I checked the Vacuum Site Overview MEDM screen and saw the telltale rise in pressure at the Y-mid station that is consistent with a warming 80K cyrogenic pump, I investigated and found that IP9 had tripped off so I restarted it. Only one of the two channels restarted. The other channel is inop. While on site I overfilled CP4 by opening the LLCV manual bypass valve 1/2 turn. LN2 was observed at the exhaust outlet in a total of 38 minutes.
C Leaving UHP GN2 regulator output set at 40 psig and flow limitied to 2 LPM. 0355 hrs. local -> Kyle leaving site (this time for real). Cheryl V. in CR
Thank you, Kyle and Gerardo, for responding. Tagging VE.
Remote logged in at around 7:00 utc, determined that IP9 had tripped, notified Kyle who was already manually filling CP4 (first time).
At the same time I noticed that I had not fully restored CP4 fill control screen and the power for LLCV was still disabled (which is very easy to miss), at this time I enabled the power and restored all values, this screen was not restored due to a wrong burtrestore file, see Patrick's earlier aLOG.