Reports until 11:46, Tuesday 03 January 2017
H1 CAL (CAL)
aaron.viets@LIGO.ORG - posted 11:46, Tuesday 03 January 2017 (32920)
New GDS filters for LHO with corrected analog AA gain
[Evan G, Alex U, Aaron V]

I have produced new filters for Hanford that incorporate the correction Evan made to the Pcal to DARM transfer function. (see https://alog.ligo-wa.caltech.edu/aLOG/index.php?callRep=32907)
The filters can be found in the calibration SVN at this location:
/ligo/svncommon/CalSVN/aligocalibration/trunk/Runs/O2/GDSFilters/H1GDS_1167485844.npz

The filters were produced using this Matlab script in SVN revision 4029:
ligo/svncommon/CalSVN/aligocalibration/trunk/Runs/O2/H1/Scripts/TDfilters/H1_run_td_filters_1167485844.m

The parameters files used (all in revision 4029) were:
ligo/svncommon/CalSVN/aligocalibration/trunk/Runs/ER10/Common/params/IFOindepParams.conf
ligo/svncommon/CalSVN/aligocalibration/trunk/Runs/ER10/H1/params/H1params.conf
ligo/svncommon/CalSVN/aligocalibration/trunk/Runs/ER10/H1/params/2016-11-12/H1params_2016-11-12.conf
ligo/svncommon/CalSVN/aligocalibration/trunk/Runs/O2/H1/params/H1_TDparams_1167485844.conf
ligo/svncommon/CalSVN/aligocalibration/trunk/Runs/ER10/H1/Scripts/CAL_EPICS/D20161122_H1_CAL_EPICS_VALUES.m

The first two plots (png files) are spectrum comparisons between CALCS and GDS. All the same options were used that are currently being used on the online calibration, including application of time-dependent corrections. The last pdf file shows kappas computed by SLM, GDS, and CALCS. It appears that Evan's correction has resolved the discrepancy we were seeing in kappa_tst and kappa_pu between GDS and CALCS. Compare the data from the summary pages from the same time period: https://ldas-jobs.ligo-wa.caltech.edu/~detchar/summary/day/20161207/cal/time_varying_factors/
However, SLM tool data does not agree at this time.
Images attached to this report
Non-image files attached to this report