CLF RIN and ISS studies

Lee, Kentaro, Nutsinee, Daniel

The A+ filter cavity will require more CLF light and demand a stronger requirement on the RIN of the CLF light to not inject noise in DARM. The requirement isn't particularly tough at 5e-5 RIN/rtHz of the CLF light carried with the squeezer field (T1800447-v6 figure 13). The laser already meets the requirement in broadband, but has lines and beam jitter noise that can take it above. Mark upgraded and installed an ISS for the 532 OPO pump as well as one for the CLF itself. Today we studied if the CLF on the SQZ path is witnessed to meet requirements with and without the new ISS's.

Bottom line is that the ISS's do help, and can meet the requirements with them even if we want to turn up the CLF more than needed. As it turns out, the pump RIN is the more important one, as the pump seems to get more noise on it going through the SHG, fiber and OPO. The OPO-transmission signal that the ISS servo feeds back to is the proper witness, but has a very low light level due to the small transmissivity at 532 of OPO-M2.

Setup:

CLF_SQZ_PATH_RIN.png shows the RIN measured from the homodyne at different configurations.

PUMP_RIN_XFER.png uses the OPO 532 REFL as a witness for the RIN during the ISS-OFF time (OPO 532 TRANS was not properly saved). It shows the direct imprint into the CLF RIN with a gain of 0.4. I have an elaborate OPO model for the filter cavity that manages to predict this factor of 0.4 at the operating NLG of 3.2: dNLG.pdf. As we increase the NLG, the RIN2RIN coupling will increase, but not severely.

CLFRINONOFF.png shows the in-loop CLFPD DC with and without the servo. The OFF-level agrees with the level expected of the laser + some acoustic excesses.

OPO_TRANS_RIN.png shows the pump RIN of the in-loop trans PD with the servo on and off. When it is off, the level corresponds roughly to the shot-noise of a 1uW 532 beam, so the diode doesn't seem to be dark-noise limited, but the sensitivity is still pretty bad.

OPO_REFL_WITNESS.png shows that with the 532 TRANS servo on and off, the REFL does not show the same reduction. While REFL has more light and lower sensing noise than TRANS, the excess shown here while the TRANS-ISS is running indicates that REFL will indeed be a poor witness to stabilize the circulating power. There must be beam jitter noise or other cavity noise driving the difference between the two.