FSS open loop transfer function
Measured the FSS open loop transfer function.  Common gain = 20.0 dB, fast gain = 15.0 dB according to the FSS MEDM screen
slider values.  Measured UGF was ~417 kHz with a phase margin of 56 deg.  Just about where we would want to be in some
regards.

    No detrimental effects from last week's outage.

OLTF.[txt, jpg] transfer function to 1 MHz, text data and image
OLTF2.[txt, jpg] transfer function to 5 MHz, text data and image