Elenna, Oli
We spent some time figuring out SDFs for the BBSS. It should be all good to go.
A couple of notable channels that will need to be figured out at some point are:
- H1:SUS-BS_M2_DRIVEALIGN_Y2L_GAIN: For the BSFM this gain was set to 11. Since we had inputs coming in for this filter we decided to commit it as a gain of 0 until we verify the gain that needs to be there
- H1:SUS-BS_M2_OLDAMP_{P,Y}_GAIN: For the BSFM the P gain had been set to 300 and the gain had been set to 650. We seemed to be doing okay with that value, but something must've changed while we were consolidating SDFs that made it so we started saturating the M2 DAC. So we turned both these gains to 0 and will need to figure them out.
I've attached a pdf with the full list of sdf diffs that we accepted. The notable channels listed above are highlighted as * and **