Jennie, Jenne, Elenna, Vicky, Erik
As referred to in this entry we took a suite of measurements on the 3rd May to determine what changing the SRCL offset, DARM offset and turning off the whitening on DCPD would have on the sensitivity.
Measurement Order: from 05/03/2023
Measurement Set | DARM Measurement Time | PCAL > DARM Measurement Time | DARM offset (mA) | Whitening? | SRCL1 OFFSET | Squeezing optimised? | DARM Measurement (1 min) | PCAL > DARM Measurement (3 mins) | Figure Folder |
1 | 1367179951 | 1367180172 | 20 | ON | -200 | YES but slightly different squeezing angle was set between PCAL and DARM measurements | /ligo/home/jennifer.wright/git/DARM_offset/2023-05-03_2009UTC_H1_DARMSPEC_1m.xml |
Ref 1
|
|
2 | 1367184708 | 1367183677 | 40 | ON | -200 | YES | /ligo/home/jennifer.wright/git/DARM_offset/2023-05-03_2131UTC_H1_DARMSPEC_1m.xml |
Ref 2
|
|
3 | 1367184497 | 1367184372 | 40 | OFF | -200 | YES | /ligo/home/jennifer.wright/git/DARM_offset/ |
ref 3
|
At low frequencies DARM looks better with 20mA offset, but we must remember that many things at low frequency are optimised for this DARM offset. See first plot for 20mA/40mA comparison.
The whitening ON/OFF (see second plot attached) does not look to have made a large difference to the sensitivity from these plots.
There is no coherence in the PCAL > DARM BB measurements we took at high frequency so to give us a good idea of the difference between 20mA and 40 mA DARM offset we will scale the DARM spectra using the one of the high frequency calibration lines.
Unfortunately, we did not spend any time in NLN with 40mA (just NLN CAL MEAS) so we will need to measure another DARM spectra in that state today.
The code I used to correct the DARM spectra with the PCAL to DARM measurements is in /ligo/home/jennifer.wright/git/DARM_offset
The contrast defect measurement I made indicated that the contrast defect is still 1.7 mW. This measurement was taken about 18 hrs into lock, so we were fully thermalized.
Added in a third plot of the power scaling in the anti-symmetric port with DARM offset changing the power at the DCPDs.
The inverse of the slope of this plot gives the loss term as in this entry.
loss term = 1/1.219 = 0.820,
the amount of light at the anti-symmetric port insensitive to DARM is 837.5 mW.