versions of the ASCII data
The indicated currents for these scans are typical of the SEM @ 1300 volts (which is the factory default). I have noticed in the past that setting the SEM voltage value in the EDIT tab does not change the value displayed in the device status screen or vice versa - so, though I set this to 1500 volts in one of those two fields, it may not have taken effect.