After we found the CO2X power has been degrading in 72943, I recalibrated the CO2 rotation stages this morning. Last done in March 66724.
Added a README.txt file to /opt/rtcds/userapps/release/tcs/h1/scripts/RS_calibration/ with instructions.
Also adjusted lscparams to output the existing values. Plan to relock with yesterdays output and then during lock, adjust CO2X to output 1.68W as the start of O4.
tcs_nom_annular_pwr = {'X': 0.95, #change back to 1.03 after alog72943 test
'Y': 1.19}