Displaying report 1-1 of 1.
Reports until 09:54, Friday 24 March 2017
H1 CAL (CAL)
aaron.viets@LIGO.ORG - posted 09:54, Friday 24 March 2017 - last comment - 05:20, Wednesday 07 June 2017(35041)
Data collected to analyze SRC detuning
I have added an (unreleased) algorithm into the GDS/DCS pipeline to compute the SRC spring frequency and Q. This algorithm was used to collect 18 hours of data on February 4 (first plot) and 18 hours of data on March 4 (second plot). The plots include kappa_c, the cavity pole, the SRC spring frequency, the SRC Q, and 4 of the coherence uncertainties (uncertainty of the 7.93 Hz line is not yet available).

The derivation this algorithm was based on is similar to what Jeff has posted ( https://dcc.ligo.org/DocDB/0140/T1700106/001/T1700106-v1.pdf ), with differences noted below:
1) The approximation made at the bottom of p4 and top of p5 was only used in the calculation of kappa_c and the cavity pole. So SRC detuning effects were not accounted for in computing S_c. However, kappa_c and the computed cavity pole were used in the calculation of S_s.
2) In eq. 18, I have a minus (-) sign instead of a plus (+) sign before EP6. S_c = S(f_1, t) has been computed this way in GDS/DCS since the start of O2 (I assume during O1 as well).
3) Similarly, in eq. 20, I have a minus sign (-) before EP12.
4) In the lower two equations of 13, I have the terms under the square root subtracted in the opposite order, as suggested by Shivaraj. (Also, I noted that the expression for Q should only depend on S(f_2, t), with no dependence on S(f_1, t). )

The smoothing (128s running median + 10s average) was done on f_s and 1/Q, since that is the way they would be applied to h(t). Therefore, the zero-crossings of 1/Q show up as asymptotes in the plot of Q. I think it would be better to output 1/Q in a channel rather than Q for this reason.

There is a noticeable ramping up of f_s at the beginning of lock stretches, and the range of values agrees with what has been measured previously.

I've noted that it is quite difficult to resolve the value of Q with good accuracy. These are some reasons I suspect:
1) Higher uncertainty of calibration measurements at low frequency can add a systematic error to the EPICS values computed at 7.93 Hz. This may be why the Q is more often negative than positive ??
2) In the calculation of S_s, the actuation strength is subtracted from the ratio of pcal and DARM_ERR. Since this is such a low frequency, the subtracted values are close to the same value in magnitude and phase. Thus, subtracting magnifies both systematic error and uncertainty.
3) The imaginary part of S_s (see eq. 13, bottom equation) in the denominator, is very close to zero, so small fluctuations (about zero, as it turns out) in 1/Q cause large fluctuations in Q.

These reasons make it difficult to measure Q with this method. The effect of these measured-Q fluctuations on S_s, the factor we would actually apply to h(t) (see eq. 22), is not enormous, so long as we apply the smoothing to 1/Q, as I have done here.
Images attached to this report
Comments related to this report
shivaraj.kandhasamy@LIGO.ORG - 09:16, Monday 27 March 2017 (35105)CAL

First attachment is a hand written note containing derivation of the equations 13 in DCC document T1700106. As Aaron mentioned above, in the derivation the order of the quantities in the sqrt function comes out to be in the opposite order  (Re[S] - abs[S]^2 instead of abs[S]^2 - Re[S]).  The second plot show  the estimation of the four sensing function quantities for 2017-01-24 calibration sweep measurement done at LHO (a-log 33604). Instead of tracking across time here we track across sweep frequencies. The top two plots in the second figure show the estimation of optical gain and cavity pole frequency assuming no detuning. We see that above ~100 Hz we get almost constant values for optical gain and cavity pole frequency suggesting detuning doesn't affect the estimation of those quantities (currently we use 331.9 Hz line at LHO for estimating optical gain and cavity pole). Substituting back the optical gain and cavity pole calculated this way, we then calculated detuning frequency and Q. The bottom two plots of the second figure show those. We see that upto ~60 Hz we can use the lines to estimate detuning frequency (currently at LHO we are running the line at 7.83 Hz). However the Q is hard to estimate, the variation is pretty large (Evan's recent a-log also indicate this 34967; Aaron also finds this to be the case). Also in the 7-10 Hz region its value seems to be negative (need to look at more data to make sure that it not just a fluctuation). With the current set of calibration lines, it seems tracking of detuning frequency would easy but estimating Q might be a little difficult.

Images attached to this comment
Non-image files attached to this comment
shivaraj.kandhasamy@LIGO.ORG - 10:21, Monday 17 April 2017 (35584)

In the second page of the derivation,  at the half way point I have unintentionally switched the notation from S_s to S_c (it should be S_s till the end of the page 2).

aaron.viets@LIGO.ORG - 05:20, Wednesday 07 June 2017 (36692)
[Daniel Finstad, Aaron Viets]

The time series and histograms attached show additional data collected using the DCS calibration pipeline from Jan 19, 2017 at 21:44:14 UTC (GPS 1168897472) until Jan 20, 2017 at 09:02:38 UTC (GPS 1168938176).
Images attached to this comment
Displaying report 1-1 of 1.