Displaying reports 2041-2060 of 89075.Go to page Start 99 100 101 102 103 104 105 106 107 End
Reports until 11:17, Monday 23 March 2026
H1 SEI (SEI)
anthony.sanchez@LIGO.ORG - posted 11:17, Monday 23 March 2026 - last comment - 16:40, Wednesday 25 March 2026(89599)
Hepi Pump Trends (Monthly)

Hepi Pump Trends (Monthly) Famis 39954:
 

On Jan 26th the HPI-PUMP_EX_PRESS1_PSI started rising in maximum pressure. and on  maybe feb 2nd it started getting worse. Seems like it has plateau-ed on Feb 16th, but the Pressure still remains elevated.

That channel back in Oct was between 92.4-92.6 PSI. 
And now it's between 91.9 - 93.5 PSI. I'm not sure if this is out of bounds for HEPI.

Images attached to this report
Comments related to this report
jim.warner@LIGO.ORG - 16:35, Wednesday 25 March 2026 (89646)

I have looked at the EX pressures a bit and I suspect that one of the pressure sensors on the EX manifold has gone bad.

On the attached trend, the top 2 traces are PRESS1 and PRESS2, bottom 2 traces are the same pressures at EY. EX PRESS1 starts seeing bigger peak to peak signal starting about 50 days ago, but PRESS2 doesn't see this increase in pressure noise. Second attach image is a zoomed in trend of the same channels. EX PRESS1 seems to have some kind of 1hz oscillation. PRESS2 might be seeing some of this same signal, but not to the same extent. The supply pressure to the VEA at EX definitely doesn't see this 1hz signal, so I think it's either not real or it's being filtered out at the manifold.

I have yet to go down to EX and test this, but I will try to do so this week.

Images attached to this comment
jim.warner@LIGO.ORG - 16:40, Wednesday 25 March 2026 (89647)

FRS ticket: https://services1.ligo-la.caltech.edu/FRS/show_bug.cgi?id=37356

H1 AOS
sophie.muusse@LIGO.ORG - posted 10:46, Monday 23 March 2026 (89598)
CHETA Updates

X-Arm electronics:

All chassis have been tested mounted in the rack. All that is left is to replace the variable power supplies currently being used for the visible laser, fan and flipper with 5V, 12V and 15V DIN mounted power supplies respectively. 

X-Arm beam profiling:

X-arm is not behaving as predicted by the model especially in the horizontal axis.
The beam was profiled using the nano scan along a 55cm length after the second lens and a q-parameter was fit using this data. 
This is compared to the model and a Jammt model using the measured QCL q and assuming a linear optical system with the lens at the same distances as in the physical set up.
For the fit q-parameter the w at the ITM was found by propagating the q at L2 to the ITM.

Additionally, during the trouble shooting process a wedge was found on the first telescope lens (L1) which isnt in the optical system and was compensating for some of the astigmatism of the QCL. The effect of this is that in the data below the beam size at the ITM is too large and astigmatic. 
Discussion and solutions of this effect will be the subject of its own log.

Vertical

Measurement/model w0 [ µm] z [mm] w at ITM [mm]
Fit from data

978.9±22.7

 -1063.1±33.8

50.09
Fintrace

857.11

-500.64 61.38
Jammt 874.53 -572 62.1

Horizontal

Measurement/model w0 [ µm] z [mm] w at ITM [mm]
Fit from data 1280.3+/-54.9

 -1813.3±83.5

38.01
Fintrace

979.42

-653.71 54.38
Jammt 1013 -256 53.1

None of the values agree within the measurement uncertainty bounds, especially in the horizontal direction. We are also observing non-gaussian beam profiles on the nano-scan, plots attached, but this is predominately in the vertical direction. This structure isnt new and appeared when matt and I were initially profiling the lens and were able to see good agreement.
The fintrace and Jammt are close but not identical which would be expected as the Jammt does not include any additional optics outside the QCL and L1 and L2. We are currently working on determining the source of this discrepancy. 

 

 

Images attached to this report
LHO VE
david.barker@LIGO.ORG - posted 10:23, Monday 23 March 2026 (89597)
Mon CP1 Fill

Mon Mar 23 10:06:23 2026 INFO: Fill completed in 6min 19secs

 

Images attached to this report
H1 SEI (SEI)
corey.gray@LIGO.ORG - posted 08:27, Monday 23 March 2026 (89592)
H1 ISI CPS Noise Spectra Check - Weekly (FAMIS #39343)

Only CPS channels which look higher by eye at high frequencies (see attached) would be the following:

  1. ITMx_ST1 V1
  2. ITMx_ST2 V1
  3. ETMx_ST1 V2
  4. ETMx_ST1 H2
  5. ETMy_ST1 H2
Non-image files attached to this report
LHO General
corey.gray@LIGO.ORG - posted 07:40, Monday 23 March 2026 - last comment - 09:17, Monday 23 March 2026(89589)
Mon Day Ops Transition

TITLE: 03/23 Day Shift: 1430-2330 UTC (0730-1630 PST), all times posted in UTC
STATE of H1: Planned Engineering
OUTGOING OPERATOR: None
CURRENT ENVIRONMENT:
    SEI_ENV state: CALM
    Wind: 4mph Gusts, 2mph 3min avg
    Primary useism: 0.02 μm/s
    Secondary useism: 0.22 μm/s 
QUICK SUMMARY:

Arm (GV5/7), FC gate valves closed (GV7 is soft-closed, waiting to cam over).  Cleanroom move this morning.  Lots of activity around the Y-Beam Manifold begins!

Comments related to this report
david.barker@LIGO.ORG - 08:10, Monday 23 March 2026 (89591)

As part of our investigation into h1ioplsc0 timing glitches since last week's h1lsc model change, I restarted all the models on h1lsc0 at 08:07 PDT this morning.

david.barker@LIGO.ORG - 09:17, Monday 23 March 2026 (89593)

I've extended the alarm bypass on GV7 for another 24 hours.

Bypass will expire:
Tue Mar 24 09:17:05 AM PDT 2026
For channel(s):
    H0:VAC-LX_GV7_ZSM179A_VALVE_ANIM
 

LHO VE
david.barker@LIGO.ORG - posted 10:30, Sunday 22 March 2026 (89588)
Sun CP1 Fill

Sun Mar 22 10:03:10 2026 INFO: Fill completed in 3min 7secs

CP1 PT114 cold cathode gauge has come back to life, I've taken it out of the alarm bypass and enabled it in VACSTAT.

Images attached to this report
LHO VE
david.barker@LIGO.ORG - posted 11:24, Saturday 21 March 2026 (89587)
Sat CP1 Fill

Sat Mar 21 10:09:41 2026 INFO: Fill completed in 9min 38secs

 

Images attached to this report
LHO VE (VE)
gerardo.moreno@LIGO.ORG - posted 18:33, Friday 20 March 2026 (89586)
HAM1 Large Ion Pump

(Jordan V., Gerardo M.)

Yesterday we tried to hand over the pumping of HAM1 to the ion pump only, it did not go good, the ion pump started having issues maintaining pressure for HAM1 volume, this ion pump showed a very unusual behavior, the pressure was going up and the output of the controller kept going down.  We tried a second controller with similar results.  The controllers just quit and turned the high voltage off themselves, self protection/preservation.  Today, we went up to the ion pump and ohmed out its feedthrough fischer connector, and to our surprised no short was found, we were convinced that the pump had shorted.  So, we took a controller chamber side and connected it with a "short" cable to the ion pump, and the pump turned on without issues.  These findings lead us to believe that the cable may be faulty/shorted/broken/no good.

The temporary setup will be used to maintain the ion pump under vacuum, and the setup will remain until another suitable cable (working one) can be acquired and installed.

Attached are a couple of photos of the temporary setup and trend of the controller/pressure at HAM1 behavior during the last couple of days.

Images attached to this report
LHO VE (VE)
gerardo.moreno@LIGO.ORG - posted 17:53, Friday 20 March 2026 (89585)
TwoLarge Gate Valves and Filter Cavity Isolation Valves are Closed

(Jordan V., Gerardo M.)

We hard closed GV5 with no issues, the gate valve hard closed at 30 psi.  Something to note is the loud clank the gate valve made, and when the valve hard closed it shook the cameras housing, probably the alignment is not what it used to be.

We tried to hard close GV7 but no luck, it soft closed only.  We slowly increased the pressure up to 35 psi, and the gate valve remained unchanged.  For the activities planed for Monday soft close is good, we will visit the GV7 on Monday.

Then, we moved to close the filter cavity tube isolation valves, we closed FCV-1, FCV-2, FCV-3, FCV-4 and FCV-5.  We also isolated the relay tube by closing RV-1 and RV-2, the filter cavity volume will likely go up in pressure.

 

LHO VE
jordan.vanosky@LIGO.ORG - posted 17:06, Friday 20 March 2026 (89584)
YBM Annulus Ion Pump (AIP) Replacement

Today we swapped the 75 l/s ion pump used for the annulus volume on the Y-Beam Manifold (YBM) . Annulus volume was isolated and the pump swapped, no issues encountered during the swap.

An aux cart was used to pump out the ion pump body before switching the IP on. After a couple hours of pumping with both the aux cart and the IP, the aux cart was isolated and the IP was able to maintain the pressure.

Closing WP 13107

H1 ISC (SUS)
elenna.capote@LIGO.ORG - posted 16:47, Friday 20 March 2026 - last comment - 13:48, Monday 23 March 2026(89583)
OMC build update for BHSS

This is a belated report of work done by me, Oli and Keita on the OMC builds for the BHSS, done yesterday March 19.

Exciting summary: build for OMCA is complete, including DCPDs and butter dish. OMCA is installed onto the BHSS platform and ready for beam alignment!

Less exciting summary: build for OMCB is halted as tombstone for DPCD A (transmission) is missing a required PCB hole for the electronic connection.

OMCA build notes:

OMCB build notes:

Images attached to this report
Comments related to this report
arnaud.pele@LIGO.ORG - 11:25, Monday 23 March 2026 (89600)

[Elenna, Camille, Arnaud]

We had a quick zoom meeting to discuss this issue. Relevant drawing is https://dcc.ligo.org/D1201273

It is likely that the backshell+PCB assembly was assembled flipped 180 degrees (items 4,5,6 on the drawing). To fix, remove the two 2-56 screws (item 14), flip the assembly and reattach it to the housing. Camille suggest slightly pressing down on the housing while removing those screws to avoid torquing the base with the bond.

corey.gray@LIGO.ORG - 11:30, Monday 23 March 2026 (89601)EPO

Tagging for EPO.

elenna.capote@LIGO.ORG - 13:48, Monday 23 March 2026 (89606)

After consultation with Arnaud and Camille, Oli and I were able to correct the problem with the PCB assembly.

On the call with Camille and Arnaud, we referred to the diagram in D1201273 and determined that part #4 is rotated 180. I could also see from this diagram that part #5 is rotated, since there was no notch visible at the top. So, for future installation, make sure the notch in part #5 is oriented upwards! This indicated to us that all that needed to happen was to unscrew the #14 screws, flip the assembly, and screw it back in (we had some concern that we might have to take apart the "sandwich" of parts between #5 and #4).

Indeed, that's all that needed to happen. I carefully unscrewed the screws on the back, flipped the assembly, and screwed it back in. I was then able to install the DCPD. We can now see the notch along the top!

Images attached to this comment
elenna.capote@LIGO.ORG - 09:56, Monday 23 March 2026 (89595)

I went back into the lab today to try to get a better photo of the problem. I wasn't able to, but I did get an ok photo of the other tombstone (on reflection of the BS) which has the correct three holes. It's hard to see, but you can use it as a reference to compare with the photo I have attached to the original post which is missing a hole. I tried looking from many angles, and I don't see even a hint of the third hole, which is evidence in support of Keita's theory that the electronics assembly is flipped 180 degrees.

Images attached to this comment
LHO General
corey.gray@LIGO.ORG - posted 16:39, Friday 20 March 2026 (89575)
Fri Ops DAY Shift Summary

TITLE: 03/20 Day Shift: 1430-2330 UTC (0730-1630 PST), all times posted in UTC
STATE of H1: Planned Engineering
INCOMING OPERATOR: n/a
SHIFT SUMMARY:

No Locking Today.  There was some Squeezer alignment checks by sheila & tony.  Gerardo & Jordan CLOSED GV5 & 7 as well as Filter Cavity mini-gate valves.
LOG:

H1 CDS
david.barker@LIGO.ORG - posted 15:54, Friday 20 March 2026 (89582)
Alarms reconfigured for GV5 and GV7 to be hard-closed.

Also, CP1 PT114 tripped during the close of GV5, it has been disabled in VACSTAT and bypassed in ALARMS.

H1 CDS
erik.vonreis@LIGO.ORG - posted 12:43, Friday 20 March 2026 (89581)
CRC errors on h1seib3 and h1seib1

At 10:19:54 PDT both h1seib3 and h1seib1 had CRC errors on DAC leg zero.  These errors probably represent missing data blocks and not data corruption.

LHO VE
david.barker@LIGO.ORG - posted 11:03, Friday 20 March 2026 (89580)
Fri CP1 Fill

Fri Mar 20 10:19:50 2026 INFO: Fill completed in 19min 46secs

 

Images attached to this report
Displaying reports 2041-2060 of 89075.Go to page Start 99 100 101 102 103 104 105 106 107 End