We are having problems powering up.
The problem happens while the rotation stage is rotating to increase power incident on the IMC. To compensate our analog CARM and IMC loops for the increase in optical gain, we decrease the analog gain using the H1:IMC-REFL_SERVO_FASTGAIN slider. This slider goes in steps of 1 dB, so it's not exactly a smooth ride, but usually it's fine. Now it is not. Dips in power everywhere after the IMC (MC2 trans, MC4 trans, PRG) are associated with some, not all, FASTGAIN steps.
It's not clear what changed. This wasn't a problem Saturday or Monday. I checked the IMC and CARM OLGs just before we power up (attached), and both looked fine.
We tried decreasing the CARM gain on H1:IMC-REFL_SERVO_IN2GAIN by 3 dB (from -22 dB to -25 dB) just before running MAXIMUM_POWER, to test if the three UGFs in the CARM loop was the problem. It was not, the decrease in CARM gain did not help with the power glitches.
(btw CARM has three UGFs because at ~20 kHz the 9 MHz sidebands become resonant in CARM, increasing the CARM plant gain. This is why the loop wins phase there.)
Anyway, the "ISS troubles" from earlier were the symptom that kills us. Jenne moved the second-loop turn on to after powerup, so the ISS no longer immediately ends the lock. Unfortunately, other lock-enders include the ADS system, whose output is not limited during powerup and is susceptible to large glitches (the impulse causes the ADS lowpasses to accumulate to push the IFO off alignment).