WP 12836
ECR E2400330
Modified List T2500232
The following SUS SAT Amps were upgraded per ECR E2400330. Modification improves the whitening stage to reduce ADC noise from 0.05 to 10 Hz.
| Suspension | Old | New | OSEM |
| ETMX L2 (PUM) | S1100146 | S1100119 | ULLLURLR |
| ETMY L2 (PUM) | S1100137 | S1100127 | ULLLURLR |
| ITMX L2 (PUM) | S1100135 | S1100118 | ULLLURLR |
| ITMY L2 (PUM) | S1000277 | S1100148 | ULLLURLR |
F. Clara, J. Kissel, O. Patane
Here's the characterization data and fit results for S1100148, assigned to ITMY L2's ULLLURLR OSEMs.
This sat amp is a UK 4CH sat amp, D0900900 / D0901284. The data was taken per methods described in T080062-v3, using the diagrammatic setup shown on PAGE 1 of the Measurement Diagrams from LHO:86807.
The data was processed and fit using ${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Scripts/
plotresponse_S1100148_ITMY_L2_ULLLURLR_20250917.m
Explicitly, the fit to the whitening stage zero and pole, the transimpedance feedback resistor, and foton design string are:
The attached plot and machine readable .txt file version of the above table are also found in ${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Results/
2025-09-17_UKSatAmp_S1100148_D0901284-v5_fitresults.txt
Per usual, R_TIA_kOhm is not used in the compensation filter -- but after ruling out an adjustment in the zero frequency (by zeroing the phase residual at the lowest few frequency points), Jeff nudged the transimpedance a bit to get the magnitude scale within the ~0.25%, shown in the attached results. Any scaling like this will be accounted for instead with the absolute calibration step, i.e. Side Quest 4 from G2501621, a la what was done for PR3 and SR3 top masses in LHO:86222 and LHO:84531 respectively.
Here's the characterization data and fit results for S1100119, assigned to ETMX L2's ULLLURLR OSEMs.
This sat amp is a UK 4CH sat amp, D0900900 / D0901284. The data was taken per methods described in T080062-v3, using the diagrammatic setup shown on PAGE 1 of the Measurement Diagrams from LHO:86807.
The data was processed and fit using ${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Scripts/
plotresponse_S1100119_ETMX_L2_ULLLURLR_20250916.m
Explicitly, the fit to the whitening stage zero and pole, the transimpedance feedback resistor, and foton design string are:
The attached plot and machine readable .txt file version of the above table are also found in ${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Results/
2025-09-16_UKSatAmp_S1100119_D0901284-v5_fitresults.txt
Per usual, R_TIA_kOhm is not used in the compensation filter -- but after ruling out an adjustment in the zero frequency (by zeroing the phase residual at the lowest few frequency points), Jeff nudged the transimpedance a bit to get the magnitude scale within the ~0.25%, shown in the attached results. Any scaling like this will be accounted for instead with the absolute calibration step, i.e. Side Quest 4 from G2501621, a la what was done for PR3 and SR3 top masses in LHO:86222 and LHO:84531 respectively.
Here's the characterization data and fit results for S1100118, assigned to ITMX L2's ULLLURLR OSEMs.
This sat amp is a UK 4CH sat amp, D0900900 / D0901284. The data was taken per methods described in T080062-v3, using the diagrammatic setup shown on PAGE 1 of the Measurement Diagrams from LHO:86807.
The data was processed and fit using ${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Scripts/
plotresponse_S1100118_ITMX_L2_ULLLURLR_20250916.m
Explicitly, the fit to the whitening stage zero and pole, the transimpedance feedback resistor, and foton design string are:
The attached plot and machine readable .txt file version of the above table are also found in ${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Results/
2025-10-14_UKSatAmp_S1100118_D0901284-v5_fitresults.txt
Per usual, R_TIA_kOhm is not used in the compensation filter -- but after ruling out an adjustment in the zero frequency (by zeroing the phase residual at the lowest few frequency points), Jeff nudged the transimpedance a bit to get the magnitude scale within the ~0.25%, shown in the attached results. Any scaling like this will be accounted for instead with the absolute calibration step, i.e. Side Quest 4 from G2501621, a la what was done for PR3 and SR3 top masses in LHO:86222 and LHO:84531 respectively.
Here's the characterization data and fit results for S1100127, assigned to ETMY L2's ULLLURLR OSEMs.
This sat amp is a UK 4CH sat amp, D0900900 / D0901284. The data was taken per methods described in T080062-v3, using the diagrammatic setup shown on PAGE 1 of the Measurement Diagrams from LHO:86807.
The data was processed and fit using ${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Scripts/
plotresponse_S1100127_ETMY_L2_ULLLURLR_20250916.m
Explicitly, the fit to the whitening stage zero and pole, the transimpedance feedback resistor, and foton design string are:
The attached plot and machine readable .txt file version of the above table are also found in ${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Results/
2025-09-16_UKSatAmp_S1100127_D0901284-v5_fitresults.txt
Per usual, R_TIA_kOhm is not used in the compensation filter -- but after ruling out an adjustment in the zero frequency (by zeroing the phase residual at the lowest few frequency points), Jeff nudged the transimpedance a bit to get the magnitude scale within the ~0.25%, shown in the attached results. Any scaling like this will be accounted for instead with the absolute calibration step, i.e. Side Quest 4 from G2501621, a la what was done for PR3 and SR3 top masses in LHO:86222 and LHO:84531 respectively.
ITMY L2 (PUM) Sat Amp S1100148 installed on 10/14/2025. Replaced on 10/16/2024 with S1100080.
https://alog.ligo-wa.caltech.edu/aLOG/index.php?callRep=87515
Note that the satamp for ETMY L2 has now been swapped back to a S1100137, which has been modified for ECR E2400330.
alog: 87722