Displaying report 1-1 of 1.
Reports until 14:12, Friday 21 June 2024
X1 SUS (SUS)
rahul.kumar@LIGO.ORG - posted 14:12, Friday 21 June 2024 - last comment - 14:09, Thursday 27 June 2024(78574)
Status update: A+ HRTS assembly and characterization work for O5

Ryan C, M Robinson, R Kumar

Given below is the summary of work done in the triples lab in the last two months regarding assembly of A+ HRTS suspension for O5 (refer to alog 76635 which gives the details about the first assembly in March 2024- assembly s/n 02). We have assembled four new suspensions (assembly s/n 03, 08, 01, 11), which brings the total number of assembled (with BOSEMs) and characterized suspension in the lab to five. All five assembly till now has been for the FREESTANDING version (only the base differs- which is as per D1900287).

In this round of assembly and testing procedure of HRTS we have made two new improvements as listed below,

1. Class B PEEK wire install tool (top mass to PUM - as shown in the picture attached) - This new PEEK part has been designed by the RAL team after we had wire breaking issues in the first round of assembly. Using this new tool we can safely and quickly install the wires between top mass and PUM outside the cage on the optical bench.

2. Vibration Isolation set up as shown in view01 and view02 - While running transfer function measurements, the HRTS suspension was strongly coupling-in with the lab/building environment (ground vibrations, doors, HVAC etc). To partially isolate the SUS, we now have an optical bench (12in by 18in, Thorlabs Nexus Breadboard) sitting on a 1inch thick viton pads (placed at the four corners of the bench). Also, Chris fabricated a hard top transparent cover for us to protect against the turbulent air flow in the lab.

Cage S/N 03 (April 2024)

Suspended masses (as per specifications):-

Top Mass = 749gm, Penultimate mass = 803gm, Dummy optic = 300gm

BOSEM S/N       OLC     offsets (-OLC/2) Gain (30,000/OLC)
S1900810 30629 15314.5 0.979464
S1900795 31353 15676.5 0.956846
S1900754 29884 14942 1.003882
S1900809 24739 12369.5 1.21266
S1900782 23231 11615.5 1.291378
S1900626 30032 15016 0.998934

Transfer function measurements for HRTS assembly S/N 03 is attached here.

Cage S/N 08 (April 2024)

Suspended masses (as per specifications):-

Top Mass = 755gm, Penultimate mass = 803gm, Dummy optic = 301gm

BOSEM S/N       OLC     offsets (-OLC/2) Gain (30,000/OLC)
S1900701 30772 15386 0.974912
S1900716 31514 15757 0.951958
S1900713 31407 15703.5 0.955201
S1900705 23915 11957.5 1.254443
S1900700 26432 13216 1.134988
S1900712 31165 15582.5 0.962618

Transfer function measurements for HRTS assembly S/N 08 is attached here.

Cage S/N 01 (May 2024)

Suspended masses (as per specifications):-

Top Mass = 752gm, Penultimate mass = 804gm, Dummy optic = 300gm

BOSEM S/N       OLC     offsets (-OLC/2) Gain (30,000/OLC)
S1900646 28515 14257.5 1.052078
S1900705 30408 15204 0.986582
S1900612 32048 16024 0.936096
S1900653 24963 12481.5 1.201779
S1900669 23010 11505 1.303781
S1900655 32286 16143 0.929195

Transfer function measurements for HRTS assembly S/N 01 is attached here.

Cage S/N 11 (May 2024)

Suspended masses (as per specifications):-

Top Mass = 750gm, Penultimate mass = 803gm, Dummy optic = 301gm

BOSEM S/N       OLC     offsets (-OLC/2) Gain (30,000/OLC)
S1900645 25810 12905 1.16234
S1900649 29934 14967 1.002205
S1900683 31808 15904 0.943159
S1900629 27963 13981.5 1.072846
S1900689 23763 11881.5 1.262467
S1900634 28063 14031.5 1.069023

Transfer function measurements for HRTS assembly S/N 11 is currently under processing and the latest plots will be attached soon.

Conclusions: - The transfer function measurements ties-up nicely to the model, however we do see some low frequency coupling and unwanted peaks. The cross coupling and unwanted peaks are due to environmental noise which is difficult to fully isolate in the lab.

I will also post a plot comparing the transfer function measurement results for all five suspension that we have tested.

Images attached to this report
Non-image files attached to this report
Comments related to this report
rahul.kumar@LIGO.ORG - 14:09, Thursday 27 June 2024 (78706)SUS

Attached below are the transfer function measurement results for Freestanding suspension SN_11. I see some strong cross coupling in Vertical dof, possible combing from R dof.

Non-image files attached to this comment
Displaying report 1-1 of 1.