Displaying reports 4281-4300 of 89074.Go to page Start 211 212 213 214 215 216 217 218 219 End
Reports until 09:52, Friday 03 October 2025
H1 CDS
david.barker@LIGO.ORG - posted 09:52, Friday 03 October 2025 - last comment - 10:46, Friday 03 October 2025(87277)
Seismic activity bar added to top of CDS Overview MEDM

I have added a seismic activity bar, similar to the wind speed bar, to the upper right of the CDS Overview. Attached shows the bar during the current earthquake.

Similar to the wind speed, the bar fills from the left, start green then gets increasingly red.

This bar is logarithmic, each point is approximately a factor of about 2.5 from the last. The range is 1.0e+01 to 1.0e+05 BLRMS.

The bar is comprised of circles ("little earths") rather than squares.

Images attached to this report
Comments related to this report
david.barker@LIGO.ORG - 10:46, Friday 03 October 2025 (87281)

DTS Environment Repositioned On CDS Overview.

To make room for the Seismic bar, I removed the DTS environment section. This has now been added back as a "system widget" in the FMCS section (see attached).

The "DTS ENV" block will turn RED if the server room temperature exceeds 80F. The block will turn MAGENTA if the dts_env_ioc GPS time falls behind by 10 minutes or more.

Clicking on the "DTS ENV" block will open the DTS Detail MEDM (attached).

Images attached to this comment
H1 General (Lockloss, SEI)
ryan.short@LIGO.ORG - posted 09:50, Friday 03 October 2025 - last comment - 13:37, Friday 03 October 2025(87278)
Lockloss @ 16:44 UTC - EQ

Lockloss @ 16:44 UTC after 12.5 hours locked - link to lockloss tool

M6.1 EQ out of Kamchatka followed by a M5.1 aftershock. I dropped observing at 16:26 UTC as it was coming in to engage the hi-gain ASC, which seemed to help for a while until the aftershock hit and peakmon reached about 2300 counts. Holding H1 in 'DOWN' until the ground settles.

Comments related to this report
ryan.short@LIGO.ORG - 13:37, Friday 03 October 2025 (87282)

H1 back to observing at 20:34 UTC. Ran an alignment after the ground motion calmed down, then while locking, two locklosses during DRMI acquisition made for a longer overall relock time. MICH_FRINGES was needed after the second DRMI lockloss.

H1 SUS (SUS)
corey.gray@LIGO.ORG - posted 08:35, Friday 03 October 2025 (87272)
Weekly In-Lock SUS Charge Measurement - FAMIS 28425

Had issues with FAMIS, so was not able to check this until today.  Looks like the SUS Charge measurements ran this past Tuesday morning. 

(A procedure note:  when I ran the printing command, it would generate the ETMx plot, but that was it.....until I saved the plot to the desktop and then closed the generated .png plot.  Then it would continue with the other plots (one at a time)---took me a bit to figure that out!)

Attached are the plots for all four Test Masses.  Closing FAMIS 28425.

Images attached to this report
LHO General
ryan.short@LIGO.ORG - posted 07:34, Friday 03 October 2025 - last comment - 09:18, Friday 03 October 2025(87271)
Ops Day Shift Start

TITLE: 10/03 Day Shift: 1430-2330 UTC (0730-1630 PST), all times posted in UTC
STATE of H1: Observing at 156Mpc
OUTGOING OPERATOR: Corey
CURRENT ENVIRONMENT:
    SEI_ENV state: CALM
    Wind: 3mph Gusts, 2mph 3min avg
    Primary useism: 0.02 μm/s
    Secondary useism: 0.21 μm/s 
QUICK SUMMARY: H1 has been locked for over 10 hours. S251003at earlier this morning at 13:23 UTC.

Comments related to this report
ryan.short@LIGO.ORG - 09:18, Friday 03 October 2025 (87273)

S251003at from this morning was retracted.

Also ran range comparison checks comparing this lock stretch to the last as our range is a bit better. Seems like DARM looks better overall above ~60Hz; Sheila says this is probably just because the squeezer is "better" today. Report attached.

Non-image files attached to this comment
H1 General
oli.patane@LIGO.ORG - posted 22:00, Thursday 02 October 2025 (87270)
Ops EVE Shift End

TITLE: 10/03 Eve Shift: 2330-0500 UTC (1630-2200 PST), all times posted in UTC
STATE of H1: Observing at 155Mpc
INCOMING OPERATOR: Corey
SHIFT SUMMARY: Observing at 155 Mpc and have been locked for over 30 minutes. We recovered from the lockloss really quickly (59 minutes!) and I didn't need to intervene.
LOG:

23:30 UTC Observing and locked for almost 4 hours
01:59 GRB-Short E606552
03:22 Lockloss
04:21 NOMINAL_LOW_NOISE
    04:24 Observing

H1 General (Lockloss)
oli.patane@LIGO.ORG - posted 20:47, Thursday 02 October 2025 - last comment - 21:24, Thursday 02 October 2025(87268)
Lockloss

Lockloss at 2025-10-03 03:22 UTC after just over 7.5 hours locked. It looks like it was caused by a very sudden localish event between 0.3 - 3 Hz (0.3-3 Hz, 1-3 Hz CS).

picket fence, build-ups, LSC, QUADs, peakmonCS movement vs LL

Images attached to this report
Comments related to this report
oli.patane@LIGO.ORG - 21:24, Thursday 02 October 2025 (87269)

10/03/2025 04:24 UTC Back to Observing

H1 General
oli.patane@LIGO.ORG - posted 20:01, Thursday 02 October 2025 (87267)
Ops EVE Midshift Update

We are still Observing and have been Locked for over 7 hours now. Nothing to report.

H1 General
oli.patane@LIGO.ORG - posted 16:32, Thursday 02 October 2025 (87266)
Ops EVE Shift Start

TITLE: 10/02 Eve Shift: 2330-0500 UTC (1630-2200 PST), all times posted in UTC
STATE of H1: Observing at 151Mpc
OUTGOING OPERATOR: TJ
CURRENT ENVIRONMENT:
    SEI_ENV state: CALM
    Wind: 12mph Gusts, 6mph 3min avg
    Primary useism: 0.03 μm/s
    Secondary useism: 0.29 μm/s 
QUICK SUMMARY: Observing at 149 Mpc and have been Locked for almost 4 hours. Secondary microseism is going back down and wind is low.

LHO General
thomas.shaffer@LIGO.ORG - posted 16:30, Thursday 02 October 2025 (87258)
Ops Day Shift End

TITLE: 10/02 Day Shift: 1430-2330 UTC (0730-1630 PST), all times posted in UTC
STATE of H1: Observing at 151Mpc
INCOMING OPERATOR: Oli
SHIFT SUMMARY: Locked for almost 4 hours. Both relocks today needed an initial alignment, and took their time to lock DRMI, but eventually got there. No calibration measurement was run today due to the lock losses around commissioning time. We will need to get one on Saturday.
LOG:

Start Time System Name Location Lazer_Haz Task Time End
22:52 LASER LVEA IS LASER HAZARD LVEA YES LVEA IS LASER HAZARD \u0d26\u0d4d\u0d26\u0d3f(\u239a_\u239a) 13:52
14:53 FAC Randy Xarm n Xarm BTE inspections 18:53
14:53 PEM Robert LVEA YES Setup measurements 16:27
17:42 FAC Kim MX n Tech clean 18:42
19:38 Cheeta Camilla Opt Lab n Optics dropoff 19:52
19:39 FAC Randy Xarm n BTE inspection, near MX 21:52
20:29 FAC Tyler Mids n Checking on and signs 21:14
21:09 ISS Keita, Jennie Opt Lab LOCAL ISS array work 00:09
21:19 PCAL Dripta PCAL Lab LOCAL Taking pictures 21:28
H1 General
thomas.shaffer@LIGO.ORG - posted 12:46, Thursday 02 October 2025 (87261)
Ops Day Mid Shift Report

We lost lock during commissioning for unknown reasons at 1800UTC. Relocking required an initial alignment, and DRMI took 12 minutes to lock, but it got there and we are now Observing at 1944UTC.

H1 ISC
elenna.capote@LIGO.ORG - posted 11:58, Thursday 02 October 2025 - last comment - 11:57, Monday 13 October 2025(87260)
Camera spot calibration

Today I began calibrating the camera servo signals into spot positions. I ran the A2L script at the nominal camera spot positions. Then, I changed the camera offset and waited for the camera servo to converge. Then I remeasured the A2L gain. Using the a2l_lookup function in /opt/rtcds/userapps/release/isc/common/scripts/decoup/BeamPosition the A2L gain can be converted into a spot position in mm. This is a slow process so it will take time to get the conversions for all.

  Camera offset A2L gain Conversion
ETMY Pitch start: -230, end: -231 start: 5.62, end: 5.92

(-23.2 mm - -21.8 mm)/-1ct 

= 1.4 mm/ct

ITMX Pitch start: -230, end: -231 start: -0.45, end: -0.56

(6.46 mm - 5.96 mm)/-1ct  

= -0.5 mm/ct

 

Comments related to this report
elenna.capote@LIGO.ORG - 10:23, Monday 06 October 2025 (87313)

I continued today following the same process stated above.

  Camera offset A2L gain Conversion
ETMX P start: -173, end: -172 start: 3.04, end: 2.92

(-9.38 mm - -9.93 mm)/ 1ct 

= 0.55 mm/ct

ETMY Y start: -349.5, end: -348.5 start: 1.41, end: 1.06

(3.90 mm - 5.18 mm) / 1ct 

= -1.28 mm/ct

ETMX Y start: -422, end: -421 start: 4.85, end: 4.68

(17.22 mm - 17.84 mm)

= -0.62 mm/ct

elenna.capote@LIGO.ORG - 11:57, Monday 13 October 2025 (87442)

I reran the ITMX beam spot calibration following the same method described above, but this time I checked how the beam spot moved on both ITMX and ITMY when I changed the BS camera offset. Turns out, they give the same answer, so that's good.

  Camera Offset A2L gain Conversion
ITMX P (ITMY P) start: -230 ct, end: -231 ct start: -0.45 (0.19), end: -0.55 (0.09)

(6.41 mm - 5.96 mm)/-1 ct

= -0.45 mm/ct

ITMX Y (ITMY Y) start: -236 ct, end: -237 ct start: 3.18 (-2.75), end: 3.06 (-2.63)

(11.26 mm - 11.70 mm)/ -1ct

= 0.44 mm/ct (-0.44 mm/ct for ITMY)

 

LHO VE
david.barker@LIGO.ORG - posted 10:39, Thursday 02 October 2025 (87259)
Thu CP1 Fill

Thu Oct 02 10:07:28 2025 INFO: Fill completed in 7min 24secs

 

Images attached to this report
H1 TCS
camilla.compton@LIGO.ORG - posted 09:02, Thursday 02 October 2025 (87257)
HWS Origin Crosses Updated on ITMs

Matt, Camilla

After TJ and Dave fixed the HWS Live GUI 87252, Matt and I checked on it and re-centered the origin crosses which are meant to mark the center of where the IFO beam is heating the ITM, before and after, sdf diff. We moved them both down but we hadn't updated them in a few years so this isn't too surprising. 

Also attached is a HWS image of the test masses comparing getting to 60W in to 180s later, plot. As usual, ITMY looks clean and ITMX has it's known point absorbers. There is some edge effects on ITMX that should be ignored.

Images attached to this report
H1 SUS
filiberto.clara@LIGO.ORG - posted 11:31, Tuesday 30 September 2025 - last comment - 15:59, Monday 13 October 2025(87225)
Sat Amps Modified: PR2 and SR2

WP 12813
ECR E2400330
Modified List T2500232

The following SUS SAT Amps were upgraded per ECR E2400330. Modification improves the whitening stage to reduce ADC noise from 0.05 to 10 Hz.

Suspension Old New OSEM
PR2 M2 S1100174 S1100159 ULLLURLR
PR2 M3 S1100093 S1100141 ULLLURLR
SR2 M2 S1100169 S1100070 ULLLURLR
SR2 M3 S1100134 S1000295 ULLLURLR

F. Clara, J. Kissel, O. Patane

Comments related to this report
oli.patane@LIGO.ORG - 15:21, Thursday 02 October 2025 (87237)

(Since Jeff is out of town, I'm putting these in for these swaps.)

Here's the characterization data and fit results for S1100159, assigned to PR2 M2's ULLLURLR OSEMs.

This sat amp is a UK 4CH sat amp, D0900900 / D0901284.

The data was taken per methods described in T080062-v3, using the diagrammatic setup shown on PAGE 1 of the Measurement Diagrams from LHO:86807.

The data was processed and fit using
    ${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Scripts/
         plotresponse_S1100159_PR2_M2_ULLLURLR_20250924.m

Explicitly, the fit to the whitening stage zero and pole, the transimpedance feedback resistor, and foton design string are
Optic    Stage    Serial_Number    Channel_Number     OSEM_Name    Zero_Pole_Hz    R_TIA_kOhm    Foton_Design
PR2      M2       S1100159         CH1                UL           0.0956:5.22      120.25    "zpk([5.22],[0.0956],1,""n"")"
                                   CH2                LL           0.0967:5.28      120       "zpk([5.28],[0.0967],1,""n"")"
                                   CH3                UR           0.0978:5.34      120       "zpk([5.34],[0.0978],1,""n"")"
                                   CH4                LR           0.0969:5.3       120       "zpk([5.3],[0.0969],1,""n"")"
                                   
  
The attached plot and machine readable .txt file version of the above table are also found in
    ${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Results/

Per usual, R_TIA_kOhm is not used in the compensation filter -- but after ruling out an adjustment in the zero frequency (by zeroing the phase residual at the lowest few frequency points), I nudged the transimpedance a bit to get the magnitude scale within the ~0.25%. shown in the attached results. Any scaling like this will be accounted for instead with the absolute calibration step, i.e. Side Quest 4 from G2501621, a la what was done for PR3 and SR3 top masses in LHO:86222 and LHO:84531 respectively.
Non-image files attached to this comment
oli.patane@LIGO.ORG - 15:19, Thursday 02 October 2025 (87262)

(Since Jeff is out of town, I'm putting these in for these swaps.)

Here's the characterization data and fit results for S1100141, assigned to PR2 M3's ULLLURLR OSEMs.

This sat amp is a UK 4CH sat amp, D0900900 / D0901284.

The data was taken per methods described in T080062-v3, using the diagrammatic setup shown on PAGE 1 of the Measurement Diagrams from LHO:86807.

The data was processed and fit using
    ${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Scripts/
         plotresponse_S1100141_PR2_M3_ULLLURLR_20250924.m

Explicitly, the fit to the whitening stage zero and pole, the transimpedance feedback resistor, and foton design string are
Optic    Stage    Serial_Number    Channel_Number     OSEM_Name    Zero_Pole_Hz    R_TIA_kOhm    Foton_Design
PR2      M3       S1100141         CH1                UL           0.0957:5.22    120.25    "zpk([5.22],[0.0957],1,""n"")"
                                   CH2                LL           0.0949:5.17    120.5    "zpk([5.17],[0.0949],1,""n"")"
                                   CH3                UR           0.0977:5.33    120.375    "zpk([5.33],[0.0977],1,""n"")"
                                   CH4                LR           0.0946:5.16    120.25    "zpk([5.16],[0.0946],1,""n"")"
                                   
  
The attached plot and machine readable .txt file version of the above table are also found in
    ${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Results/

Per usual, R_TIA_kOhm is not used in the compensation filter -- but after ruling out an adjustment in the zero frequency (by zeroing the phase residual at the lowest few frequency points), I nudged the transimpedance a bit to get the magnitude scale within the ~0.25%. shown in the attached results. Any scaling like this will be accounted for instead with the absolute calibration step, i.e. Side Quest 4 from G2501621, a la what was done for PR3 and SR3 top masses in LHO:86222 and LHO:84531 respectively.
Non-image files attached to this comment
oli.patane@LIGO.ORG - 15:24, Thursday 02 October 2025 (87263)

(Since Jeff is out of town, I'm putting these in for these swaps.)

Here's the characterization data and fit results for S1100070, assigned to SR2 M2's ULLLURLR OSEMs.

This sat amp is a UK 4CH sat amp, D0900900 / D0901284.

The data was taken per methods described in T080062-v3, using the diagrammatic setup shown on PAGE 1 of the Measurement Diagrams from LHO:86807.

The data was processed and fit using
    ${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Scripts/
         plotresponse_S1100070_SR2_M2_ULLLURLR_20250924.m

Explicitly, the fit to the whitening stage zero and pole, the transimpedance feedback resistor, and foton design string are
Optic    Stage    Serial_Number    Channel_Number     OSEM_Name    Zero_Pole_Hz    R_TIA_kOhm    Foton_Design
SR2      M2       S1100070         CH1                UL           0.0961:5.24    120.25    "zpk([5.24],[0.0961],1,""n"")"
                                   CH2                LL           0.0965:5.27    120.375    "zpk([5.27],[0.0965],1,""n"")"
                                   CH3                UR           0.0961:5.23    120.5    "zpk([5.23],[0.0961],1,""n"")"
                                   CH4                LR           0.0963:5.26    120.75    "zpk([5.26],[0.0963],1,""n"")"
                                   
  
The attached plot and machine readable .txt file version of the above table are also found in
    ${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Results/

Per usual, R_TIA_kOhm is not used in the compensation filter -- but after ruling out an adjustment in the zero frequency (by zeroing the phase residual at the lowest few frequency points), I nudged the transimpedance a bit to get the magnitude scale within the ~0.25%. shown in the attached results. Any scaling like this will be accounted for instead with the absolute calibration step, i.e. Side Quest 4 from G2501621, a la what was done for PR3 and SR3 top masses in LHO:86222 and LHO:84531 respectively.
Non-image files attached to this comment
oli.patane@LIGO.ORG - 15:26, Thursday 02 October 2025 (87264)

(Since Jeff is out of town, I'm putting these in for these swaps.)

Here's the characterization data and fit results for S1000295, assigned to SR2 M3's ULLLURLR OSEMs.

This sat amp is a UK 4CH sat amp, D0900900 / D0901284.

The data was taken per methods described in T080062-v3, using the diagrammatic setup shown on PAGE 1 of the Measurement Diagrams from LHO:86807.

The data was processed and fit using
    ${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Scripts/
         plotresponse_S1000295_SR2_M3_ULLLURLR_20250924.m

Explicitly, the fit to the whitening stage zero and pole, the transimpedance feedback resistor, and foton design string are
Optic    Stage    Serial_Number    Channel_Number     OSEM_Name    Zero_Pole_Hz    R_TIA_kOhm    Foton_Design
SR2      M3       S1000295         CH1                UL           0.0964:5.27    121.75    "zpk([5.27],[0.0964],1,""n"")"
                                   CH2                LL           0.0954:5.21    121.625    "zpk([5.21],[0.0954],1,""n"")"
                                   CH3                UR           0.0955:5.22    121.375    "zpk([5.22],[0.0955],1,""n"")"
                                   CH4                LR           0.0964:5.26    122    "zpk([5.26],[0.0964],1,""n"")"
                                   
  
The attached plot and machine readable .txt file version of the above table are also found in
    ${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Results/

Per usual, R_TIA_kOhm is not used in the compensation filter -- but after ruling out an adjustment in the zero frequency (by zeroing the phase residual at the lowest few frequency points), I nudged the transimpedance a bit to get the magnitude scale within the ~0.25%. shown in the attached results. Any scaling like this will be accounted for instead with the absolute calibration step, i.e. Side Quest 4 from G2501621, a la what was done for PR3 and SR3 top masses in LHO:86222 and LHO:84531 respectively.
Non-image files attached to this comment
H1 IOO (ISC, PSL)
jennifer.wright@LIGO.ORG - posted 10:51, Thursday 25 September 2025 - last comment - 09:27, Friday 03 October 2025(87140)
ISS measurements with new oscilloscope setting

Yesterday I went into the optics lab and re-measured the coupling between input beam motion and PD array. While taking measurements I noticed that the injection could not be seen on the AC readouts of the PDs (example) so I tuned the temperature of the laser via changing the resistance of the controller, I went from ~10kOhms up to 13 kOHms and down to 8kOhms and while I found places where the noise reduced see example of noisy trace here, I couldn't find anywhere with the controller where the trace renamed stably in the non-noisy state. I then decided to tune the pump current down from 130mA to ~100mA and eventually found a somewhat stable place. I still had to wait through some periods of noise to trigger the measurement of the PDs.

I alos increased the modulation ampltitude to 80 mVpp. The counts on the QPD LCD readout were 10672, see image.

When the laser is in its quiet state the AC PD traces should comfortably fit on the screen of the osclloscopes with a 5mV scale, with the laser noisy this is more like 100mV, I also use 100mV scale for the QPD, I didn't change it when I reduced the noise on the laser.

The noisy state for the QPD outputs is here, the quiet state is here.

For each measurement I used a capture range of 400ms on the time axis of the scope and 125 000 samples selected on the 'ACQUIRE' menu.

The final measurements are:

PD 1 - 4 measured at AC: T0012ALL.CSV

PD 5 - 8 measured at AC: T0014ALL.CSV

QPD X, Y and SUM channels measured at AC: T0013ALL.CSV

The two DC measurements are going to be averaged so I didn't wait for a quiet time to measure them.

PD 1 - 4 measured at DC: T0011ALL.CSV 

PD 5 - 8 measured at DC: T0010ALL.CSV

To save you need to click on the menu button and change the resolution to be'Full', the format to be CSV and the channels captured to be 'ALL', the file number will roll over every time you save so you don't need to enter it manually.

Images attached to this report
Comments related to this report
jennifer.wright@LIGO.ORG - 12:49, Monday 29 September 2025 (87200)

The code to produce this plot is in my optics lab code repo.

The graph of the TF from horizontal dither on the input mirror to horizontal dither across the array, shows that we are not getting much coherent modulation of the light intensity on the PDs at 100Hz which is the dither frequency. Either my code is wrong or I need to increase the dither amplitude for the mirror.

The maths to work this out was 

A time series = abs(dither in direction horizontal to bench on QPD in V)/ (motion horizontal to bench volts on QPD/mm moved horizontal to bench on QPD)

B time series = AC voltage on each PD / mean of DC voltage on each PD

 

TF = CSD (B, A) / PSD (A, A)

I used gwpy for the calculations this time.

The attached plot shows each PD in the array as a different colour with magnitude on the top and phase on the bottom.

Images attached to this comment
jennifer.wright@LIGO.ORG - 09:27, Friday 03 October 2025 (87275)

We realised in the analysis that we should be using :

H_amp on QPD = Xcos theta + Y sin theta where theta is the angle between the X axis of the QPD and the horizontal scan direction of the beam worked out from our previous calibration measurments, 14 degrees.

This gives a different value for the couplings. All the PDs other than PD6 did not have coherences 0.9 or over so I only attach the final TF for PD6. We increased the dither amplitude after this to improve the measurement of the other PDs.

 

Images attached to this comment
Displaying reports 4281-4300 of 89074.Go to page Start 211 212 213 214 215 216 217 218 219 End