Monday's 2022-04-18 evening locks without RH, vs with 0.68W RH that was turned on 1 hour before lock.
|
ITMX |
ITMY |
Lock with No RH
|
IMAGE |
IMAGE |
Lock with with 0.68W/seg RH
(turned on 1 hour before lock)
|
IMAGE |
IMAGE |
Lock with No RH = 2022-04-19 2:36 UTC 1334370978 (locked at 50W for 50m)
Lock with with 0.68W/seg RH that was turned on 1 hour before lock = 2022-04-19 04:40 UTC 1334378418 (locked at 50W n for 50m)
Ref 2022-04-19 01:25 UTC 1334366718
Friday night 1.7W/seg RH: Note that these ring heater settings were probably too hot and didn't help the IFO
alog 62642. You can see the the ring heater quickly flattens the curvature of ITMY.
ITMY:
Image of Cold IFO vs Hot IFO with no ring heater.
Image of Cold IFO vs Hot IFO with ring heater on.
IFO cold + RH cold (2022-04-15 19:00 UTC 1334084418)
IFO hot (50W input for 4hours) + RH cold (2022-04-15 23:18 UTC 1334099898)
IFO hot (50Winput in for 5hours) with RH hot (1.7W/seg for 45mins) (2022-04-16 00:28 UTC 1334104112)
ITMX:
There is an offset in spherical power due to realignment on Friday everning (
alog 62641) and only turned on at
00:04 UTC. Cannot get an image of before the RH was turned on)
Image of Cold IFO vs Hot IFO with warming ring heater.
Image of Cold IFO vs Hot IFO with ring heater on.
IFO cold + RH cold (when RH turned off after 4hours Monday AM 2022-04-18 19:00 UTC 1334343618)
IFO hot (50W input for 4hours) + RH warming (1.7W/seg for 20mins) (2022-04-16 00:04 UTC 1334102658)
IFO hot (50Winput in for 5hours) with RH hot (1.7W/seg for 45mins) (2022-04-16 00:28 UTC 1334104112)