When the gain of the ISS second loop was changed, the input offset was not adjusted. This results in the output of the 2nd loop ISS to saturate after a lock loss, until the offset slowly gets compensated by the offset loop. But, this now typically takes like 5 minutes. I adjusted the offset (H1:PSL-ISS_THIRDLOOP_OUTPUT_OFFSET) from 587.3 to 1272. We will have to check after the next lock loss that this actually works.
Just had a lock loss. Looks a lot better now.
I disabled the diffraction servo by setting H1:PSL-ISS_SECONDLOOP_REFERENCE_IN_MTRX_1_4 to 0 (from 10). This should stabilize the power after the IMC, rather than keep the pwer at the PSL AOM constant. The downside is that we may run out of range of the AOM, since we only work at 2% diffraction power.
Tagging PSL.
The diffraction servo value that Daniel has made 0 from 10 had been hard-coded into the DOWN state of the IMC_LOCK guardian, so was getting overwritten.
RyanS and I just set the value in the guardian to zero, loaded, and checked it in to the svn.
First hour trend during a lock with the diffraction compensation off: The powers are no longer drifting a significant amount after the mode cleaner.
TITLE: 04/26 Day Shift: 15:00-23:00 UTC (08:00-16:00 PST), all times posted in UTC
STATE of H1: Calibration
SHIFT SUMMARY:
LOG:
!LVEA STILL LASER HAZARD!
IFO status: Currently still locked and in NLN_CAL_MEAS to allow a Calibration sweep.
A massive reduction in SDF Diffs occured today as you can see there are much fewer diffs then there was from my last alog, which is great!!! we are many steps closer to using the OBSERVE button!
For the time being the Operator core will use UNDISTURBED bit with the OBSERVING Observatory Mode selected until we can get the proper intention bit working by reducing the SDF DIffs to zero and getting all Gaurdian node OKed.
Lockloss notes:
Lockloss info from before my shift started can be found here During the shift there weren't any locklosses.
| Start Time | System | Name | Location | Lazer_Haz | Task | Time End |
|---|---|---|---|---|---|---|
| 19:56 | LVEA | LAZSER HAZARD | LVEA | YES | THE LVEA IS LASER HAZARD | 10:57 |
| 06:52 | IFO | UNDISTURBED | UNDISTURBED | 15:26 | ||
| 15:47 | FAC | Kim | EE Lab | N | TEchnical Cleaning | 16:08 |
| 15:49 | SQZ | Camilla | Ctrl Rm | N | SQZ Angle Adjustment | 17:49 |
| 16:14 | FAC | APS | FCES | N | Contractors installing Locks & security | 22:14 |
| 16:49 | FAC | Cindy | High bay & Mech Rm | N | Vacuuming in high bay and mech room | 17:54 |
| 16:53 | SEI | Jim Mitchell | Wind-Y | N | Wind Fence repair | 19:14 |
| 17:02 | VAC | Travis & Gerardo | EY & MX | N | Putting vacuum parts in receiving areas | 17:52 |
| 17:54 | SEI | Tyler | Wind-Y | N | Wind fence repair | 18:21 |
| 17:55 | FAC | Cindy | H2 | N | Technical cleaning | 18:23 |
| 18:15 | SQZ | Camilla | CTRL Rm | N | SQZ Angle Adjustment | 18:26 |
| 18:55 | FAC | Richard | CER | N | solving other peoples noise problems | 19:15 |
| 19:02 | Commish | Gabriele | Ctrl Rm | N | Noise injection! | 19:07 |
| 19:13 | CAL | Jeff & Crew | Ctrl Rm | N | Calibration meas, PCAL Broadband | 20:12 |
| 19:50 | EE | Fil | EY | N | PEM Mag Injection cable term | 21:23 |
| 20:17 | Commish | Elena | Ctrl Rm | N | Increased CARM gain by 1db | 20:19 |
| 20:19 | SQZ | Nutsinee | Ctrl Rm | N | Reduction of Squeezing | 22:42 |
| 21:31 | SEI | Jim & Mitchel | Wind-Y | N | Wind fence repair at EY | 21:31 |
| 22:25 | Cardio | Gabriele | EX & EY | N | Running the Arms ETA: 1.5H | 00:25 |
| 22:42 | CAL | Tony & Jeff & Evan | Ctrl Rm | n | Calibration Sweep | 00:27 |
We've been seeing the sqz angle needs adjusting between the start of the lock and a few hours in 68853.
This morning I tried to optimize the SQZ angle every ~10minutes after the start of the lock (IFO in NLN at 15:47 UTC). I aimed to reduce the 3,4,5,6 BLRMS, ADF RMS at 1300Hz and the DARM plot. See Operator's SQZ Troubleshooting Wiki for some SQZ angle tuning instructions. Results attached. The angle changes around 20 degrees in the first 1h30 of the lock. We could think about adding a SQZ angle adjuster to the THERMALIZING guardian or something similar alog#68948. However the optimum angle is tricky to choose as there's a fairly large (+/- 10deg) range that gives very similar squeezing, see attached plot taken at 17:30 - 17:40 UTC (1h40 after NLN).
We could try to retake Sheila's pre-O3 squeezing verus angle alog#55760 plot overnight in the next weeks to understand this better.
Daniel suggests checking the stability of squeezing using the ADF calculated SQZ angle (H1:SQZ-ADF_OMC_TRANS_SQZ_ANG). See attached for how this increases 15deg in the first 1 hour of a lock. Nominally the ADF line is off and we don't yet have data for a clean long lock where the SQZ angle wasn't adjusted.
J. Kissel, E. Goetz, T. Sanchez
While teaching Tony how to use the command line infrastructure to gather a full set of calibration sweeps, we noticed that the default list of measurements was missing a companion PCAL measurement to the last TST stage actuation strength measurement.
Before we changed anything we ran the check of "what will the command do if I just run it," and got
$ pydarm measure
available measurements:
pcal: PCal response, swept-sine (/ligo/groups/cal/ifo/H1/templates/PCALY2DARM_SS__template_.xml)
bb : PCal response, broad-band (/ligo/groups/cal/ifo/H1/templates/PCALY2DARM_BB__template_.xml)
sens: sensing function (/ligo/groups/cal/ifo/H1/templates/DARMOLG_SS__template_.xml)
act1x: actuation X L1 (UIM) stage response (/ligo/groups/cal/ifo/H1/templates/SUSETMX_L1_SS__template_.xml)
act2x: actuation X L2 (PUM) stage response (/ligo/groups/cal/ifo/H1/templates/SUSETMX_L2_SS__template_.xml)
act3x: actuation X L3 (TST) stage response (/ligo/groups/cal/ifo/H1/templates/SUSETMX_L3_SS__template_.xml)
measurement sequence:
['bb', 'sens', 'pcal', 'act1x', 'pcal', 'act2x', 'pcal', 'act3x', 'bb']
Use --run-* option to actually execute measurement
You'll see that the list is missing a companion 'pcal' to 'act3x'.
So, we edited the default list by changing
/ligo/groups/cal/ifo/H1/
pydarm_cmd_H1.yaml
Specifically, the lines under the parameter "measurement_sequence" such that it now reads
measurement_sequence:
- bb
- sens
- pcal
- act1x
- pcal
- act2x
- pcal
- act3x
- pcal
- bb
Now, the dry-run of pydarm measure reads
$ pydarm measure
available measurements:
pcal: PCal response, swept-sine (/ligo/groups/cal/ifo/H1/templates/PCALY2DARM_SS__template_.xml)
bb : PCal response, broad-band (/ligo/groups/cal/ifo/H1/templates/PCALY2DARM_BB__template_.xml)
sens: sensing function (/ligo/groups/cal/ifo/H1/templates/DARMOLG_SS__template_.xml)
act1x: actuation X L1 (UIM) stage response (/ligo/groups/cal/ifo/H1/templates/SUSETMX_L1_SS__template_.xml)
act2x: actuation X L2 (PUM) stage response (/ligo/groups/cal/ifo/H1/templates/SUSETMX_L2_SS__template_.xml)
act3x: actuation X L3 (TST) stage response (/ligo/groups/cal/ifo/H1/templates/SUSETMX_L3_SS__template_.xml)
measurement sequence:
['bb', 'sens', 'pcal', 'act1x', 'pcal', 'act2x', 'pcal', 'act3x', 'pcal', 'bb']
Use --run-* option to actually execute measurement
I've committed this new version of the yaml to the Calibration/ifo/H1/ git repo, as of 092dc147.
it wasn't missing a pcal for act3x. It would have just used the pcal measurement taken right before act3x during processing. This cuts down the overall time needed for the calibration suite. the pcal measurements are sandwiched between the actuation and OLG ('sens') measurements for this purpose.
TITLE: 04/26 Eve Shift: 23:00-07:00 UTC (16:00-00:00 PST), all times posted in UTC
STATE of H1: Calibration
CURRENT ENVIRONMENT:
SEI_ENV state: CALM
Wind: 19mph Gusts, 12mph 5min avg
Primary useism: 0.05 μm/s
Secondary useism: 0.09 μm/s
QUICK SUMMARY:
- IFO has been locked at the start of my shift for ~7 hours
- Alert handler/CDS/DMs ok, wind looks to be picking up a bit, but nothing too bad so far
- Currently, there are calibration sweeps going on
We have started the campaign to set the Settings Definition Files (SDF) for the OBSERVE state of H1. LSC and ASC had many changes which I have looked through and then Accepted in SDF for the current Lock stretch (after some consult with TJ and Elenna). (Hitting Accept in SDF sets the current Epics channel value to be the new Set point value for future reference.) None of the LSC channels listed for the current lock are ones that are adjusted during lock stretch by Guardian. Some of these channels may change between lock stretches so will watch for these at next Observing stretches to see what is dynamic and then decide to unmonitor, etc.
In the very unlikely chance that someone needs to reset one of the settings from today based on this interface, attached are the snapshots of what was loaded from some older setpoint.
BOS had over 1800 channels so no snapshot taken (which woulnd't be useful anyways).
LSD SDF old data/pictures prior to Accepting new Set points just now.
ASC SDF old data/pictures prior to Accepting new Set points just now.
ISCEY, ISCEX, EX_ISC, and EY_ISC SDF files prior to setting new setpoint for the current lock stretch just now. Mostly dark offsets and a new filter commissioned for the first 2. Some settings which we'll watch for changes on in the second 2 snapshots.
J. Betzwieser, L. Dartez, E. Goetz, J. Kissel, J. Rollins
The details are happening all too fast in order to write them all down here but at the 10000 foot level, as of 2024-04-26 19:54 UTC, we've updated
- the primary CAL-CS variables that create CAL-DELTAL_EXTERNAL_DQ, which correspond to the interferometrically-measured free parameters (optical gain, cavity pole, actuation gains for TST, PUM, and UIM changes)
:: Values are informed by last Thursday's 2023-04-20 IFO measurement. No major change here, just pushing these to get them consistent with the rest of the pipeline. Not informed yet by the current state of the IFO, especially the TCS settings and SRCL offsets
- The "DARM model transfer function function values at calibration line frequencies" EPICs records
:: Importantly, this comes with changes that are the fall out of LHO:68880 and the solution pyDARM Merge Request 234
- The GDS-CALIB_STRAIN pipeline has new filters, and importantly its got a configuration change that now using the corner station copy of the PCAL, H1:CAL-DELTAL_REF_PCAL_DQ, as it's measure of the PCAL system rather than the end-station.
:: Again, this is the fall out of LHO:68880
Many details to come in the comments.
Here's a blow-by-blow of what happened:
- Had been locked and powered up to 76W PSL input power for ~3:30 hours
- Louis walked us through the calibration report for what would be pushed upon update from
/ligo/groups/cal/data/H1/reports/20230420T014029Z/H1_calibration_report_20230420T014029Z.pdf
- Took ISC_LOCK to NLN_CAL_MEAS
- Took a "before change" PCAL broadband injection,
/ligo/svncommon/CalSVN/aligocalibration/trunk/Runs/O3/H1/Measurements/FullIFOSensingTFs/
2023-04-26_1925UTC_H1_PCALY2DARMTF_BB_3min.xml
- While the broadband was going, took inventory of what filter modules were available for us to place the new IFO-measured free parameters in the
H1:CAL-CS_DARM_ERR
H1:CAL-CS_DARM_ANALOG_ETMX_L1
H1:CAL-CS_DARM_ANALOG_ETMX_L2
H1:CAL-CS_DARM_ANALOG_ETMX_L3
filter banks. Found that FM1 and FM2 were available in DARM_ERR, and FM4s were available in the L1, L2, and L3 banks.
- Edited the pydarm command line interface configuration / parameter file,
/ligo/groups/cal/ifo/H1/pydarm_cmd_H1.yaml
to chose these filter banks, and a new name for the filter. For the sensing function we chose "ER15_Gain" and "ER15_NoD2N" for the optical gain and cavity pole frequency in FM1 and FM2 of DARM_ERR. For the actuation strengths for the L1, L2, and L3 stages, we chose "Npct_ER15" in FM4.
- Ran the command line interface for exporting the calibration model to the front end (currently using a curated path and pydarm branch)
:: First, without actually pushing, aka a dry run, which spits out everything it's *going* to push as it stands. The first couple of lines of that output are the IFO-measured free parameters, copied here for future searchability
Hc: 3.1796e+06 :: 1/Hc: 3.1450e-07
Fcc: 4.3371e+02 Hz
Hau: 1.6072e+00 N/A :: 7.5454e-08 N/ct
Hap: 3.0579e-02 N/A :: 6.2634e-10 N/ct
Hat: 2.5464e-11 N/A :: 9.6474e-13 N/ct
These are not that different from what was installed on 2023-04-20 01:15 UTC, see LHO:68856, so we don't expect that much change here.
:: Then, we actually pushed those values,
PYTHONPATH=/ligo/home/louis.dartez/repos/pydarm pydarm export --push
which spit out to the commanding what it was changing, when it was changing it (but lacked any "done!" message)
- Went CAL-CS model's GDS_TP screen, waited ~5 seconds until the "Modified Filter File Detected" message to appear, then hit "LOAD COEFFICIENTS" (i.e. pressed the H1:FEC-117_LOAD_NEW_COEFF button), and waited another ~10 seconds for it to complete
- Went to each of the above mentioned filter banks and switched over to each of the new filters by hand; FM1 & FM2 in the DARM_ERR, and FM4s in ANALOG_ETMX_L1, ANALOG_ETMX_L2, and ANALOG_ETMX_L3.
- Took another post-calibration update broad-band injection in order to see how much better / different the calibration is/was,
/ligo/svncommon/CalSVN/aligocalibration/trunk/Runs/O3/H1/Measurements/FullIFOSensingTFs/
2023-04-26_1950UTC_H1_PCALY2DARMTF_BB_3min.xml
- Louis, on the side, created / updated new PCAL and DELTAL calibrations for the DTT template, and put them in
/ligo/home/louis.dartez/projects/20230426/
pcal_calib_dtt.txt
deltal_external_calib_dtt.txt
so, we loaded these in to the 1950UTC DTT template. Hopefully this gives us a more accurate answer for what the systematic error is without all the extra caveats we've had in recent history about "well, the DTT calibration is some old crap, that has an outdated version of the super-Nyquist corrections for DELTAL, so you can't trust the answer *exactly*." While we're not yet confident that you can trust the answer "exactly," we are confident it's at least better. To load in these files,
:: Hit "Calibration" button to open calibration pop-up window
:: Select H1:CAL-PCALY_RX_PD_OUT_DQ
:: Hit "New" to open the labeling pop-up.
. Enter in a name for the Teference (can be anything, "asdfsdf" for eg)
. Enter in a Unit of "m"
. Enter in a Time that's much earlier that now (hold down CTRL and click the down arrow once to change the year to 2022)
. Hit OK to return to the calibration pop-up
:: Unselect the check box in every other tab besides "Trans. Func."
:: Under the "Trans. Func." tab,
. Check the box
. Hit Edit... to open up the transfer function editing window
. From the toolbar, select File > Open...
. Navigate to pcal_calib_dtt.txt, and double click it to select, close the pop-up, and have it populate in the TF editing window.
. Hit "OK" to accept the new TF and close the TF editing window
:: Hit "Set" to make sure this new calibration sticks.
:: Repeat the process for H1:CAL-DELTAL_EXTERNAL_DQ using deltal_external_calib_dtt.txt
See 2023-04-26_1950UTC_CalibrationUpdate_BBInjection.png which shows the change between the BB taken after the last calibration update on 2023-04-23, just before this update on 2023-04-26 19:25 UTC, and just after at 2023-04-23 19:50 UTC. This is after we've updated the calibration within the DTT template, so for the first time in a long time, the low-frequency end of phase of this transfer function makes sense "starting" at 0 deg.
- We then went back to nominal low noise, turning the calibration lines back on so as to start measuring the TDCFs again. Even though we've moved around some delays and got rid of some DEMOD phase rotations, the TDCF answers remain about the same before and after the calibration change, as expected. This gives us confidence in what changes we made to pydarm as a result of last week's solutions to long standing mysteries about the EPICs records.
Satisfied with this "first brush" answer,
- Went to the SDF screen for the CALCS model, and accepted all the new settings (including all the changes to the "DARM model transfer function values at calibration line frequencies") -- see 2023-04-26_1950UTC_CalibrationUpdate_SDFAccept.png.
- Jamie then used the FIR filters generated in that 2023-04-20 report,
/ligo/groups/cal/data/H1/reports/20230420T014029Z/
gstlal_compute_strain_C00_filters_H1.npz
to inform and restart the GST-LAL calibration pipeline, in some way that hopefully he'll document in further comments below.
In order to update the pydarm_H1.ini parameter file value for the CALCS foton export of the sensing function, i.e. the "foton_invsensing_tf" parameter in the "[calcs]" section, I've followed the instructions in LHO:47948 to create /ligo/svncommon/CalSVN/aligocalibration/trunk/Runs/O3/H1/Measurements/Foton/ 2023-04-20_H1CALCS_InverseSensingFunction_Foton_ER15_NoD2N_tf.txt for future use.
Wed Apr 26 13:15:37 2023 INFO: Fill completed in 37secs
Super quick fill, but plot looks good.
TITLE: 04/26 Day Shift: 15:00-23:00 UTC (08:00-16:00 PST), all times posted in UTC
STATE of H1: Calibration
CURRENT ENVIRONMENT:
SEI_ENV state: CALM
Wind: 10mph Gusts, 8mph 5min avg
Primary useism: 0.02 μm/s
Secondary useism: 0.10 μm/s
QUICK SUMMARY:
!LVEA IS STILL LASER HAZARD!
Locking was straight forward. Though we did go through PRMI_ACS .
Wind-Y team has been making good progress on the wind fence today without disturbing the IFO.
To ensure that the Y End Wind Fence work doesn't break our lock Jim had me turn off sensor correction for H1:ISI-ETMY.
H1:ISI-ETMY_ST1_SENSCOR_Y_NORM_FILT4_SW1R has been engaged to minimize the amount of "Bad_stuff" that gets sent to the ISI signal.
Please see screenshot attached below for future references.
APS is here again to day to work on the door locks for Filter Cavity End Station.
IFO has been locked for over 3.5 hours in NOMINAL_LOW_NOISE to allow for commissioning.
Some SQZ Angle adjustments have been made by Camilla.
Richard went into the CER and started poking things randomly and a source of noise was discovered to be the PMT Power Supplies which were then were turned off.
The Calibration group is now taking a block of IFO time for Calibration measurements.
It's been fairly quite. very little wind, seismic motion, and the Violins look great!
Things to look forward to:
There is a need to move BIG RED the fork lift, from it's positon near the water tank to the staging building sometime in the next 24 hours. So if we lost lock there is an oppertunity to move BIG RED, so contacting Tyler after the next lock loss is a must.
We'd like to try and get the IFO into the OBSERVE State. So please try to resolve any SDF Differences that your sub systems currently have so we can try getting the IFO into the OBSERVE state. Please see Screenshot below of the SDF screen to determine if you can help resolve some SDF Diffs. Thanks! :)
Adrian, Camilla, Gabriele, Fil, Richard
While trying to track down the source of the 4 Hz bumps in DARM, Richard noticed the cosmic ray photomultiplier tube (PMT) power supply voltage display numbers were jumping around. When he reset the dial, we noticed the 4 Hz oscillations in H1:PEM-ADC_5_29_2K_OUT_DQ disappeared for a few minutes. We unplugged the power supply to the PMTs at 19:05:00 UTC and have not seen 4 Hz features in the ADC channel or DARM since. Plots of the change in ADC channel timeseries behavior when Richard first touched the power supply knobs as well as the ADC channel behavior after the power supply was unplugged are attached. A spectrum comparing the ADC power supply plugged in (blue) and unplugged (red) is attached. We had about 40 minutes to watch DARM before some commissioning tasks started and we didn't see any more bumps. Another plot is forthcoming for this...
(edit for spelling...)
We need to collect more data to be sure, but for now it looks like the 4.05 Hz bumps in DARM are gone
This is awesome and fingers crossed that it fixes things.
Just a quick note that as we try to figure out how this coupled, it seems that H1:PEM-CS_ADC_5_29_2K_OUT_DQ might not be our best witness channel because while today it saw the 1.35Hz and 4.05Hz before the CR PS unplug, it did not see those peaks during some past times when n*4.05Hz bumps were seen in h(t) such as 4/12, 4/14, and 4/16.
Some of the other PEM channels such as H1:PEM-CS_ACC_LVEAFLOOR_HAM6_Z_DQ do seem to be reliable witnesses, having seen the 1.35Hz and 4.05Hz during h(t) bump times on those previous days as well as today before CR PS unplug, but not after unplug.
There was another instance of the 4 Hz bumps at 20:50 UTC. The attached plot shows OMC-DCPD_SUM, with blue as a reference, green dotted an example of the glitch a few days ago, and orange the one now. It's complicated a little by the PCal line at 24 Hz; it could be notched if we wanted a better plot. The bump at 20 Hz does look weaker but the rest are similar to before.
Most likely one of the photomultiplier tubes has developed a minor short and is going through a charge/discharge cycle. Not surprising after ~20 years! It is probably worth testing if the discharge is in the PMTs or the HV power supply by disconnecting the HV cables from the power supply and powering it back up. Either the fluctuations in the display will stay or not.
It looks like there are fewer of those 4.05Hz bump / glitches, but the problem is still with us, some of them quite loud.
Power was restored to the PMTs around 11:00:00 local on May 16, 2023, as we believe the 4 Hz glitching is coming from the the cryyobaffle. All four displays worked when I turned the power supply back on again.
Today we made more tests of the effect of differential CO2 annular heating and the effect of thermalization on various parameters. Specifically, we measured the CARM open loop gain, PRCL open loop gain, and DARM spring at various times during power up. In parallel, the squeezing team also attempted to tune sqz during different points of thermalization. We also intended to measure frequency noise and contrast defect, but were unable to due to locklosses. See alog 68971 for details on last night's CO2 tests and the preliminary results of contrast defect and frequency noise.
The TCS setting that we tested was a differential annular CO2 step. CO2X was increased to 4W and CO2Y was left at 1.7W. The annular heating engages at the Power 25W state. The preliminary results showed marked improvement in frequency noise. The ring heaters were set to the nominal 76W settings: ETMX 1.3W, ETMY 1.2W.
CARM OLG
We measured the CARM olg during the first three hours of IFO thermalization. The results are plotted here. The show that the CARM gain drops in similar fashion to PRCL. However, we are not in danger of instability, and the gain levels off into the second hour of lock. We can increase the CARM gain as we thermalize, and increasing it will likely improve our frequency noise above 4 kHz (see 68967). Gabriele is working on implementation. He will add a comment describing the implementation.
PRCL UGF
We returned to changing the PRCL gain by hand. The PRCL gain does drop steadily as we thermalize, but at a different rate than measured by Gabriele. The UGF servo will need to be recommissioned for a few locks to collect data and then we can fit the thermalization trend.
DARM Spring
I made some measurements of the DARM spring through thermalization and also with different SRCL offsets. See the plot here. I measured both the DARM OLGTF and a PCAL to DARM transfer function to create this plot. I think the trend shows that we need less SRCL offset at the start of the lock, and then thermalization increases the SRCL detuning. For example, within the first hour, a SRCL offset of -80 was sufficient. By the third hour, we were moving towards needing something like the usual SRCL offset of -165. Jeff has commissioned low frequency line injections to track this better than I can with these measurements. I look forward to further results on this front.
Other effects
We were not able to make any well-defined measurements of these quantities, but two things were evident by eye during this lock: The frequency noise was improved, and the jitter noise was worse.
Lockloss
At the end of the third hour, we experienced an ASC lockloss. By eye, it seems to be a CSOFT P ring up, but given the amount of cross coupling it is hard to be sure. During a subsequent lock, I made some sensing matrix measurements of the ASC REFL signals at high power. There are some changes in the PRC2 P sensing that could be causing problems. Or, the problem could be something else entirely.
Unfortunately we don't have enough time right now to determine what exactly is the ASC problem and how fix it. Gabriele and I have chosen to revert the CO2X annular change to the previous setting of 1.7W (common with CO2Y). We know this setting is stable. We have commissioned the PRCL gain, various feedforward and more at this particular setting. We would also like to deliver a stable interferometer to the start of ER15 in the morning. We don't think we should give up on this setting though. We would like to use commissioning time in the next week to debug the ASC problem and regain stability with the new setting. We think this is a good change for the frequency noise. We also think that the increased jitter noise present is a clue that there are perhaps ASC offsets that could be causing the instability. We are making the CARM gain change though, as that is something that will occur independent of the CO2 change, and it will improve our frequency noise above 4 kHz.
We added some logic to the THERMALIZATION guardian to increase the CARM gain.
The gain sliders can be changed only in steps of 1 db, so we are implementing the discrete changes in the table below to follow the decrease in CARM UGF we measured over one lock
| Time from LOWNOISE_LENGTH_CONTROL [min] |
Additional REFL SERVO gain [db] |
|---|---|
| 10 | 1 |
| 25 | 2 |
| 45 | 3 |
| 80 | 4 |
| 160 | 5 |
We are increasing both LSC-REFL_SERVO_IN1GAIN and LSC-REFL_SERVO_IN2GAIN at the same time, starting from the gain they had when the THERMALIZATION guardian is asked to go to THERMALIZED
There was a problem with the new REFL servo gain adjustment. The THERMALIZATION guardian red the initial CARM gains at LOWNOISE_LENGTH_CONTROL, but at that state CARM was using only one REFL signal. So during the last lock the THERMALIZATION guardian was also keeping one of the REFL diode gains low.
I fixed the problem by writing explicitly the initial gain of the two REFL signals (6db and 6db) since those numbers are hard-coded in the LASER_NOISE_SUPPRESSION state.
I restarted the THERMALIZATION guardian and set back the CARM gain slider to 6 and 6 db
This gain change is a discrete step in an analog circuit when an output from a lower gain OpAmp is switched to a higher one. This is implemented as a binary ladder with 1dB step size, so depending on the actual gain values some elements may go down in gain while others go up. This will introduce a transient which may or may not be visible at the anti-symmetric port. The transisiton time is sub micro seconds.
Bye eye, we could see that the jitter in DARM increased with the new CO2 setting. I ran a coherence comparison of DARM with the IMC WFS DC channels during the time we were at 4W CO2X versus when we are at 1.7W CO2X. All other TCS settings are the same. Above 100 Hz, the coherence of DARM with the IMC WFS increases almost everywhere when we are on higher CO2X annular power.
The one place where jitter is worse at CO2X = 1.7W is around 50 Hz. DARM is has high coherence with PRCL there. We think that the PRCL gain was mistuned, and that PRCL also sees input jitter. We think we can mitigate that region with better tuning of the PRCL gain.
However, the increased jitter is one argument against moving to differential CO2 annular power.
Note: we are not applying any jitter subtraction at either of these times.
LIGOCAM shows H1:PEM-EX_MAG_VEA_FLLOR_X/Y/Z_DQ as disconnected. I confirmed that (i) the magnetometer was still in place (ii) the power supply was on and (iii) the magnetometer box was running. I couldn't follow the triaxial signal cables from the box to the PEM racks - the cables are bound up with a whole bundle under the beamtube near the BRS. I still don't know why they're marked as failing.
I took a look at the ASD of EX Floor magnetometer and compared it to a couple magnetometers that are marked as working. The ASD does look flatter and lower magnitude than working magnetometers.
An easy way I used to check magnetometers was to wave a fridge magnet near the magnetometer while monitoring the current power spectrum on a laptop.