J. Kissel, T. Sadecki Travis and I used a HeNe Laser, QPD, and speaker array to begin characterizing H1SUSETMY's PUM-to-TST fused silica fiber violin modes. We were able to complete the characterization of 3 of the 4 fibers, but we began to have (we believe to be) electronics problems during characterization of the last fiber, so we've left the set turned off but otherwise as is, in hopes the the gremlins will go away by tomorrow morning to characterize the final fiber. I've exported the data, and intent to process it in detail as I've done for ITMX (see LHO aLOG 39163), but for now preliminary results are as follows (under the "Answer" column): Freq resolution 62.5mHz (400 lines with 25 Hz span) +X / +Y (S1400157; Partial Suspended Fundamental: 504.5 Hz) Freq Band / Hz Answer / Hz nAvgs Exc Amp / mV FileName 1st 490 515 504.8 10 250 0003 2nd 990 1015 999.8125 10 250 0004 3rd 1460 1485 1471.3125 10 250 0005 4th 1975 2000 1980 10 250 0006 5th 2450 2475 2460.3125 10 500 0007 6th 2925 2950 2939.625 10 500 0008 -X / +Y (S1400159; Partially Suspended Fundamental: 503.0 Hz) 1st 490 515 503.625 10 500 0009 2nd 990 1015 1000.3125 10 500 0010 3rd 1460 1485 1470.375 10 500 0011 4th 1975 2000 1979.4375 10 500 0012 5th 2450 2475 2459.4375 10 500 0013 6th 2925 2950 2939.375 10 500 0014 +X / -Y (S1800486; Partially Suspended Fundamental: 513.25 Hz) 1st 500 525 512.75 10 500 0015 2nd 1000 1025 1016.8125 10 500 0016 3rd 1490 1515 1500.375 10 500 0017 4th 1975 2000 1979.5625 10 500 0018 5th 2450 2475 2459.375 10 500 0019 6th 2925 2950 2939.3125 10 500 0020 So far, the difference between fully suspended and partially suspended values look consistent with the ~0.5 Hz shift down in frequency as seen on ITMX. Note, we tried upgrading the HeNe laser to a spare fiber-coupled optical lever laser as requested by Angus in LHO aLOG 40300, but the output coupler we were given collimated the beam to much larger than we needed (a ~10mm beam for a ~3mm QPD). Also, we were seeing a large amount of glitch-like impulses while taking the ASD of the QPD SUM output. A scope, time-series reading of the QPD output showed no glitching at any time scale. Suspecting the new laser, we tried out the old, supposedly noisy one. Simultaneously, I found that the SR785 input settings were DC coupled, and switched to A-B with AC coupling. Either the laser swap or the AC coupling fixed the problem, but since we were on a roll, we didn't switch back to the new spare optical lever laser. All of today's measurements were taken with the old HeNe laser.