WP 12844
ECR E2400330
Modified List T2500232
The following SUS SAT Amps were upgraded per ECR E2400330. Modification improves the whitening stage to reduce ADC noise from 0.05 to 10 Hz.
| Suspension | Old | New | OSEM |
| MC1 M2 | S1100182 | S1100148 | ULLLURLR |
| MC1 M3 | S1100176 | S1100135 | ULLLURLR |
| MC3 M2 | S1100069 | S1100106 | ULLLURLR |
| MC3 M3 | S1100123 | S1100093 | ULLLURLR |
| PR3 M2 | S1100063 | S1000274 | ULLLURLR |
| PR3 M3 | S1100113 | S1000277 | ULLLURLR |
| MC2 M2 | S1100110 | S1100169 | ULLLURLR |
| MC2 M3 | S1000294 | S1100174 | ULLLURLR |
| SR3 M2 | S1100105 | S1100146 | ULLLURLR |
| SR3 M3 | S1100074 | S1100134 | ULLLURLR |
F. Clara, J. Kissel, O. Patane
Here's the characterization data and fit results for S1100148, assigned to MC1 M2's ULLLURLR OSEMs.
This sat amp is a UK 4CH sat amp, D0900900 / D0901284. The data was taken per methods described in T080062-v3, using the diagrammatic setup shown on PAGE 1 of the Measurement Diagrams from LHO:86807.
The data was processed and fit using ${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Scripts/
plotresponse_S1100148_MC1_M2_ULLLURLR_20251020.m
Explicitly, the fit to the whitening stage zero and pole, the transimpedance feedback resistor, and foton design string are:
The attached plot and machine readable .txt file version of the above table are also found in ${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Results/
2025-10-20_UKSatAmp_S1100148_D0901284-v5_fitresults.txt
Per usual, R_TIA_kOhm is not used in the compensation filter -- but after ruling out an adjustment in the zero frequency (by zeroing the phase residual at the lowest few frequency points), Jeff nudged the transimpedance a bit to get the magnitude scale within the ~0.25%, shown in the attached results. Any scaling like this will be accounted for instead with the absolute calibration step, i.e. Side Quest 4 from G2501621, a la what was done for PR3 and SR3 top masses in LHO:86222 and LHO:84531 respectively.
Here's the characterization data and fit results for S1100135, assigned to MC1 M3's ULLLURLR OSEMs.
This sat amp is a UK 4CH sat amp, D0900900 / D0901284. The data was taken per methods described in T080062-v3, using the diagrammatic setup shown on PAGE 1 of the Measurement Diagrams from LHO:86807.
The data was processed and fit using ${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Scripts/
plotresponse_S1100135_MC1_M3_ULLLURLR_20251020.m
Explicitly, the fit to the whitening stage zero and pole, the transimpedance feedback resistor, and foton design string are:
The attached plot and machine readable .txt file version of the above table are also found in ${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Results/
2025-10-20_UKSatAmp_S1100135_D0901284-v5_fitresults.txt
Per usual, R_TIA_kOhm is not used in the compensation filter -- but after ruling out an adjustment in the zero frequency (by zeroing the phase residual at the lowest few frequency points), Jeff nudged the transimpedance a bit to get the magnitude scale within the ~0.25%, shown in the attached results. Any scaling like this will be accounted for instead with the absolute calibration step, i.e. Side Quest 4 from G2501621, a la what was done for PR3 and SR3 top masses in LHO:86222 and LHO:84531 respectively.
Here's the characterization data and fit results for S1100106, assigned to MC3 M2's ULLLURLR OSEMs.
This sat amp is a UK 4CH sat amp, D0900900 / D0901284. The data was taken per methods described in T080062-v3, using the diagrammatic setup shown on PAGE 1 of the Measurement Diagrams from LHO:86807.
The data was processed and fit using ${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Scripts/
plotresponse_S1100106_MC3_M2_ULLLURLR_20251020.m
Explicitly, the fit to the whitening stage zero and pole, the transimpedance feedback resistor, and foton design string are:
The attached plot and machine readable .txt file version of the above table are also found in ${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Results/
2025-10-20_UKSatAmp_S1100106_D0901284-v5_fitresults.txt
Per usual, R_TIA_kOhm is not used in the compensation filter -- but after ruling out an adjustment in the zero frequency (by zeroing the phase residual at the lowest few frequency points), Jeff nudged the transimpedance a bit to get the magnitude scale within the ~0.25%, shown in the attached results. Any scaling like this will be accounted for instead with the absolute calibration step, i.e. Side Quest 4 from G2501621, a la what was done for PR3 and SR3 top masses in LHO:86222 and LHO:84531 respectively.
Here's the characterization data and fit results for S1100093, assigned to MC3 M3's ULLLURLR OSEMs.
This sat amp is a UK 4CH sat amp, D0900900 / D0901284. The data was taken per methods described in T080062-v3, using the diagrammatic setup shown on PAGE 1 of the Measurement Diagrams from LHO:86807.
The data was processed and fit using ${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Scripts/
plotresponse_S1100093_MC3_M3_ULLLURLR_20251020.m
Explicitly, the fit to the whitening stage zero and pole, the transimpedance feedback resistor, and foton design string are:
The attached plot and machine readable .txt file version of the above table are also found in ${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Results/
2025-10-20_UKSatAmp_S1100093_D0901284-v5_fitresults.txt
Per usual, R_TIA_kOhm is not used in the compensation filter -- but after ruling out an adjustment in the zero frequency (by zeroing the phase residual at the lowest few frequency points), Jeff nudged the transimpedance a bit to get the magnitude scale within the ~0.25%, shown in the attached results. Any scaling like this will be accounted for instead with the absolute calibration step, i.e. Side Quest 4 from G2501621, a la what was done for PR3 and SR3 top masses in LHO:86222 and LHO:84531 respectively.
Here's the characterization data and fit results for S1000274, assigned to PR3 M2's ULLLURLR OSEMs.
This sat amp is a UK 4CH sat amp, D0900900 / D0901284. The data was taken per methods described in T080062-v3, using the diagrammatic setup shown on PAGE 1 of the Measurement Diagrams from LHO:86807.
The data was processed and fit using ${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Scripts/
plotresponse_S1000274_PR3_M2_ULLLURLR_20251020.m
Explicitly, the fit to the whitening stage zero and pole, the transimpedance feedback resistor, and foton design string are:
The attached plot and machine readable .txt file version of the above table are also found in ${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Results/
2025-10-20_UKSatAmp_S1000274_D0901284-v5_fitresults.txt
Per usual, R_TIA_kOhm is not used in the compensation filter -- but after ruling out an adjustment in the zero frequency (by zeroing the phase residual at the lowest few frequency points), Jeff nudged the transimpedance a bit to get the magnitude scale within the ~0.25%, shown in the attached results. Any scaling like this will be accounted for instead with the absolute calibration step, i.e. Side Quest 4 from G2501621, a la what was done for PR3 and SR3 top masses in LHO:86222 and LHO:84531 respectively.
Here's the characterization data and fit results for S1000277, assigned to PR3 M3's ULLLURLR OSEMs.
This sat amp is a UK 4CH sat amp, D0900900 / D0901284. The data was taken per methods described in T080062-v3, using the diagrammatic setup shown on PAGE 1 of the Measurement Diagrams from LHO:86807.
The data was processed and fit using ${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Scripts/
plotresponse_S1000277_PR3_M3_ULLLURLR_20251020.m
Explicitly, the fit to the whitening stage zero and pole, the transimpedance feedback resistor, and foton design string are:
The attached plot and machine readable .txt file version of the above table are also found in ${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Results/
2025-10-20_UKSatAmp_S1000277_D0901284-v5_fitresults.txt
Per usual, R_TIA_kOhm is not used in the compensation filter -- but after ruling out an adjustment in the zero frequency (by zeroing the phase residual at the lowest few frequency points), Jeff nudged the transimpedance a bit to get the magnitude scale within the ~0.25%, shown in the attached results. Any scaling like this will be accounted for instead with the absolute calibration step, i.e. Side Quest 4 from G2501621, a la what was done for PR3 and SR3 top masses in LHO:86222 and LHO:84531 respectively.
Here's the characterization data and fit results for S1100169, assigned to MC2 M2's ULLLURLR OSEMs.
This sat amp is a UK 4CH sat amp, D0900900 / D0901284. The data was taken per methods described in T080062-v3, using the diagrammatic setup shown on PAGE 1 of the Measurement Diagrams from LHO:86807.
The data was processed and fit using ${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Scripts/
plotresponse_S1100169_MC2_M2_ULLLURLR_20251020.m
Explicitly, the fit to the whitening stage zero and pole, the transimpedance feedback resistor, and foton design string are:
The attached plot and machine readable .txt file version of the above table are also found in ${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Results/
2025-10-20_UKSatAmp_S1100169_D0901284-v5_fitresults.txt
Per usual, R_TIA_kOhm is not used in the compensation filter -- but after ruling out an adjustment in the zero frequency (by zeroing the phase residual at the lowest few frequency points), Jeff nudged the transimpedance a bit to get the magnitude scale within the ~0.25%, shown in the attached results. Any scaling like this will be accounted for instead with the absolute calibration step, i.e. Side Quest 4 from G2501621, a la what was done for PR3 and SR3 top masses in LHO:86222 and LHO:84531 respectively.
Here's the characterization data and fit results for S1100174, assigned to MC2 M3's ULLLURLR OSEMs.
This sat amp is a UK 4CH sat amp, D0900900 / D0901284. The data was taken per methods described in T080062-v3, using the diagrammatic setup shown on PAGE 1 of the Measurement Diagrams from LHO:86807.
The data was processed and fit using ${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Scripts/
plotresponse_S1100174_MC2_M3_ULLLURLR_20251020.m
Explicitly, the fit to the whitening stage zero and pole, the transimpedance feedback resistor, and foton design string are:
The attached plot and machine readable .txt file version of the above table are also found in ${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Results/
2025-10-20_UKSatAmp_S1100174_D0901284-v5_fitresults.txt
Per usual, R_TIA_kOhm is not used in the compensation filter -- but after ruling out an adjustment in the zero frequency (by zeroing the phase residual at the lowest few frequency points), Jeff nudged the transimpedance a bit to get the magnitude scale within the ~0.25%, shown in the attached results. Any scaling like this will be accounted for instead with the absolute calibration step, i.e. Side Quest 4 from G2501621, a la what was done for PR3 and SR3 top masses in LHO:86222 and LHO:84531 respectively.
Here's the characterization data and fit results for S1100146, assigned to SR3 M2's ULLLURLR OSEMs.
This sat amp is a UK 4CH sat amp, D0900900 / D0901284. The data was taken per methods described in T080062-v3, using the diagrammatic setup shown on PAGE 1 of the Measurement Diagrams from LHO:86807.
The data was processed and fit using ${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Scripts/
plotresponse_S1100146_SR3_M2_ULLLURLR_20251020.m
Explicitly, the fit to the whitening stage zero and pole, the transimpedance feedback resistor, and foton design string are:
The attached plot and machine readable .txt file version of the above table are also found in ${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Results/
2025-10-20_UKSatAmp_S1100146_D0901284-v5_fitresults.txt
Per usual, R_TIA_kOhm is not used in the compensation filter -- but after ruling out an adjustment in the zero frequency (by zeroing the phase residual at the lowest few frequency points), Jeff nudged the transimpedance a bit to get the magnitude scale within the ~0.25%, shown in the attached results. Any scaling like this will be accounted for instead with the absolute calibration step, i.e. Side Quest 4 from G2501621, a la what was done for PR3 and SR3 top masses in LHO:86222 and LHO:84531 respectively.
Here's the characterization data and fit results for S1100134, assigned to SR3 M3's ULLLURLR OSEMs.
This sat amp is a UK 4CH sat amp, D0900900 / D0901284. The data was taken per methods described in T080062-v3, using the diagrammatic setup shown on PAGE 1 of the Measurement Diagrams from LHO:86807.
The data was processed and fit using ${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Scripts/
plotresponse_S1100134_SR3_M3_ULLLURLR_20251020.m
Explicitly, the fit to the whitening stage zero and pole, the transimpedance feedback resistor, and foton design string are:
The attached plot and machine readable .txt file version of the above table are also found in ${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Results/
2025-10-20_UKSatAmp_S1100134_D0901284-v5_fitresults.txt
Per usual, R_TIA_kOhm is not used in the compensation filter -- but after ruling out an adjustment in the zero frequency (by zeroing the phase residual at the lowest few frequency points), Jeff nudged the transimpedance a bit to get the magnitude scale within the ~0.25%, shown in the attached results. Any scaling like this will be accounted for instead with the absolute calibration step, i.e. Side Quest 4 from G2501621, a la what was done for PR3 and SR3 top masses in LHO:86222 and LHO:84531 respectively.