At 8pm tonight, I'll have a script turn ITMX RH up from 0W to 0.44W/segment and ETMX down from 0.44W/segment to 0W. Using 'python3 /tcs/common/scripts/power_adj_scripts/ring_heater_on.py ITMX -s 1357704018 -p 0.4'.
Keeping total amount of power in X-arm the same in the hope of avoiding ringing up PIs.
Tomorrow morning Sheila will log-in and check if we are locked / able to lock with these settings. If not she will turn them back to the nominal (ITMX at 0W, ETMX at 0.44W/seg) as Robert has tests Saturday PM that need a locked IFO. Thank you Sheila!
We have left the ring heaters in the ITMX at 0.44W/seg, ETMX at 0W settings as the IFO seemed happy and locking over the weekend.
Attaching ndscope plots where circulating powers are visible.
Attached are the demodulated frequency injection lines from this RH test and several days of locks afterwards. This RH change made the 25 and 4500 Hz coupling worse but did improve the 201Hz. Although around 200Hz there is quite a lot of structure in the frequency noise coupling so unsure how much weight should be put on the 200Hz demodulation line as we could just be seeing this zero feature more around.