Fil, Dean, Richard, Rich
After removing the newly implemented revised piezo driver two days ago, some modifications were made to see if there was any improvement to the broadening of noise in DARM around 180Hz. We put the revised chassis back in and the DARM noise is back at 180Hz.
Attached is a detailed description of the state of the changes including:
1. Noise spectra of the bias HV driver as measured on the bench
2. Power supply rejection vs frequency of the bias HV driver
3. Simplified description of the topology used.
Koji reminded me that this type of power line harmonic noise associated with the OMC PZT driver has been seen at LLO, and mitigated by the addition of a ferrite toroid to the shutter trigger cable. There was also mention of unplugging the Picomotor driver while it was in an unused state. These changes apparently improved multiple lines in DARM. It's a good read. See:
https://alog.ligo-la.caltech.edu/aLOG/index.php?callRep=41374
This should be investigated at LHO prior to removing (again) the modified OMC Piezo driver.