Displaying report 1-1 of 1.
Reports until 18:36, Wednesday 11 October 2023
H1 PEM (DetChar)
robert.schofield@LIGO.ORG - posted 18:36, Wednesday 11 October 2023 (73403)
Broad 120 Hz peak no longer in DARM after PSL periscope tuning and shutdown of SF2

On Tuesday I tuned the periscope in the PSL by adjusting weights and Tyler and I shut down SF2, which was responsible for the new 120 Hz HVAC peak ( https://alog.ligo-wa.caltech.edu/aLOG/index.php?callRep=73378 ). Figure 1 shows the reduction of the periscope peak from tuning, and from tuning plus the fan shutdown. The fan shutdown had the greatest effect.

Figure 2 shows the disappearance of the peak from the summary pages, as well as the 3+ MPc increase in range from the fan change only.

Non-image files attached to this report
Displaying report 1-1 of 1.